VISIOMED THERMOFLASH LX-26 User manual

MEDICAL INFRARED NO CONTACT THERMOMETER
- LX-26 MODEL
ENGLISH
USER MANUAL

ThermoFlash LX-26 User Manual – English V 15 200904 2
CONTENTS
I S
AFETY PRECAUTIONS
………………………
3
II I
NTRODUCTION
………………………
4
III P
RECAUTIONS BEFORE USE
………………………
4
IV O
PERATING
P
RINCIPLE
………………………
5
T
HE DIFFERENT METHODS OF TEMPERATURE MEASUREMENT
……………………… 5
N
ORMAL TEMPERATURES ACCORDING TO MEASUREMENT METHOD
……………… 6
A
DVANTAGES OF TEMPORAL ARTERY
(
TA
)
TEMPERATURE
… … ……………… 7
N
ORMAL TEMPERATURE ACCORDING TO AGE
……………………… 7
P
RACTICAL CONSIDERATIONS WHEN TAKING A TEMPERATURE
……………………… 7
H
OW TO TAKE A TEMPERATURE
……………………… 8
C
ONSTRAINTS
……………………… 8
V D
ESCRIPTION OF THE
T
HERMO
F
LASH
LX-26
………………………
9
VI F
UNCTIONS
………………………………………… ………………………
11
VII U
SE
……………………………………………
…………………………
11
VIII S
ETTING
&
F
UNCTION OF MENU
………………………………………
12
1
C
HOOSING THE TEMPERATURE UNIT
–
F1
M
ENU
……………………… 12
2
ALARME SETUP
–
F2
M
ENU
……………………………………………… 12
3
C
HOICE OF DISPLAY MODE
:
B
ODY OR
S
URFACE
T
EMP
–
F3
M
ENU
13
4
T
OTAL DIFFERENCE
–
F4
M
ENU
…………………………………………… 14
5
B
UZZER
O
N
/
O
FF
–
F5
M
ENU
…………………………………………… 14
6
E
XITING THE SETTING MODE
……………………………………………… 14
7
D
ATA
M
EMORY
……………… ………………………………………… 14
8
C
HANGING THE BATTERIES
…………………………………… ……… 15
IX T
ECHNICAL
C
HARACTERISTICS
&
PRECISION
…………………………
15
X R
EMARKS
………………………………………… ……………………………
16
XI A
CCESSORIES SUPPLIED
………………………………………… ………
16
XII T
ROUBLESHOOTING
………………………………………………… 17

ThermoFlash LX-26 User Manual – English V 15 200904 3
I SAFETY PRECAUTIONS
-Follow the maintenance advice stipulated in this instruction
manual
-This device may be used for personal home use
-This device must only be used for the purposes described in this
instruction manual
-This device must only be used in an ambient temperature range
of between 10 and 40 °C
-This device must always be kept in a clean, dry area
-Do not expose this thermometer to electric shocks
-Do not expose this thermometer to extreme temperature
conditions of >50°C or > - 20°C
-Do not use this device in relative humidity higher than 85%
-The protective glass over the lens is the most fragile part of the
thermometer
-Do not touch the glass of the infrared lens with your fingers
-Clean the glass with a cotton bud lightly moistened with 70°
alcohol
-Do not expose the thermometer to sunlight or to water
-Do not use this device outside
-Never drop the device
-Should a problem occur with your device, please contact your
retailer Do not attempt to repair this device yourself
IMPORTANT SAFEGUARDS
For your safety, ONLY use the charger supplied
Do not dismantle the device or the rechargeable base station
T
HE MANUFACTURER RESERVES THE RIGHT TO ALTER THE SPECIFICATIONS
OF THE PRODUCT WITHOUT PRIOR NOTIFICATION

