8
quality factor (Q) in 4-digit.
△%: The instrument selects range according to the set value on BCD, and the result is
output after each measurement. The measurement speed is fast up to 8 times/sec. So the
mode is applied to fast measure batches of devices. The output result is displayed in
3-digit percentage, and quality factor in 4-digit. And ∆%=[(Lx-L)/L]×100%, where L is
nominal value, and Lx is measurement value.
In %△mode, nominal inductance is set in the mode of L×10nuH. For example, when the
nominal inductance is 47 mH, 47mH=47×103uH=0.47×105uH,
The set value should be transformed into two-digit decimal and one-digit power, as the
following figure shows:
2.2.3.2 If L and △% are synchronously off, the instrument is in the state of measuring
BCD switch reading. The set inductance nominal value (L), high limit value (L-), low limit
value (L-), and quality factor limit value (Q-) are displayed, then press L or △% mode key.
The instrument will automatically enter the measurement state after this measurement
completes.
2.2.4 Sorting function
2.2.4.1 At the time of producing inductor or incoming inspection, a large number of
inductors of the same model need measuring. And it needs to be known whether the
component’s parameter is in some particular range, instead of its detailed value. This is
sorting. What the instrument is asked to do is to get the result fast and simply. TH2773A
provides the functions of one-bin inductance sorting and quality factor qualification sorting.
In all the modes of inductance direct reading (L), percentage deviation (△%), and quality
factor (Q), sorting function is valid.
2.2.4.2 Percentage limit and quality factor limit setting
When setting bin limit, inductance qualified bin that data on BCD switch denote is in the
form of percentage, and the default high limit is positive, low limit negative.
For example, the inductance bin limit is set from -12.3%to 45.6%, shown as the following
figure:
And quality factor limit is set to 10.2, shown as the following figure: