manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. YOKOGAWA
  6. •
  7. Multimeter
  8. •
  9. YOKOGAWA 73201 User manual

YOKOGAWA 73201 User manual

■ To avoid electric shock!
• Check the testing leads before use and do not use deteriorated or damaged
ones. Check the continuity of the testing leads.
• Do not use the instrument if there is any damage to the casing or when the
casing is removed.
• Disconnect the instrument from the circuit under test before opening the casing
to replace the batteries or for any other reason.
• Avoid using the instrument if it has been exposed to rain or moisture or if your
hands are wet.
■ To avoid electrical shock or fire!
Do not use the instrument in an atmosphere where any flammable or explosive
gas is present.
a) If the model is 73201 or 04, place the function switch
in the position.
If the model is 73202 or 03, place the function switch
in the Ω / position, and then press the SELECT key.
(The symbol appears on the display.)
b) Plug the black testing lead into the COM input terminal
and the red testing lead into the V•Ω• input terminal.
c) Connect the testing leads to the circuit under check.
If the circuit is continuous (no more than approximately
50 Ω), the buzzer sounds.
d) When the test is complete, place the function switch in
the OFF position and turn off the multimeter.
User's
Manual
Model 73201/02/03/04
Digital Multimeters
IM 73201E
7th Edition Dec. 2011(YMI)
When operating the instrument, be sure to observe the cautionary notes
given below to ensure correct and safe use of the instrument. If you use
the instrument in any way other than as instructed in this manual,
the instrument’s protective measures may be impaired. Yokogawa is by
no means liable for any damage resulting from use of the instrument in
contradiction to these cautionary notes.
WARNING
CAUTION
Indicates a hazard that may result in the loss of life or
serious injury of the user unless the described instruction
is abided by.
Indicates a hazard that may result in an injury to the user
and/or physical damage to the product or other equipment
unless the described instruction is abided by.
This symbol indicates that the operator must refer to an explanation
in the instruction manual in order to avoid the risk of injury or death of
personnel or damage to the instrument.
■ To avoid damage to instrument or electric shock!
The restrictions on the maximum voltage level for which the 73201, 02, 03 and
04 multimeters can be used, depend on the measurement categories specified
by the safety standards. These category specifications are formulated to protect
operators against transient impulse voltages in power lines.
Category of Testing leads
With caps:
1000V 10A CAT.III / 600V 10A CAT.IV
Without caps:
1000V 10A CAT.II / 600V 10A CAT.II
Double Insulation
This symbol indicates double insulation.
Alternating current
This symbol indicates AC voltage/current.
Direct current
This symbol indicates DC voltage/current.
Fuse
This symbol indicates a fuse.
Earth TERMINAL
This symbol indicates ground.
The instrument and this manual also use the following safety symbols:
Caution
Store this manual in a safe
place for future reference.
1) Function Switch
Turns off the power or selects the measurement mode.
OFF: Turns off the power.
V: AC voltage measurement
V: DC voltage measurement
Ω/ : Resistance measurement and continuity check (73202/03)
Ω : Resistance measurement (73201/04)
: Continuity check (73201/04)
: Diode test
: Capacitor check (73202/03)
µA: DC/AC current measurement in micro-amperes
mA: DC/AC current measurement in milli-amperes (73201/02/03)
A: DC/AC current measurement in amperes
2) SELECT key (73202/03) or / key (73201)*
This function is not supported on the 73204 multimeter.
This key is enabled only if the multimeter is in one of the following
measurement modes.
• Ω/ : In this mode, the key selects between the resistance
measurem ent and continuity test.
• : In this mode, the key adjusts the stray capacitance of the
testing leads and the multimeter itself to zero.
