Woodpecker UDS-E User manual

HW-5L

Contents
1 The installation and components of equipment .................................................1
2 Product function and operation ..........................................................................5
3 Sterilization and maintenance ............................................................................8
4. Precaution .......................................................................................................11
5. After service ....................................................................................................13
6. Symbol instruction ..........................................................................................13
7. Environmental protection ................................................................................14
8. Manufacturer’s right .......................................................................................14
9. European authorized representative.................................................................14
10. EMC - Declaration of conformity..................................................................15
11. Statement .......................................................................................................18
TABLE OF OPERATING POWER OF THE TIPS.............................................19

1
1 The installation and components of equipment
1.1 Instruction
Guilin Woodpecker Medical Instrument Co., Ltd. is a professional
manufacturer in researching, developing and producing ultrasonic scalers. The
product is mainly used for teeth cleaning and also an indispensable equipment for
teeth disease prevention and treatment. The new product UDS-E LED ultrasonic
scaler has scaling, perio, endo and auto-water supply functions with the following
features:
1.1.1 Optical handpiece, more convenient for clinical operation.
1.1.2 With auto-water supply function, easier operation.
1.1.3 The inner water pipe is made from antiseptic material, clinical uids,
such as Hydrogen peroxide, chlorhexidine and sodium Hypochlorite, etc,
can be used perfectly under the mode of auto-water supply system so that the
performance of perio and endo is improved dramatically.
1.1.4 The handpiece is detachable and can be autoclaved to the temperature of
134°C and the pressure of 0.22MPa.
1.1.5 Automatic frequency tracking ensures that the machine always works on
the best frequency and performs more steadily.
1.1.6 Controlled by PC, easy operation and more ecient for scaling.
These features make UDS-E LED become a new generation product in the
world dental market.
1.2 Components
1.2.1 The components of machine are listed in the packing list.
1.2.2 Product performance and structure
Ultrasonic scaler is composed of electrocircuit, water way and ultrasonic
transducer.
1.2.3 Scope of application
Ultrasonic scaler UDS-E LED is used for the dental calculus elimination and
root canal treatment.
1.3 The main technical specications
Power supply input: 220 - 240V~ 50Hz/60Hz l50mA
Main unit input: 24V~ 50Hz/60Hz l.3A
Output primary tip vibration excursion: ≤100μm
Output half-excursion force: <2N
Output tip vibration frequency: 28kHz±3kHz
Output power: 3W to 20W
Main unit fuse: T1.6AL 250V
Power supply fuse: T0.5AL 250V
Weight of main unit: 1.17kg
Weight of power supply: 1 kg

2
Operating mode: Continuous operation
Type of protection against electric shock: Class II equipment
Degree of protection against electric shock: Type BF applied part
Applied part of the equipment: handpiece and scalling tip
Degree of protection against harmful ingress of water: Ordinary equipment
Protection degree against water (used on the foot switch): IPX1
Degree of safety of application in the presence of a Flammable Anaesthetic
Mixture with air or with Oxygen or Nitrous Oxide: Equipment not suitable for
being used in the presence of a ammable anaesthetic mixture with air or with
oxygen or nitrous oxide.
1.4 Installation of the main components
Sketch map for installation and connectiom.
1.4.1 The front and back sketch map of the main unit
HW-5L
Figure 1

3
Figure 2
1.4.2 Sketch map for connection of foot switch, power supply and main unit.
Figure 3
1.4.3 Water way connection
Sketch map for water bottle installation

4
HW-5L
Figure 4
1.4.4 Sketch map for connection of detachable handpiece
Figure 5
1.4.5 Sketch map for how to install tip and endochuck with wrench

5
Figure 6
2 Product function and operation
2.1 Scaling function
2.1.1 Operation
a) Open the packing box, make sure that all the parts and accessories are
complete according to the packing list. Take the main unit out of the box and put
it on a stable plane.
b) Turn the water control knob to the max based on symbol as shown in
3.5.2 [ note l ].
c) Insert the plug of the foot switch to its socket (See Figure 3).
d) Water way installation:
Install the bottle reversely to the connector with adequate puried water (See
Figure 4).
e) Screw the scaling tip tightly to the handpiece by torque wrench, then
connect the handpiece and the connector of cable correctly.
f) Insert the plug of the power supply to its socket , then get through to the
power.
g) Switch on the main unit, then the scaling indicator and the rst ve power
indicators shine, auto-water indicator is on.

