Trumpf TruSystem 7500 User manual

TruSystem 7500 Operating Table Column
Instruction manual

We at Trumpf Medical are constantly developing our products.
Therefore, we reserve the right to make changes to the format, equipment, and technology at any time.
Reprinting, copying or translating this document, in whole or in part, is forbidden without the express written
permission of TRUMPF Medizin Systeme GmbH + Co. KG.
TRUMPF Medizin Systeme GmbH + Co. KG expressly reserves all rights under copyright law.
Within the bounds of the legal requirements, the manufacturer is responsible for the technical safety characteristics
of this apparatus only if the maintenance, repairs, and modifications to this apparatus are performed by him or
by someone appointed by him and in accordance with his instructions.
Valid for CE marked products (see model plate):
–CE mark: This is a Class I medical device according to the Council Directive 93/42/EEC
concerning medical devices and is compliant with the Directive version currently in force at
the time of product sale including its changes.
–Conformity: The manufacturer declares the conformity of this product with the essential
requirements of the Council Directive 93/42/EEC concerning medical devices according
to Annex I, as well as the implementation of an assessment procedure required for Class I
product conformity under Annex VII and documents this with the CE mark.
Manufacturer and distributor TRUMPF Medizin Systeme GmbH + Co. KG
Carl–Zeiss–Straße 7–9
07318 Saalfeld
Germany
www.trumpfmedical.com
Phone +49 3671 586–0
Fax +49 3671 586–41165
Technical customer service Phone +49 3671 586–41911
Fax +49 3671 586–41175
E-mail [email protected]
Translation of the original German instruction manual
Revision level: 02/11/2017
Document number: 4990113_030_01 © TRUMPF Medizin Systeme GmbH + Co. KG

4990113_030_01 — 2069106 — 2017–11–02 3
This instruction manual applies to the following sales units:
Designation Material number
Mobile operating table column
Mobile column TruSystem 7500 1717023
Mobile column TruSystem 7500 U 1730720
TruSystem 7500 Hybrid (MC) 1773204
Operating table column, floor mounted (fixed installation)
Stationary column TruSystem 7500 1717020
Stationary column TruSystem 7500 U 1730731
TruSystem 7500 Hybrid (SC) 1854086
TruSystem 7500 Hybrid Plus (SC) 1854088
TruSystem 7500 Hybrid MR (SC) 2064886
TruSystem 7500 Hybrid MR IMRIS 2067886
Operating table column, floor structure (fixed installation)
Floor mounting column TruSystem 7500 1717021
Floor mounting column TS 7500 U 1730732
TruSystem 7500 Hybrid (FC) 1854085
TruSystem 7500 Hybrid Plus (FC) 1854087
TruSystem 7500 Hybrid MR (FC) 2064837
External power supply unit
Power supply pack 2065993

44990113_030_01 — 2069106 — 2017–11–02
This page is intentionally empty.