(The display shows the
CAL
symbol.)
• µA/mA/A: In this mode, the key selects between the DC and AC modes.
* The 73201 multimeter comes with the / key instead of the SELECT
key. This key is used only to select between the DC and AC modes in the
current measurement mode.
3) RANGE key * This function is not supported on the 73204 multimeter.
Allows the operator to select the measuring range manually (the display
shows the
R • H
symbol).
To return to the normal auto-ranging mode, hold down this key for at
least one second until the display shows “AUTO.”
4) AUTO-H key
A • H
5) Display View with All Elements Turned On
A • H
R • H
CAL
+ –
Lit when in DC-mode measurement
Lit when in AC-mode measurement
Polarity indicator lit when the polarity is negative
Lit when in diode test
Lit when in continuity check
AUTO HOLD indicator
Manual range indicator
Auto range indicator
AUTO POWER OFF indicator
Unit for capacitance measurement
Unit for voltage measurement
Unit for current measurement
Unit for resistance measurement
Lit when stray capacitance is adjusted to zero
Lit when the batteries are low
Symbol and Unit
AUTO
AUTO POWER OFF
nF, µF
mV, V
µA, mA, A
Ω, kΩ, ΜΩ
Description
Internal Wiring
Outlet
T
Fixed Equipment,
etc.
Equipment
Disposing the Product
Safety Precautions Components Measuring Instructions
This Product complies with the WEEE Directive (2002/96/EC) marking
requirement.
The affixed product label (see below) indicates that you must not discard
this electrical/electronic product in domestic household waste.
Product Category
With reference to the equipment types in the WEEE directive Annex 1,
this product is classified as a “Monitoring and Control instrumentation”
product.
To return unwanted products within the EU area, contact your local
Yokogawa Europe B. V. office. Do not dispose in domestic household waste.
Yokogawa Meters & Instruments Corporation
International Sales Dept.
Tachihi Bld. No.2, 6-1-3, Sakaecho, Tachikawa-shi,Tokyo 190-8586 Japan
Phone: 81-42-534-1413, Facsimile: 81-42-534-1438
YOKOGAWA CORPORATION OF AMERICA (U.S.A.)
Phone: 1-800-888-6400 Facsimile: 1-770-254-0928
YOKOGAWA EUROPE B. V. (THE NETHERLANDS)
Euroweg 2, 3825 HD, Amersfoort, THE NETHERLANDS
Phone : 31-88-4641000 Facsimile : 31-88-4641111
YOKOGAWA AMERICA DO SUL LTDA. (BRAZIL)
Phone: 55-11-5681-2400 Facsimile: 55-11-5681-4434
YOKOGAWA ENGINEERING ASIA PTE. LTD. (SINGAPORE)
Phone: 65-6241-9933 Facsimile: 65-6241-2606
YOKOGAWA MEASURING INSTRUMENTS KOREA CORPORATION (KOREA)
Phone: 82-2-551-0660 to -0664 Facsimile: 82-2-551-0665
YOKOGAWA AUSTRALIA PTY. LTD. (AUSTRALIA)
Phone: 61-2-8870-1100 Facsimile: 61-2-8870-1111
YOKOGAWA INDIA LTD. (INDIA)
Phone: 91-80-4158-6000 Facsimile: 91-80-2852-1441
YOKOGAWA SHANGHAI TRADING CO., LTD. (CHINA)
Phone: 86-21-6239-6363 Facsimile: 86-21-6880-4987
YOKOGAWA MIDDLE EAST B. S. C (C) (BAHRAIN)
Phone: 973-17-358100 Facsimile: 973-17-336100
YOKOGAWA ELECTRIC CIS LTD. (RUSSIAN FEDERATION)
Phone: 7-495-737-7868 Facsimile: 7-495-737-7869 IM3E-2011.7
CAT. III
Measurement Category
Measurement category Description
Maximum Input Voltage
CAT. II
CAT. I
CAT. II
CAT. III
CAT. IV
600 V
300 V
I
II
III
IV
For measurements performed on circuits
not directly connected to MAINS.
For measurements performed on circuits
directly connected to the low voltage installation.
For measurements performed
in the building installation.
For measurements performed at the
source of the low-voltage installation.
Remarks
Appliances, portable
equipment, etc.
Distribution board,
circuit breaker, etc.
Overhead wire, cable
systems, etc.
<5> Display
Input terminal
<1> Function switch
<4> AUTO.H key
<2> SELECT key (73202/03)
or ( / = key (73201)
<3> RANGE key
*(2) and (3) are not supported on the
73204 multimeter.
• Do not use the multimeter near noise-emitting equipment or where there may
be a sudden temperature change. Otherwise, the instrument may produce an
unstable reading or errors.
• Do not wipe the instrument using any solvent (chemicals) such as benzine or