6
h) According to need, press water button to choose auto-water or DRY mode.
The indicator of choosen mode is on.
i) Select a suitable scaling tip as you need, screw it on the handpiece tightly by
the torque wrench (See Figure 6).
j) Step on the foot switch, the tip begins to vibrate, and the LED lamp on the
top of the handpiece shines. Release the foot switch, the LED lamp keep shining
for 10 seconds.
k) The normal frequency is extremely high. Under the normal working state
of the scaling tip, a light touch and a certain to-and-fro motion will eliminate the
tartar without heating. Overexertion and long-time lingering are forbidden.
l) Vibration intensity: Adjust the vibration intensity as you need, generally
increase the power by pressing the Increasing Power Switch to a higher grade.
According to patient's dierent sensitivity and the rigidity of the gingival tartar,
adjust the vibration intensity during the clinical treatment.
m) Water volume adjust: Step on the foot switch, and the tip begins to vibrate,
then turn the water control knob to form ne spray to cool down the handpiece
and clean the teeth.
n) The handpiece can be handled in the same gesture as a pen in hand.
o) During the clinical treatment, be sure not to make the end of tip touch the
teeth vertically and not to make the tip overexert on the surface of the teeth in
case of hurting the teeth and damaging the tip.
p) After nishing operation, keep the machine working for 30 seconds on the
water supply condition in order to clean the handpiece and the scaling tip.
q) Unscrew the scaling tip and pull out handpiece, then sterilize them.
Notice:
①Don't pull out the handpiece when the foot switch is stepped on and the
machine is producing ultrasonic vibration.
②Please don't choose dry mode unless needed in the treatment, please don't
make the machine work under dry mode for a long time. The teeth will be hurt if
the handpiece and tips get overheat during the treatment.
2.1.2 Instruction for main components of detachable handpiece (See Figure 5):
a) Nipple: The nipple can be removed. You can screw out the nipple and clean
the pole with alcohol termly.
b) Decorative ring: can be disassembled and cleaned with alcohol regularly,
can be autoclaved under the high temperature and pressure.
c) Handpiece: The main part of the whole handpiece, can be autoclaved under
the high temperature and pressure.
d) The connector of the cable: Connect the handpiece with the water source
and power supply of the main unit.
e) LED lamp、Light pipe: Clean them with puried water and sterilize them
under the high temperature of 134°C and high pressure of 0.22Mpa.
Notice: Keep dry when the detachable handpiece connects to the connector of

7
the cable.
2.1.3 Instruction of torque wrench (See Figure 6)
a) The torque wrench's structure is designed in special way which can control
the strength of the scaling tip's installation properly and correctly. It also can
guarantee the operator screw or unscrew the scaling tip eectively and keep their
hands away from being scratched.
b) Operation
① Take the scaling tip into the torque wrench, operate as showed in gure 6.
②Tip installation: Hold the handpiece, turn the tip toward direction as
showed in gure 6 with the torque wrench. Turn one more circles when the tip
stops, then the tip is installed.
③Tip uninstallation: Hold the handpiece, turn the wrench toward anti-
clockwise direction.
④Sterilize it in sterilizer after each treatment.
⑤The torque wrench must be cooled naturally after sterilization to avoid
scald when using next time.
⑥Keep the torque wrench in a cool, dry and ventilated place and keep it
clean.
2.2 Auto-water supply function
2.2.1 Usage process
a) Pull out the water bottle vertically.
b) Open the lid, ll the bottle with adequate puried water and tighten the lid.
c) Clean the joint and interface of water bottle.
d) Place the bottle with water upside down, insert the joint into the interface
vertically (See Figure 4).
e) Press "WATER" to choose auto-water supply.
2.2.2 Notice
a) Make sure the air hole and water entrance are unblocked (See Figure 7).
Figure 7 Figure 8
b)Check if the cushion inside the lid is on
good condition. If it becomes deformed or falls o, reinstall or change it.
c) Tighten the lid.

