UINICS SP2441 Series User manual

【 Operation Manual 】
Compact digital meter
MODEL : SP2441 Series
(Pulse input type)
Series name Output
Input
Function
SP2441
Sensor power : DC12V 50mA MAX
Power source : DC power supply(DC24V)
Preset output : NPN open collector(×2)
Input signal : NPN open collector(×1)
Color : Black
P1 Preset output one points
(1c contact relay output)
AI Analog current output (DC4-20mA)
AV3
Analog voltage output (DC1-5V)
AV4
Analog voltage output (DC0-5V)
AV5
Analog voltage output (DC0-10V)
F Voltage pulse input
F2 Current pulse input
! Caution
For professional use only or designed for use by a licensed electrician only.
! Caution
Check if the label (model name) of the unit and your desired product
specification correspond before use.
【 The 3rd edition 21 Aug. 2020 】
@SP2441CE(3)-E

P
recautions
Please read this Operation Manual including the following precautions carefully to ensure
s
1. Perform wiring with the power off. There is a risk of
2. Do not touch the terminals while power is being supplied. There is a risk of electric shock.
3. Do not disassemble or touch the inside of the product. There is a risk of electric shock or
4. Do not use the product where t
5. In the event of a product failure or abnormality, build an emergency stop or
1. Use the power supply
2. Do not use in the following environments.
3. Do not allow metal powder, dust, water, chemicals, oil, etc., to enter
4. Check regular
5. If the product is malfunctioning or has fire, smoke, overheating, abnormal noise, etc.,
6
7. Keep the product and wiring away from noise sources.
8.
9
10
1
1
Product Description
recautions
Please read this Operation Manual including the following precautions carefully to ensure
s
afe
use of your meter.
Warning
1. Perform wiring with the power off. There is a risk of
2. Do not touch the terminals while power is being supplied. There is a risk of electric shock.
3. Do not disassemble or touch the inside of the product. There is a risk of electric shock or
fire.
4. Do not use the product where t
5. In the event of a product failure or abnormality, build an emergency stop or
fail-
safe system to ensure safety.
Caution
1. Use the power supply
2. Do not use in the following environments.
・
A place exposed to metal powder, dust, water, chemicals, oil, etc.
・
A place with corrosive gas
・
Outdoor use and places exposed to direct sunlight
・
Locations wher
・
Temperature and humidity outside the rated range
・
Locations subject to vibration or impact
3. Do not allow metal powder, dust, water, chemicals, oil, etc., to enter
the product. There is a risk of failure or fire.
4. Check regular
5. If the product is malfunctioning or has fire, smoke, overheating, abnormal noise, etc.,
immediately turn off the power and stop using it.
6
.
Install the switch or circuit
in an emergency.
Please indicate that it is a device shut
7. Keep the product and wiring away from noise sources.
8.
If there is a possibility of invasion the lightning surges,
install countermeasure parts
9
. It can be used almost at the same time as the power is turned on, but itrequires
30 minutes of power to satisfy all performance requirements.
10
. When
cleaning, wipe with a dry cloth
such as b
1
.
If the waterproof packing is used in a deteriorated state, the waterproof and
dustproof function will be impaired. Inspect and trade it periodically.
Product Description
With the panel mounted, the front has been protected by grade IP66. dust and
water resistant products.
recautions
Please read this Operation Manual including the following precautions carefully to ensure
use of your meter.
Warning
:
Content
1. Perform wiring with the power off. There is a risk of
2. Do not touch the terminals while power is being supplied. There is a risk of electric shock.
3. Do not disassemble or touch the inside of the product. There is a risk of electric shock or
4. Do not use the product where t
5. In the event of a product failure or abnormality, build an emergency stop or
safe system to ensure safety.
Caution
:
Contents that may cause minor injury or property damage.
1. Use the power supply
2. Do not use in the following environments.
A place exposed to metal powder, dust, water, chemicals, oil, etc.
A place with corrosive gas
Outdoor use and places exposed to direct sunlight
Locations wher
Temperature and humidity outside the rated range
Locations subject to vibration or impact
3. Do not allow metal powder, dust, water, chemicals, oil, etc., to enter
the product. There is a risk of failure or fire.
4. Check regular
ly for any malfunctions or abnormalities.
5. If the product is malfunctioning or has fire, smoke, overheating, abnormal noise, etc.,
immediately turn off the power and stop using it.
Install the switch or circuit
in an emergency.
Please indicate that it is a device shut
7. Keep the product and wiring away from noise sources.
If there is a possibility of invasion the lightning surges,
install countermeasure parts
. It can be used almost at the same time as the power is turned on, but itrequires
30 minutes of power to satisfy all performance requirements.
cleaning, wipe with a dry cloth
such as b
enzine, thinner or alcohol.
If the waterproof packing is used in a deteriorated state, the waterproof and
dustproof function will be impaired. Inspect and trade it periodically.
Product Description
With the panel mounted, the front has been protected by grade IP66. dust and
