UV MEDICO UV222 User manual

(Filtered Far UV-C)
Technical & Installaon Manual
Downlight

UV222 Technical & Installation Manual2
CAUTION UV LIGHT
THE PRODUCT AS DESCRIBED HEREIN CONTAINS COMPONENTS THAT EMIT
FILTERED FAR UV LIGHT.
DO NOT USE THE PRODUCT BEFORE CAREFULLY READING THIS MANUAL.
ALWAYS KEEP THIS MANUAL WITH THE PRODUCT AND CONSULT IT REGULARLY.
ATTENTION
Lumière UV émise
par cet appareil.
Évitez de regarder
directement la
lampe en cours
d’ulisaon.
CAUTION
UV Light emied
from this device.
Avoid looking directly
at the light.
UV
IEC/TR 62471-2:2009

3
1. USING THIS MANUAL – GENERAL
This manual (hereinaer referred to as the “Manual”) is intended to be used as work of
reference for professional, well trained and authorised users to be able to safely install and
use the product menoned on the cover of this document (hereinaer referred to as the
“Product”). The observaon of the Manual does not dispense the user from the obser-
vance of the local safety instrucons and regulaons other than provided by the Manual.
Such safety instrucons and regulaons must be observed at all me.
For informaon about specic adjustments, maintenance or repair jobs which are not dealt
with in this manual, please contact your Product supplier or UV Medico A/S.
If you suspect that the Product is damaged or detect a potenal or actual danger when
using the Product, immediately disconnue use of the Product and contact your supplier of
the Product.
Contents
1. USING THIS MANUAL – GENERAL 3
1.1 BASIC SAFETY INSTRUCTIONS 4
2. DISCLAIMERS 5
3. SAFETY INSTRUCTIONS AND WARNINGS 5
3.1. User manual 5
3.2. Intended use and installaon 5
3.3. Safety features of the Product 5
3.4. Modicaon 6
4. GENERAL CHARACTERISTICS AND TECHNICAL SPECIFICATIONS 6
5. CONSTRUCTION 7
6. INSTALLATION OF THE PRODUCT 8
6.1. Choose the appropriate locaon 8
6.2. Unboxing and inspecon 8
6.3. Product installaon 9
6.4. Understanding the unit’s operaon 10
6.5. Seng up the Product 10
6.6. Understanding the funcon of the signal LED 19
6.7. Replacements 19
7. PRODUCT FAILURE 20
8. CLEANING AND MAINTENANCE 20
9. LIGHT SOURCE REPLACEMENT 21
10. REPAIRS 21
Warranty 21
NOTES 22

UV222 Technical & Installation Manual4
1.1. BASIC SAFETY INSTRUCTIONS
To minimise the risk of injury, the following basic safety precauons should be observed
with each use of the Product:
Do not remove the scker on the front of the unit unl the Product is commissioned by
a qualied installer.
Do not put any object in close front of the surface of the Product.
Do not put plasc or glass in front of the Product. It does not transmit UV.
Make sure the Product has an unobstructed path towards the area to be disinfected.
Do not use ammable substances near the lamp module.
This is anelectronic device. Do not use ammable or combusble gases or liquids, such
as oxygen, hydrogen, or chemicals, near the Product. Failure to obey may result in re or
explosion.
Do not turn on any Product with a broken front glass.
If the front glass is broken, disconnect the Product immediately and and contact an
authorised installer.
Do not disassemble or modify the lamp module.
Disassembly or modicaon may result in re, electric shock, or injury.
Turn o the electrical power during lamp module maintenance.
Do not remove any safety labelling or parts.
Do not touch the lamp module during operaon.
Otherwise, it may result in re, electric shock, or injury.
Ground the device or metal parts of the housing that are close to the lamp module.
Otherwise, it may result in re, electric shock, or injury.
Do not use the unit for purposes other than disinfecon.
The unit is designed for disinfecon. Using it for any other purpose may result in unexpect-
ed accidents.

