Unitron Z12 User manual


Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 2
CONTENTS
SAFETY NOTES...................................................................................................................3
CARE AND MAINTENANCE ................................................................................................3
INTRODUCTION ..................................................................................................................4
UNPACKING AND COMPONENTS .....................................................................................4
COMPONENT DIAGRAMS ..................................................................................................5
ASSEMBLY DIAGRAM.........................................................................................................6
DETAILED ASSEMBLY.....................................................................................................7-9
OPERATION
ILLUMINATION........................................................................................................................10
REFLECTOR............................................................................................................................10
FOCUSING TENSION..............................................................................................................10
DIOPTER & FOCUSING ..........................................................................................................11
INTERPUPILLARY DISTANCE................................................................................................11
VIEWING ANGLE.....................................................................................................................11
CLICK/CLICK STOP.................................................................................................................12
EYEGUARDS...........................................................................................................................12
APERTURE DIAPHRAGM .......................................................................................................13
LIGHT PATH ...........................................................................................................................13
MOUNTING A MICROSCOPY CAMERA (OPTIONAL) ...........................................................13
POLARIZER/ANALYZER .........................................................................................................14
TROUBLESHOOTING...................................................................................................15-16
MAINTENANCE..................................................................................................................17
SERVICE............................................................................................................................17
WARRANTY .......................................................................................................................17

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 3
SAFETY NOTES
1. Open the shipping carton carefully to prevent any accessory, i.e. objectives or eyepieces, from dropping
and being damaged.
2. Do not discard the molded shipping carton; the container should be retained should the microscope ever
require reshipment.
3. Keep the instrument out of direct sunlight, high temperature or humidity, and dusty environments. Ensure
the microscope is located on a smooth, level and firm surface.
4. If any specimen solutions or other liquids splash onto the stage, objective or any other component,
disconnect the power cord immediately and wipe up the spillage. Otherwise, the instrument may be
damaged.
5. All electrical connectors (power cord) should be inserted into an electrical surge suppressor to prevent
damage due to voltage fluctuations.
6. LAMP REPLACEMENT -- CAUTION: the glass housing of the lamp may be extremely hot.
DO NOT attempt to change the lamp before it is completely cooled or without wearing
adequate skin protection.
7. FUSE REPLACEMENT -- For safety when replacing the fuse (ONLY replace with the same
size, type and rating of original fuse), be sure the main switch is in the off position, disconnect
the power cord from outlet, and replace the fuse. Reconnect the power cord and turn unit on.
8. Confirm that the input voltage indicated on your microscope corresponds to your line voltage. The use of a
different input voltage other than indicated will cause severe damage to the microscope.
CARE AND MAINTENANCE
1. Do not attempt to disassemble any component including eyepieces, objectives or focusing assembly.
2. Keep the instrument clean; remove dirt and debris regularly. Accumulated dirt on metal surfaces should be
cleaned with a damp cloth. More persistent dirt should be removed using a mild soap solution. Do not use
organic solvents for cleansing.
3. The outer surface of the optics should be inspected and cleaned periodically using an air stream from an
air bulb. If dirt remains on the optical surface, use a soft cloth or cotton swab dampened with a lens
cleaning solution (available at camera stores). All optical lenses should be swabbed using a circular
motion. A small amount of absorbent cotton wound on the end of a tapered stick such as cotton swabs or
Q-tips, makes a useful tool for cleaning recessed optical surfaces. Avoid using an excessive amount of
solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may
pick up grease making cleaning more difficult.
4. Store the instrument in a cool, dry environment. Cover the microscope with the dust cover when not in
use.
5. UNITRON®microscopes are precision instruments which require periodic preventative maintenance to
maintain proper performance and to compensate for normal wear. An annual schedule of preventative
maintenance by qualified personnel is highly recommended. Your authorized UNITRON® distributor can
arrange for this service.

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 4
INTRODUCTION
Congratulations on the purchase of your new UNITRON®microscope. UNITRON®microscopes are
engineered and manufactured to the highest quality standards. Your microscope will last a lifetime if used
and maintained properly. UNITRON®microscopes are carefully assembled, inspected and tested by our
staff of trained technicians in our New York facility. Careful quality control procedures ensure each
microscope is of the highest quality prior to shipment.
UNPACKING AND COMPONENTS
Your microscope arrived packed in a molded shipping carton. Do not discard the carton: the carton
should be retained for reshipment of your microscope if needed. Avoid placing the microscope in dusty
surroundings or in high temperature or humid areas as mold and mildew will form. Carefully remove the
microscope from the EPE foam container by its arm and base and place the microscope on a flat,
vibration-free surface. Check the components against the packing list included with your microscope.

