Vega vegaflex 83 User manual

Operating Instructions
TDR sensor for continuous level and
interface measurement of liquids
VEGAFLEX 83
Four-wire Modbus
Converter version in second chamber
Polished rod probe
Document ID: 41843

2
Contents
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
Contents
1 About this document
1.1 Function ........................................................................................................................... 4
1.2 Target group ..................................................................................................................... 4
1.3 Symbols used................................................................................................................... 4
2 For your safety
2.1 Authorised personnel ....................................................................................................... 5
2.2 Appropriate use................................................................................................................ 5
2.3 Warning about incorrect use............................................................................................. 5
2.4 General safety instructions............................................................................................... 5
2.5 CE conformity................................................................................................................... 5
2.6 NAMUR recommendations .............................................................................................. 6
2.7 Environmental instructions ............................................................................................... 6
3 Product description
3.1 Conguration.................................................................................................................... 7
3.2 Principle of operation........................................................................................................ 8
3.3 Packaging, transport and storage................................................................................... 11
3.4 Accessories and replacement parts ............................................................................... 11
4 Mounting
4.1 General instructions ....................................................................................................... 14
4.2 Mounting instructions ..................................................................................................... 14
5 Connecting to power supply and bus system
5.1 Preparing the connection ............................................................................................... 21
5.2 Connection..................................................................................................................... 22
5.3 Wiring plan ..................................................................................................................... 24
5.4 Double chamber housing with DISADAPT ..................................................................... 25
5.5 Supplementary electronics............................................................................................. 26
5.6 Switch-on phase............................................................................................................. 26
6 Set up the sensor with the display and adjustment module
6.1 Adjustment volume......................................................................................................... 28
6.2 Insert display and adjustment module............................................................................ 28
6.3 Adjustment system......................................................................................................... 29
6.4 Parameter adjustment - Quick setup .............................................................................. 30
6.5 Parameter adjustment - Extended adjustment................................................................ 30
6.6 Saving the parameter adjustment data........................................................................... 48
7 Setting up sensor and Modbus interface with PACTware
7.1 Connect the PC.............................................................................................................. 50
7.2 Parameter adjustment with PACTware............................................................................ 51
7.3 Set instrument address .................................................................................................. 53
7.4 Set up with the quick setup............................................................................................. 54
7.5 Saving the parameter adjustment data........................................................................... 55
8 Diagnostics and servicing
8.1 Maintenance .................................................................................................................. 56
8.2 Diagnosis memory ......................................................................................................... 56
8.3 Status messages............................................................................................................ 57
8.4 Rectify faults................................................................................................................... 61

3
Contents
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
8.5 Exchanging the electronics module................................................................................ 64
8.6 Exchanging the rod ........................................................................................................ 65
8.7 Exchanging the seal ....................................................................................................... 66
8.8 Software update ............................................................................................................. 68
8.9 How to proceed if a repair is necessary.......................................................................... 69
9 Dismount
9.1 Dismounting steps.......................................................................................................... 70
9.2 Disposal ......................................................................................................................... 70
10 Supplement
10.1 Technical data ................................................................................................................ 71
10.2 Basics Modbus............................................................................................................... 78
10.3 Modbus register ............................................................................................................. 79
10.4 Modbus RTU commands ............................................................................................... 82
10.5 Levelmaster commands ................................................................................................. 85
10.6 CongurationoftypicalModbushosts ........................................................................... 88
10.7 Dimensions .................................................................................................................... 92
10.8 Industrial property rights................................................................................................. 96
10.9 Trademark ...................................................................................................................... 96
Safety instructions for Ex areas
TakenoteoftheExspecicsafetyinstructionsforExapplications.
These instructions are attached as documents to each instrument
with Ex approval and are part of the operating instructions manual.
Editing status: 2016-04-13

4
1 About this document
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
1 About this document
1.1 Function
This operating instructions manual provides all the information you
need for mounting, connection and setup as well as important instruc-
tionsformaintenanceandfaultrectication.Pleasereadthisinforma-
tion before putting the instrument into operation and keep this manual
accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained specialist
personnel.The contents of this manual should be made available to
these personnel and put into practice by them.
1.3 Symbols used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
• List
The dot set in front indicates a list with no implied sequence.
→ Action
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat-
teries and accumulators.