paint thinner. Otherwise, the front panel may be damaged or discolored.
When cleaning the instrument, use a dry cloth.
• Do not leave the multimeter exposed to direct sunlight or in a hot and humid
location such as the inside of a car, for any prolonged length of time.
To avoid damage to Instrument or equipment.
• Before starting measurement, check to which mode the function switch is set
and make sure the testing leads are plugged into the terminals for the desired
mode of measurement.
• Temporarily remove the testing leads from the device under test before
operating the function switch.
To avoid damage to instrument!
Turn off the power to the circuit under test before starting measurement in order
to prevent any excessive voltage from being applied to the multimeter.
To avoid damage to instrument!
Turn off the power to the circuit under test before starting measurement in order
to prevent any excessive voltage from being applied to the multimeter.
• AC Voltage Measurement ( V)
a) Set the function switch from the OFF position to the
V position.
b) Plug the black testing lead into the COM input termi -
nal and the red testing lead into the V•Ω• input
terminal.
c) Connect the testing leads to the circuit under test
and then read the multimeter when it stabilizes.
d) When measurement is complete, set the function switch
back to the OFF position and turn off the multimeter.
• DC Voltage Measurement (
• Resistance Measurement (Ω)
V)
a) Set the function switch from the OFF position to the
V position.
b) Plug the black testing lead into the COM input terminal
and the red testing lead into the V•Ω• input terminal.
c) Connect the testing leads to the circuit under test
and then read the multimeter when it stabilizes.
d) When measurement is complete, set the function switch
back to the OFF position and turn off the multimeter.
NOTE
Do not mistake the following for a malfunction!
If the 400 mV range is selected and the testing leads are left open-circuited,
the multimeter may give a certain reading. This does not affect your measurement.
a) Place the function switch in the Ω position (73201/04)
or Ω/ position (73202/03).
b) Plug the black testing lead into the COM input terminal
and the red testing lead into the V•Ω• input terminal.
c) Connect the testing leads to the circuit under test and
then read the multimeter when it stabilizes.
d) When measurement is complete, place the function
switch in the OFF position and turn off the multimeter.
• Continuity Check ( )
V/
4/10
/
4/10
V
Black testing lead
Red testing lead
Black testing lead
Red testing lead
Black testing lead
Red testing lead
Black testing lead
Red testing lead
/
4/10
/
4/10
WARNING
WARNING
CAUTION
CAUTION
CAUTION
Set up AUTO HOLD function. (the display shows the symbol).
CAUTION
CAT. III
CAT. I
CAT. IV
Distribution
board
CAT. II
Entrance
Cable
• Capacitor Check ( )
*This function is not supported on the 73201/04 multimeters.
a) Set the function switch from the OFF position to the
position.
b) Plug the black testing lead into the COM input terminal
and the red testing lead into the V•Ω•  input terminal.
c) Connect the testing leads to the diode and then read
the multimeter when it stabilizes.
Input
Resistance
Maximum
Input Voltage
Range
0.5%+1 600 V>100MΩ
11MΩ
10M Ω
0.5%+1 0.3%+ 1
0.75%+1
Accuracy
73201
Resolution
73202/04 73203
Input
Resistance
Range
4 V
40 V
400 V
600 V
1%+5 0.75%+5
11 MΩ, <50 pF 600 V
10 MΩ, <50 pF
Accuracy (40-500 Hz)
Resolution
Input Voltage
73201 73202 73203/04
CAL
A • H
A • H
A • H
a) Set the function switch from the OFF position to the
position.
b) Plug the black testing lead into the COM input terminal
and the red testing lead into the μA • mA • input
terminal.
c) Press the SELECT key to adjust the stray capacitance
to zero (thedisplay shows the CAL symbol).
d) Connect the testing leads to the circuit under check
and then read the multimeter when it stabilizes.
e) When measurement is complete, set the function
switch back to the OFF position and turn off the