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 5
COMPONENTS DIAGRAM
Trinocular
Viewing Head
Aperture
Diaphragm
Control
Coarse/Fine Focus
Adjustment Knobs
Eyepiece
Stage Insert
Plate
Camera/Photo
Port
Focus
Drive
Tension
Adjustment Collar
Zoom
Knob
Lock Knob
Stage Clip
Light Path
Selection Slider
Objective
Base
Reflector
Adjustment
Knob
Microscope
Body

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 6
ASSEMBLY DIAGRAM
The diagram below shows how to assemble the various components. The numbers indicate the order of
assembly. Use the 1.5mm and 3mm hex wrenches that are supplied with your microscope when
required. Be sure to keep this wrench for changing out components or making adjustments.
When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching
any parts or touching glass surfaces.
①Focus Drive
②Microscope Body
③Objective
④Trinocular Viewing Head
⑤Eyepieces
⑥Stage Insert Plate
⑦ Stage Clips
⑧Illuminator Light Guide
⑨Illuminator Power Cord
④
①
②
③
⑤
⑥⑦⑧
⑨

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 7
DETAILED ASSEMBLY
Mount the Focus Drive (Fig. 1-2)
Loosen the lock knob ①of the focus drive ②. Position the
focus drive as shown and gently slide it all the way down
onto the column pole ③until the bottom of the focus drive is
flush with the column base ④, Fig. 2.
Tighten the lock knob.
WARNING: the focus drive MUST be mounted correctly as
shown in Fig. 2. Mounting it improperly will result in the
microscope tipping over which may cause damage.
Mount the Microscope Body (Fig. 3)
Remove the access cap ①on the side of the focus group
and loosen the screw inside with a 3mm hex wrench.
Carefully slide the microscope body ②onto the dovetail
interface ③of the focus group as shown.
Tighten the lock screw on the focus group and replace the
access cap.
Install the Objective (Fig. 4)
Carefully position and screw the objective ①into the
bottom ②of the microscope body by rotating it clockwise as
shown until it stops.
DO NOT force it on or overtighten it.
①
③
Fig. 1
②
Fig. 2
Correct
X
Wrong!
④
Fig. 3
Fig. 4
①
③
②
①
②

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 8
DETAILED ASSEMBLY - continued
Mount the Viewing Head (Fig. 5-6)
Loosen the lock screw ①on the front of the microscope
body with the 3mm mm hex wrench.
With the eyepiece tubes ②facing forward (as shown in
Fig. 6), carefully slide the dovetail ③(Fig. 5) on the bottom
of the viewing head ④(Fig. 5) onto the dovetail mount on
the top of the microscope body by aligning the orientation
groove ⑤ with the orientation pin ⑥of the microscope
body. Make sure it is seated properly (it should be level and
not wobble), then tighten the lock screw.
Install the Eyepieces (Fig. 6)
Remove the eyetube caps ⑦and insert the eyepieces ⑧
into the eyepiece tubes ②.
Lock in the eyepieces by tightening the thumbs screws ⑨
on each eyepiece tube.
Mount/Unmount the Eyepiece Reticle (Fig. 7)
To mount the reticle: unscrew the mounting ring ①from
the eyepiece by turning it counterclockwise.
Clean the reticle ②and insert it into the mounting ring with
the inscription side as shown.
Screw the mounting ring back onto the eyepiece.
To remove the reticle: unscrew the mounting ring from the
eyepiece, then carefully take out the reticle and wrap it in
clean, soft, lint-free paper for storage.
Screw the mounting ring back onto the eyepiece.
Install the Stage Insert Plate (Fig. 8)
Gently push the stage insert plate ①with the edge tilted
toward the spring on the inside of the hole in the base and
set it into place.
To remove the stage insert plate, press down with one finger
near the spring – it will pop up on the opposite side and you
can then lift it out.
①
⑤
Fig. 5
③
Fig. 6
Fig. 7
Fig. 8
①
②
④
⑥
⑦
⑧
⑨
Inscription Side
①
②