5
2 For your safety
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual must
be carried out only by trained specialist personnel authorised by the
plant operator.
During work on and with the device the required personal protective
equipment must always be worn.
2.2 Appropriate use
VEGAFLEX 83 is a sensor for continuous level measurement.
Youcannddetailedinformationabouttheareaofapplicationin
chapter "Product description".
Operational reliability is ensured only if the instrument is properly
usedaccordingtothespecicationsintheoperatinginstructions
manual as well as possible supplementary instructions.
2.3 Warning about incorrect use
Inappropriate or incorrect use of the instrument can give rise to
application-specichazards,e.g.vesseloverllordamagetosystem
components through incorrect mounting or adjustment. Also the pro-
tectivecharacteristicsoftheinstrumentcanbeinuenced.
2.4 General safety instructions
This is a state-of-the-art instrument complying with all prevailing
regulations and guidelines. The instrument must only be operated in a
technicallyawlessandreliablecondition.Theoperatorisresponsible
for the trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new regula-
tions.
The safety instructions in this operating instructions manual, the na-
tional installation standards as well as the valid safety regulations and
accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbi-
traryconversionsormodicationsareexplicitlyforbidden.
The safety approval markings and safety tips on the device must also
be observed.
2.5 CE conformity
ThedevicefullsthelegalrequirementsoftheapplicableECguide-
lines.ByaxingtheCEmarking,weconrmsuccessfultestingofthe
product.

6
2 For your safety
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
YoucanndtheCECerticateofConformityinthedownloadsection
of our homepage.
Electromagnetic compatibility
Instruments in four-wire or Ex-d-ia version are designed for use in an
industrial environment. Nevertheless, electromagnetic interference
from electrical conductors and radiated emissions must be taken into
account, as is usual with class A instruments according to EN 61326-
1.Iftheinstrumentisusedinadierentenvironment,theelectromag-
netic compatibility to other instruments must be ensured by suitable
measures.
2.6 NAMUR recommendations
NAMUR is the automation technology user association in the process
industry in Germany.The published NAMUR recommendations are
acceptedasthestandardineldinstrumentation.
ThedevicefullstherequirementsofthefollowingNAMURrecom-
mendations:
• NE 21 – Electromagnetic compatibility of equipment
• NE53–Compatibilityofelddevicesanddisplay/adjustment
components
• NE107–Self-monitoringanddiagnosisofelddevices
For further information see www.namur.de.
2.7 Environmental instructions
Protection of the environment is one of our most important duties.
That is why we have introduced an environment management system
with the goal of continuously improving company environmental pro-
tection.Theenvironmentmanagementsystemiscertiedaccording
to DIN EN ISO 14001.
Pleasehelpusfullthisobligationbyobservingtheenvironmental
instructions in this manual:
• Chapter "Packaging, transport and storage"
• Chapter "Disposal"

7
3 Product description
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
3 Product description
3.1 Conguration
Thetypelabelcontainsthemostimportantdataforidenticationand
use of the instrument:
2
1
13
14
15
16
12
11
5
3
6
4
7
8
9
10
Fig. 1: Layout of the type label (example)
1 Instrument type
2 Product code
3 Approvals
4 Power supply and signal output, electronics
5 Protection rating
6 Probe length
7 Process and ambient temperature, process pressure
8 Material, wetted parts
9 Hardware and software version
10 Order number
11 Serial number of the instrument
12 Symbol of the device protection class
13 ID numbers, instrument documentation
14 Reminder to observe the instrument documentation
15 NotiedauthorityforCEmarking
16 Approval directives
The type label contains the serial number of the instrument. With it
youcanndthefollowinginstrumentdataonourhomepage:
• Product code (HTML)
• Delivery date (HTML)
• Order-specicinstrumentfeatures(HTML)
• Operating instructions and quick setup guide at the time of ship-
ment (PDF)
• Order-specicsensordataforanelectronicsexchange(XML)
• Testcerticate(PDF)-optional
Go to www.vega.com "VEGATools" and "Instrument search". Enter
the serial number.
Alternatively, you can access the data via your smartphone:
• Download the smartphone app "VEGATools" from the "Apple App
Store" or the "GooglePlayStore"
Type label
Serial number - Instru-
ment search

