VWR 5000I User manual

EN - English ......... 1
FR - Français ......... 15
ES - Español ......... 31
IT - Italiano ......... 47
DE - Deutsch ......... 63
PT - Português ......... 79
NL - Nederlands ......... 95
NO - Norsk ......... 99
DA - Dansk .........103
SV - Svenska .........107
FI - Suomi .........111
HU - Magyar .........115
PL - Polski .........119
Revision 6
29/12/15
Instruction Manual
Incubating Orbital Shakers
European Catalogue Numbers:
Large Capacity Shaker
EU/UK/CH Plug: 444-7084
Model 5000I Incubating Orbital Shaker
EU/UK/CH Plug: 444-7109
Model 5000IR Incubating/Refrigerating Orbital Shaker
EU/UK/CH Plug: 444-7112

1
Package contents
Incubating Orbital Shaker
Non-skid rubber mat
234cm detachable power cord
Instruction manual
Warranty
VWR International warrants that this product will be free from defects in material and
workmanship for a period of two (2) years from date of delivery. If a defect is present,
VWR will, at its option and cost, repair, replace, or refund the purchase price of this
product to the customer, provided it is returned during the warranty period. This war-
ranty does not apply if the product has been damaged by accident, abuse, misuse,
or misapplication, or from ordinary wear and tear. If the required maintenance and
inspection services are not performed according to the manuals and any local regula-
tions, such warranty turns invalid, except to the extent, the defect of the product is not
due to such non-performance.
Items being returned must be insured by the customer against possible damage or
loss. This warranty shall be limited to the aforementioned remedies. IT IS EXPRESSLY
AGREED THAT THIS WARRANTY WILL BE IN LIEU OF ALL WARRANTIES OF
FITNESS AND IN LIEU OF THE WARRANTY OF MERCHANTABILITY.
Compliance with local laws and regulations
The customer is responsible for applying for and obtaining the necessary regulatory
approvals or other authorizations necessary to run or use the Product in its local envi-
ronment. VWR will not be held liable for any related omission or for not obtaining the
required approval or authorization, unless any refusal is due to a defect of the product.
table of contents
Package Contents . . . . . . . . . . . . 1
Warranty . . . . . . . . . . . . 1
Installation . . . . . . . . . . . . 2
Maintenance & Servicing . . . . . . . . . . . . 2
Intended Use . . . . . . . . . . . . 2
Equipment Disposal . . . . . . . . . . . . 2
Environmental Conditions . . . . . . . . . . . . 2
Safety Instructions . . . . . . . . . . . . 3
Warning . . . . . . . . . . . . 3
Standards & Regulations . . . . . . . . . . . . 3
Control Panel . . . . . . . . . . . . 4
Specifications . . . . . . . . . . . . 5-7
Operating Instructions . . . . . . . . . . . . 8-11
Technical Service . . . . . . . . . . . . 11
Troubleshooting . . . . . . . . . . . . 12
Accessories . . . . . . . . . . . . 13-14
EC Declaration of Conformity . . . . . . . . . . . . 123-124
EN
Legal Address of Manufacturer: VWR International bvba • Researchpark Haasrode 2020 • Geldenaakesbaan 464 • B-3001 Leuven • + 32 16 385011 • http://be.vwr.com
Country of Origin: United States

2
InstallatIon
Upon receiving the VWR Incubating Orbital Shaker, check to ensure that no dam-
age has occurred during shipment. It is important that any damage that occurred in
transport is detected at the time of unpacking. If you do find such damage the carrier
must be notified immediately.
After unpacking, it requires two (2) people to lift the Incubating Orbital Shaker, from
the bottom, to place on a level bench or table, away from explosive vapors. Ensure
that the surface on which the unit is placed will withstand typical heat produced by the
unit. Always place the unit on a sturdy work surface.
The Incubating Orbital Shaker is supplied with a power cord that is inserted into the
IEC connector on the back of the unit first, then it can be plugged into a properly
grounded outlet. The 230V unit plugs into a 230 volt, 50/60 Hz source.
MaIntenance & servIcIng
The Incubating Orbital Shaker is built for long, trouble-free, dependable service.No lubri-
cation or other technical user maintenance is required. However at least every three
(3) months you should:
• Unplug the unit.
• Remove any accumulated dirt from the base and tray.
• Check all accessible items to make sure they are properly tightened.
The unit should be given the care normally required for any electrical appliance. Avoid
wetting or unnecessary exposure to fumes. Spills should be removed promptly. DO
NOT use a cleaning agent or solvent on the front panel or lid which is abrasive or
harmful to plastics, nor one which is flammable. Always ensure the power is discon-
nected from the unit prior to any cleaning. If the unit ever requires service, contact your
VWR representative.
Intended Use
The Incubating Orbital Shaker is designed for a variety of shaking applications.
eqUIPMent dIsPosal
This equipment must not be disposed of with unsorted waste. It is
your responsibility to correctly dispose of the equipment at life-cycle-
end by handing it over to an authorized facility for separate collection
and recycling. It is also your responsibility to decontaminate the
equipment in case of biological, chemical and/or radiological con-
tamination, so as to protect the persons involved in the disposal and
recycling of the equipment from health hazards.
For more information about where you can drop off your waste of equipment, please
contact your local dealer from whom you originally purchased this equipment. By doing
so, you will help to conserve natural and environmental resources and you will ensure
that your equipment is recycled in a manner that protects human health.
envIronMental condItIons
Operating Conditions: Indoor use only.
Temperature: 15 to 32°C
Humidity: maximum 80% relative humidity, non-condensing
Altitude: 0 to 2000 M above sea level
Non-Operating Storage:
Temperature: -20 to 65°C
Humidity: maximum 80% relative humidity, non-condensing
Installation Category II and Pollution Degree 2 in accordance with IEC 664.
EN