water resistant products.
Please read this Operation Manual including the following precautions carefully to ensure
use of your meter.
Content
s
that may cause death or serious injury.
1. Perform wiring with the power off. There is a risk of
2. Do not touch the terminals while power is being supplied. There is a risk of electric shock.
3. Do not disassemble or touch the inside of the product. There is a risk of electric shock or
4. Do not use the product where t
5. In the event of a product failure or abnormality, build an emergency stop or
safe system to ensure safety.
Contents that may cause minor injury or property damage.
1. Use the power supply
voltage and load within the specified range.
2. Do not use in the following environments.
A place exposed to metal powder, dust, water, chemicals, oil, etc.
A place with corrosive gas
Outdoor use and places exposed to direct sunlight
Locations wher
e condensation occurs
Temperature and humidity outside the rated range
Locations subject to vibration or impact
3. Do not allow metal powder, dust, water, chemicals, oil, etc., to enter
the product. There is a risk of failure or fire.
ly for any malfunctions or abnormalities.
5. If the product is malfunctioning or has fire, smoke, overheating, abnormal noise, etc.,
immediately turn off the power and stop using it.
Install the switch or circuit
breaker in a position where it can be
Please indicate that it is a device shut
7. Keep the product and wiring away from noise sources.
If there is a possibility of invasion the lightning surges,
install countermeasure parts
. It can be used almost at the same time as the power is turned on, but itrequires
30 minutes of power to satisfy all performance requirements.
cleaning, wipe with a dry cloth
enzine, thinner or alcohol.
If the waterproof packing is used in a deteriorated state, the waterproof and
dustproof function will be impaired. Inspect and trade it periodically.
Product Description
With the panel mounted, the front has been protected by grade IP66. dust and
water resistant products.
Please read this Operation Manual including the following precautions carefully to ensure
that may cause death or serious injury.
1. Perform wiring with the power off. There is a risk of
2. Do not touch the terminals while power is being supplied. There is a risk of electric shock.
3. Do not disassemble or touch the inside of the product. There is a risk of electric shock or
4. Do not use the product where t
here is flammable gas or ignited material.
5. In the event of a product failure or abnormality, build an emergency stop or
safe system to ensure safety.
Contents that may cause minor injury or property damage.
voltage and load within the specified range.
2. Do not use in the following environments.
A place exposed to metal powder, dust, water, chemicals, oil, etc.
A place with corrosive gas
Outdoor use and places exposed to direct sunlight
e condensation occurs
Temperature and humidity outside the rated range
Locations subject to vibration or impact
3. Do not allow metal powder, dust, water, chemicals, oil, etc., to enter
the product. There is a risk of failure or fire.
ly for any malfunctions or abnormalities.
5. If the product is malfunctioning or has fire, smoke, overheating, abnormal noise, etc.,
immediately turn off the power and stop using it.
breaker in a position where it can be
Please indicate that it is a device shut
7. Keep the product and wiring away from noise sources.
If there is a possibility of invasion the lightning surges,
install countermeasure parts
such as a
. It can be used almost at the same time as the power is turned on, but itrequires
30 minutes of power to satisfy all performance requirements.
cleaning, wipe with a dry cloth
,
enzine, thinner or alcohol.
If the waterproof packing is used in a deteriorated state, the waterproof and
dustproof function will be impaired. Inspect and trade it periodically.
With the panel mounted, the front has been protected by grade IP66. dust and
Please read this Operation Manual including the following precautions carefully to ensure
that may cause death or serious injury.
1. Perform wiring with the power off. There is a risk of
2. Do not touch the terminals while power is being supplied. There is a risk of electric shock.
3. Do not disassemble or touch the inside of the product. There is a risk of electric shock or
here is flammable gas or ignited material.
5. In the event of a product failure or abnormality, build an emergency stop or
Contents that may cause minor injury or property damage.
voltage and load within the specified range.
2. Do not use in the following environments.
A place exposed to metal powder, dust, water, chemicals, oil, etc.
Outdoor use and places exposed to direct sunlight
e condensation occurs
Temperature and humidity outside the rated range
Locations subject to vibration or impact
3. Do not allow metal powder, dust, water, chemicals, oil, etc., to enter
the product. There is a risk of failure or fire.
ly for any malfunctions or abnormalities.
5. If the product is malfunctioning or has fire, smoke, overheating, abnormal noise, etc.,
immediately turn off the power and stop using it.
breaker in a position where it can be
Please indicate that it is a device shut
-
off device.
7. Keep the product and wiring away from noise sources.
If there is a possibility of invasion the lightning surges,