Content
4990113_030_01 — 2069106 — 2017–11–02 5
Content
1 Important information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.1 Proper use of the equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.2 Contraindications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
1.3 Product combination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
1.4 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
1.4.1 Safety during operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
1.4.2 Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
1.4.3 High-frequency (HF) surgical equipment and defibrillators . . . . . . . . . . . . . . . 18
1.4.4 MRI safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
1.4.5 Maintenance/Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
1.4.6 Protection against infection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
1.4.7 Environmental protection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
1.5 Explanation of Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
1.6 Key performance feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
1.7 Service life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2 Unpacking and setting up the operating table column. . . . . . . . . . . . . . 22
3 Overview of operating table column . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.1 Mobile operating table column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.2 Fixed installation operating table column with internal power supply. . . . . . . 26
3.3 Fixed installation operating table column without internal power supply . . . . 27
3.4 Labels on the product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.5 Audible signals (factory setting) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4 Control units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.1 Malfunction caused by other devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.2 Column keypad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.2.1 Overview of functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.2.2 Visual indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.3 Remote control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.1 Internal power supply (not available for TruSystem 7500 Hybrid MR) . . . . . 42
5.1.1 mobile operating table column. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.1.2 Fixed installation operating table column. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.2 External power supply (line connection) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.2.1 Mobile operating table column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.2.2 Fixed installation operating table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.3 External power supply (available for TruSystem 7500 Hybrid MR). . . . . . . . 45
5.4 Charging the operating table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.4.1 mobile operating table column. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.4.2 Fixed installation operating table column. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6 Equipotential bonding of the mobile operating table column . . . . . . . . 46
7 Switch on the operating table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Content
64990113_030_01 — 2069106 — 2017–11–02
8 Switch off the operating table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
8.1 mobile operating table column. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
8.2 Fixed installation operating table column with internal power supply . . . . . . .47
8.3 Fixed installation operating table column without internal power supply . . . . .47
9 Special features of the operating table column . . . . . . . . . . . . . . . . . . . 48
9.1 Networks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
9.2 External control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
10 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
11 Load capacity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
12 Operate fixed installation operating table column. . . . . . . . . . . . . . . . . 53
12.1 Key release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
12.2 Selecting the head position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
12.3 Level position. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
12.4 Tilt. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
12.5 Anti-Trendelenburg/Trendelenburg . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
12.6 Lift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
12.7 Rotate fixed installation operating table column . . . . . . . . . . . . . . . . . . . . . . . .59
12.8 Endolight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
12.9 Localize wireless TruSystem 7500 remote control . . . . . . . . . . . . . . . . . . . . . .60
12.10 Emergency mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
12.11 Further functionalities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
13 Configuration options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
14 Transfer (shuttle operation) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
14.1 Operating tabletop (without operation table column) . . . . . . . . . . . . . . . . . . .66
14.1.1 Transfer from the shuttle to the operating table column. . . . . . . . . . . . . . . . . . .66
14.1.2 Transfer of the operating table column to the shuttle. . . . . . . . . . . . . . . . . . . . .67
14.2 Mobile operating table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
14.2.1 Transfer to the shuttle. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
14.2.2 Set down operating table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
14.3 Changing direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
14.4 Emergency transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
15 Care, cleaning and disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
15.1 Operating table column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
15.2 External power supply unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
15.3 Cleaning shuttle. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
16 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
16.1 Conditions for storage and transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
16.2 Conditions for operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
16.3 Operating table column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
16.4 External power supply unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
16.5 Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78

Content
4990113_030_01 — 2069106 — 2017–11–02 7
17 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
18 Maintenance and repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
19 Decommissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
19.1 Mobile operating table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
19.2 Fixed installation operating table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
19.2.1 Operating table column with internal power supply . . . . . . . . . . . . . . . . . . . . 82
19.2.2 Operating table column without an internal power supply . . . . . . . . . . . . . . . 82
20 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
21 EMC notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
22 Information on ordering additional equipment for the operating
table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
23 Radio license . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88

Content
84990113_030_01 — 2069106 — 2017–11–02
This page is intentionally empty.

Important information
4990113_030_01 — 2069106 — 2017–11–02 9
1 Important information
The quality management system at TRUMPF Medizin Systeme
GmbH + Co. KG has been certified in accordance with the
currently applicable EN ISO 13485 standard.
Prior to using the products, read this instruction manual carefully from
the beginning to the end in order to familiarize yourself with the
product and its functions in a step-by-step manner. Keep the
operating instructions near the product so that information can be
found at a later point in time. The instruction manual is an integral
part of the product and must be handed over when changing
location or personnel. Furthermore, the instruction manual must be
accessible to all product users at all times.
The operating table column is safe to use. Residual risks are
addressed in the respective sections of this manual. Pay close
attention to the information contained in this manual to reduce risk.
Information on the installation of a fixed installation operating table
column can be found in the installation instructions for the operating
table column.
The illustrations in these operating instructions are intended for
general understanding and may differ from the actual design of the
product. No claims can be derived from the illustrations in the
operating instructions.
Glossary The following terms and abbreviations are used in this instruction
manual:
CAUTION
Risk of injury for patients and personnel.
The operating table column is only intended for use in medical
facilities and may only be used/operated by trained medical staff.
Term Explanation
ITALIC text Descriptions of keys, functions, displays,
and device parts
Numbers in brackets
[50]
Number code in graphics (allocation of
graphic - text)
[a10] Labeling of displays on a control unit
[i10] Labeling of functions (keys) on a control
unit
[10] Labeling of parts
# Material number
Button Button on the remote control touch
screen