8
d) Clean the joint and interface of water bottle before each clinical operation.
e) After each clinical operation with clinical liquid, change a bottle with
puried water, turn the water supply to max, make the machine work with auto-
water supply for 30 seconds in order to keep the water way and spare parts clean
and durable.
f) When the water is lower than water level
lower limit, please add water (See Figure 8).
2.3 Endo function
2.3.1 Usage process
a) Fix endochuck to handpiece by endo wrench (See Figure 6).
b) Unscrew the screw cap on the endochuck.
c) Put the ultrasonic le into the hole in the front of endochuck.
d) Screw the screw cap with endo wrench to tight up the ultrasonic le.
e) Press option key, turn to endo function.
f) When ultrasonic scaler turns into endo function, only the first power
indicator is on and the power is at the lst grade. Put the ultrasonic le into the
patient's root canal slowly, step on the foot switch to start endo treatment. During
the treatment, turn up the power gradually according to the needs.
2.3.2 Notice
a) When xing endochuck,it must be screwed down.
b) The screw cap on the endochuck must be screwed down.
c) Don't press it too much when the ultrasonic le in root canal.
d) Don't step on the foot switch until the ultrasonic le is in root canal.
e) The power range of endo treatment is advised from the lst to the 5th grade.
3 Sterilization and maintenance
3.1 Sterilization of detachable handpiece
3.1.1 Autoclaved under high temperature, pressure, time:
a) 134°C, 2.0bar~2.3bar (0.20MPa~0.23MPa), 4min.
b) Pull out the handpiece and unscrew the scaling tip and endochuck after each
operation.
c) Pack the handpiece with sterile gauze or sterile bag before sterilizing.
d) Reuse handpiece after it cools naturally to avoid scalding hands.
3.1.2 Notice
a) Dry the cleaning liquid in the handpiece with compressed air before
sterilization.
b) Be sure that the scaling tip has been unscrewed from the handpiece and it
cannot be sterilized with others.
c) Please notice whether the outer of the handpiece is damaged during
the treatment or sterilization, don't smear any protective oil on the surface of
handpiece.
d) There are two waterproof "o" rings at the end of handpiece. Please lubricate

9
them with dental lube frequently, as sterilization and repeated pulling and
inserting will reduce their working life. Change a new one once it is damaged or
worn excessively.
e) The following sterilizing methods are forbidden:
①Put handpiece into any liquid for boiling.
②Dip handpiece in disinfectors such as iodine, alcohol and glutaraldehyde.
③Put handpiece into oven or microwave oven for baking.
3.2 Sterilization of scaling tips and endochuck
All the scaling tips and endochuck can be autoclaved under high temperature
and pressure.
3.3 Sterilization of torque wrench and endo wrench
3.3.1 The torque wrench and endo wrench can be sterilized under high
temperature and pressure.
3.3.2 The following sterilization ways for torque wrench are forbidden.
a) Soaked in solution.
b) Dip in iodine, alcohol or glutaraldehyde.
c) Torrefy in oven or microwave oven.
Notice: We are not responsible for any damage of the torque wrench directly
or indirectly made by any way in the above items.
3.4 Cleaning of tips, endochuck, torque wrench and endo wrench
The scaling tip, endochuck, torque wrench and endo wrench can be cleaned
by ultrasonic cleaner.
3.5 Sterilization and cleaning of LED lamp and Light pipe
Clean the LED lamp and Light pipe with purified water and sterilize them
under high temperature and high pressure after every operation.
3.6 Troubleshooting and notes
3.6.1 Troubleshooting
Fault Possible cause Solutions
The scaling tip doesn't
vibrate and there is no
water owing out when
stepping on the foot
switch.
The power plug is in loose
contact.
Make the plug insert to the
socket well.
The foot switch is in loose
contact.
Insert the foot switch to its
socket tightly.
The fuse in the main unit is
broken.
Contact our dealers or us.