8
3 Product description
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
• Scan the Data Matrix code on the type label of the instrument or
• Enter the serial number manually in the app
Theinstrumentcontainstwodierentelectronicsinitshousingcham-
bers:
• The Modbus electronics for power supply and communication with
the Modbus-RTU
• The sensor electronics for the actual measuring tasks
2
1
Fig. 2: Position of Modbus and sensor electronics
1 Modbus electronics
2 Sensor electronics
This operating instructions manual applies to the following instrument
versions:
• Hardware from 1.0.0
• Software from 1.2.0
• OnlyforinstrumentversionswithoutSILqualication
This electronics version can be determined via the product code on
the type label as well as on the electronics.
• Standard electronics:Type FX80HK.-
The scope of delivery encompasses:
• Sensor
• Documentation
– Quick setup guide
– Testcerticatemeasuringaccuracy(optional)
– Supplementary instructions "GSM/GPRSradiomodule"
(optional)
– Supplementary instructions manual "Heating for display and
adjustment module" (optional)
– Supplementary instructions manual "Plug connector for con-
tinuously measuring sensors" (optional)
– Ex-specic"Safety instructions" (with Ex versions)
– Ifnecessary,furthercerticates
3.2 Principle of operation
The VEGAFLEX 83 is a level sensor with polished rod probe for
continuous level or interface measurement, particularly suitable for
applications in the food processing and pharmaceutical industry.
Electronics design
Scope of this operating
instructions manual
Versions
Scope of delivery
Application area

9
3 Product description
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
Optionally an autoclaved version with separable housing is available.
High frequency microwave pulses are guided along a steel cable or
a rod. Upon reaching the product surface, the microwave pulses are
reected.Therunningtimeisevaluatedbytheinstrumentandoutput-
ted as level.
d
h
1
Fig. 3: Level measurement
1 Sensorreferenceplane(sealsurfaceoftheprocesstting)
d Distancetothelevel(HARTvalue1)
h Height - Level
Probe end tracking
To increase sensitivity, the probe is equipped with probe end tracking.
In products with a low dielectric constant, this function is very helpful.
This is the case, for example, in plastic granules, packing chips or in
vesselswithuidizedproducts.
Between a dielectric constant of 1.5 and 3, the function switches on, if
required. As soon as the level echo can no longer be detected, probe
end tracking is automatically activated. The measurement is contin-
ued with the last calculated dielectric constant.
The accuracy thus depends on the stability of the dielectric constant.
If you measure a medium with a dielectric constant below 1.5, probe
end tracking is always active. In this case, you have to enter the
dielectric constant of the medium. A stable dielectric constant is very
important here.
High frequency microwave impulses are guided along a steel cable or
rod. Upon reaching the product surface, a part of the microwave im-
pulsesisreected.Theotherpartpassesthroughtheupperproduct
andisreectedbytheinterface.Therunningtimestothetwoproduct
layers are processed by the instrument.
Functional principle -
level measurement
Functional principle - in-
terface measurement

10
3 Product description
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
TS
d1
h1
h2
d2
1
L2
L1
L3
Fig. 4: Interface measurement
1 Sensorreferenceplane(sealsurfaceoftheprocesstting)
d1 Distancetotheinterface(HARTvalue1)
d2 Distancetothelevel(HARTvalue3)
TS Thicknessoftheuppermedium(d1-d2)
h1 Height - Interface
h2 Height - Level
L1 Lower medium
L2 Upper medium
L3 Gasphase
Upper medium (L2)
• The upper medium must not be conductive
• The dielectric constant of the upper medium or the actual distance
to the interface must be known (input required). Min. dielectric con-
stant:1.6.Youcanndalistofdielectricconstantsonourhome
page: www.vega.com.
• The composition of the upper medium must be stable, no varying
products or mixtures
• Theuppermediummustbehomogeneous,nostratications
within the medium
• Min. thickness of the upper medium 50 mm (1.97 in)
• Clear separation from the lower medium, emulsion phase or detri-
tus layer max. 50 mm (1.97 in)
• If possible, no foam on the surface
Lower medium (L1)
• The dielectric constant must be 10 higher than the dielectric
constant of the upper medium, preferably electrically conductive.
Example: upper medium dielectric constant 2, lower medium at
least dielectric constant 12.
Gas phase (L3)
• Air or gas mixture
• Gas phase - dependent on the application, gas phase does not
always exist (d2 = 0)
Prerequisites for inter-
face measurement

