Vega VEGASCAN 693 User manual

Operating Instructions
VEGASCAN 693
15-channel HART signal conditioning
instrument
in out

Contents
1About this document
1.1Function ............................. 4
1.2Target group .......................... 4
1.3Symbolism used ....................... 4
2For your safety
2.1Authorised personnel .................... 5
2.2Appropriate use........................ 5
2.3Warning about misuse ................... 5
2.4CE conformity ......................... 5
2.5Safety information for Ex areas............. 6
2.6Environmental instructions ................ 6
3Product description
3.1Configuration.......................... 7
3.2Principle of operation.................... 8
3.3Adjustment ........................... 8
3.4Storage and transport ................... 9
4Mounting
4.1General instructions..................... 10
4.2Mounting information .................... 10
5Electrical connection
5.1Preparing the connection ................. 12
5.2Connection steps....................... 13
5.3Wiring plans .......................... 14
6Setup with the integrated indicating and adjustment
unit
6.1Adjustment system ..................... 16
6.2Setup procedure ....................... 17
6.3Menu schematic ....................... 25
7Setup with PACTware™
7.1Connecting the PC ..................... 31
7.2Parameter adjustment with PACTware™...... 34
7.3Setup web server/e-mail,remote enquiry...... 35
8Application examples
8.1Inventory enquiry via network and browser .... 36
8.2Inventory enquiry via modem and Visual VEGA.37
2VEGASCAN 693 -15-channel HART signal conditioning instrument
Contents
29252-EN-060513

9Maintenance and fault rectification
9.1Maintenance .......................... 39
9.2Fault rectification ....................... 39
9.3Instrument repair ....................... 41
10 Dismounting
10.1Dismounting procedure .................. 42
10.2Disposal ............................. 42
11 Supplement
11.1Technical data......................... 43
11.2Dimensions ........................... 46
Note:
The operating instructions manual "RS232/Ethernet connec-
tion"as well as the additional instructions manual "Modbus-
TCP,VEGA ASCII protocol"is attached to instructions with
RS232/Ethernet interface.Here you will find further information
for setup.
VEGASCAN 693 -15-channel HART signal conditioning instrument 3
Contents
29252-EN-060513

1About this document
1.1Function
This operating instructions manual has all the information you
need for quick setup and safe operation.Please read this
manual before you start setup.
1.2Target group
This operating instructions manual is directed to trained,
qualified personnel.The contents of this manual should be
made available to these personnel and put into practice by
them.
1.3Symbolism used
Information,tip,note
This symbol indicates helpful additional information.
Caution:If this warning is ignored,faults or
malfunctions can result.
Warning:If this warning is ignored,injury to persons and/or
serious damage to the instrument can result.
Danger:If this warning is ignored,serious injury to persons
and/or destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
lList
The dot set in front indicates a list with no implied sequence.
àAction
This arrow indicates a single action.
1Sequence
Numbers set in front indicate successive steps in a procedure.
4VEGASCAN 693 -15-channel HART signal conditioning instrument
About this document
29252-EN-060513

2For your safety
2.1Authorised personnel
All operations described in this operating instructions manual
must be carried out only by trained specialist personnel
authorised by the operator.For safety and warranty reasons,
any internal work on the instruments must be carried out only
by personnel authorised by the manufacturer.
2.2Appropriate use
VEGASCAN 693 is a universal signal conditioning instrument
and power supply unit for connection of up to 15 HART
sensors.
2.3Warning about misuse
Inappropriate or incorrect use of the instrument can give rise to
application-specific hazards,e.g.vessel overfill or damage to
system components through incorrect mounting or adjustment.
2.4General safety instructions
The VEGASCAN 693 is a high-tech instrument requiring the
strict observance of standard regulations and guidelines.The
user must take note of the safety instructions in this operating
instructions manual,the country-specific installation standards
(e.g.the VDE regulations in Germany)as well as all prevailing
safety regulations and accident prevention rules.
2.5CE conformity
VEGASCAN 693 is in CE conformity with EMC (89/336/EWG)
and LVD (73/23/EWG).
Conformity has been judged acc.to the following standards:
lEMC:
-Emission EN 61326:1997 (class A)
-Susceptibility EN 61326:1997/A1:1998
lLVD:EN 61010-1:2001
VEGASCAN 693 is designed for use in an industrial environ-
ment.Nevertheless,electromagnetic interference from elec-
trical conductors and radiated emissions must be taken into
account,as is usual with a class Ainstrument acc.to
VEGASCAN 693 -15-channel HART signal conditioning instrument 5
For your safety
29252-EN-060513