3
safety InstrUctIons
Please read the entire instruction manual before operating the Incubating Orbital
Shaker.
WARNING! DO NOT use the Incubating Orbital Shaker in a hazardous atmo-
sphere or with hazardous materials for which the unit was not designed. Also,
the user should be aware that the protection provided by the equipment may
be impaired if used with accessories not provided or recommended by the
manufacturer, or used in a manner not specified by the manufacturer.
Exercise extreme caution when moving the unit, there is a risk of pinching or
crushing fingers.
Always operate unit on a level surface for best performance and maximum
safety.
DO NOT lift unit by the tray, front bezel or lid.
CAUTION! To avoid electrical shock, completely cut off power to the unit by
disconnecting the power cord from the unit or unplug from the wall outlet.
Disconnect unit from the power supply prior to maintenance and servicing.
Spills should be removed promptly. DO NOT immerse the unit for cleaning.
DO NOT operate the unit if it shows signs of electrical or mechanical damage.
CAUTION! The caution hot indicator light warns that the temperature of the air in
the chamber is above 40°C. The light will illuminate and remain lit when the
temperature of the air in the chamber reaches approximately 40°C. When the
heat is turned off, the caution hot indicator light will stay lit until the temperature
of the air in the chamber is less than 40°C.
Earth Ground - Protective Conductor Terminal
Alternating Current
standards & regUlatIons
VWR international hereby declares under its sole responsibility that the products
comply with the requirements of the following directives and associated standards:
Associated EU directives:
EMC directive 2004/108/EC
LVD directive 2006/95/EC
ROHS directive 2011/65/EU
WEEE directive 2002/96/EC
Safety standards:
EN 61010
EN 61010-2-051
EN 61010-2-010
UL 61010-1
CAN/CSA C22.2 No. 61010-1-04
EMC standards:
EN 45501 EN 61000-4-3
EN 55022A EN 61000-4-4
EN 6100-3-2 EN 61000-4-5
EN 6100-3-3 EN 61000-4-6
EN 61000-4-1 EN 61000-4-11
EN 61000-4-2
EN

4
The front panel of the Incubating Orbital Shaker contains all the controls and displays
needed to operate the unit.
A. Standby button/standby indicator light: The standby indicator light will illuminate
when the unit is plugged in. The unit will be in standby mode. Press the standby
button to activate the temperature, speed and time functions. The standby indicator
light will shut off and the temperature, speed and time displays will illuminate. Press
the stand-by button again and the unit will once again be in standby mode.
B. Temperature display: Displays the actual/set-point temperatures in conjunction
with the actual/set-point indicator lights. C. Up/down arrows for set-point control.
On/off button starts/stops the heating function. D. The heat indicator light will be
illuminated when the unit is heating.
E. Actual indicator light: Illuminates when the temperature displayed is the actual
temperature of the air in the chamber.
F. Set-point indicator light: Illuminates when the set-point temperature is displayed.
G. Caution hot indicator light: Illuminates when the air temperature of the chamber
is above 40°C (104°F).
H. Speed display: Displays the speed of the shaker. I. Up/down arrows for set-point
control. On/off button starts/stops shaking function. J. The speed indicator light will
be illuminated when the unit is shaking.
K. Time display: Displays accumulated time (continuous mode) or how much time
is remaining (timed mode). The display range is from 0 to 9999 minutes in one (1)
second increments. The display will indicate minutes and seconds until the timer
reaches 99 minutes and 59 seconds (99:59), then the display will automatically
display minutes up to 9999. L. Up/down arrows for set-point control. On/off button
starts/stops the time function.
control Panel
H.
L.I.
D.
A.
K.
C.
J.
E.
F.
G. B.
EN

5
Overall dimensions (L x W x H): 64,8 x 35,6 x 40,6cm
Interior dimensions (L x W x H): 34 x 30,5 x 24,1cm
Tray dimensions (L x W): 33 x 27,9cm
Electrical (50/60 Hz): 230 volts, 5 amps, 450 watts
Fuses: 5mm x 20mm, 5 amp quick acting
Temperature range: ambient +5°C to 65°C
Temperature uniformity: ±0,5°C at 37°C
Speed range: 15 to 500rpm
Speed accuracy:
Above 100rpm ±1% of set speed
Below 100rpm ±1rpm
Timer: 1 second to 9999 minutes
(increased in 1 second increments)
Orbit: 19mm
Maximum weight capacity: 15,9kg
Controls: see page 4
Tray material: aluminum
Ship weight: 37,7kg
European Catalogue Number: EU/UK/CH plug: 444-7084
sPecIfIcatIons - large caPacIty IncUbatIng shaker
EN