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 9
DETAILED ASSEMBLY - continued
Install the Stage Clips (Fig. 9)
Insert the stage clip post of the stage clip ①into the hole ②
on the top of the base as shown; repeat for other stage clip.
Connect the LED Illuminator & Power Supply
(Fig. 10)
Make sure the power switch on the illuminator ① is set to
“OFF”.
Align the end of the flat part of the illuminator light guide ②
with the lock screw ③ on the socket ④on the front of the
illuminator and slide it in and tighten the lock screw.
Insert the other end of the illuminator light guide into the
socket ⑤ on the back of the microscope and tighten the lock
screw on the top of the socket.
Insert one end ⑥of the power cord on the power supply ⑦
into the socket ⑧ on the back of the illuminator.
Plug the other end ⑨into the corresponding socket on the
other end of the power supply.
Plug the other end of the power supply into a grounded (3-
prong) outlet.
NOTE:Always use the power cord that is provided with your
microscope; using a different power cord may damage your
microscope. Should you need a replacement, contact your
authorized UNITRON dealer or call UNITRON at 1-631-543-
2000 for a dealer nearest you.
①
②
Fig. 9
③
Fig. 10
①
②
④
⑤
⑥
⑦
⑧
⑨

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 10
OPERATION
Adjusting the Illumination (Fig. 11)
Turn the power switch ①to “ON”.
NOTE: For longer lamp life always turn the light
intensity control knob ②to the lowest illumination
intensity setting possible before turning the power on or
off.
The light level may need adjustment depending upon
the specimen density and objective magnification.
Adjust the light intensity for comfortable viewing by
turning the light intensity control knob clockwise to
increase brightness. Turn counterclockwise to
decrease brightness.
Adjusting the Reflector (Fig. 12)
The reflector has two sides: one is a plane mirror, the
other is sandblasted aluminum. Both sides reflect light
but the plane mirror is provides a stronger reflection.
You can change the reflector type or the reflection
intensity by rotating the reflector adjustment knob ①.
The reflector adjustment knob can also be moved from
front to back to achieve different lighting effects of the
reflector.
Adjusting the Focusing Tension (Fig. 13)
If the feel is very heavy when focusing with the focusing
knobs ①, or the specimen leaves the focus plane after
focusing, adjust the tension with the tension adjustment
ring ②.
Turn the tension adjustment ring clockwise to tighten or
counterclockwise to loosen according to user
preference.
②
①
Fig. 11
Fig. 12
Fig. 13
①
①
②
③

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 11
OPERATION (continued)
Adjusting the Diopter & Focus (Fig. 14)
Set the dioptor rings of both eyepieces to “0” position.
(Do this when users change, because different users
will have different diopter settings.)
Place an easy-to-observe specimen on the stage plate,
i.e., a coin.
Rotate the zoom knob to the highest magnification, then
turn the focusing knob to focus the specimen.
Rotate the zoom knob to the lowest magnification,
looking only into the left eyepiece, adjust the diopter
ring on left eyepiece to focus the specimen.
Repeat procedure for the right eyepiece.
Adjusting Interpupillary Distance (Fig. 15)
To adjust the interpupillary distance, hold the left and
right eyetubes while observing a specimen. Rotate the
eyetubes around the central axis until the fields of view
of both eyetubes coincide completely. A complete circle
should be seen in the viewing field when viewing the
specimen slide. An improper adjustment will cause
operator fatigue and will disrupt the objective
parfocality.
Adjusting the Viewing Angle (Fig. 16)
To adjust the viewing angle, hold the eyepiece tubes
with both hands and move them up or down to the
position that feels most comfortable.
The angle can be adjusted from 5° to 45°.
Left Diopter
Adjustment Ring
Right Diopter
Adjustment Ring
Fig. 14
Fig. 16
Fig. 15

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 12
OPERATION (continued)
Changing the Click/Click Stop Feature
(Fig. 17)
The microscope is equipped with a Click/Click Stop
feature that enables the user to set the magnification
value indicated on the zoom knob to stop at each fixed
magnification and will make a “click” sound, or to finely
adjust the zoom magnification.
To turn on the Click feature to click at each
magnification value, using a 3mm hex wrench, turn the
Click/Click Stop screw ①clockwise as shown by the
arrow in Fig. 17.
To turn on the Click Stop feature to finely adjust the
zoom magnification, using a 3mm hex wrench, turn the
Click/Click Stop screw ①counterclockwise.
NOTE: Be careful not to overturn the screw for either
position as it may result in damage to the housing and
internal mechanisms of the microscope.
Adjusting the Eyepiece Eyeguards
(Fig. 18)
The eyepieces come standard with roll-
down eyeguards.
The eyeguards should be rolled down ①for users
wearing eyeglasses to prevent them from being too far
from observing through the eyepieces. Or, not rolled
down ②for those who don’t wear glasses to prevent
extraneous light from entering the eyepiece which may
interfere with specimen observation.
Click
Click Stop
Fig. 17
Fig. 18
①