ThermoFlash LX-26 User Manual – English V 15 200904 4
II INTRODUCTION
The Visiomed Thermo
Flash
LX-26 thermometer without contact
has been developed using latest infrared technology This
technology allows temporal artery (TA) temperature to be taken at a
distance of about 5cm away from the forehead
Precise, Instantaneous and without Contact, the Thermo
Flash
LX-26 is, to date, the most suitable thermometer for no risk
temperature measurement It has been demonstrated that this
method of TA temperature measurement is more precise than
tympanic thermometry and better tolerated than rectal
thermometry
(1)
However, as with other types of thermometer, it is essential to use
the Thermo
Flash
LX-26 properly in order to obtain reliable and
stable results
You are therefore advised to read this instruction manual and the
safety precautions carefully before use
(1) Greenes D, Fleisher G Accuracy of a Noninvasive Temporal Artery Thermometer for
Use in Infants Arch Pediatr Adolesc Med 2001;155:376
III PRECAUTIONS BEFORE USE
The Thermo
Flash
LX-26 is pre-set at the factory.
It is not necessary to calibrate the device when
starting it up.
In order to obtain reliable and stable results, you are advised each
time there is a significant change in the ambient temperature due to
a change in environment, to allow the Thermo
Flash
LX-26 to
acclimatise to this ambient temperature for 15 to 20 minutes before
using it
It is important to allow a one minute interval between two
measurements

ThermoFlash LX-26 User Manual – English V 15 200904 5
IV OPERATING PRINCIPLE
All objects, solid, liquid or gas, emit energy by radiation The
intensity of this energy depends on the temperature of the object
The Thermo
Flash
LX-26 infrared thermometer is therefore able to
measure the temperature of a person by the energy the person
emits This measurement can be taken thanks to an external
temperature probe on the device which permanently analyses and
registers the ambient temperature Therefore, as soon as the
operator holds the thermometer near the body and activates the
radiation sensor, the measurement is taken instantly by detection of
the infrared heat generated by the arterial blood flow Body heat
can therefore be measured without any interference from the heat
of the surrounding environment
T
E DIFFERENT MET ODS OF TEMPERATURE MEASUREMENT
Core temperature
Core temperature is the most precise measurement and involves
measuring the temperature in the pulmonary artery by means of a
catheter equipped with a thermal probe which can read the
temperature
in situ
The same method is employed for probes
measuring the oesophageal temperature However, such invasive
temperature measurement methods require specific equipment and
expertise
Rectal thermometry
Rectal temperature adjusts slowly in comparison to the evolution of
the body’s internal temperature It has been demonstrated that
rectal temperature remains raised long after the internal
temperature of the patient has started to drop and vice versa
Furthermore, rectal perforations have been known to occur as a
result of this method and without appropriate sterilisation
techniques, rectal thermometry can spread germs often found in
faeces

ThermoFlash LX-26 User Manual – English V 15 200904 6
Oral thermometry
Oral temperature is easily influenced by recent ingestion of food or
drinks and by breathing through the mouth To measure oral
temperature, the mouth must remain closed and the tongue
lowered for three to four minutes which is a difficult task for young
children to accomplish
Axillary (armpit) temperature
Although it may be easy to measure axillary temperature, it has
been proven that it does not provide an accurate measurement of
the child’s internal temperature To take this type of temperature,
the thermometer must be wedged tightly over the axillary artery
Despite the low sensitivity and relative inaccuracy of axillary
temperature in detecting fever, this method is recommended by
The
American Academy of Pediatrics
as a screening test for fever in
newborns
Tympanic thermometry
In order to obtain a precise temperature reading, good command of
the measurement technique is required The thermometer probe
must be placed as close as possible to the warmest part of the
external ear canal An incorrectly placed probe could lead to a false
temperature reading
NORMAL TEMPERATURES ACCORDING TO MEASUREMENT MET OD
M
EASUREMENT
M
ET OD
N
ORMAL TEMP
°
R
ECTAL
36 6°C
–
38°C
O
RAL
35 5°C
–
37 5°C
A
XILLARY
34 7°C
–
37 3°C
A
URICULAR
35 8°C
–
38°C
T
EMPORAL
(T
ERMO
F
LAS
)
35 8°C
–
37 8°C
The temperature of the human body varies throughout the day It
can also be influenced by numerous external factors: age, sex, type
and thickness of skin…