such as a
lightning arrestor in outside.
. It can be used almost at the same time as the power is turned on, but itrequires
30 minutes of power to satisfy all performance requirements.
d
o not use organic solvents
If the waterproof packing is used in a deteriorated state, the waterproof and
dustproof function will be impaired. Inspect and trade it periodically.
With the panel mounted, the front has been protected by grade IP66. dust and
Please read this Operation Manual including the following precautions carefully to ensure
that may cause death or serious injury.
1. Perform wiring with the power off. There is a risk of
electric shock or fire.
2. Do not touch the terminals while power is being supplied. There is a risk of electric shock.
3. Do not disassemble or touch the inside of the product. There is a risk of electric shock or
here is flammable gas or ignited material.
5. In the event of a product failure or abnormality, build an emergency stop or
Contents that may cause minor injury or property damage.
voltage and load within the specified range.
A place exposed to metal powder, dust, water, chemicals, oil, etc.
Outdoor use and places exposed to direct sunlight
Temperature and humidity outside the rated range
3. Do not allow metal powder, dust, water, chemicals, oil, etc., to enter
ly for any malfunctions or abnormalities.
5. If the product is malfunctioning or has fire, smoke, overheating, abnormal noise, etc.,
immediately turn off the power and stop using it.
breaker in a position where it can be
off device.
7. Keep the product and wiring away from noise sources.
If there is a possibility of invasion the lightning surges,
lightning arrestor in outside.
. It can be used almost at the same time as the power is turned on, but itrequires
30 minutes of power to satisfy all performance requirements.
o not use organic solvents
If the waterproof packing is used in a deteriorated state, the waterproof and
dustproof function will be impaired. Inspect and trade it periodically.
With the panel mounted, the front has been protected by grade IP66. dust and
Please read this Operation Manual including the following precautions carefully to ensure
that may cause death or serious injury.
electric shock or fire.
2. Do not touch the terminals while power is being supplied. There is a risk of electric shock.
3. Do not disassemble or touch the inside of the product. There is a risk of electric shock or
here is flammable gas or ignited material.
5. In the event of a product failure or abnormality, build an emergency stop or
Contents that may cause minor injury or property damage.
voltage and load within the specified range.
A place exposed to metal powder, dust, water, chemicals, oil, etc.
3. Do not allow metal powder, dust, water, chemicals, oil, etc., to enter
5. If the product is malfunctioning or has fire, smoke, overheating, abnormal noise, etc.,
breaker in a position where it can be
operated immediately
lightning arrestor in outside.
. It can be used almost at the same time as the power is turned on, but itrequires
30 minutes of power to satisfy all performance requirements.
o not use organic solvents
If the waterproof packing is used in a deteriorated state, the waterproof and
dustproof function will be impaired. Inspect and trade it periodically.
With the panel mounted, the front has been protected by grade IP66. dust and
Please read this Operation Manual including the following precautions carefully to ensure
electric shock or fire.
2. Do not touch the terminals while power is being supplied. There is a risk of electric shock.
3. Do not disassemble or touch the inside of the product. There is a risk of electric shock or
here is flammable gas or ignited material.
5. In the event of a product failure or abnormality, build an emergency stop or
Contents that may cause minor injury or property damage.
voltage and load within the specified range.
A place exposed to metal powder, dust, water, chemicals, oil, etc.
3. Do not allow metal powder, dust, water, chemicals, oil, etc., to enter
5. If the product is malfunctioning or has fire, smoke, overheating, abnormal noise, etc.,
operated immediately
lightning arrestor in outside.
. It can be used almost at the same time as the power is turned on, but itrequires
If the waterproof packing is used in a deteriorated state, the waterproof and
dustproof function will be impaired. Inspect and trade it periodically.
With the panel mounted, the front has been protected by grade IP66. dust and
Please read this Operation Manual including the following precautions carefully to ensure
2. Do not touch the terminals while power is being supplied. There is a risk of electric shock.
3. Do not disassemble or touch the inside of the product. There is a risk of electric shock or
5. If the product is malfunctioning or has fire, smoke, overheating, abnormal noise, etc.,
operated immediately
. It can be used almost at the same time as the power is turned on, but itrequires
If the waterproof packing is used in a deteriorated state, the waterproof and
With the panel mounted, the front has been protected by grade IP66. dust and
Please read this Operation Manual including the following precautions carefully to ensure
2. Do not touch the terminals while power is being supplied. There is a risk of electric shock.
3. Do not disassemble or touch the inside of the product. There is a risk of electric shock or