6
Overall dimensions (L x W x H): 81,5 x 67,6 x 59,7cm
Interior dimensions (L x W x H): 52,3 x 63 x 43,2cm
Tray dimensions (L x W): 45,7 x 45,7cm
Electrical (50/60 Hz): 230 volts, 8 amps, 800 watts
Fuses: 5mm x 20mm, 8 amp quick acting
Temperature range: ambient +5°C to 65°C
Temperature uniformity: ±0,5°C at 37°C
Speed range: 15 to 500rpm
Speed accuracy:
Above 100rpm ±1% of set speed
Below 100rpm ±1rpm
Timer: 1 second to 9999 minutes
(increased in 1 second increments)
Orbit: 25mm
Maximum weight capacity: 22,7kg
Controls: see page 4
Tray material: aluminum
Ship weight: 97,5kg
European Catalogue Number: EU/UK/CH plug: 444-7109
sPecIfIcatIons - Model 5000I IncUbatIng shaker
EN

7
sPecIfIcatIons - Model 5000Ir IncUbatIng/refrIgeratIng shaker
EN
Overall dimensions (L x W x H): 104,4 X 67,6 X 59,7cm
Interior dimensions (L x W x H): 52,3 x 63 x 43,2cm
Tray dimensions (L x W): 45,7 x 45,7cm
Electrical (50/60 Hz): 230 volts, 10 amps, 800 watts
Fuses: 5mm x 20mm, 10 amp quick acting
Temperature range: ambient -15°C to 65°C
Temperature uniformity: ±0,5°C at 37°C
Speed range: 15 to 500rpm
Speed accuracy:
Above 100rpm ±1% of set speed
Below 100rpm ±1rpm
Timer: 1 second to 9999 minutes
(increased in 1 second increments)
Orbit: 25mm
Maximum weight capacity: 22,7kg
Controls: see page 4
Tray material: aluminum
Ship weight: 117,9kg
European Catalogue Number: EU/UK/CH plug: 444-7112

8
Please follow the procedure below if setting the 5000IR unit’s temperature at/near the
room’s ambient temperature.
For optimum temperature control in the chamber of the 5000IR unit, the upper ambient
control limit is to be set a minimum of 3°C above the measured room temperature.
Example: Measured (ambient) temperature is 22°C, the upper ambient control limit
should be set at a minimum of 25°C.
With the uinit at the heat/cool setting as described below, and in cool mode (“cool”
on time display), a temperature setting is displayed in the temperature window. The
default temperature setting of the 5000IR is 28°C. This is the factory set upper ambient
control limit. This setting can be adjusted between 20°C and 32°C by the end user.
• Any unit set point above the upper limit setting will heat only
• Any unit set point below 20°C will cool only
• Any set point between 20°C and the selected upper limit setting, up to 32°C,
the chiller and heater will both control chanber temperature
Heat/Cool Setting Procedure to Adjust Upper Ambient Control Limit
1. Put unit in standby mode.
2. Press and hold the speed down and time down time arrows simultaneously, now
press the standby button. Release all buttons.
3. Time display will not read “cool” and the temperature display will read 28°C. If
the time display reads “heat” press the time up/down arrow until the word “cool”
appears in the display.
4. Now you can set the upper ambient control limit. Using the temperature up/
down arrows adjust this temperature to be 3°C above the measured ambient
temperature.
5. Once the desired temperature is selected, press the standby button to return to
normal operation.
6. This procedure does not require recalibrating the unit.
Additional 5000IR Notes:
• 5000IR units have a toggle switch on the left side. This switch must be in the
“on” position (depress “I”) for the unit to function properly.
• For 5000IR units, make sure the chiller’s hose is positioned to drain freely (into
a sink when possible).
• Do not allow the end of the chiller hose to become submerged.
• If chiller hose is positioned in a drain container, the container should have a
volume of at least 2 gallons if the unit will be unattended for 48 hours.
5000Ir UPPer aMbIent control lIMIt
EN

9
The Incubating Orbital Shaker has been designed for the temperature, speed and
time functions to work independently of one another. The temperature and speed can
be reset without resetting the timer and the timer can be stopped and started without
interrupting the heating and shaking functions.
1. Getting ready:
a. Plug the power cord into a properly grounded outlet. The standby indicator light
will illuminate, verifying power to the unit.
b. Press the standby button to move the unit from standby mode. The standby
indicator light will turn off and the temperature, speed and time displays will
illuminate, displaying the previously used settings.
Note: 5000IR units have a toggle switch on the left side. This switch must be in the
“on” position (depress “I”) for the unit to function properly.
Note: See the procedure on page 8 for 5000IR units that will have the chamber
temperature set at/near the ambient temperature.
2. Setting temperature:
a. Press the up/down arrows below the temperature display until you reach the
desired temperature. When you release the button, the display will blink off and
then on indicating the new set temperature has been accepted.
b. Press the on/off button to start the heating function. The indicator light below the
temperature display will illuminate to indicate the heating function is in use and
remain lit until heating has ceased.
c. Temperature adjustments can be made without interrupting heating by using
the up/down arrows below the temperature display. After the change has been
made and you release the button, the display will blink off and then on indicating
the new set temperature has been accepted.
d. To stop the heating function, press the on/off button below the temperature
display. The heat indicator light will turn off.
CAUTION HOT indicator:
The caution hot indicator light warns that the temperature
of the air in the chamber is above 40°C. The light will
illuminate and remain lit when the temperature of the
air in the chamber reaches approximately 40°C. When
the heat is turned off, the caution hot indicator light will
stay lit until the temperature of the air in the chamber is
less than 40°C.
3. Setting speed:
a. Press the up/down arrows below the speed display until you reach the desired
speed. When you release the button, the display will blink off and then on indi-
cating the new set speed has been accepted.
b. Press the on/off button to start the shaking function. The indicator light below
the speed display will illuminate and blink until the set-point is reached. Once
the set-point is reached the light will stop blinking and remain lit until shaking
has ceased. The micro-processor controlled ramping feature slowly increases
speed until the set-point is reached which helps to avoid splashing, and provides
excellent low end control.
c. Speed adjustments can be made without interrupting shaking by using the up/
down arrows below the speed display. After the change has been made and you
release the button, the display will blink off and then on indicating the new set
speed has been accepted.
d. To stop the shaking function, press the on/off button below the speed display.
The speed indicator light will turn off.
oPeratIng InstrUctIons
EN