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 13
OPERATION (continued)
Adjusting the Aperture Diaphragm
(Fig. 19)
Adjust the aperture diaphragm by the aperture
diaphragm control ①.
To enlarge the aperture diaphragm, slide the control to
the left until the desired contrast of the observed image
is achieved.
To decrease the aperture diaphragm, slide the control to
the right until the desired contrast of the observed image
is achieved.
Selecting the Light Path (Fig. 20)
The Z12 is outfitted with a binocular viewing head with
one photo port for HDMI/digital imaging. You must
select the appropriate light path for observing
specimens.
The light path is set to 100% to the binocular eye-pieces
as the default setting at our facilities where the light path
selection slider ①is pushed all the way in.
Pull the light path selection slider all the way out to send
100% to the top photo port for HDMI/digital imaging and
documentation.
Mounting a Microscopy Camera (Optional)
(Fig. 21)
Loosen the lock screw ①on the camera/photo port until it
is flush with the inside of the portl remove the dust cap ②.
Remove the dust caps from both ends of the c-mount ③
and from your camera lens and attach the top opening ④
on the c-mount to the threaded mount on your camera.
Attach the bottom dovetail mount of the c-mount (with the
mounted camera) onto the camera/photo port and tighten
the lock screw ⑤.
Focus the image in the binocular view then pull the light
path selection lever out ⑥. If the image is not clear,
loosen the lock screw ⑤and adjust the focusing screw ⑦
until the image is clear on the monitor, then tighten the lock
screw ⑤.
①
Fig. 19
Fig 20
①
②
①
②
③
⑥
④
⑤
⑦
Fig 21

Z12 ZOOM STEREO ICAL MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 14
OPERATION (continued)
Using the Polarizer/Analyzer (Fig. 22)
The simple polarizer includes the polarizer and
analyzer.
Remove the stage insert plate ①(see Fig. 8, p. 3).
Insert the polarizer ②with the flat edge closest to
the column (or aligned with the spring on the inside of
the hole in the base), and screw it in place using the
3mm hex wrench. Replace the stage insert plate.
Slide the analyzer ③ onto the bottom outside
diameter of the objective ④ and tighten the analyzer
lock screw ⑤.
With the 360° rotating analyzer, you can change the
orthogonal state of the polarized light by rotating the
analyzer adjustment ring ⑥.
①
Fig. 22
②
⑥
③
⑤
④

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 15
TROUBLESHOOTING
Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a
problem occurs, please review the following list and take remedial action as needed. If you cannot solve the
problem after checking the entire list, please contact your local dealer for assistance.
OPTICAL
PROBLEM CAUSE SOLUTION
The illumination is on, but the
field of view is dark.
The bulb is burnt out. Replace it with a new one
The brightness is set too low
Set it to the appropriate position
The field diaphragm is not open enough or
closed
Open the field diaphragm
Too many filters are in the optical path
Reduce them to the minimum required
number
The edge of the field of view is
obscured or not evenly
illuminated.
The nosepiece is not in the located position
Turn the nosepiece into the position where
you can hear it engaged
Dirt or dust has accumulated on the lens
(objective or eyepieces)
Clean the lens
Light path selection slider is not in the right
position
Pull it into the right position
The color filter, polarizer or analyzer is not
inserted fully
Push it in all the way
Dirt or dust is visible in the field
of view
Dirt/dust on the specimen
Replace with a clean specimen
Dirt/dust on the eyepiece Clean the eyepieces
The image is not clear
Cover glass on the specimen slide
Clean the specimen
The specimen is not vertical to the objective
Adjust it
The nosepiece is not in the correct position
Turn the nosepiece into the correct position
The aperture diaphragm is not open correctly
Adjust it
The light path selection slider is not in the
correct position
Push it into the correct position
Visibility is poor
Image is not sharp
Contrast is poor
Details are indistinct
The objective is not correctly engaged in the
light path Turn the nosepiece into the engaged position
the aperture diaphragm is opened or stopped
down too far in brightfield observation
adjust the aperture diaphragm properly
The lens (condenser, objective, eyepieces)
are dirty Clean it thoroughly
One side of the image is dark,
blurred or it moves while
focusing
The objective is not in the center of the light
path
Insure the nosepiece is in the “clicked”
position
The specimen is not correctly positioned on
the stage Reposition the specimen on the stage
The eyes tire easily; the right
field of view doesn’t
superimpose with the left
Interpupillary distance is incorrect Adjust the interpupillary distance
Diopter adjustment is incorrect Adjust the diopter
The eyepiece for the right eye is different from
the left one Use the same eyepieces