Contents
1. About confirmation of an attachment. and a guaranteed period ・・・・・・・・ 1
2. Specifications ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2 - 4
3. Mounting meter ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 5
4. Connecting terminal boards ・・・・・・・・・・・・・・・・・・・・・・・・ 6 - 7
5. Construction of input/output circuit ・・・・・・・・・・・・・・・・・・・・8 - 9
6. Setting the dip switch ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 10
7. Names and functions of components on front ・・・・・・・・・・・・・・ 11 - 12
8. The setting menu ・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 13 - 14
9. Initial setting values and initialization ・・・・・・・・・・・・・・・・・・・・・・ 15
10. Content and setting the each mode ・・・・・・・・・・・・・・・・・・・ 16 - 32
≪1. Operating method (the mode setting)≫・・・・・・・・・・・・・・・・・・ 16
≪2. Content of the each mode and set value≫ ・・・・・・・・・・・・17 - 32
Mode 1 "Ratemeter : Setting of scaling data" ・・・・・・・・・・・・・・・・・ 17
Calculation example of scaling data (setting example) ・・・・・・・・ 18
Mode 2 "Ratemeter : Setting of Exp. value, least significant digit,
the unit time and the decimal point" ・・・・・・・・・・・・・・・・ 19
Mode 3 "Ratemeter : Setting of the sampling time" ・・・・・・・・・・・・・・・・ 20
Mode 4 "Ratemeter : Setting of the moving average range" ・・・・・・・ 20 - 21
Mode 5 "Ratemeter : Setting the auto-zero time" ・・・・・・・・・・・・・・・ 21
Mode 6 "Setting of EXT input and measurement indication" ・・・・・・・・・ 22
Mode 7 "Totalizer : Setting of scaling data" ・・・・・・・・・・・・・・・・・・ 23
Mode 8 "Totalizer : Setting of Exp. value, reset dwell time, overflow
indication and the decimal point" ・・・・・・・・・・ 23 - 24
Mode 9 "Setting of OUT1 preset output" ・・・・・・・・・・・・・・・・ 25 - 26
【 Chart at the timing of the presetting output 】・・・・・・・・・・27
Mode A "Setting of OUT2 preset output" ・・・・・・・・・・・・・・・・・・・ 28
Mode b "Totalizer : Setting of the synchronization pulse output" ・・・・・・・ 29
Mode C "Analog output (option) : Setting of measurement choice and
the output digit" ・・・・・・・・・・ 30 - 31
Mode d "Analog output(option) : Setting of maximum output indication"・・・・ 32