10
6. Turning unit off:
a. To turn the unit off, press the standby button. The temperature, speed and time
displays will be blank, the standby indicator light will illuminate. The Incubating
Orbital Shaker should be kept in standby mode when not in use. To completely
cut off power to the unit, disconnect the power cord from the unit or unplug from
the wall outlet.
OPERATING TIPS
Opening the lid on the Incubating Orbital Shaker will cause the unit to pause shaking
and/or heating. Close the lid and the unit will resume shaking and/or heating at cur-
rent settings.
Centering your sample and even weight distribution on the tray helps with
balance and stability.
As a safety feature, a built-in program will shut power off to the motor if the tray is pre-
vented from rotating, or the unit is overloaded beyond its recommended weight capacity.
The shaker will automatically restart after a power interruption. Built-in memory main-
tains the last used temperature, speed and time settings during a power interruption.
LOAD SENSING FUNCTION
The Incubating Orbital Shaker is equipped with a load sensing function that can be
activated by the user. This function provides protection against improper positioning
of load and maximum load being exceeded. When activated, the unit will automatically
sense improper load conditions and slow to a safe running speed, then display that
speed followed by an E04 error message on the speed display. The unit will also beep
three (3) times every 60 seconds until the error is reset by pressing the speed on/off
button. To activate the load sensing function use the following steps:
1. Place the unit in standby mode.
2. Press and hold the speed on/off button and press the standby button. The unit will
beep two (2) times, confirming the function is enabled.
4. Setting time to zero (0:00) and continuous mode: Accumulated time.
a. Press and hold the on/off button below the time display. After three (3) seconds
the display will indicate the previous set time.
b. Simultaneously press both the up and the down arrows, the display will indicate
zero (0:00). The unit time is now set to zero (0:00) minutes. Alternately, you can
use the up/down arrows to get to zero (0:00).
c. Press the on/off button below the time display. The display will indicate accu-
mulated time. The up/down arrows will become inactive. To stop timer, press the
on/off button again. IMPORTANT: This will NOT interrupt the shaking function.
Press the on/off button below the speed display to interrupt the shaking function.
d. To reset, press and hold the on/off button below the time display. After three (3)
seconds the display will indicate the previous set time, which was zero (0:00).
5. Setting timed mode: Programmed time.
a. Press the up/down arrows below the time display until you reach the desired
time.
b. Start this function by pressing the on/off button below the time display. The unit
will run for the selected time, the up/down arrows will become inactive while
the timer is running. The unit will stop shaking when time display reaches zero
(0:00). Four (4) audible beeps will indicate the count down function is complete.
The time display will default back to the set time. To repeat for the same time,
simply press the on/off button again.
c. To interrupt an automatic timing cycle before it is completed, press the on/
off button below the time display. The display will flash off and on to indicate
the time function is on “hold”. IMPORTANT: This will NOT interrupt the shak-
ing function. Press the on/off button below the speed display to interrupt the
shaking function. Restart the timer by pressing the on/off button below the
time display. Unit will continue counting down to zero (0:00). When the display
reaches zero (0:00), you will hear the four (4) audible beeps that indicate the
count down function is complete and the shaking function will cease.
oPeratIng InstrUctIons (cont’d)
EN

