Specifications
AQ2170/AQ2170H
Sensor
Specifications
Model AQ2170 AQ2170H
Wavelength Range
(nm)
850/1300/1310/1490/1550/
1625/1650 nm
1310/1490/1550/1625/
1650 nm
Detector InGaAs
Applicable Fibers Type Light emitted from SM( ITU-T
G.652), GI( 50/125 μm), GI(
62.5/125 μm) optical fiber
Light emitted from SM( ITU-T
G.652), optical fiber
Optical Connector FC, SC, LC, ϕ1.25 ferrule, ϕ2.5 ferrule
Power Range(dBm) -70 to +10 dBm -50 to +26 dBm
Noise Level(dBm) -60 dBm -40 dBm
Uncertainty under
standard conditions1
±5% ±5%
General
Specifications
Displayed Resolution 0.01dB(>-60dBm),0.1dBm(≤-
60dBm)
0.01dB(>-40dBm),0.1dBm(≤-
40dBm)
Displayed Unit Absolute values:dBm, mW, μW Relative values: dB
Modulation Mode CW, CHOP(270 Hz, 1 kHz, 2 kHz)
Measurement Interval Approximately150 ms
Measurement mode Absolute value measurement
Relative value measurement:
Relative value measurement relative to a set reference
value
Relative value measurement based on a displayed
measurement
Power supply Dry and Rechargeable Battery AAA
Battery life2Approximately 40 hours
Power saving Power turns off automatically when there is no key input for
approximately 10 minutes (this function can be disabled)
Battery check Battery indicator
Environmental
conditions
Operating temperature range:-10~50ºC
Operating humidity range:20~80%
Storage temperature range:-20~+70ºC(no condensation)
Storage humidity range:5~95%(no condensation)
Dimensions and
weight3
63(W)×116(H)×35(D) mm
Approximately160 g
EMC Emissions Compliant standards:
EN 61326-1 Class B, EN 55011 Class B Group1
EMC standards of Australia and New Zealand EN55011
Class B, Group 1,
Korea Electromagnetic Conformity Standard ( 한국
전자파적합성기준 )
Immunity
EN61326-1 Table2 (for use in industrial locations), EN61326-2-2
Environmental standard
Compliant standard: EN50581 Monitoring and instruments
1 23 ± 2°C ambient temperature reference conditions/Wavelength setting of Power Meter : 1310
nm/Wavelength of light source to be measured : 1310 nm/Spectral width of the light source to be
measured : Less than 5 nm/Power of measured light source : -10 dBm/Modulation : CW/Use Fiber
: SM fiber (ITU-T G652)/Connector Type : FC/Does not include the polarization dependence and
polarization conditions/1 year aging/Including uncertainty in attaching and removing the adapter
2 Continuation Measurement, using alkaline dry cells, at 23ºC ±2ºC
3 A protector is removed.
Note : Specifications are at 23ºC±2ºC unless otherwise noted.Optical connector: FC/PC
AQ2180/AQ2180H
Sensor
Specifications
Model AQ2180 AQ2180H
Wavelength Range
(nm)
850/1300/1310/1490/1550/
1625/1650 nm
1310/1490/1550/1625/
1650 nm
Detector InGaAs
Applicable Fibers Type Light emitted from SM( ITU-T
G.652), GI( 50/125 μm), GI(
62.5/125 μm) optical fiber
Light emitted from SM(
ITU-T G.652), optical fiber
Optical Connector FC, SC, LC, ϕ1.25 ferrule, ϕ2.5 ferrule
Power Range(dBm) -70 to +10 dBm -50 to +26 dBm
Noise Level(dBm) -60 dBm -40 dBm
Uncertainty under
standard conditions1
±5% ±5%
General
Specifications
Displayed Resolution 0.01 dB, 0.01 dBm, 0.0001 μW
Displayed Unit Absolute values:dBm, mW, μW Relative values: dB
Modulation Mode CW, CHOP(270 Hz, 1 kHz, 2 kHz)
Measurement Interval Approximately150 ms
Measurement mode Absolute value measurement
Relative value measurement:
Relative value measurement relative to a set reference
value
Relative value measurement based on a displayed
measurement
Memory function 999 records
Interface USB-B(mini)
Power supply Dry and Rechargeable Battery AA
Battery life2Approximately 40 hours
Power
saving
Power turns off automatically when there is no key input for
approximately 10 minutes (this function can be disabled)
Battery
check
Battery indicator
Environmental
conditions
Operating temperature range:-10~50ºC
Operating humidity range:20~80%
Storage temperature range:-20~+70ºC (no condensation)
Storage humidity range:5~95%(no condensation)
Dimensions and
weight3
76(W)×153(H)×43(D) mm
Approximately280 g
EMC Emissions Compliant standards:
EN 61326-1 Class B, EN 55011 Class B Group1
EMC standards of Australia and New Zealand EN55011
Class B, Group 1,
Korea Electromagnetic Conformity Standard ( 한국
전자파적합성기준 )
Immunity
EN61326-1 Table2 (for use in industrial locations), EN61326-2-2
Environmental standard
Compliant standard: EN50581 Monitoring and instruments
1 23 ± 2°C ambient temperature reference conditions/Wavelength setting of Power Meter : 1310
nm/Wavelength of light source to be measured : 1310 nm/Spectral width of the light source to be
measured : Less than 5 nm/Power of measured light source : -10 dBm/Modulation : CW/Use Fiber
: SM fiber (ITU-T G652)/Connector Type : FC/Does not include the polarization dependence and
polarization conditions/1 year aging/Including uncertainty in attaching and removing the adapter