EN 61326.IfVEGASCAN 693 is used in a different environ-
ment,the electromagnetic compatibility to other instruments
must be ensured by suitable measures.
2.6Safety information for Ex areas
Please note the Ex-specific safety information for installation
and operation in Ex areas.These safety instructions are part of
the operating instructions manual and come with the Ex-
approved instruments.
2.7Environmental instructions
Protection of the environment is one of our most important
duties.That is why we have introduced an environment
management system with the goal of continuously improving
company environmental protection.The environment man-
agement system is certified acc.to DIN EN ISO 14001.
Please help us fulfil this obligation by observing the environ-
mental instructions in this manual:
lChapter "Storage and transport"
lChapter "Disposal"
6VEGASCAN 693 -15-channel HART signal conditioning instrument
For your safety
29252-EN-060513

3Product description
3.1Configuration
The scope of delivery encompasses:
lSignal conditioning instrument VEGASCAN 693
lSocket
lCoded pins and bridges
lRS232 modem connection cable (optional)
ldocumentation
-this operating instructions manual
-operating instructions manual "RS232/Ethernet con-
nection"
-Additional instruction manual "Modbus-TCP,VEGA
ASCII protocol"(optional)
-Ex specific safety instructions (with Ex versions), if
necessary further certificates
VEGASCAN 693 consists of the following components:
lVEGASCAN 693 signal conditioning instrument with
indicating and adjustment unit in the front
lCommunication interface for VEGACONNECT (I²C)
lRS232 or Ethernet interface
lSocket
VEGASCAN 693
12 34
910 1112
17 18 20 21 22
56 78
13 14 15 16
23 24 25 26 27 28
Ser. No.
12345678
on
OK
ESC
COM
2
3
4
1
6
5
Fig.1:VEGASCAN 693
1Ex separating chamber with Ex version
2VEGASCAN 693
3Indicating and adjustment unit
4Communication interface for VEGACONNECT (I²C)
5RS232 or Ethernet interface
6Socket
Scope of delivery
Components
VEGASCAN 693 -15-channel HART signal conditioning instrument 7
Product description
29252-EN-060513

3.2Principle of operation
VEGASCAN 693 is a universal signal conditioning instrument
for a number of applications such as level,gauge and process
pressure measurement.At the same time,it can serve as
power supply unit for connected sensors.VEGASCAN 693 is
designed for connection of 15 independent VEGA HART
sensors (5with Ex applications). Hence 15 (5with Ex)
independent measurements can be carried out.
The measured values can be retrieved through one of the
integrated interfaces (RS232/Ethernet)via modem or network
and displayed with the web browser or Visual VEGA.
Furthermore it is possible to transmit measured values or
messages by e-mail.The use of VEGASCAN 693 is
particularly suitable of stocktaking,VMI (Vendor Managed
Inventory)and remote enquiry.
VEGASCAN 693 signal conditioning instrument can power the
connected sensors and also evaluate their measuring signals.
The requested parameter is indicated in the display and,in
addition,outputted to the integrated interface and the web
server for further processing.On request,the measured values
can be sent event or time-controlled via e-mail to different
addressees.
Wide-range power supply unit with 20 …253 VAC/DC for
world-wide use.
You can find detailed information on the power supply in the
"Technical data"in the "Supplement".
3.3Adjustment
VEGASCAN 693 can be adjusted with the following adjust-
ment media:
lthe integrated indicating and adjustment unit
lan adjustment software acc.to FDT/DTM standard,e.g.
PACTware™and a Windows PC
The entered parameters are generally saved in VEGASCAN
693,when operated with PACTware™optionally also on the
PC.
Information:
When using PACTware™and corresponding VEGA DTM,
additional settings can be carried out which are not or only
partly possible with the integrated indicating and adjustment
Area of application
Physical principle
Supply
8VEGASCAN 693 -15-channel HART signal conditioning instrument
Product description
29252-EN-060513