11
This process may be repeated at the same set-point, multiple times for fine tuning if
desired.
The unit will now use the biased offset for that specific temperature setting and
increase or decrease temperature accordingly to bring the temperature to set
temperature. The decimal point of the display will flash to indicate a biased off-
set is being used. All other temperature settings will use the standard internal
calibration. This offset will be stored in memory and retained until reset.
To restore unit to factory setting:
Press and hold the standby button while pressing the temperature down arrow once.
The reset will be confirmed with two (2) audible beeps. Press the standby button to exit
calibration mode and return to normal heating.
Speed Calibration Procedure:
This procedure is used to self calibrate the incubating shakers. The tray should be free
of any samples, vessels, and accessories prior to calibrating.
1. Turn unit on. Speed and time displays will be illuminated.
2. Press and hold the standby button and momentarily press the speed on/off button.
The speed display should read “CAL”.
3. The unit will be run for approximately one (1) minute and automatically calibrate.
RS-232 SERIAL PORT
The RS-232 serial port provides two-way communications for data logging and unit
control.
3. To restore normal operation, remove AC power to the unit for ten (10) seconds and
then restore.
If the E04 error occurs be sure the load is within the maximum specification and properly
balanced (centered on tray) and/or reduce sample size/speed before restarting the
unit. If the E04 error occurs due to acceptable sample vibration or another vibration
source, the vibration sensing function can be disabled as described above.
BEEPER PREFERENCE (muting audible alarm)
To silence beeper operation (except for error codes), with the unit in standby mode,
press and hold the time on/off button and press the standby button. Release the
standby button first, and then release the on/off button. To restore normal beeper
operation, remove AC power to unit for ten (10) seconds and then restore.
SINGLE POINT CALIBRATION PROCEDURE
This procedure is used to fine tune and calibrate the Incubating Orbital Shaker at a
specific temperature setting. This process may be repeated for up to three (3) sepa-
rate set-points. If a fourth calibration set-point is entered, the first set-point entered
will be overwritten.
1. Turn unit on.
2. Set desired temperature.
3. Stabilize one (1) hour or more, measuring the temperature with a temperature
probe/thermometer.
4. Press and hold standby button, then press the temperature up arrow once. The
unit will beep two (2) times, confirming calibration mode. The display will now be
flashing.
5. Press the temperature up/down arrows until the display matches the temperature
probe/thermometer.
6. Press standby button to exit calibration mode and return to normal heating.
oPeratIng InstrUctIons (cont’d)
technIcal servIce
For more information or technical assistance contact your local VWR representative
or visit www.vwr.com.
Web resources: Visit VWR’s website at www.vwr.com for:
• Complete technical service contact information
• Access to VWR’s Online Catalogue, and information about accessories and
related products
• Additional product information and special offers
EN

12
Problem Cause Solution
Unit fails to power on Missing or blown fuse Check and replace fuse if necessary
If problem persists, please contact your VWR representative for repair.
Unit is excessively noisy Rattling or ticking sounds may indicate a loose screw
on the tray
Ensure that tray is secured tightly.
If problem persists, please contact your VWR representative for repair.
Unit not shaking at proper speed -Perform speed calibration on page 11
If problem persists, please contact your VWR representative for repair.
Unit not heating to proper temperature -Perform single point calibration on page 11
If problem persists, please contact your VWR representative for repair.
E1 RTD open or ±1ºC temperature deviation from set-
point (after unit has stabilized)
This error connot be fixed by the end user.
If problem persists, please contact your VWR representative for repair.
E2 RTD shorted or temperature below 0ºC This error cannot be fixed by the end user.
If problem persists, please contact your VWR representative for repair.
E3
Mechanical obstruction
Drive system failure
Ceased bearing
Drive belt broken
Remove mechanical obstruction.
If problem persists, the reason may be the drive system. Please contact your VWR repre-
sentative for repair.
E4
Improper positioning of load
or
Maximum load exceeded
Ensure the load is evenly distributed and does not exceed the maximum load capacity for
the unit. See “Load Sensing Function” on page 10-11.
Verify load has not exceeded the maximum allowable weight. Reduce load if necessary.
If error code resumes, contact your VWR representative for repair.
E6 Over temperature error This error cannot be fixed by the end user.
Please contact your VWR representative for repair.
E7 Default sensor Error (5000IR units only)
Error will shut unit off if set temperature is 20°C or below. Unit is still operational with a set
temperatue above 20°C. Error will be indicated every 5 minutes until repaired.
Please contact your VWR representative for repair.
troUbleshootIng
To clear error press the standby button
EN

13
EN
accessorIesaccessorIes
PLATFORMS
Platform can be used for mounting Erlenmeyer flask/media bottle clamps and test tube racks. Slides over-top included tray.
Tighten with adjustment screws. Stainless steel construction.
descrIPtIon dIMensIons (l xW) cat. no.
Platform 33 x 28cm 444-7087
Platform 47,5 x 47,5cm 444-0697
33 x28cM 47,5 x47,5
tUbe PlatforM PlatforM
descrIPtIon caPacIty caPacIty caPacIty cat. no.
Half Size Test Tube Racks
10 to 13mm Test Tube Rack 63 2 4 444-0278
14 to 16mm Test Tube Rack 48 2 4 444-0279
18 to 20mm Test Tube Rack 35 2 4 444-0280
22 to 25mm Test Tube Rack 24 2 4 444-0281
15mL Centrifuge Tube Rack 35 2 4 444-0283
50mL Centrifuge Tube Rack 12 2 4 444-0284
1,5 to 2mL Micro-Tube Rack 70 2 4 444-0276
Full Size Stationary Test Tube Racks
10 to 14mm Test Tube Rack 48 N/A 3 444-7037
16 to 20mm Test Tube Rack 33 N/A 3 444-7038
21 to 25mm Test Tube Rack 21 N/A 3 444-7039
Full Size Pivoting Test Tube Racks
13mm Test Tube Rack 90 N/A 2 444-2955
16mm Test Tube Rack 60 N/A 2 444-2956
20mm Test Tube Rack 40 N/A 2 444-2957
25mm Test Tube Rack 24 N/A 2 444-2958
30mm Test Tube Rack 21 N/A 2 444-2959
TEST TUBE RACKS - Requires Platform accessory.
Installation of Half Size Test Tube Racks:
1. Unit should be in standby mode.
2. Remove non-skid rubber mat. Slide Platform over-top included tray.
3. Using the hardware provided, attach brackets to the Platform. Position brackets as
shown. The proper spacing between the brackets is 19,7cm.
4. Align the test tube rack side-to-side with the brackets on the Platform.
5. Press the test tube rack in place, you will hear the test tube rack clip into position. No
mounting screws are required.
Installation of Full Size Pivoting Test Tube Racks (Model 5000I/5000IR Only):
1. Unit should be in standby mode.
2. Line up the 4 openings on the bottom of the outer rack with the four compatible
openings on the Universal Platform. Secure with the four screws provided.
3. If a tilt angle is desired, loosen wing knobs, tilt inner holder and tighten knobs.
4. To maximize the platforms capacity for holding racks in the stationary position, remove
the outer holder and secure the inner holder to the Universal Platform using the four
openings on the bottom of the inner holder.
Installation of Full Size Stationary Test Tube Racks (Model 5000I/5000IR Only):
1. Line up the two openings on the sides of the test tube rack with the two compatible
openings on the Universal Platform.
2. Place the washer on top of the opening on the test tube rack, place the clamp knob
through the washer and rack and hand tighten the clamp knob to the platform.