11
3 Product description
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
The instrument is always preset to the application "Level measure-
ment".
For the interface measurement, you can select the requested output
signal with the setup.
3.3 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
based on ISO 4180.
The packaging of standard instruments consists of environment-
friendly, recyclable cardboard. For special versions, PE foam or PE
foil is also used. Dispose of the packaging material via specialised
recycling companies.
Transport must be carried out in due consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or con-
cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
• Not in the open
• Dry and dust free
• Not exposed to corrosive media
• Protected against solar radiation
• Avoiding mechanical shock and vibration
• Storage and transport temperature see chapter "Supplement -
Technicaldata-Ambientconditions"
• Relative humidity 20 … 85 %
3.4 Accessories and replacement parts
The display and adjustment module PLICSCOM is used for measured
value indication, adjustment and diagnosis. It can be inserted into the
sensor and removed at any time.
Youcanndfurtherinformationintheoperatinginstructions"Display
andadjustmentmodulePLICSCOM" (Document-ID 27835).
The interface adapter VEGACONNECT enables the connection of
communication-capable instruments to the USB interface of a PC. For
parameter adjustment of these instruments, the adjustment software
PACTware with VEGA-DTM is required.
Youcanndfurtherinformationintheoperatinginstructions"Interface
adapterVEGACONNECT" (Document-ID 32628).
Output signal
Packaging
T
ransport
Transport inspection
Storage
Storage and transport
temperature
PLICSCOM
VEGACONNECT

12
3 Product description
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
The VEGADIS 81 is an external display and adjustment unit for VEGA
plics®sensors.
For sensors with double chamber housing the interface adapter
"DISADAPT" is also required for VEGADIS 81.
Youcanndfurtherinformationintheoperatinginstructions"VE-
GADIS81" (Document-ID 43814).
The adapter "DISADAPT" is an accessory part for sensors with dou-
ble chamber housings. It enables the connection of VEGADIS 81 to
the sensor housing via an M12 x 1 plug.
Youcanndfurtherinformationinthesupplementaryinstructions
"AdapterDISADAPT" (Document-ID 45250).
PLICSMOBILE T61 is an external GSM/GPRS radio unit for transmis-
sion of measured values and for remote parameter adjustment of
plics®sensors. Adjustment is carried out via PACTware/DTM and the
integrated USB connection.
Youcanndfurtherinformationinthesupplementaryinstructions
"PLICSMOBILET61" (Document-ID 37700).
The protective cover protects the sensor housing against soiling and
intense heat from solar radiation.
Youwillndadditionalinformationinthesupplementaryinstructions
manual "Protective cover" (Document-ID 34296).
Screwedangesareavailableindierentversionsaccordingtothe
following standards: DIN 2501, EN 1092-1, BS 10, ASME B 16.5,
JIS B 2210-1984, GOST 12821-80.
Youcanndadditionalinformationinthesupplementaryinstructions
manual "FlangesaccordingtoDIN-EN-ASME-JIS".
The electronics module VEGAFLEX series 80 is a replacement part
forTDRsensorsofVEGAFLEXseries80.Thereisadierentversion
available for each type of signal output.
Youcanndfurtherinformationintheoperatinginstructionsmanual
"ElectronicsmoduleVEGAFLEXseries80".
The electronics module Modbus is a replacement part for TDR sen-
sors of VEGAFLEX series 80.
Youcanndfurtherinformationintheoperatinginstructionsmanual
"ElectronicsmoduleModbusVEGAFLEXseries80".
The display and adjustment module can be optionally replaced by a
display and adjustment module with heating function.
You can use this display and adjustment module in an ambient tem-
perature range of -40 … +70 °C.
Youcanndfurtherinformationintheoperatinginstructions"Display
and adjustment module with heating" (Document-ID 31708).
VEGADIS 81
DISADAPT
PLICSMOBILE T61
Protective cap
Flanges
Electronics module
Electronics module,
Modbus
Display and adjustment
module with heating

13
3 Product description
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
If the standard sensor housing is too big or in case of strong vibra-
tions, an external housing can be used.
Then the sensor housing is made of stainless steel. The electronics is
located in the external housing which can be mounted in a distance of
up to 10 m (147 ft) to the sensor by using a connection cable.
Youcanndadditionalinformationintheoperatinginstructions
manual "Externalhousing" (Document-ID 46802).
If you mount the VEGAFLEX 83 in a bypass tube or standpipe, you
have to avoid contact to the bypass tube by using a spacer at the
probe end.
Youcanndadditionalinformationintheoperatinginstructions
manual "Centering".
If you are using an instrument with rod version, you can extend the
rodprobeindividuallywithrodextensionsofdierentlengthsorseg-
mentitfordicultinstallationsituations.
All extensions used must not exceed a total length of 4 m (13.12 ft).
The extensions are available in the following lengths:
Rod ø 8 mm (0.315 in)
• Basic segment: 450 mm (17.72 in)
• Rod segments: 450 … 480 mm (17.72 … 18.9 in)
• End segment: 26 … 480 mm (1.02 … 18.9 in)
Youcanndfurtherinformationintheoperatinginstructionsmanual
"RodandcablecomponentsVEGAFLEX80series".
External housing
Centering
Rod components

