HMD70U/Y
Operating Manual HMD70U/Y-U162en-1.4
41996-08-21
TECHNICAL DATA
Relative humidity
Measurement range 0...100 %RH
Accuracy at 20 °C
10 4020 605030 8070 %RH
%RH
3
2
1
0
-1
-2
-3
090 100
Temperature dependence
-20 200
2.0
1.5
1.0
0.5
0
-0.5
-1.0
-1.5
-2.0
8040 60 °C10-10 30 50 70
%RH
Response time (90%)
at +20 °C in still air 15 s with membrane
filter
Humidity sensor HUMICAP180
Temperature (Y model only)
Measurement range -20...+80 °C
Accuracy
-10
0.8
0.6
0.4
0.2
0
-0.2
-0.4
-0.6
-0.8
°C
0-20 10 20 30 40 50 60 70 80 °C
Linearity better than 0.1 °C
Temperature sensor Pt 1000 IEC 751class B
General
Supply voltage range depends on the selected output
signal. When AC supply is used, an isolated source is
recommended. The current outputs require an
optional output module, part no. 18945HM.
DC AC
0...1 V 10...35 V 9...24 V
0...5 V 14...35 V 12...24 V
0...10 V 19...35 V 16...24 V
0...20 mA (RL= 0 Ω) 10...35 V 11...24 V
0...20 mA (RL= 500 Ω) 20...35 V 17...24 V
Factory setting 0...1V. Other outputs selectable by
jumper connections. An output change causes an error
which is less than 0.5 %RH without recalibration.
Operating temperature range:
electronics -5...+55 °C
sensor head -40...+80 °C
Storage temperature range -40...+80 °C
Housing:
sensor head stainless steel
electronics housing cast aluminium
Cable glands
bushing for 7...10 mm (PG9)
cable (housing IP65/
NEMA 4),
part no 18941HM
or armoured cable glands part no 10528HM
Sensor protection:
standard membrane filter
(part no. 17039HM)
option stainless steel sintered
filter
(part no. HM46670)
Connections screw terminals
0.5...1.5 mm2
Electromagnetic compatibility
The emission and immunity tests have been performed
according to standards EN50081-1 and EN50082-1.
Emissions:
Test Setup according to Performance
Radiated
interference EN55022 class B
Immunity:
Test Setup according to Performance
Electrostatic
discharge IEC 801-2:1991 criteria B
Electrical fast
transients IEC 801-4:1988 criteria B
RF-radiated
fields IEC 801-3:1984 criteria A
*GSM-field
immunity ENV50204:1995 criteria A
(* additional test)
GUARANTEE
Vaisala issues a guarantee for the material and
workmanship of this product under normal operating
conditions for one year from the date of delivery.
Exceptional operating conditions, damage due to
careless handling or misapplication will void the
guarantee.