14
EN
accessorIes
33 x 28cM 47,5 x47,5
descrIPtIon PlatforM caPacIty PlatforM caPacIty cat. no.
Stainless Steel Flask Clamps
10mL Erlenmeyer Flask Clamp 60 113 444-7040
25mL Erlenmeyer Flask Clamp 25 64 444-7041
50mL Erlenmeyer Flask Clamp 13 32 444-7042
125mL Erlenmeyer Flask Clamp 10 20 444-7043
250mL Erlenmeyer Flask Clamp 9 20 444-7044
500mL Erlenmeyer Flask Clamp 7 13 444-7045
1L Erlenmeyer Flask Clamp 4 8 444-7030
2L Erlenmeyer Flask Clamp N/A 5 444-7031
2.8L Erlenmeyer Flask Clamp N/A 2 444-7032
4L Erlenmeyer Flask Clamp N/A 4 444-7033
6L Erlenmeyer Flask Clamp N/A 2 444-7034
FLASK CLAMPS - Requires Platform accessory
33 x 28cM 47,5 x47,5
descrIPtIon PlatforM caPacIty PlatforM caPacIty cat. no.
Stainless Steel Media Bottle Clamps
500mL Media Bottle Clamp 5 16 444-7035
1L Media Bottle Clamp 2 10 444-7036
PVC Flask Clamps
125mL Erlenmeyer Flask Clamp 10 20 444-2960
250mL Erlenmeyer Flask Clamp 8 18 444-2961
500mL Erlenmeyer Flask Clamp 5 12 444-2962
1L Erlenmeyer Flask Clamp 2 8 444-2963
2L Erlenmeyer Flask Clamp N/A 4 444-2964
Installation of Flask Clamps:
1. Unit should be in standby mode.
2. Remove non-skid rubber mat. Slide Platform over-top included tray.
3. Line up the hole(s) in the flask clamp to the hole(s) in the Platform. Using the screw(s) provided, hand tighten into place using a flathead screwdriver. DO NOT over tighten.
4. Insert flask/media bottle into clamp and wrap the spring around the neck of the clamp to hold secure.
accessorIesaccessorIes
PLATFORMS
Platform can be used for mounting Erlenmeyer flask/media bottle clamps and test tube racks. Slides over-top included tray.
Tighten with adjustment screws. Stainless steel construction.
descrIPtIon dIMensIons (l xW) cat. no.
Platform 33 x 28cm 444-7087
Platform 47,5 x 47,5cm 444-0697

EN - English ......... 1
FR - Français ......... 15
ES - Español ......... 31
IT - Italiano ......... 47
DE - Deutsch ......... 63
PT - Português ......... 79
NL - Nederlands ......... 95
NO - Norsk ......... 99
DA - Dansk .........103
SV - Svenska .........107
FI - Suomi .........111
HU - Magyar .........115
PL - Polski .........119
Révision 6
29/12/15
Manuel d’Instruction
Agitateur Incubateur Orbital
Les numéros de catalogue européenne:
Grande capacité Shaker
Prise EU/UK/CH: 444-7084
Agitateur incubateur orbital, modèle 5000I
Prise Euro: 444-7109
Agitateur incubateur orbital, modèle 5000IR
Prise Euro: 444-7112