5
2. DISCLAIMERS
UV Medico A/S with registered oce at Søren Frichs Vej 40E, 8230 Åbyhøj - Denmark, or
any of its aliates, does not accept any liability for material damage, including damage to
the Product or personal injury caused by non-observance by the user of the safety instruc-
ons in the Manual or the applicable laws. As such UV Medico A/S shall among others not
be liable for any damage or injury if the Product is used contrary to or for purposes other
than its intended use, is installed, used, maintained or repaired negligently or contrary to
the instrucons menoned in the Manual or on the Product; also in the case the Product is
modied, and when safety devices and features are disregarded or overridden.
3. SAFETY INSTRUCTIONS AND WARNINGS
3.1. User manual
Every person using the Product must read and understand the contents of the Manual
and familiarise themselves with it before any use of the Product, and must strictly observe
the instrucons set out herein. If you are an employee or are acng on behalf of another
person, your employer/that person must instruct you in accordance with the Manual and
ensure that all instrucons set out herein are complied with.
Do not use the Product if that is not the case.
3.2. Intended use and installaon
The Product is intended exclusively for air and surface disinfecon. Using the Product for
other purposes is contrary to its intended use and is done at own risk.
To ensure safe use of the Product, its installaon, including all its safety features and
devices, and its maintenance and repair must be carried out correctly by a qualied
installer. All of the Product safety features and devices must be mounted and may not be
removed but for the purpose of maintenance and repair.
The Product is not intended to be used by minors. Any use of the Product by minors,
regardless of whether or not under the supervision of an adult, is at own risk.
3.3. Safety features of the Product
Each Product is equipped with the following safety features:
• Moon sensor: To detect moon.
• Lidar sensor: To measure distance to the occupancy of the room.
• Mulcolour LED: To signal the operaon and state of the unit.

UV222 Technical & Installation Manual6
It is strictly forbidden to override these safety devices or to use the Product if the safety
features are not fully present or disfuncon.
The safety features should be regularly checked for their proper funconing, and if
required, be immediately repaired or replaced.
The pictograms, warnings and instrucons aached to the Product are part of its safety
features. They must not be covered or removed and must be present and legible during the
enre use of the Product.
Immediately replace or repair missing, damaged or illegible pictograms, warnings and
instrucons.
3.4. Modicaon
No modicaon of the product is allowed. All warranty and liability are void if any modica-
ons are done to the Product.
4. GENERAL CHARACTERISTICS AND TECHNICAL
SPECIFICATIONS
The Product is designed to emit Far UV-C radiaon with a wavelength of 222 nm, in order
to eliminate viruses, bacteria, fungi, and other pathogens, from air and surfaces.
The Far UV-C lamps are designed for +10 000 hours of ecient operaon. Far UV-C lamps
used longer than this, may no longer emit the required level of ultraviolet radiaon to en-
sure the desired germicidal ecacy on harmful microorganisms.
Technical data
Light source Kr-Cl Excimer lamp
Wavelength 222 nm
UV output power (60°) 115 mW (typical)
UV output power (100°) 70 mW (typical)
Input voltage 85-305V AC 50/60 Hz
Max power consumpon 17 W
Weight 1.6 kg (3.5 lbs)
Dimensions Ø 178 mm x 129 mm (Ø 7 in x 5 in)
Power lead 3 x 0.75 mm² (18AWG)
Operang temperature 0 to +50°C (32° to 122° F)
Ambient humidity 5-90% RH Non condensing

7
5. CONSTRUCTION
141 mm
[5.5 in]
99 mm
[3.9 in]
141 mm [5.5 in]
220 mm [8.7 in]
Ø
178 mm
[7.0 in]
3 mm
[0.1 in]
9 mm
[0.4 in]
Ø
150 mm
[5.9 in]

UV222 Technical & Installation Manual8
6. INSTALLATION OF THE PRODUCT
Perform the following steps for the installaon of the Product.
Outside every room where Far UV-C light is used, there must be informaon signs in clear
line of sight.
Installaon must be done according to the local regulaons by a qualied installer.
6.1. Choose the appropriate locaon
Choosing the appropriate locaon is important for the proper funconing and safe use of
the Product.
A DIALux simulaon can help you to choose the right posion for disinfecon.
Install the Product in a protected area where the temperature does not fall below 0° C
(32° F) and the humidity level is low. The Product funcons ideally in a temperature range
from 0°C to +50° C (32° to 122° F).
Select a mounng locaon that is close to surface/area that you want to disinfect.
6.2. Unboxing and inspecon
Read and save all noces, warnings and safety instrucons received with the Product.
Open the box and carefully take the unit out of the box. Take the unit out of the plasc
bag. Do not remove the protecon label on the front of the Product. The protecon label is
intended to avoid unexpected UV radiaon. Make a visual inspecon of the unit.
Do not power on the unit unl installed by a qualied installer!
CAUTION: Do not touch the glass of the lamp without gloves.
Doing so may result in damage to the lamp. Oil from ngerprints will
permanently etch the glass of the lamp and weaken the structure.