14
4 Mounting
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
4 Mounting
4.1 General instructions
Protect your instrument against moisture ingress through the following
measures:
• Use the recommended cable (see chapter "Connectingtopower
supply" of the operating instructions)
• Tighten the cable gland
• Whenmountinghorizontally,turnthehousingsothatthecable
gland points downward
• Loop the connection cable downward in front of the cable gland
This applies mainly to outdoor installations, in areas where humidity is
expected (e.g. through cleaning processes) and on cooled or heated
vessels.
Metric threads
In the case of instrument housings with metric thread, the cable
glands are screwed in at the factory. They are sealed with plastic
plugs as transport protection.
You have to remove these plugs before electrical connection.
NPT thread
In the case of instrument housings with self-sealing NPT threads, it is
not possible to have the cable entries screwed in at the factory. The
free openings for the cable glands are therefore covered with red dust
protection caps as transport protection.The dust protection caps do
notprovidesucientprotectionagainstmoisture.
Prior to setup you have to replace these protective caps with ap-
proved cable glands or close the openings with suitable blind plugs.
Make sure that all parts of the instrument exposed to the process are
suitable for the existing process conditions.
These are mainly:
• Active measuring component
• Processtting
• Process seal
Process conditions are particularly:
• Process pressure
• Process temperature
• Chemical properties of the medium
• Abrasionandmechanicalinuences
Youcannddetailedinformationontheprocessconditionsinchapter
"Technicaldata" as well as on the type label.
4.2 Mounting instructions
Mount VEGAFLEX 83 in such a way that the distance to vessel instal-
lations or to the vessel wall is at least 300 mm (12 in). In non-metallic
Protection against mois-
ture
Cable glands
Suitability for the process
conditions
Installation position

15
4 Mounting
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
vessels, the distance to the vessel wall should be at least 500 mm
(19.7 in).
During operation, the probe must not touch any installations or the
vessel wall. If necessary, fasten the probe end.
In vessels with conical bottom it can be advantageous to mount the
sensor in the center of the vessel, as measurement is then possible
nearly down to the lowest point of the bottom. Keep in mind that
measurement all the way down to the tip of the probe may not be pos-
sible.The exact value of the min. distance (lower dead band) is stated
in chapter "Technicaldata".
Fig. 5: Vessel with conical bottom
Plastic vessel/Glass vessel
The guided microwave principle requires a metallic surface on the
processtting.Therefore,inplasticvessels,etc.,useaninstru-
mentversionwithange(fromDN50)orplaceametalsheet
(ø>200mm/8in)beneaththeprocessttingwhenscrewingitin.
Makesurethattheplatehasdirectcontactwiththeprocesstting.
When installing rod or cable probes in vessels without metal walls,
e.g.inplasticvessels,themeasuredvaluecanbeinuencedby
strongelectromagneticelds(emittedinterferenceaccordingto
EN 61326: class A). In this case, use a probe with coaxial version.
Type of vessel

16
4 Mounting
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
1 2
Fig. 6: Installation in non-metallic vessel
1 Flange
2 Metal sheet
Ifpossible,avoidsockets.Mountthesensorushwiththevesseltop.
If this is not possible, use short sockets with small diameter.
Higher sockets or sockets with a bigger diameter can generally be
used.They can, however, increase the upper blocking distance (dead
band). Check if this is relevant for your measurement.
In such cases, always carry out a false signal suppression after instal-
lation.Youcanndfurtherinformationunder"Setup procedure".
h
d
dh
DN25 ...DN150 ≤ 150 mm (5.91")
≤ 100 mm (3.94")> DN150 ...DN200
Fig. 7: Mounting socket
Whenweldingthesocket,makesurethatthesocketisushwiththe
vessel top.
Mounting socket