11. The mode protect function ・・・・・・・・・・・・・・・・・・・・・・・・・ 33
12. Calling up and modifying the offset value setting ・・・・・・・・・・・・・・・ 34
13. Calling up and modifying the preset value setting
・
・・・・・・・・・・・・・・ 35
14. Adjusting the analog output (option)・・・・・・・・・・・・・・・・・・・ 36 - 37
15. External dimensions
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 37
16. About a noise countermeasure ・・・・・・・・・・・・・・・・・・・・・・・・ 38
17. Troubleshooting
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 39

- 1 -
1. About confirmation of an attachment and a guaranteed period
About confirmation of attachments.
When you received as a product, please confirm whether it includes
the following.
(1)SP2441 (The chosen specification) ・・・・・・・・・・・・・・・・・・1
(2)Installation adapter(Attachment) ・・・・・・・・・・・・・・・・・・・・1
(3)SP2441 Operation manual (Digest version) ・・・・・・・・・・・・・・1
(4)Rubber packing : Color (Black)(Attachment)・・・・・・・・・・・・・・・1
(5)Unit label(Attachment) ・・・・・・・・・・・・・・・・・・・・・・・・・1
If there are the mistaking parts and the missing parts, please inform a dealer
or us.(There is a case that you don't attach by convenience.)
About a guaranteed period and a guaranteed area.
1. Guaranteed period
The period a product guarantees is 4 years from a delivered day.
2. Guaranteed area
If we trouble by responsibility in whole guaranteed period, it's repaired without
charge at our factory. But if a product conflicted in the following matter,
it isn't a guarantee target. Please understand.
(1) Case of outside of the product specifications.
(2) Case of User-conducted alterations and modifications of the unit
(3) Case of besides our responsibility.
(4) Case of safekeeping and transportation beyond the product specification
condition.
(5) Case of an accident.

- 2 -
2. Specifications
【 Standard specifications 】
Items Specifications
Measurement
Type
Ratemeter/Totalizer
Method
Cycle calculation method
Display
Display
Red LED : 5 digits, Character height : 7mm(range : 0 - 99999)
Display switching Rate meter/Totalizer
Rate
meter
Display accuracy
±0.05%rdg.±1digit
( at Sampling time for 0.5 second or more )
Scaling
1×10
-9
- 9999(per pulse)
Indication area
0 – 99999
Overflow indication
"99999" flashing, endless, or 10x shift.
Time unit
10
-1
, 10
-2
, 10
-3
, 10
-4
, nothing
Sampling time
Rate reading averaged by 0.1-100.0 sec.
Moving average
Display moving average.
Averaged by 2 to 19 indication values.
Pulses moving average
Averaged by 2 to 19 input pulses.
Auto zero time
After an input signal is stopped, the reading turns to zero after
time that can be set 0.1 – 99.9 sec.
Least significant digit
Real, fixed at 0, or 0/5
Totalizer
Measurement
accuracy
±0 (Scaling is set to 1.)
Scaling
1×10
-9
- 9999(per pulse)
Indication area
0 – 99999
Decimal point
10
-1
,10
-2
,10
-3
,10
-4
,nothing
Overflow Indication
"99999" flashing, endless, or 10x shift.
O
ffset
Value after a reset at the reach of 0 - 99999. (selectable)
Sensor
input
Input signal
NPN open collector pulse input or No-voltage contact.
*Sensor conditions:
ON residual voltage 2.0V or less, OFF leakage current 1.5mA or less.
Can open and close load current of 10.0mA.
Input response frequency
(at 50% duty)
LO : 0.01Hz - 50Hz
HI : 0.01Hz - 10kHz
Sensor power DC 12V(±10%) 50mA (max.)
EXT
input
Input method NPN open collector input or No-voltage contact. (×1)
* It works by turning it ON for 50ms or more.
Function selection
Reset / Hold / Inhibit / Indication change
Synchronize
totalizer of
pulse output
Output method NPN open collector output. (×1)
* It's impossible to use at the time of P1 option choice.
Maximum rating DC30V 50mA.(max)
Output digits 1 - 4 digits.
Output width 0.01 - 1.99 sec.
* Maximum output frequency is 50Hz.

