
Soleil-Babinet Compensator Chapter 5: Optional LabVIEW Software and Serial Support
Rev E, Sept. 14, 2018 Page 12
5.3.4. Calibrating the Compensator
In order to calibrate the compensator, you will need two linear polarizers mounted in rotation stages and a calibration
laser. Usually a HeNe laser is used as the calibration laser but other lasers could be used as long as the spectral
linewidth is not too large.
Although this procedure looks complex, after performing it once or twice you will become familiar with it to the point
that calibrations will be quick and easy.
Thorlabs carries a complete line of polarizers, optic mounts, rotation stages, and other accessories. Please visit our
website, www.thorlabs.com, or contact technical support using the information at the end of this manual for more
information.
1. Set the polarizers, compensator, and test laser as shown in figure 2.
2. With the compensator removed from the setup, rotate one of the two waveplates until the two polarizers are
exactly crossed at 90°. It helps to place a target screen after the last polarizer so that the transmitted beam can be
safely observed. The polarizers are crossed when this beam vanishes.
3. Re-insert the compensator back into the setup. The beam will reappear since the compensator will introduce
an arbitrary elliptical polarization which is not extinguished completely by the second polarizer.
4. Loosen the rotation lock screw and rotate the compensator until the beam disappears again. At this point, the
fast axis is aligned with the transmission axis of one of the two polarizers. Tighten the lock screw.
5. Check the tip and tilt adjustment by observing the reflected beam and adjust the kinematic adjustment screws
on the back plate until the compensator is perpendicular to the beam propagation. If necessary, repeat step 4 to
ensure the fast axis is still in alignment with the polarizer.
6. The compensator must now be rotated so that the fast axis is 45° to the linear polarizer. The compensator has
spring-loaded detent stops that allow the compensator to be easily positioned. Firmly grasp the compensator optics
housing and rotate the compensator to 45°. If the compensator does not move readily, the detent spring tension
screw may need to be loosened first. At this point, the compensator is aligned and ready for calibration.
7. Adjust the micrometer clockwise and find the starting Null point. This is the first micrometer position closest to
0.0000 that the beam extinguishes. At this point the retardance is exactly zero waves.
8. Press the Set Starting Null button to capture this location. The screen will update the text box above the button
with the current micrometer value. Optional: If you prefer, you can press the ZERO button on the micrometer to set
this position to 0.0000, however, this is not necessary since the software will do the necessary calculations.
9. Adjust the micrometer counterclockwise to find the second null. At this point, the retardance is exactly one
wave.
10. Press the Set Ending Null button to capture this location. At this point, the software calculates the distance of
one full wave of retardance.
WARNING
Do not null the micrometer display from this point on. Doing so will invalidate the calibration and
you will need to repeat this procedure.
11. If you haven’t done so already, enter the wavelength of the calibration laser in the “Enter Calibration Source
Wavelength” input box. This can be done either before or after the above steps.
At this point, the compensator is accurately calibrated and ready to be used.