
The slew rate prompt allows the test developer to automatically set the ramp rate, so that the output voltage linearly increases in
amplitude up to the test voltage. The slew rate can be set from 1 volt per second to 2000 volts per second. If 100 V/S is the
desired ramp rate, then press enter. If another rate is desired, then use the number keys to input a new value and then press enter.
After the slew rate has been set, the 944i will prompt for a test frequency. This advanced feature, which enables the user to test
AC dielectric strength over a wide range of test frequencies, is unique to the Vitrek 944i. The frequency prompt is as follows:
SELECT TEST FREQUENCY
60 HZ 50 HZ _ _ _ HZ
Should the user desire a test frequency of 60 Hz, press enter. If 50 Hz is the desired test frequency, use the right cursor to
highlight the 50 Hz option and then press enter. When an alternate test frequency is required, use the cursors to highlight the _ _ _
Hz option and then use the number keys to fill in the frequency of interest and then press enter. The frequency range of the 944i is
from 50 to 100 hertz.
Following frequency selection, the current sense selection menu will appear as shown below:
SELECT CURRENT SENSE
REAL APPARENT TOTAL
The Vitrek 944i uses internal phase angle circuitry to determine if the leakage current being measured is a result of a resistive load
(real), reactive load (apparent) or both (total). In most applications total current is the desired sense mode, so "total" is highlighted
when the current sense prompt pops-up. To select "total" press the enter key.
In many applications involving devices with EMI filters, the current being drawn by the filter capacitor is of no interest to the
user. To ignore the current flow caused by the reactive component, select the "real" current sense mode. To determine the
amount of current flow caused by the reactive component, select "apparent" current sense.
Due to the nature of phase measurement, it is difficult to accurately determine the phase of low amplitude signals. To
prevent erroneous phase measurement, the 994i will default to "total" current sense when phase signals are below the
minimum threshold. Once the current sense mode has been selected, the 944i will prompt the user to set a current limit with
the following message:
ENTER MAX CURRENT
10.0 mA
This value sets the maximum amount of leakage current which the 944i will source without failing the DUT and shutting down the
test. Any current which exceeds this limit will cause the unit to automatically stop the test. The default value is 10.0 mA. Any
value between 1nA and 35mA may be used. To change the value of the current limit, use the number keypad. To change the
units from m (milli) to u (micro) or n (nano), use the up/down cursor arrows. Once the desired maximum current limit is
displayed, press enter. After the maximum current limit is set the next prompt is:
ENTER MIN CURRENT
0.0_ mA
The minimum current limit is provided as a convenience to assure that the load is actually connected. Any current measured
below this limit will cause the unit to automatically stop the test. The default value is 0.0 mA. When the minimum current limit
is set to zero, the minimum current limit becomes inactive. Any value between 0 nA and 35 mA may be used as a minimum
current limit. To change the value of the current limit, use the number keypad. To change the units from m (milli) to u (micro) or
n (nano), use the up/down cursor arrows. Once the desired current limit is displayed, press enter. The next prompt is:
ENTER TEST TIME
10.0 SECONDS
The test time is the amount of time that the 944i will hold the test voltage once it has arrived at that voltage. This period is often
referred to as dwell time. During the dwell time the unit continuously monitors leakage current against both min and max current
limits, while holding the output constant. If at any time during either the ramp or dwell period, a current limit is exceeded, the test
is aborted and the DUT is failed. Use the number keypad to overwrite the default test time with the desired time and then press
enter. This completes the test parameter selection for automatic AC testing. The next prompt is as follows:
SELECT MODE
RUN ADD CHANNL STORE