2 Continuation Measurement, using alkaline dry cells, at 23ºC ±2ºC
3 A protector is removed.
Note : Specifications are at 23ºC ±2ºC unless otherwise noted.Optical connector: FC/PC
AQ4280A/AQ7280B/AQ4280C
Light Source
Specifications
Model AQ4280A AQ4280B AQ4280C
Light emitter LD
Compliant fiber SM (ITU-T G.652)
Center wavelength11310/1550±20 nm 1310/1550±20 nm
1490±10 nm
1310/1550±20 nm
1490/1625±10 nm
Spectral width1, 2 <5 nm(1310 nm) <5nm (1310 nm,
1490 nm)
<5 nm(1310 nm,
1490 nm,1625 nm)
<10 nm(1550 nm) < 10 nm(1550 nm) < 10 nm(1550 nm)
Optical output level3-5 dBm±1 dB -5 dBm±1 dB -5 dBm±1 dB
Level stability
(15min)1, 3
<±0.05 dB <±0.05 dB
(1310/1550 nm)
<±0.05 dB
(1310/1550 nm)
<±0.1 dB (1490 nm) <±0.1 dB (1490/1625
nm)
Modulation Mode CW, CHOP(270 Hz, 1 kHz, 2 kHz)
Optical Connector FC, SC, ST
General
Specifications
Power supply Dry and Rechargeable Battery AA
Battery life4Approximately 25 hours
Power saving Power turns off automatically when there is no key input for
approximately 10 minutes (this function can be disabled)
Battery check Battery indicator
Environmental
conditions
Operating temperature range:-10~50ºC
Operating humidity range:20~80%
Storage temperature range:-20~+70ºC (no condensation)
Storage humidity range:5~95%(no condensation)
Laser Class CLASS1(IEC 60825-1)
Dimensions and
weight5
76(W)×153(H)×43(D) mm, Approximately 300 g
Safety standard IEC 60825-1 Class1(Laser), FDA 21CFR1040.10(Laser)
EMC Emissions Compliant standards:
EN 61326-1 Class B, EN 55011 Class B Group1
EMC standards of Australia and New Zealand EN55011 Class
B, Group 1,
Korea Electromagnetic Conformity Standard ( 한국
전자파적합성기준 )
Immunity
EN61326-1 Table2 (for use in industrial locations), EN61326-2-2
Environmental
standard
Compliant standard: EN50581 Monitoring and instruments
1 Modulation Mode:CW, ambient temperature: 23 ±2ºC
2 RMS (2σ, -20 dB)
3 In the case of the PC/FC 2m code connection
4 Continuation Measurement, using alkaline dry cells, at 23ºC ±2ºC
5 A protector is removed.
Note : Specifications are at 23ºC ±2ºC unless otherwise noted.Optical connector: FC/PC
Operation
AQ2170/AQ2170H
1
2
3
4
5
ON
OFF
>2s PER M
dBm/
dB/mw
REF
>2s SET
No. Key Function
Switches instrument on/off. Long keypress over 2
seconds while powering on is to activate the instrument
without Auto-off function.
Selects measurement wavelength in sequence of
850/1300/1310/1490/1550/1625/1650nm
(AQ2170H:1310/1490/1550/1625/1650
Switches measurement units among dBm, dB and mW.
Short keypress to display reference level of present test
wavelength. Long Key press over 2seconds to set a new
reference level of present test wavelength.
Switches backlighting on/off.
REF
nm
5
3
2
1
4
AQ2180/AQ2180H
1
2
3
5
6
ON
OFF
>2s PER M
dBm/
dB/mw
REF
>2s SET
No. Key Function
Switches instrument on/off. Long keypress over 2 seconds while
powering on is to activate the instrument without Auto-off
function.
Selects measurement wavelength in sequence of 850/1300/
1310/1490/1550/1625/1650 nm
(AQ2180H:1310/1490/1550/1625/1650) and to Long key press
over 2 seconds activate auto-wavelength recognition(TWIN). In
memory, browse the forward records and Long key press over
2seconds browse the forward 20 record.
Short key press to display reference value of present test
wavelength.
Long key press over 2seconds to set a new reference value of
present test wavelength.In memory,browse the afterward records
and Long key press over 2 seconds browse the afterward 20
record.
Switches backlighting on/off.
4
Switches measurement unit among dBm,dB and mW. In memory,
delete current record and Long key press over 2 seconds delete
alls record.
Long key press over 2seconds to store the current test value.
Short keypress to display the record.
6
5
3
2
4
1
AQ4280A/AQ4280B/AQ4280C
1
2
3
5
6
ON
OFF
>2s PER M
No. Key Function
Switches Instrument on/off. Long Key press over 2 seconds
while powering on to activate the instrument without Auto-off
function.
Select CH1 , CH2(Only AQ4280C), CH1 and CH2 can emit
light independently.
CH which is not used should stop light emitting.
SINGLE: Auto-wavelength recognition is off.
TWIN: Auto-wavelength recognition is on.
Switches backlighting on/off.
4
Modulated Wavelength Shifting Key: Switches modulated
wavelength and continuous wavelength.
AQ4280A:1310nm, 1550nm
AQ4280B:1310nm, 1490nm, 1550nm
AQ4280C:1310nm, 1550nm + 1490nm, 1625nm
shift
CW/Hz
TWIN
AUTO -OF F
270Hz
1kH z
2kH z
nm
SIN GLE
TWI N
6
3
5
4
2
IM AQ2170-91EN