DESCRIPTION
Carl Zeiss Microscope systems Fluar Illuminator Z mot.
12 435180-9060-701 05/2012
2.3 Microscope systems
Fig. 2 Axio Zoom.V16 with coarse/fine drive
Axio Zoom.V16 with manual coarse/fine drive
AMicroscope body (Axio Zoom.V16) with
control unit (HIP)
BBinocular photo tube
CEyepiece
DObjective
FIntermediate tube (Fluar Illuminator Z mot.)
GCamera adapter with microscope camera
(AxioCam)
HReflected light illumination (slit-ring light with
cold light source CL 6000 LED)
PStand base with coarse/fine drive
Fig. 3 Axio Zoom.V16 with focusing motor
Axio Zoom.V16 with focusing motor
AMicroscope body (Axio Zoom.V16) with
control unit (HIP)
BBinocular photo tube
CEyepiece
DObjective
ENosepiece
FIntermediate tube (Fluar Illuminator Z mot.)
GCamera adapter with microscope camera
(AxioCam)
HReflected light illumination (slit-ring illuminator
with cold light source CL 6000 LED)
IUV/visual glare shield
KStand base with focusing motor with control unit
(MaRC)
LTransmitted light illumination (transmitted light
illuminator S)
MSpecimen stage (gliding stage)
NSYCOP 3 system control panel
OEMS 3 controller with optional foot pedal S
The items of microscope equipment shown here
are examples and may differ from those
actually provided!
2.4 Technical data
Dimensions
Fluar Illuminator Z mot. (length x width x height) ................................................31 mm x 277 mm x 92 mm
HXP 200 C illuminator (width x height x depth) ............................................... 130 mm x 210 mm x 290 mm
Weight
Fluar Illuminator Z mot. ......................................................................................................................3 kg
HXP 200 C illuminator........................................................................................................................5 kg
Power supply
Fluar Illuminator Z mot. ..............................................................over CAN bus connection from microscope
HXP 200 C illuminator ............................................................................. 100 V AC to 240 V AC, 50/60 Hz
Power consumption
HXP 200 C illuminator ...........................................................................................................max. 320 VA
Mains fuse
HXP 200 C illuminator ....................................................................................................... 2x T4A/H/250V
Ambient conditions
Storage (in packaging)
Permissible ambient temperature.......................................................................................... +10 to +40 °C
Permissible relative humidity ............................................................ max. 75 % at +35 °C (non-condensing)
Transport (in packaging):
Permissible ambient temperature........................................................................................... -40 to +70 °C
Operation
Permissible ambient temperature.......................................................................................... +10 to +40 °C
Permissible relative humidity ..................................................................................................... max. 75 %
Air pressure...............................................................................................................800 hPa to 1060 hPa
Degree of pollution................................................................................................................................ 2
Optical risk group classification acc. to DIN EN 62471:2009
Axio Zoom.V16 microscope system with:
HXP 200 C ............................................................................................... Risk group 1 acc. to DIN EN 62471:2009
CL 9000 LED....................................................................................... LED risk group 1 acc. to DIN EN 62471:2009
Transillumination top 450 mot............................................................ LED risk group 1 acc. to DIN EN 62471:2009
Transillumination base 300 ................................................................. LED risk group 1 acc. to DIN EN 62471:2009
Radiation apertures
Axio Zoom.V16 microscope system with:
HXP 200 C, CL 9000 LED........................................................................... vertically downwards from the objective
Transillumination top 450 mot. ..............................................vertically upwards from the transmitted light adapter
Transillumination base 300........................................................ vertically upwards from the transmitted light base