ThermoFlash LX-26 User Manual – English V 15 200904 7
ADVANTAGES OF TEMPORAL ARTERY
(
TA
)
TEMPERATURE
Infrared arterial temperature can be measured using a device
placed on the forehead, in the temporal artery region It has been
demonstrated that this relatively new method of measuring
temperature is more precise than tympanic thermometry and better
tolerated than rectal thermometry
The Thermo
Flash
LX-26 thermometer has been designed to
produce an instant forehead temperature reading without any
contact with the temporal artery As this artery is quite close to the
surface of this skin and therefore accessible and given the blood
flow is permanent and regular, it allows precise measurement of the
temperature This artery is linked to the heart by the carotid artery
which is directly linked to the aorta It forms part of the main trunk
of the arterial system The efficiency, speed and comfort of taking a
temperature from this area make it ideal compared with other
temperature measurements methods
NORMAL TEMPERATURES ACCORDING TO AGE
Age °C °F
0-2 years 36 4-38 0 97 5-100 4
3-10 years 36 1-37 8 97 0-100 0
11-65 years 35 9-37 6 96 6-99 7
> 65 years 35 8-37 5 96 4-99 5
P
RACTICAL CONSIDERATIONS W EN TAKING A TEMPERATURE
-In order to ensure that precise and accurate temperature
measurements are obtained, it is essential that each user has
received adequate information on and training in the temperature
measurement technique when using such a device
-It is essential to remember that although procedures such as
taking a temperature may be simple they must not be trivialised
-Temperature should be taken in a neutral context The patient
must not have undertaken vigorous physical activity prior to

ThermoFlash LX-26 User Manual – English V 15 200904 8
taking his/her temperature and the room temperature must be
moderate
-Be aware of physiological variations in temperature which must
be taken into consideration when evaluating the results:
temperature increases by 0 5C° between 6am and 3pm Women
have a temperature that is higher, on average, by around 0 2C°
Their temperature also varies in accordance with their ovarian
cycle It rises by 0 5C° in the second half of the cycle and at the
early stages of pregnancy
-When sitting, temperature is lower by about 0 3 to 0 4C° than
when standing
OW TO TAKE A TEMPERATURE
Aim at the FORE EAD, over
the right temporal region,
from a distance of about
5cm, press the thermometer’s
measurement button and the
temperature is instantly
displayed
The reliability of the
measurement cannot
be guaranteed if the
temperature is measured over
another part of the body (e g
arm, torso…)
C
ONSTRAINTS
Please observe the following before any temperature measurement
to ensure a stable and reliable result:
- Push back hair from the forehead
- Wipe away any perspiration from the forehead
- Avoid any drafts (e g from nasal specs, air conditioning…)
- Allow a 1 minute interval between two measurements
- Each time there is a significant change in the ambient
temperature due to a change in environment, to allow the
Thermo
Flash
LX-26 to acclimatise to this ambient
temperature for at least 15 minutes before using it

ThermoFlash LX-26 User Manual – English V 15 200904 9
V DESCRIPTION OF THE ThermoFlash LX-26

ThermoFlash LX-26 User Manual – English V 15 200904 10

ThermoFlash LX-26 User Manual – English V 15 200904 11
VI FUNCTIONS
1 Specially designed to take the body temperature of a person
regardless of the room temperature
2 Quick and reliable results as it use the HEIMMANN infrared
detection system
3 Sound alarm if temperature is exceeded
4 Memorization of the 32 last measures
5 LCD back-lighted digital screen
6 Data displayed in Celsius or Fahrenheit
7 Automatic stop (energy saver)
8 Small, convenient, easy to use
A
DDITIONAL USES
:
ThermoFlash LX-26 can also be used to measure the temperature of
a baby-bottle or bath, or room temperature (by using the Surface
Temp function) This function is in accordance with the Directive
89/336/EEC Electromagnetic Compatibility
VII USE
1 Install battery
2 For the first use or when inserting
new batteries wait between 10
minutes for the warm-up of the
apparatus and when inserting the
new batteries
3 Aim towards the forehead (see the
diagram below for the ThermoFlash
positioning), from a distance of 5 cm
(2 in), press the measuring key, the
temperature is displayed in 1 second
The temperature can also be taken
behind the ear lobe
4 Before taking the temperature,
make sure to remove hair and
perspiration from the forehead
You can also take the temperature
behind the ear lobe

ThermoFlash LX-26 User Manual – English V 15 200904 12
VIII CONFIGURATION AND FUNCTION OF MENU
1. C
OOSING T E TEMPERATURE UNIT
–
F1
F
UNCTION
Press PROG SETTING button for 3 seconds, the screen displays: F1
Select ‘-’ for degrees Celsius, ‘+’ for degrees Fahrenheit
Confirm by pressing PROG SETTING button
2. A
LARM SETUP
–
F2
M
ENU
Press PROG SETTING button for 3 seconds, the screen displays: F1
Press twice PROG SETTING button to get F2
Select ‘+’ to increase the threshold by 0 1°C (0 1°F), ‘-’ to reduce it
by 0 1°C (0 1°F)
Confirm by pressing PROG SETTING button
Note: The alarm threshold default value is 38°C (100.4
°F)
.