9
6.3. Product installaon
The unit must be inserted on a suspended ceiling or similar surface that can t the unit and
support its weight, preferably near electrical wires with grounding.
The diametre of the opening should be between 150 and 155 mm.
141 mm
[5.5 in]
99 mm
[3.9 in]
141 mm [5.5 in]
220 mm [8.7 in]
Ø178 mm
[7.0 in]
3 mm
[0.1 in]
9 mm
[0.4 in]
Ø150 mm
[5.9 in]
141 mm
[5.5 in]
99 mm
[3.9 in]
141 mm [5.5 in]
220 mm [8.7 in]
Ø178 mm
[7.0 in]
3 mm
[0.1 in]
9 mm
[0.4 in]
Ø
150 mm
[5.9 in]
Once the Product is connected to a power source, push the side springs and insert in the
opening made on the ceiling panel.

UV222 Technical & Installation Manual10
6.4. Understanding the unit’s operaon
The maximum permissible UV-C exposure shall not exceed the ACGIH TLV and NIOSH REL
of 22,0 mJ/cm² for an 8 h day, 40 h work week exposure to UV radiaon at 222 nm.
(UV Medico A/S expects posive changes into the ACGIH that will result in a higher expo-
sure than allowed at this moment. If this changes, UV Medico A/S will update the soware
of the unit to align with the new threshold of exposure).
To achieve the correct radiaon within these limits, following informaon will be pro-
grammed into the Product before the start-up of this installaon.
6.5. Seng up the Product
The installer will need a Microso Windows computer with the UVMEDICO applicaon
installed. Contact your distributor to get access to the latest soware and rmware for the
Product.
To launch the app, go to the installed folder and open the le:
UV Medico Conguraon App.exe
The app includes two dierent set up modes: simple and advanced.
For the simple mode, follow the instrucons that will appear on the screen.
For the advanced mode, see below:
There is a USB-C connecon
A
on the
side of the unit to connect to a computer
with a USB cable.
Connect your computer to the unit and
it will show up automacally on the
applicaon.
When the app is launched, following
steps needs to be taken:
A

11
1) If the Product was already installed, you can read out the seng of the unit by press-
ing “Get” all sengs will be shown in the applicaon.
B
2) Check if the rmware version is the latest. If not, click on “Update FW”.
C
B
C

UV222 Technical & Installation Manual12
Select the correct version
A
and click on “Open” to upload it. Then conrm by clicking
“Yes.”
B
When nished installing, the rmware should be up to date.
A
B

13
3) Set the correct hour by pressing “Set Time and Zone”.
C
The date and hour are now the same as your computer.
4) Dene the ming of the use (duty cycle) of the room where the Product is installed:
“On me”: Time from which the room will be occupied by people.
D
“O me”: Time from which the room will not be occupied.
E
5) Add the “Maximum Occupant Time”: how long the room is occupied during the duty
cycle.
F
D
E
C
F

UV222 Technical & Installation Manual14
6) Add the pre and post cleaning cycle, if desired.
A
The unit will then light up before
and aer the duty cycle.
7) Change the intensity of the LED signal to the desired output.
B
8) Select mode.
C
• Normal: The Product will follow the sengs as provided.
• Only on when movement: The Product only will turn on when the Pir sensor detects
movement.
• Always on, only ‘Safety o distance’ is enabled: The Product will only turn o when an
object gets closer than the selected safety o distance.
B
A
C

15
9) Add 3 distances:
a. “Distance to eye height”: Closest distance of the unit to the eye of a human (even
with people walking around).
D
b. “Safety distance”: Closest distance people are allowed to get to the lamp before
it turns itself o. The unit will turn itself back on aer 10 sec. when/if the safety
distance is regained.
E
c. “Primary surface distance”: Distance from the unit to the main surface to
disinfect.
F
All measurements are in cenmetres.
E
F
D

UV222 Technical & Installation Manual16
10) Save setup into template by pressing “Save Setup”.
A
Name the template. This can be recalled whenever you have similar sengs.
A
B
11) Press “Set” to upload the sengs into the Product.
B
Once this informaon is saved and programmed into the unit, it is ready to test/start up.