Important information
10 4990113_030_01 — 2069106 — 2017–11–02
CT Computed tomography system
DGHM Deutsche Gesellschaft für Hygiene und
Mikrobiologie (German Society for
Hygiene and Microbiology)
Trumpf Medical TRUMPF Medizin Systeme GmbH +
Co. KG
interventional radiology Therapeutic operations are performed
under image control (such as CT, MRI).
Medical device Device classified as a medical device as
per Directive 93/42/EEC
MRI Magnetic resonance imaging
Operating table Operating tabletop on an operating
table column (operating table system)
The section segments attached to the
operating tabletop are also part of the
operating table.
Mobile operating table The operating table can be transported
by a shuttle.
Fixed installation
operating table
The operating table column is
permanently installed in the operating
room.
Section segment Add-on component for further furnishing
of the operating table
The section segment is fastened to
coupling points on the operating table
and extends the space for positioning
the patient.
PEMS programmable electrical medical
systems
Column keypad Operating keypad on the operating
table column
Shuttle Transporter for the operating table or the
operating tabletop
Transfer (shuttle
operation)
Transfer a mobile operating table or an
operating tabletop on a shuttle.
Set down a mobile operating table or
transfer an operating tabletop to an
operating table column.
Term Explanation

Important information
4990113_030_01 — 2069106 — 2017–11–02 11
1.1 Proper use of the equipment
The devices described in this manual are for human medical use
only. The operating table column may only be used/operated by
trained personnel in a controlled and responsible manner! Training
of operating and nursing personnel may only be done by the
manufacturer or other persons authorized by the manufacturer.
The operating table column may only be used with other products
from chapter 1.3 on page 13. The operating table is intended for
the following use: Patient positioning during surgery, from the
induction of anesthesia through the actual surgery to recovery from
anesthesia. Information on the proper use of products used in
combination with the operating table column is provided in the
instruction manuals of the specific products.
VAH Verbund für Angewandte Hygiene e. V.
(German Registered Association for
Applied Hygiene)
Accessories, accessory
components
Object designed by the manufacturer to
enable or support the intended use of the
operating table.
Product name description
Hybrid (Plus) Operating table column with integration
interface to be used with an imaging
system, for example, computed
tomography
Hybrid MR Operating table column has a separate
power supply unit
Operating table column with integration
interface for use with an imaging system,
for example magnetic resonance
tomography
SB, FC The operating table column is firmly
mounted on a floor mounting plate in the
operating room.
SF, SC The operating table column is firmly
mounted on a floor installation plate in
the operating room.
SM, MC Mobile operating table column
U Supply voltage and network frequency
of the operating table column comply
with the US American state of the art.
Term Explanation

Important information
12 4990113_030_01 — 2069106 — 2017–11–02
The operating table column TruSystem 7500 Hybrid can be used
with an imaging system. This is subject to compatibility having been
confirmed in advance by Trumpf Medical and the manufacturer of
the imaging system.
The positioning of the patient on the operating table is in
accordance with general practice and doctrine e.g. described in
specialist literature.
Comply with the operating instructions for the proper use of the
operating table column. Any other use of the operating table
column is regarded as improper use. The supplier/manufacturer is
not liable for damage to property or injury to persons resulting from
improper use.
1.2 Contraindications
• Do not transport objects, devices, or materials on the operating
table.
• The operating table column is not approved for use with shuttles
types “1, 2, 3”.
WARNING
Hazard to the patient!
Changes to the medical device are prohibited! The manufacturer
will not be held liable if changes are made to the operating
tabletop.
WARNING
Risk of personal injury and material damage when
permissible loads are exceeded!
Do not exceed the permitted load capacity for the operating table.
If the permissible load is exceeded, the mobile operating table
may tip over and cause serious injuries to the patient or staff. In
general, overloading the operating table can lead to a failure of
electrical functions and cause material damage to mechanical
parts!