unit.When using an adjustment software,you either need one
of the integrated interfaces (RS232/Ethernet)or the interface
converted VEGACONNECT.
Further instructions for setting up the web server and e-mail
functions are stated in the online help of PACTware™or the
VEGASCAN 693 DTM as well as the operating instructions
manual "RS232/Ethernet connection".
3.4Storage and transport
Your instrument was protected by packaging during transport.
Its capacity to handle normal loads during transport is assured
by a test acc.to DIN EN 24180.
The packaging of standard instruments consists of environ-
ment-friendly,recyclable cardboard.For special versions PE
foam or PE foil is also used.Dispose of the packaging material
via specialised recycling companies.
lStorage and transport temperature see "Supplement -
Technical data -Ambient conditions"
lRelative humidity 20 …85 %
Packaging
Storage and transport tem-
perature
VEGASCAN 693 -15-channel HART signal conditioning instrument 9
Product description
29252-EN-060513

4Mounting
4.1General instructions
Each series 600 instrument consists of the actual signal
conditioning instrument as well as a plug-in socket for carrier
rail mounting.Because it has protection class IP 30 or IP 20,
the instrument is intended to be used in switching cabinets.
4.2Mounting information
The plug-in socket is constructed for carrier rail mounting acc.
to EN 50022.Power supply is connected to terminals 17 and
18.For neighbouring series 600 signal conditioning instru-
ments,it is possible to continue connection L1and Ndirectly
via the supplied bridges.
Danger:
The bridges must never be used with single instruments or at
the end of a row of instruments.If this rule is not heeded,there
is a danger of coming into contact with the operating voltage or
causing a short circuit.
A VEGASCAN 693 in Ex version is an auxiliary,intrinsically
safe instrument and must not be installed in hazardous areas.
Before setup,the Ex separating chamber must be attached (as
shown below)with Ex versions.Safe operation can be only
ensured if the operating instructions manual and the EU type
approval certificate are observed.VEGASCAN 693 must not
be opened.
All signal conditioning instruments are provided with different
gaps depending on type and version (mechanical coding).
The plug-in socket is provided with coded pins that can be
inserted to prevent accidental interchanging of the various
instrument types.
With a VEGASCAN 693 in Ex version,the supplied coded pins
(type coded pin and Ex coded pin)must be inserted by the
user acc.to the below chart.
Installation location
Mounting
Instrument coding
10 VEGASCAN 693 -15-channel HART signal conditioning instrument
Mounting
29252-EN-060513

A o
B o
C o
1 o
2 o
3 o
o
o
o
7 o
8 o
9 o
o
o
12 o
N
L1
VEGA A o
B o
C o
1 o
2 o
3 o
4o
6 o
7 o
8 o
9 o
10 o
11 o
12 o
N
L1
VEGA
1234
5678
5o
1
4 4
2
3
Fig.2:Socket VEGASCAN 693
1Ex separating chamber
2Ex coding with Ex version
3Type coding for VEGASCAN 693
4Bridges for looping the power supply
VEGASCAN 693 -15-channel HART signal conditioning instrument 11
Mounting
29252-EN-060513

5Electrical connection
5.1Preparing the connection
Always observe the following safety instructions:
lConnect only in the complete absence of line voltage
lIf overvoltage surges are expected,overvoltage arresters
should be installed
Tip:
We recommend VEGA overvoltage arresters B61-300 (power
supply VEGASCAN 693)and B62-36G(sensor supply).
In hazardous areas you should take note of the appropriate
regulations,conformity and type approval certificates of the
sensors and power supply units.
The power supply can be 20 …253 VAC,50/60 Hz.
Power supply of VEGASCAN 693 is connected with standard
cable acc.to the national installation standards.
Standard two-wire cable with screening can be used for
connecting the sensors.The screening is absolutely neces-
sary to ensure interference-free operation with HART sensors.
Connect the cable screen on both ends to ground potential.In
the sensor,the screen must be connected directly to the
internal ground terminal.The ground terminal outside on the
sensor housing must be connected to the potential equal-
isation (low impedance).
If potential equalisation currents are expected,the screen
connection on the VEGASCAN 693 must be made via a
ceramic capacitor (e.g.1nF,1500 V). The low frequency
potential equalisation currents are thus suppressed,but the
protective effect against high frequency interference signals
remains.
Take note of the corresponding installation regulations for Ex
applications.In particular,make sure that no potential equal-
isation currents flow over the cable screen.In case of
grounding on both sides this can be achieved by the use of a
capacitor or a separate potential equalisation.
Note safety instructions
Take note of safety
instructions for Ex
applications
Select power supply
Select connection cable
Cable screening and ground-
ing
Select connection
cable for Ex applica-
tions
12 VEGASCAN 693 -15-channel HART signal conditioning instrument
Electrical connection
29252-EN-060513