16
contenU de l’eMballage
Agitateur incubateur orbital
Tapis en caoutchouc antidérapant
Cordon d’alimentation détachable de 234 cm
Manuel d’instruction
garantIe
VWR International garantit ce produit pièces et main-d’œuvre pour une durée de deux (2)
ans à compter de la date de livraison. En cas de vice, VWR pourra, à sa discrétion et à ses
frais, réparer, remplacer ou rembourser au client le prix d’achat du produit, à condition qu’il
lui soit retourné au cours de la période de garantie. Cette garantie n’est pas applicable si
le dommage provient d’un accident, d’une utilisation abusive ou incorrecte, d’une mauvaise
application ou de l’usure normale du produit. Cette garantie deviendrait non valide dans le
cas où les services de maintenance et de vérification requis ne seraient pas exécutés con-
formément aux manuels et réglementations locales, sauf exception si le défaut du produit
n’est pas imputable à cette non exécution.
Il est recommandé au client d’assurer les éléments retournés contre les risques éventuels
d’endommagement ou de perte. Cette garantie se limite aux réparations susmentionnées.
IL EST EXPRESSÉMENT CONVENU QUE LA PRÉSENTE GARANTIE SE SUBSTITUE À
TOUTES LES GARANTIES DE CONFORMITÉ ET DE VALEUR MARCHANDE.
Conformité à la législation et aux réglementations locales
Le client est chargé de la demande et de l’obtention des approbations réglementaires
et autres autorisations nécessaires à l’utilisation ou à l’exploitation du Produit dans
l’environnement local. VWR ne saura être tenu responsable de toute omission ou non
obtention des approbations ou autorisations requises, sauf exception si le refus est dû à un
défaut du produit.
table des MatIères
Contenu de l’emballage . . . . . . . . . . . . 16
Garantie . . . . . . . . . . . . 16
Installation . . . . . . . . . . . . 17
Maintenance et réparation . . . . . . . . . . . . 17
Utilisation prévue . . . . . . . . . . . . 17
Élimination du matériel . . . . . . . . . . . . 17
Conditions ambiantes . . . . . . . . . . . . 17
Consignes de sécurité . . . . . . . . . . . . 18
Avertissement . . . . . . . . . . . . 18
Normes et réglementations . . . . . . . . . . . . 18
Panneau de commande . . . . . . . . . . . . 19
Spécifications . . . . . . . . . . . . 20-22
Consignes d’utilisation . . . . . . . . . . . . 23-27
Service technique . . . . . . . . . . . . 27
Dépannage . . . . . . . . . . . . 28
Accessoires . . . . . . . . . . . . 29-30
Déclaration de Conformité CE . . . . . . . . . . . . 123-124
FR
Legal Address of Manufacturer: VWR International bvba • Researchpark Haasrode 2020 • Geldenaakesbaan 464 • B-3001 Leuven • + 32 16 385011 • http://be.vwr.com
Country of Origin: United States

17
InstallatIon
Dès réception de Agitateur incubateur orbital VWR, vérifiez l’absence de dommages sur-
venus pendant le transport. Il est important que tout dommage résultant du transport soit
détecté lors du déballage. Informez immédiatement le transporteur en cas de dommage.
Après le déballage, deux (2) personnes sont nécessaires pour soulever l’agitateur incuba-
teur orbital afin de la placer sur un plan de travail ou une table de niveau, à l’écart de toute
vapeur explosive. Assurez-vous que la surface sur laquelle le système est placé résiste
à la chaleur généralement produite par le système. Placez toujours l’appareil sur un plan
de travail solide.
Agitateur incubateur orbital est fourni avec un cordon d’alimentation qu’il vous faut tout
d’abord insérer dans le connecteur IEC situé au dos de l’appareil avant de le brancher dans
une prise correctement mise à la terre. L’appareil de 230 V se branche dans une source
d’alimentation de 230 volts, 50/60 Hz.
MaIntenance et réParatIon
Votre Agitateur incubateur orbital est construit pour fonctionner longtemps, de façon effi-
cace et sans problème. Aucun graissage ou tout autre entretien n’est requis de la part de
l’utilisateur. Toutefois, vous devez, au moins tous les trois (3) mois:
• débrancher l’appareil.
• nettoyer la saleté accumulée sur la base et le plateau.
• vérifier tous les éléments possibles afin de vous assurer qu’ils sont
bien serrés.
Votre appareil doit être traité avec le même soin que n’importe quel autre appareil élec-
trique. Évitez de le mouiller ou de l’exposer inutilement à des émanations. Essuyez prompt-
ement tout liquide renversé. Pour nettoyer le paneau avant, N’utilisez PAS de nettoyant ou de
solvant abrasif ou pouvant endommager le plastique, ni de produit inflammable. Assurez-vous
toujours que l’alimentation est débranchée avant d’effectuer tout nettoyage. Si l’appareil doit
être réparé, contactez votre représentant VWR.
UtIlIsatIon PrévUe
L’agitateur incubateur orbital est conçu pour une variété d’applications d’agitation.
élIMInatIon dU MatérIel
Cet appareil ne peut pas être mis au rebut avec les déchets non triés. Il
vous incombe de le mettre correctement au rebut à la fin de sa vie utile
en le confiant à une entreprise agréée pour sa collecte et son recyclage.
Il vous incombe également de le décontaminer en cas de contamination
biologique, chimique et/ou radiologique, de façon à protéger la santé des
personnes impliquées dans la mise au rebut et le recyclage de l’appareil.
Pour plus de renseignements sur les endroits où vous pouvez déposer votre appareil
usagé, veuillez contacter le concessionnaire auprès duquel vous avez effectué l’achat. De
cette façon, vous contribuerez à protéger les ressources naturelles et environnementales
et assurerez que votre appareil est recyclé de façon sanitaire.
condItIons aMbIantes
Conditions de fonctionnement: Pour une utilisation à l’intérieur uniquement.
Températures: 15 à 32°C
Humidité: 80 % HR maximum, sans condensation
Altitude: jusqu’à 2000 m au-dessus du niveau de la mer
Stockage:
Températures: -20 à 65 °C
Humidité: 80 % HR maximum, sans condensation
Installation de Catégorie II et palier de pollution 2 suivant la norme CEI 664.
FR