- 3 -
Preset
output
Output method NPN open collector output.(×2)
Maximum rating DC30V 50mA(max.)
Comparative method
The upper limit, lower (immediately) and lower (delay).
Output mode Compared with the indication value by preset value.
Comparison, hold and 1 shot.
Judgment
prohibition time
The presetting output function is disabled for the specified time
interval following power ON or reset. Up to 1
–
60 sec, 9 stages.
〔Does not work with lower(delay) mode.〕
Others
Error display
function
When abnormality occurred to a sensor power supply, "Err01" and a measured
value are indicated alternately.
Data backup
Each mode set value and totalized value are saved on FRAM.
(The memory writable number of times
is within 100,000 times, retention about
10 years.)
Mode protect Prohibit mode setting change.
Warm up time After turning on the power, more than 30 minutes.
Power supply DC24V(
±
10%)
Power consumption
3W max
Temperature/humidity
conditions
-10 - 50℃ 30 - 80%RH (No-condensation)
Dimensions and weight
W48
×
H24
×
D63mm (An installation adapter isn't included.)Approx 40g
Color Black
Flame retardant grade
UL94 V-0
Material of the case
Polycarbonate
Safety class IP66 (front)
Isolation Power supply and input/output are not isolated
EMC
EN61326-1
EN55011(Group1 ClassA), EN61000-4-2, EN61000-4-3, EN61000-4-4,
EN61000-4-5, EN61000-4-6

- 4 -
【
Option specifications
】
≪Relay output : P1 option≫
Preset output
Output method
Relay c contact output(
×
1)
Maximum rating AC220V 0.12A (Resistance load)
DC
30V
1A (Resistance load)
≪Analog output : AI/ AV3 - 5 option≫
Analog output
Output signal
[AI] DC4 - 20mA Load impedance 500Ω or less.
*Connect to the power supply-side and the floating input
[AV3] DC1 - 5V Load impedance 2kΩ or more.
[AV4] DC0 - 5V Load impedance 2kΩ or more.
[AV5] DC0 - 10V Load impedance 2kΩ or more.
Accuracy
Within ±0.3%F.S. for indicated value.(at 23℃)
Temperature
characteristic
±100ppm/℃:
0 - 50℃
±120ppm/℃ :
-10 - 0℃
Response time
Approx 20ms(that is the time an output change gets to 90%)
Maximum
resolution
13000
≪Sensor input : F option , F2 option≫
Sensor input Input signal
〔F〕 Voltage pulse input
LOW : 2.0 V or less, HI : 3.8 - 30.0 V
〔F2〕Current modulation pulse input
LOW : 8 mA or less, HI : 15 - 20 mA

- 5 -
3. Mounting meter
How to mount meter
1.
Cut the panel to insert the meter from the front.
2.
Slide installation adapter from the rear to fix the body.
At this time, if the body is not secured tightly, fasten screws a little more.
・
Put the meter on a panel with a thickness of 1.0
–
4.0mm.
Fig.3-1
Panel cutout dimensions and pitch for mounting two or more meters.
Fig.3-2
(
Unit : mm)
* When using by protection against dust and waterproofing in forebody (IP66),
please use rubber packing.
DISP

- 6 -
4. Connecting terminal boards
Terminal boards
Fig.4-1
【 Standard type 】
Fig.4-2
【 P1 option type 】
Terminal pitch : 3.5mm (Phoenix : SMKDS1/12-3.5)
Wire : AWG30 – 16 (SQ Conversion : 0.05 – 1.3mm
2
)
Length of stripped wire : 5.0mm

- 7 -
<
Connection diagrams>
Sensor another power supply use.
Fig.4-3
Sensor power supply use.
Fig.4-4
DC two-line pulse output sensor
Fig.4-5
! < Caution >
Always turn the power OFF before commencing any wiring work.
! < Caution >
Please confirm the specification.
! < Caution >
DC power supply connections
If the + and - are connected in reverse by mistake, the internal protective circuit
is activated to stop the reverse current flow. In this case, disconnect, then reconnect
correctly normal operation.
! < Caution >
The input/output wiring scheme varies with sensor type. Please refer to the
connection diagrams (Fig. 4-3, 4-4, 4-5) above for wiring details, to avoid damage
to the sensor or input/output circuits.
! < Caution >
When making a connection to the terminal board, make sure the
lead wire is fully and firmly inserted.
! < Caution >
Please tighten a screw of the terminal stand surely.
! < Caution >
Please don't use a sensor power supply for the use of anything but a sensor.