Important information
4990113_030_01 — 2069106 — 2017–11–02 13
1.3 Product combination
Use of the operating table column is permissible in combination with
the following Trumpf Medical products. The instruction manuals for
the products used must be observed and followed for their use.
Remote controls The following remote controls can be used on the operating table
column:
Shuttles The following shuttles are authorized for the operating table
columns:
CAUTION
Hazard to the patient!
The patient may be positioned on the operating tabletop only lying
down. Extremities must not extend beyond the end of the operating
table in a longitudinal direction.
Improper loading may damage the operating tabletop or cause
the mobile operating table to tip over! When the tabletop has been
moved in a longitudinal or transverse direction, the patient may not
climb onto or off the operating tabletop over the extended tabletop
side. Patients may only get on or off in the area of the support
column. Do not sit, perch or kneel on the end of the tabletop.
Designation Material
number
TS 7500 wired remote control 2003862
TS 7500 wireless remote control 2003863
TS 7500 SL wireless remote control
included in the packs:
– TruSystem 7500 SensorLine Premium 1997423
– TruSystem 7500 SensorLine 1947993
TS 5500 IR remote control 1637573
TS 5500 wired remote control 1637572
Designation Material
number
Shuttle 1.6 1249145
Shuttle 1.6 360 1418348
Shuttle 2.6 1249146
Shuttle 2.6 360 1418349
Shuttle 3.6 1249147

Important information
14 4990113_030_01 — 2069106 — 2017–11–02
Operating tabletops The following operating tabletops and variants can be used on the
operating table columns:
Before using an operating tabletop on the operating table column,
check the instruction manual to determine which shuttles are
authorized for the operating tabletop.
Please observe the specific instruction manual for compatibility with
the operating tabletop. Please contact Trumpf Medical if you have
any questions concerning compatibility.
Shuttle 3.6 360 1489491
Shuttle 2.7 1254362
Shuttle 2.7 360 1459862
Shuttle 3.7 1254363
Shuttle 3.7 360 1459863
Power Shuttle (H) 1774603
Designation Material
number
Designation (see label on the operating tabletop)
Operating tabletop U14
Operating tabletop U24
Operating tabletop U26
Operating tabletop ST14
Operating tabletop ST26
Operating tabletop SQ14
Operating tabletop Carbon X-TRA 7500
Operating tabletop Carbon X-TRA 7500 short
Operating tabletop Carbon FloatLine
Operating tabletop MR Neuro
Operating tabletop Carbon Spine
Operating tabletop PTS Combi Suite
Operating tabletop L
Operating tabletop LK
Operating tabletop OT
Operating tabletop pediatrics
Operating tabletop GM

Important information
4990113_030_01 — 2069106 — 2017–11–02 15
1.4 Safety instructions
Safety instructions is defined as measures to protect the user and
patient from dangers that might occur through the use of the
operating table.
The medical device is subject to special precautions regarding
electromagnetic compatibility (EMC). Installation and
commissioning must be done in accordance with the EMC notes
specified in chapter 21 on page 85.
In critical situations during an operation, changing the patient's
position by means of operating table movements (level position,
Trendelenburg) may be necessary to facilitate lifesaving measures.
In the rare event that an operating table fails during surgery,
alternate measures may need to be taken to reposition the patient.
These measures include supporting the patient with additional pads
or transferring the patient from the operating table. When using the
operating table column in an imaging system, free access to the
patient must be ensured by first driving away or swiveling away the
imaging system.
1.4.1 Safety during operation
• These operating instructions must be read and understood in full
by the operating personnel, especially the chapter “Safety
instructions”, prior to start-up or, if appropriate, communicated by
means of an in-house training session taking the level of
professional training into account. The instruction manual must be
strictly observed and made available at the location of use.
• Only use the operating table with the defined products
(chapter 1.2) for the purpose set out in the “Proper use of the
equipment” (chapter 1.1).
• The values defined under “Technical data” (chapter 16) must be
observed for the operating table.
• All work with or on the operating table (set-up, operation,
maintenance, decommissioning, transport, and disposal) may
only be performed by trained staff.
• Risk of material damage to the electronics due to condensation.
When transitioning from a cold to a warm environment or vice
versa, moisture can form inside the operating table and cause a
WARNING
Patient hazard due to malfunction!
Given the present state of the art on the market, the possibility of an
operating table malfunction cannot be fully excluded. While very
unlikely, there does exist a possibility that the motorized table
functions may fail during surgery.