18
consIgnes de sécUrIté
Veuillez lire le manuel d’instruction dans sa totalité avant d’utiliser Agitateur incubateur
orbital.
AVERTISSEMENT ! NE vous servez PAS de Agitateur incubateur orbital dans
un environnement dangereux ou avec des matériaux dangereux pour lesquels
cet appareil n’a pas été conçu. D’autre part, sachez que la protection offerte par
l’appareil devient obsolète si celui-ci est utilisé avec des accessoires non fournis ou
recommandés par le fabricant ou s’il est utilisé de façon non appropriée.
Extrême prudence lors du déplacement de l’unité, il ya un risque de pincement ou
d’écrasement des doigts.
Utilisez toujours l’appareil sur une surface de niveau pour assurer une perfor-
mance optimale et une sécurité maximale.
NE soulevez PAS l’appareil par le plateau ou le couvercle.
AVERTISSEMENT Pour éviter tout risque d’électrocution, coupez complètement
l’alimentation de l’appareil, débranchez le cordon d’alimentation de l’appareil ou
de la prise murale. Débranchez l’alimentation avant toute procédure d’entretien
et de dépannage.
Essuyez promptement tout liquide renversé. N’immergez PAS l’appareil pour le
nettoyer.
NE faites PAS fonctionner l’appareil s’il semble avoir subi des dommages élec-
triques ou mécaniques.
AVERTISSEMENT Le témoin Caution Hot signale que la température ambiante
de la chambre est supérieure à 40°C. Le témoin s’allume et reste allumé lorsque
la température ambiante de la chambre atteint environ 40°C. Une fois le chauffage
éteint, le témoin Caution Hot reste allumé jusqu’à ce que la température ambiante
de la chambre soit inférieure à 40°C.
Terre - Borne du conducteur de protection
Courant alternatif
norMes et régleMentatIons
Par la présente, VWR International déclare sur l’honneur que les produits sont con-
formes aux exigences des directives et des normes suivantes.
Directives UE connexes:
Directive EMC 2004/108/EC
Directive LVD 2006/95/EC
Directive ROHS 2011/65/EU
Directive WEEE 2002/96/EC
Normes de sécurité:
EN 61010
EN 61010-2-010
EN 61010-2-051
UL 61010-1
CAN/CSA C22.2 No. 61010-1-04
Normes EMC:
EN 45501 EN 61000-4-3
EN 55022A EN 61000-4-4
EN 6100-3-2 EN 61000-4-5
EN 6100-3-3 EN 61000-4-6
EN 61000-4-1 EN 61000-4-11
EN 61000-4-2
FR

19
Le panneau avant de Agitateur incubateur orbital réunit toutes les commandes et tous
les affichages nécessaires au fonctionnement de l’appareil.
A. Touche/témoin veille: Le témoin veille s’allume lorsque l’appareil est branché.
L’appareil passe en mode veille. Appuyez sur la touche veille pour activer les fonc-
tions de température, de vitesse et de minuterie. Le témoin veille s’éteint et les
affichages de la durée, de la température et de la vitesse s’allument. Appuyez sur
la touche veille à nouveau et l’appareil repasse en mode veille.
B. Affichage Temperature (température): Indique la température réelle/de consigne
en association avec les témoins actual/setpoint (réelle/consigne). C. Flèches haut/
bas pour le contrôle des valeurs de consigne. La touche marche/arrêt active et dés-
active la fonction de chauffage. D. Le témoin de vitesse s’allume lorsque l’appareil
est en train de chauffer.
E. Témoin actual (valeur réelle): S’allume lorsque la température affichée est la
température ambiante réelle de la chambre.
F. Témoin point de réglage: S’allume lorsque la température de réglage s’affiche.
G. Témoin Caution Hot (Attention plaque chaude): S’allume lorsque la température
ambiante de la chambre est supérieure à 40°C.
H. Affichage Speed (vitesse): Indique la vitesse de l’agitateur. I. Flèches haut/bas
pour le contrôle des valeurs de consigne. La touche marche/arrêt active et désac-
tive la fonction d’agitation. J. Le témoin de vitesse s’allume lorsque l’appareil est
en train d’agiter.
K. Affichage Time (durée): Indique la durée accumulée (mode continu) ou le temps
restant (mode temporisé). La plage d’affichage est de 0 à 9999 minutes par incré-
ments d’une (1) seconde. L’affichage indique les minutes et les secondes jusqu’à
ce que la minuterie atteigne 99 minutes et 59 secondes (99:59), puis il affiche
automatiquement les minutes, jusqu’à 9999. L. Flèches haut/bas pour le contrôle
des valeurs de consigne. La touche marche/arrêt active et désactive la fonction
de minuterie.
PanneaU de coMMande
H.
L.I.
D.
A.
K.
C.
J.
E.
F.
G. B.
FR
This manual suits for next models
1
Table of contents
Languages:
Other VWR Paint Sprayer manuals