10
Fault Possible cause Solutions
The scaling tip doesn't
vibrate but there is water
owing out when stepping
on the switch.
The tip is in loose contact. Screw the tip on the
handpiece tightly (See Figure
6).
The connect plug between the
handpiece and the circuit board
is in loose contact.
Contact our dealers or us.
Something wrong with the
handpiece.
Send the handpiece to our
company to repair.
Something wrong with the
cable.
Contact our dealers or us.
The scaling tip vibrates
but there is no spray
when stepping on the foot
switch.
The water control knob is not
on.
Turn on the water contrl
knob [note 1].
The vibration of the tip
becomes weak.
The tip hasn't been screwed on
to the handpiece tightly.
Screw the tip on the
handpiece tightly (See Figure
6).
The tip is loose because of
vibration.
Screw on the tip tightly (See
Figuree 6).
The coupling between the
handpiece and the cable isn't
dry.
Dry it by the hot air.
The tip is damaged [note 2]. Change a new one.
There is water seeping
from the coupling between
the handpiece and cable.
The waterproof "O"ring is
damaged.
Change a new waterproof
"O"ring.
The u-le doesn't vibrate. The screw is loose. Tighten it.
Endochuck is damaged. Change a new one.
There is noise coming
from the endochuck.
The screw is loose. Tighten it.

11
Fault Possible cause Solutions
LED light don't work
Poor contact Contact tightly
Something wrong with LED
light
Change a new one
LED lamp installed
backwards
Please install the "+" of
the LED lamp to the "+" of
the handpiece
There is no water coming
out from the handpiece
(automatic water supply
mode).
There is air in the water pipe. Turn the water control to the
Max, reinsert the bottle.
If the problem still can't be solved, please contact with local dealer or
manufacturer.
3.6.2 Notes
[Note l] Turn the water control knob as symbol showed. It comes to the min,
on the contrary direction, it comes to the max.
[Note 2] If the scaling tip has been screwed on tightly and there is ne spray
too, the following phenomena show that the scaling tip is damaged:
a) The vibrating intensity and the water spray degree become weak obviously.
b) During treatment, it produces the sound like "buzz" from the scaling tip.
4. Precaution
4.1 Notice when using equipment
4.1.1 Keep the scaler clean before and after operation.
4.1.2 The handpiece, scaling tip, torque wrench, endo wrench and endochuck
must be sterilized before each treatment.
4.1.3 Don't screw or unscrew the scaling tip and endochuck when stepping on
the foot switch.
4.1.4 The scaling tip must be fastened and there must be ne spray or drip
coming out from the tip when operating.
4.1.5 Change a new one when the tip and ultrasonic le are damaged or worn
excessively.
4.1.6 While scaler working ,the heat of scaling tip will become higher if there
is no water owing out.Please keep the water ow smoothly.
4.1.7 Don't twist or rub the tip and endochuck.
4.1.8 Ensure the connector of handpiece and the socket of the cable dry before