5.2Connection steps
Tip:
Before the actual setup,each HART sensors must be
assigned an address (see chapter "Setup"). When assigning
an address,only one sensor should be connected to
VEGASCAN 693.Depending on the installation location of the
sensors,it can be advantageous to carry out this addressing
before installing and connecting the sensors.This can be
conveniently carried out e.g.in the workshop.You just need a
24 Volt power supply as well as an indicating and adjustment
module PLICSCOM or the adjustment software PACTware™
with VEGACONNECT.
For electrical connection,proceed as follows:
1Snap socket without VEGASCAN 693 onto the carrier rail
2Connect sensor cable to terminal 1/2(active input)or 3/4
(passive input), provide a screening
3When using several sockets,loop the power supply by
means of bridges
4Connect power supply (switched off)to terminal 17 and 18
5If necessary,connect relays or other outputs
6Insert VEGASCAN 693 into the plug-in socket and screw it
down tightly
The electrical connection is finished.
Before setting up Ex versions,make sure the Ex separating
chamber is plugged onto the left side of the housing (above the
sensor terminals). The pins for type and Ex coding must also
be inserted correctly.
Information:
lOn the active input (terminals 1/2), VEGASCAN 693
provides the power supply for the connected sensors.
Power supply and measured value transmission are
carried out via the same two-wire cable.This operating
mode is intended for sensors without separate power
supply (sensors in two-wire version).
lOn the passive input (terminals 3/4), the sensors are not
powered,only the measured value is transmitted.This
input is for connection of transmitters with their own,
VEGASCAN 693 -15-channel HART signal conditioning instrument 13
Electrical connection
29252-EN-060513

separate power supply (sensors in four-wire version). The
connection and operation in Ex ia is not permitted on the
passive input.
Note:
VEGASCAN 693 is designed for connection of up to 15 HART
sensors (5with Ex). Because they are accessed via different
addresses in the HART multidrop mode,all sensors must be
connected to the same sensor input.These are either
terminals 1/2(active input)or terminals 3/4(passive input).
Mixed operation of active and passive input is not possible.
Since this is a digital bus system,only one two-wire cable
should lead to the sensors.Adistributor can then be
connected directly forward of the sensors.As an alternative,
the connection cable can be looped through the second cable
entry on the sensor housing.
5.3Wiring plans
17 18 19 20 21 22
910 1112
1
2
34
23 24 25 26 27 28
13 14 15 16
56 78
L1 N
+
-
+-
+-
5
3
1
4
2
6
7
8
Fig.3:Connection example VEGASCAN 693 with two-wire sensors
1Measured data input with sensor supply (active input)
2Measured data input (passive input), not in Exia
3Internal fail safe relay
4Distributor
5Power supply of VEGASCAN 693
6HART two-wire sensor with Multidrop address 1
7HART two-wire sensor with Multidrop address 2
8HART two-wire sensor with Multidrop address 3
Wiring plan for two-wire sen-
sors
14 VEGASCAN 693 -15-channel HART signal conditioning instrument
Electrical connection
29252-EN-060513

17 18 19 20 21 22
910 1112
1
2
34
23 24 25 26 27 28
13 14 15 16
56 78
L1 N
+
-
+-
+-
5
6
7
8
9
1
2
4
3
Fig.4:Connection example,VEGASCAN 693 with four-wire sensors
1Measured data input with sensor supply (active input)
2Measured data input (passive input), not in Exia
3Internal fail safe relay
4Distributor
5Power supply of VEGASCAN 693
6HART four-wire sensor with Multidrop address 1
7HART four-wire sensor with Multidrop address 2
8HART four-wire sensor with Multidrop address 3
9Power supply for four-wire sensors
Wiring plan for four-wire sen-
sors
VEGASCAN 693 -15-channel HART signal conditioning instrument 15
Electrical connection
29252-EN-060513