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 16
TROUBLESHOOTING (continued)
MECHANICAL PART
PROBLEM CAUSE SOLUTION
The coarse adjustment knob is
too difficult to rotate The tension adjustment ring is tightened too
much Loosen it
The image goes out of focus
during observation or the stage
drops by itself The tension adjustment collar is too loose Tighten it
ELECTRICAL SYSTEM
PROBLEM CAUSE SOLUTION
The lamp doesn’t light
No power to the lamp Check the power cord is connected correctly
The lamp is not installed correctly Install it correctly
The lamp burns out Replace the lamp
The light intensity is not
enough
The light intensity control knob is not set
properly Adjust the light intensity control knob
The wrong lamp is used Replace the lamp with the correct one
The lamp keeps burning out The wrong lamp is used Replace the lamp with the correct one
The lamp flickers or the
brightness is not stable
The lamp will burn out soon Replace the lamp
The power supply doesn’t connect well Connect the power supply correctly

Z12 ZOOM STEREO MICROSCOPE
UNITRON 73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 17
MAINTENANCE
Please remember to never leave the microscope with any of the objectives or eyepieces removed and
always protect the microscope with the dust cover when not in use.
SERVICE
UNITRON®microscopes are precision instruments which require periodic servicing to keep them
performing properly and to compensate for normal wear. A regular schedule of preventative maintenance
by qualified personnel is highly recommended. Your authorized UNITRON®distributor can arrange for
this service. Should unexpected problems be experienced with your instrument, proceed as follows:
1. Contact the UNITRON®distributor from whom you purchased the microscope. Some
problems can be resolved simply over the telephone.
2. If it is determined that the microscope should be returned to your UNITRON®distributor or to
UNITRON®for warranty repair, pack the instrument in its original Styrofoam shipping carton. If you no
longer have this carton, pack the microscope in a crush-resistant carton with a minimum of three inches
of a shock absorbing material surrounding it to prevent in-transit damage. The microscope should be
wrapped in a plastic bag to prevent Styrofoam dust from damaging the microscope. Always ship the
microscope in an upright position; NEVER SHIP A MICROSCOPE ON ITS SIDE. The microscope or
component should be shipped prepaid and insured.
LIMITED MICROSCOPE WARRANTY
This microscope is warranted to be free from defects in material and workmanship for a period of five (5) years for
mechanical and optical components and one (1) year for LED bulb and electrical components from the date of
invoice to the original (end user) purchaser. This warranty does not cover damage caused in-transit, misuse,
neglect, abuse or damage resulting from improper servicing or modification by other than UNITRON®approved
service personnel. This warranty does not cover any routine maintenance work or any other work, which is
reasonably expected to be performed by the purchaser. Normal wear is excluded from this warranty. No
responsibility is assumed for unsatisfactory operating performance due to environmental conditions such as
humidity, dust, corrosive chemicals, deposition of oil or other foreign matter, spillage or other conditions beyond the
control of Unitron Ltd. This warranty expressly excludes any liability by UNITRON Ltd. for consequential loss or
damage on any grounds, such as (but not limited to) the non-availability to the End User of the product(s) under
warranty or the need to repair work processes. Should any defect in material, workmanship or electronic
component occur under this warranty contact your UNITRON®distributor or UNITRON®at (631) 543-2000. This
warranty is limited to the continental United States of America. All items returned for warranty repair must be sent
freight prepaid and insured to Unitron Ltd., 73 Mall Drive, Commack, NY 11725 – USA. All warranty repairs will be
returned freight prepaid to any destination within the continental United States of America. For all foreign warranty
repairs, return freight charges are the responsibility of the individual/company who returned the merchandise for
repair.
UNITRON®is a registered trademark of UNITRON, Ltd., Commack, NY 11725
Table of contents
Other Unitron Microscope manuals

Unitron
Unitron Z645 Series User manual

Unitron
Unitron TCS Pro 500 User manual

Unitron
Unitron System 274 User manual

Unitron
Unitron EXAMET-4 User manual

Unitron
Unitron System 374 User manual

Unitron
Unitron LX Microscopes MIDAS User manual

Unitron
Unitron MEC2 User manual

Unitron
Unitron Z850 Series User manual

Unitron
Unitron Versamet 4 User manual

Unitron
Unitron 12100 Series User manual