- 8 -
5. Construction of input/output circuit
1. Senser input :
NPN open collector pulse input
Fig.5-1
2. Senser input : Voltage
pulse input
Fig.5-2
3. EXT input :
NPN open collector input
Fig.5-3

- 9 -
4. Analog output :
Voltage output (AV3 - 5) Current output (AI)
Fig.5-4 Fig.5-5
5. Preset output :
Synchronization pulse output
(NPN open collector output) 6. P1 Option relay output
Fig.5-6 Fig.5-7

- 10 -
6. Setting the dip switch
The input response frequency, NPN open collector pulse input, and voltage pulse
input, is selected by setting the dip switch.
DSW1-1
DSW1-2
OFF⇔ON
NPN open collecter pulse input ON -
Voltage pulse input OFF -
Input response frequency
0.01Hz - 50Hz (LOW) - ON
Input response frequency
0.01Hz~10kHz (HI) - OFF
※As default setting,
NPN open collector pulse input.
Input response frequency is HI (0.01Hz - 10kHz).
Fig.6-1
2
1
O
N

- 11 -
7. Names and functions of components on front
Fig.7-1
1. Display unit(Red)
Measurement state : A measured value is indicated.
Setting state:
A・・・・・
Mode No. is indicated.
B to E・・・The present set value is indicated.
: When preset value setting indicates the preset value.
: When offset value setting indicates the offset value.
: It indicates “L-oFF” and “L-on” at the time of mode protect
function setting.
: AnA, An-1-2 and the present bit value are indicated at the
time of analog output adjustment mode setting.
2. Mode Key
Turning on
: When a power supply is supplied while is pushing
down Mode Key, a test mode functions is called.
(A stop of the test mode function is power supply off.)
Measurement state : When Shift Key is pressing for 2 sec. or more,while is
pushing down Mode Key, mode setting is called.
: When Mode Key is pressing for 2 sec. or more, preset
value is called.
: When Display Key is pressing for 2 sec. or more,
while is pushing down Mode Key, offset value setting is called.
Setting state : Switch mode No. (1→2→3……9→A→b→C→d→1)
: OUT1,2 is switched at the time of preset value setting.
: An-1-2 is switched at the time of analog output
adjustment mode setting.
A B CD E
1.Display unit(Red)
2. 3. 4. 5.
8.Totalizer LED(Green)
6.Preset output "OUT1" LED(Orange)
7.Preset output "OUT2" LED(Orange)
MODE

- 12 -
3. Shift Key
Measurement state : Shift Key is used to call the mode setting.
(It's on for more than 2 sec with Mode Key.)
: Shift Key is used to call the mode protect function,
or used to change it.
(It's on for more than 2 sec. → The current mode protect function
state is indicated. → It's on for more than 8 sec just as it is.
→ Change.《L-oFF ⇔ L-on》)
Setting state :
Shift Key moves the flash figure when each setting, to the right.
: The output bit values decreases at the time of analog output
adjustment mode setting.
4. Display Key
Turning on :When a power supply is supplied keeping pushing
an analog output adjustment mode setting.
(A stop of the analog output adjustment mode setting is power supply off.)
Measurement state : Display Key is used to call the offset value setting.
(It's on for more than 2 sec with Mode Key.)
: Display Key is used to switch the ratemeter/totalizer.
(Need to set "Mode No.6".)
Setting state : Display Key changes the flash figure when each setting.
: The output bit values increases at the time of analog
output adjustment mode setting.
5. Reset Key
Turning on :Throw power supply while Reset Key pressed to call
initialization setting.
Measurement state:Resets the totalizer if offset value is set, change the
reading to an offset value.
It also resets the presetting output.
(When using EXT input (terminals No.6 - 7), need to set "Mode No.6".)
Setting state : Reset Key is used to register and make it measurement status.
: The bit value of "An-1 and 2" is registered at the time of
analog output adjustment mode setting.
6,7. Preset output LED (Orange)
Measurement state : "OUT1,2" turns on during generating preset output.
: "OUT1" LED turns on during generating a
synchronization pulse.
Setting state : The presetting output LED lights up at the time of the
preset output value setting.
8. Totalizer LED ( or "×10 LED") (Green)
Measurement state : It lights up at totalizer.
: When totalizer is being "x 10", it'll change to a flash.
(Need to set "Mode No.8".)
Setting state : It lights up at totalizer.
RST
DISP
▲

- 13 -
8. The setting menu
≪The test mode function≫
Fig.8-1
Power supply OFF
Test mode
being pressed.
Turn power supply on with