6Setup with the integrated indicating and
adjustment unit
6.1Adjustment system
The integrated indicating and adjustment unit is used for
measured value display,adjustment and diagnosis of VEGA-
SCAN 693 as well as the connected sensors.The indication
and adjustment is carried out via the clear,graphic-capable
display with background lighting as well as four keys.The
adjustment menu with selectable language is clearly structured
and enables easy setup.
Certain adjustment options are not possible or are partially
restricted with the integrated indicating and adjustment unit,e.
g.the settings for the e-mail server.For such applications,the
use of PACTware™with appropriate DTMs is recommended.
VEGASCAN 693 Ser. No.
12345678
on
OK
ESC
COM
1
2
3
6
5
4
Fig.5:Indicating and adjustment elements
1LC display
2Adjustment keys
3Communication interface for VEGACONNECT
4Status indication operation
5Status indication fail safe relay
6Status indication interface activity
l[OK]key:
-move to the menu overview
-confirm selected menu
-edit parameter
-save value
l[->]key to select:
-menu change
Function
Indicating and adjustment el-
ements
Key functions
16 VEGASCAN 693 -15-channel HART signal conditioning instrument
Setup with the integrated indicating and adjustment unit
29252-EN-060513

-list entry
-editing position
l[+]key:
-modify value of a parameter
l[ESC]key:
-interrupt input
-jump to the next higher menu
Note:
Approx.10 minutes after the last pressing of a key,an
automatic reset to measured value indication is triggered.Any
values not confirmed with [OK]will not be saved.
6.2Setup procedure
Setup encompasses mainly the selection of the application,
assignment of HART addresses and adjustment of measure-
ment loops.Scaling the measured value to the requested size
and unit (possibly with the application of a linearization curve),
as well as setting an integration time for smoothing the
measured value,are additional setup steps.
For instruments with Ethernet interface,the instrument must be
provided with the host name and IP addr./Subnet mask.If
necessary,the e-mail/web server can also be configured with
PACTware™.
VEGASCAN 693 can process measured values of several
connected HART sensors.All measured values are trans-
mitted as digital HART signals on the same cable (bus). An
analogue 4…20 mAtransmission is not possible.An own,
unique address (address range 1-15)must be assigned first of
all to each connected sensor.This mode is also called HART
multidrop mode.Addressing can be done directly on each
HART sensor via the respective adjustment unit or an
appropriate adjustment software.As an alternative,the setting
of the sensor address can be also carried out via the
VEGAMET menu under "Service -Sensor address".
Note:
When addresses are being assigned,only one sensor with a
particular address must be connected on the bus.If this is not
the case,the sensor cannot be accessed and it is not possible
to assign an address.
Setup
Set HART address
VEGASCAN 693 -15-channel HART signal conditioning instrument 17
Setup with the integrated indicating and adjustment unit
29252-EN-060513

After being switched on,VEGASCAN 693 first of all carries out
a short self-check.The following steps are carried out:
lInternal check of the electronics
lindication of the instrument type,firmware version as well
as the instrument TAG (instrument name)
lthe output signals jump briefly to the set fault current
Then the current measured values will be displayed and
outputted.
The measured value indication displays three measured
values each in one common window.The digital display value,
the measurement loop designation (meas.loop TAG)and the
unit are shown.By pushing the [>]key,you can move to the
next three measured values.
32.6
%
TAG-No.1
■
àBy pushing [OK]you move from the measured value
indication to the main menu.
▶
Device settings
Meas.loops
Display
Diagnostics
Service
Info
àSelect the menu item "Device settings"with [->]and
confirm with [OK].
You can assign an unambiguous name to VEGASCAN 693 via
the Device-TAG.This function is recommended when several
VEGAMETs are implemented and a good documentation of
larger system is required.
Device-TAG
Device Name
àEnter the requested values via the appropriate keys and
save your settings with [OK].
Switch-on phase
Measured value indication/
Main menu
Device settings -Device-TAG
18 VEGASCAN 693 -15-channel HART signal conditioning instrument
Setup with the integrated indicating and adjustment unit
29252-EN-060513