ThermoFlash LX-26 User Manual – English V 15 200904 13
3. C
OICE OF DISPLAY MODE
:
B
ODY OR
S
URFACE
T
EMP
–
F3
M
ENU
The ThermoFlash LX-26 is specially designed to take the body
temperature of a human being
For this, use the BODY mode
Measurement range for Body mode: 32°C – 42 9°C (90°F – 109°F)
You can also use the ThermoFlash LX-26 to measure the
temperature of an area or an object, a food, a liquid or a room
temperature
For this, use the SURFACE TEMP mode
Measurement range for Surface TEMP mode: 0°C – 60°C (32°F -
140°F)
Press PROG SETTING button for 3 seconds, the screen displays: F1
Press PROG SETTING button twice to get F3
Select 0 by pressing ‘-’ button to get the BODY mode
Select 1 by pressing ‘+’ button to get the SURFACE TEMP mode
Confirm by pressing PROG SETTING button
Note: The ThermoFlash LX-26 is automatically set to BODY
Important: The area temperature differs from the
internal body temperature To obtain the internal
temperature always use the BODY mode
Please make sure to select the BODY mode for an
internal temperature reading and the SURFACE TEMP
mode for an external area reading (bottle, bath,
room…)

ThermoFlash LX-26 User Manual – English V 15 200904 14
4. T
OTAL DIFFERENCE
–
F4
M
ENU
To adjust the total variation of your Thermoflash LX-26
Press PROG SETTING button for 3 seconds, the screen displays: F1
Press PROG SETTING button three times to get F4
Select ‘+’ to increase the difference by 0 1°C (0 1°F), ‘-’ to reduce it
by 0 1°C (0 1°F)
Confirm by pressing PROG SETTING button
If in doubt, you are advised to leave the total variance at +0 8°C
(original setting)
Each time there is a significant change in the ambient temperature
due to a change in environment, to allow the Thermo
Flash
LX-26 to
acclimatise to this ambient temperature for at least 15 minutes
before using it
5. B
UZZER
O
N
/O
FF
–
F5
M
ENU
Press PROG SETTING button for 3 seconds, the screen displays: F1
Press four times PROG SETTING button to get F5
Select ‘+’ to open the buzzer (a sound icon is displayed on the
LCD screen), press ‘-’ to stop it (the icon disappear)
Confirm by pressing PROG SETTING button
6. E
XITING T E SETTING MODE
Press PROG SETTING button until the screen turns off
7. D
ATA
M
EMORY
To display the last temperature measurement, press simultaneously
the ‘-’ and ‘+’ button
You will obtain the last temperature measurement To switch on the
last measurement, press the '+' button
The number indicated in interval of two measures corresponds to
the number of measurements
Press on the button '-' for review
To exit data memory, press the measure key The Thermoflash
switches off in 5 seconds

ThermoFlash LX-26 User Manual – English V 15 200904 15
8. C
ANGING T E BATTERIES
Display: when the LCD screen displays ‘Battery’, the battery is used
Operation: Open the lid and change the batteries, taking great care
with the correct positioning A mistake with this could cause damage
to the apparatus and compromise the guarantee of your
ThermoFlash LX-26
Never use rechargeable batteries Use only batteries for single
usage
IX TECHNICAL CHARACTERISTICS
1 Normal conditions of use
Operating temperature: 10° C
῀
40° C (50°
῀
140° F)
Humidity rate: ≤ 85%
2 Power: DC3 V (2 batteries AA)
3 Size: 196 x 150 x 50 mm – 7 7 x 5 9 x 1 9 in (L x W x H)
4 Weight: 220g
5 Display Resolution: 0 1° C (0 1°F)
6 Measuring range:
In body mode: 32°C – 42 9°C (90°F – 109°F)
In Surface Temp mode: 0°C
῀
60°C (32°F to 140°F)
7 Precision: From 36°C to 39°C (
96 8°F to 102 2°F
) = +/-0 2°C/F
8 Consumption: ≤ 50mW
9 Accuracy: ± 0 3° C (0 54° F)
10 Measuring distance: 5 cm – 8 cm (2 in – 3 14 in)
11 Automatic stop: 5 sec