multimeter.
• Diode Test (  )
• Current Measurement (μA/mA/A)
*This fucntion is not supported on the 73204 multimeter.
Black testing lead
Red testing lead
Black testing lead
Red
testing lead
NOTE
Do not mistake the following for a malfunction!
Zero calibration is only effective when the 20nF range is selected.
Figure 1 Forward-bias
Diode Test
Figure 2 Reverse-bias
Diode Test
Red testing
lead
Black testing
lead
Red testing
lead
Black testing
lead
/
4/10
To avoid damage to Instrument or equipment.
Before starting measurement, be sure to discharge the capacitor under check.
CAUTION
CAUTION
To avoid damage to instrument!
• Check to which mode the function dial is set before starting measurement.
• A current of 11 to 20 A can also be measured if the time interval is kept within
30 seconds. The buzzer will sound if the interval exceeds 30 seconds. If this
happens, immediately stop measurement. To continue measurement, wait for
2 minutes or more when restart.
The 732 series of multimeters can automatically retain the measured
value when the testing leads are handled as described below.
a) Press the AUTO.H key. (The display shows the symbol.)
b) Connect the testing leads to the object under test.
c) When the reading stabilizes, the buzzer sounds.
d) Remove the testing leads from the object under test.
e) The multimeter now shows the measured value that it retains.
f) You can repeat steps b) to e) as many times as you like as long as the
display shows the symbol.
d) When the test is complete, set the function switch
back to the OFF position and turn off the multimeter.
(1) Forward-bias Diode Test
Connect the black testing lead to the cathode and
the red testing lead to the anode (see Figure 1).
Silicon diodes should give a reading of approxi-
mately 0.5 V and light-emitting diodes a reading
between 1.5 V and 2.0 V. Note that readings
close to 0 V represent a short-circuit and the "---"
symbol indicates an opencircuit.
(2) Reverse-bias Diode Test
Connect the black testing lead to the anode and
the red testing lead to the cathode (see Figure 2).
Normally, the display shows the "---" symbol,
indicating that the diode under test is normal.
The diode is defective if the display gives a
certain voltage level.
a) Set the function switch from the OFF position to either the μA, mA or A
position. (If the magnitude of the current being measured is not known,
select the A position.)
b) When measuring AC current, press the SELECT key to select the AC
mode. (The display shows the “ ” symbol to indicate AC-mode
measurement.)
c) Plug the black testing lead into the COM input terminal and the red testing
lead into the A input terminal. If the current is in the order of milli-amperes
or less, plug the red testing lead into the μA•mA• input terminal.
d) Connect the testing leads to the circuit under test and then read the
multimeter when it stabilizes.
e) When measurement is complete, set the function switch back to the
OFF position and turn off the multimeter.
Black testing lead
Red testing lead
Black testing lead
Red testing lead
AUTO HOLD Function Specifications
AUTO POWER OFF Function
Battery Replacement
1. General Specifications
• AC Current Measurement ( A)
* This function is not supported on the 73204 multimeter.
• Resistance Measurement (Ω)
NOTE
Do not mistake the following for a malfunction!
• In DC voltage measurement, the AUTO HOLD function is only available for
ranges greater than the 4 V range.
• This function is not available for current-mode measurement.
• In a capacitor check, the AUTO HOLD function requires a few seconds before
it takes effect.
• The AUTO HOLD function cannot be applied to unstable signals.
The multimeter automatically turns off if no key is pressed for a period of
20 minutes. The multimeter will beep for approximately one minute to alert
the operator before the automatic power-off function takes effect.
(Pressing any key while the multimeter is beeping postpones the power-off
time. Pressing any key once after the power to the multimeter is automati-
