IOP 011B.9
Effective Date: 1/09/2017
Page 3 of 4
f. Shake off excess KCl and rinse the stainless steel thoroughly with
distilled water to prevent corrosion. Reinstall the sensor guard. Store
the sensor in a humid environment such as the calibration chamber
between measurements and when not in use.
II. Calibration
A. Ensure that the sponge inside the instrument calibration chamber is wet. Insert the probe
into the calibration chamber.
B. Turn the instrument on by pressing the ON/OFF button on the front of the instrument.
Wait a few minutes for the dissolved oxygen and temperature readings to stabilize.
C. Use two fingers to press and release the two “up” and “down” arrow keys at the same
time.
D. The LCD will prompt you to enter the local altitude in hundreds of feet. Use the arrow
keys to increase or decrease the altitude (example: entering the number 12 indicates 1200
feet). When the proper altitude appears on the LCD, press the ENTER key once to view
the calibration value and a second time to move to the salinity compensation procedure.
E. The LCD will prompt you to enter the approximate salinity of the water you are about to
analyze. Use the arrow key to decrease the salinity compensation to 0 (fresh water
value). When 0 appears on the LCD, press the ENTER key.
F. It may be necessary to re-calibrate the instrument each time the YSI Model 55 is turned
off. Complete all calibrations at a temperature as close as possible to the sample
temperature.
III. Oxygen measurement
A Once the calibration is complete, the only keys which will remain operational are the
MODE key, the LIGHT key, and the ON/OFF key.
B. Place the probe in the water. When measuring oxygen concentration of stream water,
immerse the probe in slow moving water away from riffles or highly aerated areas.
C. Slowly move the probe to maintain movement of water across the membrane. Stream
current furnishes sufficient movement when measuring the oxygen content of stream
water.
D. Allow sufficient time for the probe to stabilize before measuring temperature and
dissolved oxygen concentration.
E. Move between the mg/L and % air saturation mode by pressing the MODE key; the
LIGHT key activates the back-light on the LCD.
F. The ON/OFF key turns the instrument off.
IV. Storage
A. Turn the ON/OFF switch to OFF.
B. Add a small quantity of deionized water to the sponge in the calibration chamber.
C. Rinse probe with deionized water.