17
4 Mounting
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
1 2
Fig.8:Socketmustbeinstalledush
1 Unfavourable installation
2 Socketush-optimuminstallation
Before beginning the welding work, remove the electronics module
from the sensor. By doing this, you avoid damage to the electronics
through inductive coupling.
Donotmounttheinstrumentsinorabovethellingstream.Makesure
thatyoudetecttheproductsurface,nottheinowingproduct.
Fig.9:Mountingofthesensorwithinowingmedium
The reference plane for the measuring range of the sensors is the
sealingsurfaceofthethreadorange.
Keep in mind that a min. distance must be maintained below the refer-
ence plane and possibly also at the end of the probe - measurement
in these areas is not possible (dead band).The length of the cable
can be used all the way to the end only when measuring conductive
Welding work
Inowingmedium
Measuring range

18
4 Mounting
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
products.Theseblockingdistancesfordierentmediumsarelisted
in chapter "Technicaldata". Keep in mind for the adjustment that the
default setting for the measuring range refers to water.
Theprocessttingmustbesealedifthereisgaugeorlowpressurein
the vessel. Before use, check if the seal material is resistant against
the measured product and the process temperature.
Themax.permissiblepressureisspeciedinchapter"Technical
data" or on the type label of the sensor.
Incaseofdicultinstallationconditions,theprobecanalsobe
mounted laterally. For this, adapt the rod with rod extensions or
angled segments.
To compensate for the resulting changes in signal runtime, let the
instrument determine the probe length automatically.
The determined probe length can deviate from the actual probe
length when using curved or angled segments.
If internal installations such as struts, ladders, etc. are present on the
vessel wall, the measuring probe should be mounted at least 300 mm
(11.81 in) away from the vessel wall.
Youcanndfurtherinformationinthesupplementaryinstructionsof
the rod extension.
If there is a risk of the rod probe touching the vessel wall during
operation due to product movements or agitators, etc., the measuring
probeshouldbesecurelyxedattheextremelowerend.
For this, use an additional plastic sleeve (PTFE, PPS, PEEK etc.), to
protect the probe against damages.
Keep in mind that below the fastening, a measurement is not possible.
1
2
3
1
2
3
Fig. 10: Fasten the probe
1 Measuring probe
2 Retaining sleeve
3 Plasticsleeve(PTFE,PPS,PEEKetc.)
Pressure
Lateral installation
Fasten

19
4 Mounting
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
Incaseofdicultinstallationconditions,forexampleinasocket,the
probe can be suitably adapted with a rod extension.
To compensate for the resulting changes in signal runtime, let the
instrument determine the probe length automatically.
Youcanndfurtherinformationinthesupplementaryinstructionsof
the rod and cable components.
Foruseinanautoclave,e.g.forsterilization,theVEGAFLEX83is
available as autoclaved version.
Henceyoucanseparatethehousingfromtheprocesstting.
Under very harsh ambient conditions, the autoclavable version can be
optionally also combined with an external housing.
Open the slotted nut with a hook wrench and remove the housing in
an upward direction.
Thesideoftheprocessttingmustbecoveredwithalidafterthe
housing is removed. Screw the enclosed lid with slotted nut onto the
instrumentsideoftheprocessttingandtightenthenutwithatorque
of 20 Nm.
Make sure that no liquid or contamination penetrates into the housing
or the process side.
Afterautoclaving,screwthelidoagainandplacethehousingverti-
callyonthesideoftheprocesstting.Tightentheslottednutwitha
torque of 20 Nm.
Rod extension
Autoclaved version

20
4 Mounting
VEGAFLEX 83 • Four-wire Modbus
41843-EN-160517
112 mm
(4.84")
DIN DN25 DN32 DN40 / 1" 11/2"
L
ø 54 mm
(2.13")
ø w
183 mm
(7.20")
65,3 mm
(2.57")
ø w
ø 50,5
DIN DN50 / 2" ø 64
ø 91DIN DN65 / 3"
21
3
Fig. 11: Autoclaved version
1 Groovenut
2 Processtting
3 Coverwithgroovenut
Other manuals for vegaflex 83
11
Table of contents
Other Vega Measuring Instrument manuals

Vega
Vega VEGABAR 17 User manual

Vega
Vega VEGASON 61 User manual

Vega
Vega VEGACAP 69 User manual

Vega
Vega VEGABAR 66 User manual

Vega
Vega VEGACAP 66 User manual

Vega
Vega VEGAPULS 6X User manual

Vega
Vega VEGACAL 65 User manual

Vega
Vega INFO-1 User manual

Vega
Vega VEGACAP 35 User manual

Vega
Vega VEGASWING 63 User manual