cally turned off switches the multimeter on again.)
To cancel the automatic power-off function, hold down the SELECT key
and then set the function switch from the OFF to the position of any
desired measurement mode. (The AUTO POWER OFF indication turns off
when the function is canceled.)
To enable the function once again, temporarily switch the function switch
back to the OFF position, and then select the desired measurement mode.
If the batteries fall below the normal operating voltage, the symbol
turns on. If this happens, replace the batteries with new ones (AAA-size
[ANSI] batteries: 2pcs.).
To replace the batteries:
Fuse Replacement
If a current greater than the rated value flows when the multimeter is in the
current-measurement range, a protection fuse may blow. If this happens,
replace that fuse. The multimeter contains the following two types of fuses:
• Type F05 250 V/500 mA fuse
• Type F02 250 V/15 A fuse
To replace the fuse(s):
b)
c)
a)
To avoid electric shock!
• Be sure to disconnect the multimeter from the circuit under test before replacing
the batteries.
• Replace both batteries at the same time making sure to position them with the
correct polarities.
To avoid electric shock!
• Be sure to disconnect the multimeter from the circuit under test before replacing
the fuse(s).
• Do not operate the multimeter with the casing left open.
• In order to avoid damage to the multimeter or any possible accident, use fuses
of the specified rating.
a) Remove the three screws on the back of the casing.
b) Open the casing.
c) Take the battery holder out of the casing.
d) Replace the batteries with new ones and install the
battery holder back into the casing.
e) Close the casing and fasten it with the three screws.
a) Remove the three screws on the back of the casing.
b) Open the casing.
c) Remove the blown fuse from the fuse holder.
d) Install a new fuse in the holder.
e) Close the casing and fasten it with the
three screws.
F02:15 A/250 V
protection fuse
View with the Casing Open
F05:500 mA/250 V
protection fuse
• Measurement functions: AC voltage, DC voltage, AC current
(73201/02/03), DC current (73201/02/03),
resistance, continuity, diode and capacitance
(73202/03)
• Additional functions: Auto hold, manual range selection, over-range
alarm, and Auto power off.
• Display: LCD display that is capable of indicating a
significant reading of up to 4300 counts along
with the indications of the unit and function. It
shows the negative polarity only; no indication is
given for positive polarity. The display also has
the OL or over-range and lowbattery alarm
indicators.
Note: The most significant reading is 210 counts
for the diode test and 2300 counts for the
capacitor check.
• Range selection: Manual or automatic
• Sampling: 2 times/sec
• Operating temperature and humidity ranges:
0 to 50°C (Accuracy guaranteed range: 23±5°C)
Where the range is 0 to 40°C for humidity of
80%RH or less and 40 to 50°C for humidity of
70%RH or less
• Storage temperature and humidity ranges:
-20°C to 60°C; 70% RH maximum
• Battery life: Approximately 600 hours (when continuously
operated on alkaline batteries)
• Power supply: AAA-size batteries (ANSI) .........................2pcs.
• External dimensions: Approx.74 (W) × 155 (H) × 31 (D) mm
(excluding protrusions)
• Weight: Approx. 240 g (including batteries)
• Safety standards: EN 61010-1, EN 61010-031
(AC/DC 300V CAT.III, AC/DC 600 V CAT. II,
Pollution degree2) Operable altitude : 2000m
max above sea level, Indoor use
• EMC standards: EN 55011 Group 1 Class B
EN 61326-1, EN 61326-2-2
• Effect of EMS immunity: Accuracy range of reading: (Rated accuracy of
each range) ×5 for the strength of a radio-
frequency electromagnetic field of 3 V/m
• Accessories: Batteries (housed in the instrument) ..........2pcs.
Testing leads..............................................1set
Spare fuse: F05 (500 mA/250 V) ................1
F02 (15 A/250 V) .....................1
User’s manual ..........................................1