Important information
16 4990113_030_01 — 2069106 — 2017–11–02
short circuit. If this occurs, place the operating table in a room for
some time (at least 12 hours) with the ambient temperature and
humidity prescribed (see chapter 16.2 on page 75) and allow
it to warm/cool prior to use.
The batteries must be completely recharged after a significant
change in temperature or humidity.
• Check that all electrical and mechanical functions/parts of the
operating table, including accessories, are undamaged and in
good working order before using them.
Defective or damaged products must not be used!
• Hazard to the patient due to improper attachment! Check that
the section segments and accessories are securely fastened to
the operating table before each use. Spontaneous movement
may occur and loose elements may slip out! Loosen control units
for adjustment purposes only. Clamp them down immediately
thereafter! Always check that fastening elements are secured
properly.
• Risk of crushing by moving parts! Familiarize yourself with its
operating functions before using the product!
• Note the labels on the product (see chapter 3.4 on page 28)!
All labels on the device must be complete and legible! Replace
damaged or loose labels! Labels may not be changed,
removed or covered with other objects.
• For the proper and safe use of the operating tabletop and
additional equipment, follow the instructions for the relevant
products.
Transport/
Repositioning
• Occupational safety for personnel: Given their heavy weight,
when lifting or carrying the section segments and accessories, or
when manually moving the operating table with or without a
patient, take extra care not to strain your back! Work with an
additional person if necessary. Never couple or uncouple
multiple section segments or heavy, unwieldy accessories at the
same time! Hold the component securely and do not drop it!
• Secure the patient on the operating table (e.g. by using belts)
and provide active support to the patient when making
adjustments to the operating table.
• Risk of collision when adjusting the operating table! Perform all
patient repositioning in a controlled and responsible manner!
Pay close attention to all electrically supported operating table
functions and movements up to the final position. This is to rule
out any hazard to the patient and material damage to the
operating table, equipment or furnishings. Potential collisions
must be prevented by aborting the function!
Make sure that no surgical draping, hoses or other objects come
in contact with movable parts and get caught by them.

Important information
4990113_030_01 — 2069106 — 2017–11–02 17
Pay particular attention to the operating table when lowering,
since the operating tabletop may collide with objects lying
below it, or it may even make contact with the floor if set to an
extreme position. In some circumstances, there is a possible
crushing hazard for persons.
• When moving to the level position on the operating table, larger
patient incline positions can occur compared to the initial
position. Only select the function Level position (transfer position)
if collisions cannot occur. Monitor the movement to the level
position up to the final position.
Only in case of mobile operating table column:
• Danger of collisions when transporting the operating table. Be
careful to avoid collisions with nearby persons or furniture and to
prevent damage to the table!
1.4.2 Electrical safety
• Position the operating table such that the detachable plug or
device couplers are easily accessible.
• The operating table may only be connected to a power supply
with a protection conductor. This prevents the risk of electric
shock.
• Get a qualified electrician to check the electrical safety of the
operating table and the power supply every year.
We recommend an annual general safety inspection by
Trumpf Medical Technical Customer Services.
• Always keep the wireless remote control in an operational
(charged) state so that it is ready for use immediately!
mobile operating
table column
• When used on an electrically conductive floor, the operating
table may be used only if connected to an equipotential
bonding cable. The equipotential bonding cable connection
location on the operating table is in accordance with
IEC 60601-1 (DIN EN 60601-1/VDE 0750-1).
• An extension cable cannot be used with the mains power cable
at the operating table, otherwise, in case of a fault, the permitted
patient leakage current threshold for cardiac surgery (applied
part CF according to IEC 60601-1) will be exceeded. The fault
occurs when the protection conductor is interrupted by a
damaged mains power cable. The operating table must only be
connected to the power supply using the original mains power
cable from Trumpf Medical.
In addition, the supply voltage must be observed. The operating
table must only be connected to the supply voltage indicated on
the device label.