ThermoFlash LX-26 User Manual – English V 15 200904 16
T ERMOFLAS PRECISION
From 34°C to 35 9°C =
±
0 3°C
From 93 2°F to 96 6°F =
±
0 3°F
From 36°C to 39°C =
±
0 2°C
From 96 8°F to 102 2°F =
±
0 2°F
From 39°C to 42 5°C =
±
0 3°C
From 102 2°F to 108 5°F =
±
0 3°F
According to ASTM Standard
E1965-1998 (2003)
The ThermoFlash LX-26 can take temperature readings below 32°C
or above 42 9°C (90°F to 109°F) but precision is not guaranteed
outside of this range
LONGEVITY USE
The ThermoFlash is guaranteed for 40 000 readings
X ADVICE
-The protective glass over the lens is the most important and
fragile part of the thermometer, please take great care of it
-Clean the glass with cotton fabric, wet with water or 70°
alcohol
-Do not use other batteries than mentioned batteries, do not
recharge non rechargeable batteries, do not throw in fire
-Remove the batteries when thermometer is not used for an
extended period of time
-Do not expose the thermometer to sunlight or water
-An impact will damage the product
XI ACCESSORIES SUPPLIED
User Manual in English
Fast learning user manual
Guarantee card
Carrying bag
Batteries supplied (AA)

ThermoFlash LX-26 User Manual – English V 15 200904 17
XII TROUBLESHOOTING
If you have one of the following problems while using your
ThermoFlash, please refer to this breakdown service guide to help
resolve the problem If the problem persists please contact our
customer service at contact us@visiomed-lab com
T
HE SCREEN DISPLAYS TEMPERATURE SUPERIOR TO
95°
The temperature is in Fahrenheit change the measurement to
Celsius (by pressing the Program Setting button to get the F1
function key)
T
HE SCREEN DISPLAYS THE BODY TEMPERATURE INFERIOR TO
32°C
(89 6°F)
To take a body temperature the function key F3 must be on Body
mode If you’re on Surface Temp mode the 32°C (89 6°F)
temperature displayed is showing the external temperature that
your body releases
T
HE SCREEN DISPLAYS THE MESSAGE
HI
When using the ThermoFlash LX-26 the
message HI can show on the screen
The analysis is above the measurement range
selected, either superior to 42 9°C (109°F)in
Body Mode or superior to 60°C (140°F) in
Surface Temp Mode

ThermoFlash LX-26 User Manual – English V 15 200904 18
T
HE SCREEN DISPLAYS THE MESSAGE
LO
When using the ThermoFlash LX-26 the
message Lo can show on the screen
The temperature analyzed is under the
measuring range selected, either less than 32°C
(90°F) in Body Mode or less than 0°C (32°F) in
Surface Temp Mode
This message displays in various cases – please find below a list of
the main cases:
Reasons for LO
message display
Advice
Temperature reading hampered
by hair, perspiration…
Make sure that there is no
obstruction prior to taking a
temperature
Temperature hampered by an
air flux
Make sure there is no air flux as
this could interfere with the
infrared system
Temperature readings too close
together, the ThermoFlash did
not have the chance to boot
itself
Respect the pause of 15
seconds minimum between two
readings – 1 minute pause is
advised
The measuring distance is too
far
Please respect the measuring
distance (between 5 and 8 cm –
2 in and 3 14 in)
Other manuals for THERMOFLASH LX-26
2
Other VISIOMED Thermometer manuals

VISIOMED
VISIOMED ThermoFlash EVOLUTION PRO LX-261E User manual

VISIOMED
VISIOMED THERMOSOFT VM-DS400 User manual

VISIOMED
VISIOMED LX-361T User manual

VISIOMED
VISIOMED bewell connect MyThermo Pocket BW-MTX1 User manual

VISIOMED
VISIOMED easyScan VM-ZX1 User manual

VISIOMED
VISIOMED bewell connect MyThermo Pocket BW-MTX1 User manual

VISIOMED
VISIOMED EASYSCAN FTX3 User manual

VISIOMED
VISIOMED THERMOFLASH LX-260T EVOLUTION User manual

VISIOMED
VISIOMED easyScan VM-ZXP1 User manual

VISIOMED
VISIOMED ThermoFlash LX-26E EVOLUTION User manual