2. Accuracy (Electrical Specification)
• DC Voltage Measurement ( V)
Response: 1.5 sec maximum for 400 mV range or 1 sec maximum for other ranges
• AC Voltage Measurement ( V )
(Mean-value detection and rms-value calibration)
(Mean-value detection and rms-value calibration)
Response: 2 sec maximum
Test conditions: 23 ±5°C at 80% RH maximum
Accuracy: ±(percentage of reading + number of LSD reading)
Note: Each response noted below is a value measured in the
Range Hold mode (manual range setting).
Maximum
g) To cancel this function, press the key once again.
*1, *2 These ranges may produce a readout error equivalent to several times their resolution.
*3 A current of 11 to 20 A can also be measured if the time interval is kept within 30
seconds. The buzzer will sound if the interval exceeds 30 seconds.
Response: 1 sec maximum
• DC Current Measurement ( A)
* This function is not supported on the 73204 multimeter.
Voltage
Drop
<0.17 mV/μA
Maximum
Input Current
400 mA
The input is
protected by a
500 mA/250 V fuse
10 A The input is
protected by a
15 A/250 V fuse
<3 mV/mA
<0.04 V/A
Accuracy
73201 73202 73203
Range Resolution
μA
mA
A
400 μA*1
4000 μA
40 mA*2
400 mA
10 A*3
0.1 μA
1 μA
0.01 mA
0.1 mA
0.01 A
1% + 2
2% + 2
*1, *2 These ranges may produce a readout error equivalent to several times their resolution.
*3 A current of 11 to 20 A can also be measured if the time interval is kept within 30 seconds.
The buzzer will sound if the interval exceeds 30 seconds.
Response: 2 sec maximum
Response: 1 sec maximum for ranges lower than the 400 kΩ range, 5 sec maximum for the
4 MΩ range, and 15 sec maximum for the 40 MΩ range
• Continuity check ( )
Response: 0.2 sec maximum (for a buzzer response)
• Diode Test (
Response: 1 sec maximum
• Capacitor check ( )
Response: 1 sec maximum
µA
m A
A
Range Resolution 73201
Accuracy (40-500 Hz) Voltage
Drop
Maximum
Input Current
400 μA*1
4000 μA
40 mA*2
400 mA
10 A*3
0.1 μA
1 μA
0.01 mA
0.1 mA
0.01 A
2% + 20
2% + 5
2% + 20
2% + 5
2.5% + 20
<0.17 mV/μA
<3 mV/mA
<0.04 V/A
400 mA
The input
is protected by a
500 mA/250 V fuse
10 A
The input is
protected by a
15 A/250 V fuse
73202 73203
Range Resolution Accuracy
73201/02/03/04
0.75%+2
0.75%+1
2%+1
5%+2
Measuring
Current
<1.0 mA
<0.5 mA
<70 μA
<7 μA
<0.7 μA
<70 nA
Open-loop
Voltage
<3.4 V
<1.0 V
<0.7 V
Input Protective
Voltage
600 V
Range Resolution Range of operation
73201/02/03/04
Open-circuit
Voltage
Input Protective
Voltage
400 Ω 0.1 Ω The buzzer turns on for
resistances lower than 50 ±20 Ω.
600 V<3.4 V
Range
2 V
Resolution
0.01 V
Accuracy
73201/02/03/04
1% + 1 (for measuring currents
smaller than 1.0 mA)
Open-circuit
Voltage
<3.4 V
Input Protective
Voltage
600 V
Range Resolution Accuracy
73201/04 73202 73203
Protection Fuse
20 nF
200 nF
2 μF
20 μF
200 μF
0.01 nF
0.1 nF
0.001 μF
0.01 μF
0.1 μF
This function
is not
available.
2% + 5 typical
Readings in the 20nF
range are values after zero
calibration has been
completed.
By means of a
500 mA/250 V fuse
CAUTION
CAUTION
)
IM 73201E <P2>
400 mV
4 V
40 V
400 V
600 V
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
0.001 V
0.01 V
0.1 V
1 V
400 Ω
4 kΩ
40 kΩ
400 kΩ
4 MΩ
40 MΩ
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
/
4/10
/
4/10
mA
mA
/
4/10
A
A
"Measures for Administration of the Pollution Control of
Electronic Information Products" of the People's
Republic of China
产品中有毒有害物质或元素的名称及含量
部件名称
有毒有害物质
铅
(Pb) 汞
(Hg) 镉
(Cd) 六价铬
(Cr (VI)) 多溴联苯
(PBB) 多溴二苯醚
(PBDE)
框架(塑料) × × × × ○ ○
线路板 ASSY × × × × ○ ○
导线 × × × × ○ ○
电池 × × × × ○ ○
○:表示该部件的所有均质材料中的有毒有害物质的含量均在 SJ/T11363-2006 标准中所规定的限量
  以下
×:表示该部件中至少有一种均质材料中的有毒有害物质或元素的含量超过 SJ/T11363-2006 标准所规
定的限量要求
。
环保使用期限 :
该标识适用于2006 年 2月 28日颁布的《电子信息产品污染控制管理办法》
以及SJ /T11364 – 2006《电子信息产品污染控制标识要求》中所述,在中华人民
共和国销售的电子信息产品的环保使用期限
。
只要您遵守该产品相关的安全及使用注意事项,在自制造日起算的年限内,则不
会因产品中有害物质泄漏或突发变异,而造成对环境的污染或对人体及财产产生
注)该年数为“环保使用期限"
,并非产品的质量保证期。零件更换的推荐周期,
请参照使用说明书
。
恶劣影响
。
This User’s Manual explains the Prevention of Pollution Control of
Electronic Equipment Method in China. This manual is valid only in China.

