SECTION 4 –INPUT SPECIFICATIONS
Throughout this section, a voltage on a single terminal or probe input refers to the voltage between the specified
terminal or probe input and the 4700 chassis ground.
The accuracy specifications are valid for a period of 1 year at ambient temperatures within ±5C of calibration
temperature (add 0.004% of reading per C outside of this). The accuracy specifications include the measurement
uncertainties in the standards and methods employed by Vitrek during calibration of the 4700 and probes; to
obtain the accuracy relative to the standards and methods used in the actual calibration of the 4700 or probes
reduce all DC accuracy specifications by 0.0075% of reading and all AC accuracy specifications by 0.025% of
reading.
The accuracy specifications for probes assume the use of the probe with the specific 4700 used for its’ calibration.
When this is not the case, add 0.01% of reading to all probe accuracy specifications.
MAXIMUM INPUT VOLTAGES
HVL-150 Probe 140KVdc, 100KVrms (<65Hz), 150KVpk continuous, fully specified
6500KV.Hz max above 65Hz.
30KV per millisecond max signal slew rate above 50KV.
10% overload for <1 second without damage.
HVL-100 Probe 100KVdc, 75KVrms, 110KVpk continuous, fully specified
15% overload for <1s, no permanent damage
HVL-100G Probe 100KVdc, 110KVpk continuous, fully specified
15% overload for <1s, no permanent damage
100KV/ms max rate of change
HVL-70 Probe 70KVdc, 50KVrms, 75KVpk continuous, fully specified
15% overload for <1s, no permanent damage
HVL-70G Probe 70KVdc, 75KVpk continuous, fully specified
15% overload for <1s, no permanent damage
65KV/ms max rate of change
HVP-35/HVL-35 Probe 35KVdc, 30KVrms, 45KVpk continuous, fully specified
15% overload for <1s, no permanent damage
HVL-35G Probe 35KVdc, 45KVpk continuous, fully specified
15% overload for <1s, no permanent damage
50KV/ms max rate of change
DIRECT Terminal DC & ≥15Hz : 10KVdc, 10KVrms, 15KVpk continuous, fully specified
<15Hz : 7KVrms continuous, fully specified
20% overload for <1s, no permanent damage
COMMON Terminal 600Vdc, 600Vrms, 800Vpk continuous, fully specified
2KVdc, 2KVrms, 3KVpk for <1s, no permanent damage
Decrease maximum voltages by 1%/C above 35C ambient temperature and by 5%/1000ft above 5000ft ASL
altitude.