- 14 -
≪The setting menu≫
Fig.8-2
Turn power supply on.
Power supply OFF
Measurement state
Initialization
Turn power supply on
with
1. 10 0 0
3.
2.
0
3
0
2.
1 1
1
7.
6.
5.
4.
0 0
0 0
0
0 0.
b.
C.
d. 10
0
A.
9.
8.
00
0 0
3 0
0 5
0 0
00
0 0
10
99 9 9 9
9999 9
+
<The preset value setting>
<The mode setting>
0
00 2.
0000 0
<The offset value setting>
+
0
0
F
o
_
L F
n
o
_
L
<The mode protect function>
_
An1
n
A A
Turn power supply on with
<Adjusting the analog output>
00 0 0
※After adjusting the setting,use
A2
0 0
n
0
_
0
0
being pressed.
being pressed.
to register it.
<Operation reset>
0
※When is pressed for more than 2 seconds,
a display returns to measurement indication.

- 15 -
9. Initial setting values and initialization
If the specifications desired by the user are requested prior to shipment,
the meter will be set these settings.
Otherwise, the regular factory settings are shown below.
Value setting of each mode
Table.9-1
Mode No.
Initial setting
Notes
Mode contents
A B
C
D
E
B
C
D
E
1. 1
0
0
0
Setting of scaling data
2. 3
0
1
1
Setting of Exp. value, least significant digit,
the unit time and the decimal point
3. 0
2.
0
-
Setting of the sampling time
4. 0
0
0
-
Setting of the moving average range
5. 0
2.
0
-
Setting the auto-zero time
6. 0
0
-
-
Setting of EXT input and measurement indication
7. 1
0
0
0
Setting of scaling data
8. 3
0
1
0
Setting of Exp. value,reset time,
overflow indication and the decimal point
9. 0
0
0
0
Setting of OUT1 preset output
A. 0
0
0
0
Setting of OUT2 preset output
b. 0
0.
0
5
Setting of the synchronization pulse-output
C. 0
0
-
-
Analog output:Setting of measurement choice
(option) and the output digit
d. 1
0
0
0
Analog output:Setting of maximum output
(option) indication
Presetting output set value Table.9-2
Preset value
Initial setting
Notes
A
B C
D
E A
B C
D
E
OUT1 9
9
9
9
9
OUT2 9
9
9
9
9
* With the P1 option, there is no OUT2 preset value setting.
Offset value Table.9-3
Offset
Initial setting
Notes
A
B C
D
E A
B C
D
E
Indication 0
0
0
0
0
〔Initialization〕
Throw power supply in with Reset Key pressed to initialize the settings.
After the initialization, the set values will be as shown in "Table.9-1",
"Table.9-2" and "Table.9-3".
Counter data and the mode protect function are also cleared.
! < Caution >
* Since an initialization changes all existing setting values to the initial
setting values, be sure to record all the setting values before an
initialization.
* In case the computer froze when unusual functioning occurred with
the normal operation, initialize according to the above procedure and
set the desired value again.

- 16 -
10. Mode setting
≪1. Operating method (the mode setting)≫
When doing mode setting, please operate as follows.
Operation key
Indication Procedure
+
A B C D E
1.1 0 0 0
While pushing down Mode Key, press
Shift Key for 2 sec, or more.
"1" appears in display A, the value setting
for mode No.1 is shown.
A B C D E
1.1→0→0→0
└--←--┘
A figure of flash indication is shifted.
Each time the key is pressed, a flash figure
is shifted to the right.
A B C D E
1.1 1 0 0
↑
0→9
Changes the value of the flashing digit.
Each time Display Key is pressed, the number
goes up by one. (0→1→……→9→0→……)
* In situation, doesn't indicate by a setting
figure, up to nine.
A B C D E
2.3 0 1 1
↑
1 - d
Changes the value of the mode No.
Each time Mode Key is pressed, the number goes
up by one. (1→2→……→d→1→2……)
Set the desired value with the Shift Key and
Display Key.
A B C D E
measured value
After adjusting the setting, use Reset Key
to register it.
The display returns to the readings
following registration.
! < Caution >
* Please make the mode protect function "L-off" at the mode setting.
If it's a condition of "L-on", it can't be changed.
About the contents of the mode protect function, please refer to,
"11. The mode protect function".
MODE
MODE
DISP
RST
▲
▲
Table of contents
Other UINICS Measuring Instrument manuals