This manual suits for next models

3

Other YOKOGAWA Multimeter manuals

YOKOGAWA 73101 User manual

YOKOGAWA

YOKOGAWA 73101 User manual

YOKOGAWA TY520 User manual

YOKOGAWA

YOKOGAWA TY520 User manual

YOKOGAWA 734 01 User manual

YOKOGAWA

YOKOGAWA 734 01 User manual

YOKOGAWA 90050B User manual

YOKOGAWA

YOKOGAWA 90050B User manual

YOKOGAWA 732 01 User manual

YOKOGAWA

YOKOGAWA 732 01 User manual

YOKOGAWA 733 01 User manual

YOKOGAWA

YOKOGAWA 733 01 User manual

YOKOGAWA CA450 User manual

YOKOGAWA

YOKOGAWA CA450 User manual

YOKOGAWA CW10 User manual

YOKOGAWA

YOKOGAWA CW10 User manual

YOKOGAWA 73201 User manual

YOKOGAWA

YOKOGAWA 73201 User manual

YOKOGAWA DM7560 User manual

YOKOGAWA

YOKOGAWA DM7560 User manual

YOKOGAWA TY710 User manual

YOKOGAWA

YOKOGAWA TY710 User manual

YOKOGAWA 7555 User manual

YOKOGAWA

YOKOGAWA 7555 User manual

YOKOGAWA CA450 User manual

YOKOGAWA

YOKOGAWA CA450 User manual

Popular Multimeter manuals by other brands

Unit UT151D operating manual

Unit

Unit UT151D operating manual

Unit UT107 operating manual

Unit

Unit UT107 operating manual

BASETech BT-11 operating instructions

BASETech

BASETech BT-11 operating instructions

HoldPeak 770A Operator's manual

HoldPeak

HoldPeak 770A Operator's manual

HoldPeak 770HD Operator's manual

HoldPeak

HoldPeak 770HD Operator's manual

Keysight U1271A quick start guide

Keysight

Keysight U1271A quick start guide

APPA 61 user manual

APPA

APPA 61 user manual

Maxwell MX-25 302 user guide

Maxwell

Maxwell MX-25 302 user guide

Draper DCM1B Instructions for use

Draper

Draper DCM1B Instructions for use

RS PRO IDM-8351 instruction manual

RS PRO

RS PRO IDM-8351 instruction manual

Kopp GDT-292 A owner's manual

Kopp

Kopp GDT-292 A owner's manual

GW Instek GDM-8246 Short instructions

GW Instek

GW Instek GDM-8246 Short instructions

Elma Instruments 6800 user manual

Elma Instruments

Elma Instruments 6800 user manual

ZOTEK ZT219 user manual

ZOTEK

ZOTEK ZT219 user manual

Craftsman 82139 owner's manual

Craftsman

Craftsman 82139 owner's manual

Owon XDM1000 Series user manual

Owon

Owon XDM1000 Series user manual

AMES 64021 Owner's manual & safety instructions

AMES

AMES 64021 Owner's manual & safety instructions

Amprobe AM-420 user manual

Amprobe

Amprobe AM-420 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.