Important information
18 4990113_030_01 — 2069106 — 2017–11–02
• Ensure that the power cable and equipotential bonding cable
are not crushed or driven over. Stop using cables if damaged. If
you have any concern about the safety of the power cables and
equipotential bonding conductors, work only with the internal
(rechargeable) power supply until you have replaced the cable
in question.
Remove cables prior to relocating the table.
• Always keep the operating table column in an operational
(charged) state so that it is ready for use immediately! Fully
charge the operating table column prior to first use.
Fixed installation
operating table
column
• In an emergency, disconnect the operating table column from
the power supply using the separate on/off switch (e.g. on the
operating panel).
1.4.3 High-frequency (HF) surgical equipment and defibrillators
The operating table is suitable for the use of HF surgical
equipment, defibrillators, and defibrillator monitors. Please
follow the manufacturer's instruction manual and safety
instructions for the equipment.
Trumpf Medical operating tables are electrically conductive in
accordance with the current version of the IEC 60601-1
regulations. When high-frequency surgery devices, defibrillators,
and defibrillator monitors are used, the patient runs the risk of burns
if safety precautions are not followed.
• Risk of burns to patient! Position the patient on the operating
table so that he or she is insulated from metal parts (operating
table, accessories) and conductive pads or hoses.
• Make sure the patient does not come into contact with damp
towels or pads! Only use dry materials!
• The electrically motorized operating table functions can be
interrupted when high-frequency surgical devices are used
simultaneously.

Important information
4990113_030_01 — 2069106 — 2017–11–02 19
1.4.4 MRI safety information
Tests have confirmed that the operating table column Hybrid MR is
MR conditional to the MR environment with the approved operating
tabletops and the cable remote control TS 7500 (#2003862) and
can be used safely for MRI up to a maximum of 3 Tesla.
1.4.5 Maintenance/Repair
• It is prohibited to position a patient on the operating table if the
table is undergoing maintenance or repair.
• Keep a log book for this medical equipment! All repairs and
maintenance are documented here.
• Every repair must comply with the safety regulations according
to IEC 60601-1 (DIN EN 60601-1/VDE 0750-1) and
DIN EN 62353 (VDE 0751-1).
• Within the limits of the legal requirements, the manufacturer is
only responsible for the technical safety characteristics of this
device, if the maintenance, repairs and modifications to this
apparatus are performed by Trumpf Medical service technicians
or by personnel authorized and trained by Trumpf Medical.
• Trumpf Medical will not be held liable for damage of any kind
arising from the failure to perform inspections or as a result of
inadequate maintenance, or of modifications to the product.
CAUTION
Failure to observe the specific procedures for handling
MR conditional products can result in injury to persons
involved.
Please observe the following points to ensure safety in an MRI
environment:
– Please observe the MRI safety information for the operating
table top, the shuttle and the remote control.
– Only use pads specifically designed for the operating table top
(MR safe).
– Do not use the operating tabletop with the MRI until the shuttle
has been transferred to the operating table column (MR
conditional).
– Hang the remote control on the side rail of the operating
tabletop by the “PLACE PENDANT HERE” sticker (MR
conditional).
– Remove the shuttle from the room immediately after the
operating tabletop has been transferred (MR Unsafe).
– Only use authorized accessories.
– It is essential to follow the safety guidelines of the MR device
manufacturer.

Important information
20 4990113_030_01 — 2069106 — 2017–11–02
1.4.6 Protection against infection
• Follow all rules and regulations concerning cleaning /
disinfection (see chapter 15 on page 73). Use only the cleaning
materials and disinfectants described in these operating
instructions!
• Only devices and kits that have been cleaned and disinfected
may be handed over to service technicians for maintenance and
repair work.
1.4.7 Environmental protection
• Dispose of any remaining cleaning and disinfection materials
and any associated residues in a safe and environmentally-
friendly manner.
• Use the recycling options for batteries that have become
unserviceable.
• Consult this document regarding environmentally-friendly
product disposal (see chapter 20 on page 84).
1.5 Explanation of Symbols
Important information in this instruction manual is marked with
symbols and keywords. Keywords such as DANGER, WARNING
or CAUTION indicate the level of danger involved. The symbols
emphasize the message visually. Additional symbols can indicate
injury hazards or danger to life and limb. The measures to prevent
dangers arising must be implemented without fail.
DANGER
Risk of death (for example, electric shock)!
Indicates an imminent danger that will result in death or serious
injuries if the appropriate precautionary measures are not taken.
WARNING
Risk of death!
Indicates an imminent danger that may result in death or serious
injuries if the appropriate precautionary measures are not taken.
CAUTION
Risk of injury (for example, crushing)!
Refers to a possible danger that can lead to slight to moderate
injury or damage to the equipment if the appropriate precautionary
measures are not taken.
This manual suits for next models
14
Table of contents
Other Trumpf Medical Equipment manuals