For instruments with integrated Ethernet interface also a host
name and the IP address must Subnet Mask for integration in
the network must be entered.These specifications are
available from your network administrator.These settings are
only effective after a restart of VEGASCAN 693.
LAN/Internet
IP address:
192.168.200.200
Subnet mask:
255.255.255.0
Change?
àEnter the values via the appropriate keys and save your
settings with [OK].Disconnect briefly the supply voltage so
that the modified settings become effective.
Date and time can be entered in this menu item for instruments
with integrated RS232/Ethernet interface.These time settings
are buffered in case of power failure for approx.3days.
Time/Date
8:40
28.09.2004
àEnter the values via the appropriate keys and save your
settings with [OK].
Because VEGASCAN 693 can read the measured values of
up to 15 sensors (5with Ex), the individual sensors must be
assigned to the measurement loops.After the addresses of the
HART sensors are assigned,a list with the existing sensors
can be prepared and displayed.Now you can assign the
requested sensor to each measurement loop.
For transmission,VEGASCAN 693 must be informed which
sensor value should be used for further processing.Depend-
ing on the sensor type this can be distance,pressure,interface
or temperature.If a sensor can deliver two measured values,
e.g.pressure and temperature with VEGABAR sensors,a
separate measurement loop must be assigned for each input
variable.When HART sensors of other manufacturers are
connected,the options PV (Primary Value)and SV (Secondary
Value)will be available.The parameter to be transmitted is
stated in the operating instructions manual of the respective
sensor manufacturer.
Device settings -Host Name/
IP addr.
Device settings -Time/Date
Measurement loop -Input
VEGASCAN 693 -15-channel HART signal conditioning instrument 19
Setup with the integrated indicating and adjustment unit
29252-EN-060513

Input 1
VEGAPULS (addr.1)
Distance
Change input?
Via the parameter you inform VEGASCAN 693 of the type of
measurement.The following options are available:
lLevel
lProcess pressure (only with VEGABAR,VEGAWELL,D-
series)
lTemperature (only with VEGABAR,VEGAWELL)
lUniversal (for sensors of other manufacturers)
Parameter
Level
▼
Information:
Keep in mind that some settings must be carried out
individually several times,because they are specifically
required for each measurement loop.This applies e.g.to the
single measurement loops,the displayed values in the display
as well as the outputs.
The unit of measurement must be selected before starting the
adjustment.Depending on the connected instrument this can
be e.g.m(d), ft(d), bar,psi,°Cor %.
Units of measurement 1
m(d)
▼
The following illustrations and examples relate to the min./max.
adjustment of a radar sensor.
Min.adjustment 1
0.00 %
≙
32.00 m(d)
9.245 m(d)
Meas.loop -Parameter
Meas.loop -Adjustment
20 VEGASCAN 693 -15-channel HART signal conditioning instrument
Setup with the integrated indicating and adjustment unit
29252-EN-060513
Other manuals for VEGASCAN 693
1
Table of contents
Other Vega Measuring Instrument manuals

Vega
Vega VEGABAR 54 User manual

Vega
Vega VEGAPULS 67 User manual

Vega
Vega VEGAFLEX 62 User manual

Vega
Vega VEGAPULS 61 User manual

Vega
Vega VEGAPULS 62 User manual

Vega
Vega VEGABAR 81 User manual

Vega
Vega VEGABOX 03 User manual

Vega
Vega VEGABAR 38 User manual

Vega
Vega LMPVL5TN User manual

Vega
Vega VEGACAL 64 User manual

Vega
Vega VEGACAP 66 User manual

Vega
Vega VEGAVIB 62 Installation instructions

Vega
Vega VEGACAL 63 User manual

Vega
Vega VEGAPULS 64 User manual

Vega
Vega VEGAPULS WL 61 User manual

Vega
Vega VEGAPULS 61 User manual

Vega
Vega WEIGHTRAC 31 User manual

Vega
Vega VEGAPULS 63 User manual

Vega
Vega VEGABAR 52 User manual

Vega
Vega VEGACAL 65 User manual