17
The unit will calculate the radiaon automacally to stay within the limits of the ACGIH like
the example below:
If during the duty cycle a person comes closer to the unit than the programmed distance
to the eye, Product will automacally stop, and it will resume the acvity 10 sec. aer this
security distance is cleared.
If during the duty cycle, something/someone comes closer to the unit than the pro-
grammed safety o distance, the Product will automacally stop, and it will resume the
acvity 10 sec. aer the security distance is cleared.
If movement in the room is detected during the pre or post disinfecon cycle, the unit will
start its duty cycle again, and will resume its pre or post disinfecon cycle aer no more
movement is detected.
If the informaon programmed into the Product when installed is dierent than the
use/occupancy of the room, UV Medico A/S cannot be held liable for any injuries or misuse
of these units.
If the room changes of use, occupancy or any change that has an inuence on the risks of
exposure, the installer needs to be contacted so the Product is reprogrammed.
222 nm Irradiance
Time
Occupied meslot Post
disinfe.
Pre
disinfe.
TLV

UV222 Technical & Installation Manual18
11) Tesng the Product.
Right click on the photo of the Far UV-C unit
A
and a pop up of the test mode will show
up.
Here you can quickly light up the Product and test the signal LED.
When you acvate the 222 LightSource
B
, it will turn on the Product with no safety. This
is only to test the source, (the box “Enable“ will be automacally cked).
When you go to LED Mode you have 3 possibilies to choose:
• “Normal Mode”: Work as te Product is programmed.
• “LED O”: Turns the LED o.
• “LED RGBW O”: LED test cycle.
When you unck the “Enable” box it goes back to normal conguraon/use.
A
B

19
6.6. Understanding the funcon of the signal LED
LED MEANING
No light No power.
Red connuous Error: no feedback from the Product light source.
Cyan connuous Pre disinfecon / aer disinfecon light source on.
Magenta connuous Light source is on.
Blue connuous In-cycle UV o.
Green connuous Out of cycle.
Flashing red Object too close.
Flashing magenta Pre disinfecon / aer disinfecon per seen movement.
Red-blue-green (loop) Recommended change of Far UV-C light source.
6.7. Replacements
The Product has an internal baery and fuse that can be replaced if broken or empty.
Take the unit o the installaon panel locate the USB-C connector
C
. The panel to re-
move is the one to the le. Unscrew the 4x screws
D
and take o the cover
E
. You will
see the fuse (T250 V 500 mA) and baery (3 V CR2032). Replace only with correct fuse
and baery to prevent damage.
Baery 3 V
CR2032
Fuse
T250 V 500 mA
C
D
D
D
E

UV222 Technical & Installation Manual20
7. PRODUCT FAILURE
FAILURE TO DO
Unit does not light up Check incoming power.
Check cable damage.
Replace fuse.
Check if the Product is correctly installed.
Unit ickers Check if the Product is correctly installed.
If not, contact your local distributor.
White ash Power cycle the unit.
Does not hold sengs Replace baery.
If the failure is unsolved or recurring, please return the unit to an ocial UV Medico dealer/
service centre for repair or replacement (see point 10. REPAIRS).
8. CLEANING AND MAINTENANCE
Clean the external surfaces with a so cloth and non-aggressive cleaning agents only.
Before performing any maintenance always unplug the Product and disconnect the power
supply.
CAUTION: Turn o and lock out ALL unit power before
performing any cleaning. Do not wash the Product with a
pressure washer or by immersion.
Other manuals for UV222
5
Table of contents
Other UV MEDICO Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Atmos
Atmos S 201 Thorax operating instructions

inhealth
inhealth Blom-Singer Instructions for use

TASKA Prosthetics
TASKA Prosthetics TASKA-XXXXXX Appropriate Use Guidelines

Arjo
Arjo Enterprise ENT-ACC64 Instructions for use

Bauerfeind
Bauerfeind CaligaLoc manual

DHG
DHG RoMedic EasyRoll Sliding Sheet Instructions for use