12
installing the handpiece.
4.1.9 Don't pull the cable forcibly in case of the handpiece falling o from the
cable.
4.1.10 Don't knock or rub the handpiece.
4.1.11 Please put the power plug into the socket easy to pull out, to make sure
it can be pull out in emergency.
4.1.12 The power supply is considered as a part of ME equipment.This device
can only be equipped with the special power supply of Guilin Woodpecker
Medical Instrument Co., Ltd.
4.1.13 The power supply is NOT waterproof. Please keep it dry and away from
the water.
4.1.14 After operation, turn o power, and pull out the plug.
4.1.15 We are only responsible for the safety on the following conditions:
I ) The maintenance, repair and modication are made by the manufacturer or
the authorized dealer.
II ) The changed components are original of "WOODPECKER" and are
operated according to instruction manual.
4.1.16 The internal screw thread of the scaling tips produced by some
manufacturers maybe coarse, rusty and collapsed. This will damage the external
screw threat of the handpiece irretrievably. Please use "WOODPECKER" brand
scaling tips.
4.1.17 This model is only matched the power supply of our company.
4.1.18 Please select a suitable power when using dierent type of tips (refer to
"TABLE OF OPERATING POWER OF THE TIPS").
4.2 Contraindication
4.2.1 The hemophilia disease patient is not allowed to use this equipment.
4.2.2 The patients or doctors with heart pacemaker are forbidden to use this
equipment.
4.2.3 The heart disease patient, pregnant woman and children should be
cautious to use the equipment.
4.3 Storage and maintenance
4.3.1 The equipment should be handled carefully and lightly. Be sure that it is
far from the vibration, and installed or kept in a cool, dry and ventilated place.
4.3.2 Don't store the machine together with the articles that are combustible,
poisonous, caustic, and explosive.
4.3.3 This equipment should be stored in a room where the relative humidity
is 10% ~ 93%, atmospheric pressure is 70kPa to106kPa, and the temperature is
-20°C ~ +55°C.
4.3.4 If not use for a long time, please make the machine get through to the

13
power and water once per month for ve minutes.
4.4 Transportation
4.4.1 Excessive impact and shake should be prevented in transportation. Lay
it carefully and lightly and don't invert it.
4.4.2 Don't put it together with dangerous goods during transportation.
4.4.3 Avoid solarization and getting wet in rain or snow during transportation.
4.5 Working condition
4.5.1 Environment temperature: +5°C to +40°C
4.5.2 Relative humidity: 30% ~75%
4.5.3 Atmosphere pressure: 70kPa to 106kPa
4.5.4 A temperature of the water at the inlet:not higher than +25°C
5. After service
We oer one year free repair to the equipment according to the warranty card.
The repair of the equipment should be carried out by our professional
technician. We are not responsible for any irretrievable damage caused by the not
professional person.
6. Symbol instruction
Trademark Ordinary equipment
Alternating current Drip-proof
Date of manufacture Manufacturer
Class II equipment Type BF applied part
Foot switch interface Used indoor only
Adjustment for the water
ow Can be autoclaved
24VAC power supply socket Water entrance pressure
CE marked product Follow Instructions for Use
Caution Recover

14
Keep dry Handle with care
Add water Mode of auto-water system
Appliance compliance WEEE directive
Atmospheric pressure for
storage Power switch
Temperature limitation for
storage
Humidity limitation for
storage
Authorised Representative in the EUROPEAN COMMUNITY
7. Environmental protection
Please dispose according to the local laws.
8. Manufacturer’s right
We reserve the rights to change the design of the equipment, the technique,
ttings, the instruction manual and the content of the original packing list at any
time without notice. If there are some differences between blueprint and real
equipment, take the real equipment as the norm.
9. European authorized representative

15
10. EMC - Declaration of conformity

16

17

18
The device has been tested and homologated in accordance with EN 60601-1-2
for EMC. This does not guarantee in any way that this device will not be eected
by electromagnetic interference Avoid using the device in high electromagnetic
environment.
11. Statement
All rights of modifying the product are reserved to the manufacturer without
further notice. The pictures are only for reference. The nal interpretation rights
belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO., LTD.
The industrial design, inner structure, etc, have claimed for several patents by
WOODPECKER, any copy or fake product must take legal responsibilities.
Other manuals for UDS-E
1
Table of contents
Other Woodpecker Media Converter manuals
Popular Media Converter manuals by other brands

Haiwei
Haiwei H2 user manual

elco
elco TSE 485/WiFi user manual

B&B Electronics
B&B Electronics 485ABOIC quick start guide

EDWANZ group
EDWANZ group Deutronic DVC Series Installation and safety instructions

Matrix Electronic Technology
Matrix Electronic Technology mini-i Series user manual

Ascent Communication Technology
Ascent Communication Technology EC2200 user guide