manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. 3B Medical
  6. •
  7. Medical Equipment
  8. •
  9. 3B Medical RESmart BPAP User manual

3B Medical RESmart BPAP User manual

This manual suits for next models

1

Other 3B Medical Medical Equipment manuals

3B Medical Luna GII User manual

3B Medical

3B Medical Luna GII User manual

3B Medical Luna Series User manual

3B Medical

3B Medical Luna Series User manual

3B Medical LUNA II User manual

3B Medical

3B Medical LUNA II User manual

3B Medical Stratus 5 User manual

3B Medical

3B Medical Stratus 5 User manual

3B Medical Line II QX Series User manual

3B Medical

3B Medical Line II QX Series User manual

3B Medical Lumin LM3000EU User manual

3B Medical

3B Medical Lumin LM3000EU User manual

3B Medical Cirrus 5 User manual

3B Medical

3B Medical Cirrus 5 User manual

3B Medical Luna G3 BPAP 25A User manual

3B Medical

3B Medical Luna G3 BPAP 25A User manual

3B Medical Aer X User manual

3B Medical

3B Medical Aer X User manual

3B Medical LUNA G3 User manual

3B Medical

3B Medical LUNA G3 User manual

Popular Medical Equipment manuals by other brands

Itamar Medical WatchPAT Operation manual

Itamar Medical

Itamar Medical WatchPAT Operation manual

PROTEOR 1P650 KNEE Assembling and adjustment instructions

PROTEOR

PROTEOR 1P650 KNEE Assembling and adjustment instructions

Liberate Medical VentFree user manual

Liberate Medical

Liberate Medical VentFree user manual

Dicarre DA20 user guide

Dicarre

Dicarre DA20 user guide

Stryker Medical M Series Operation manual

Stryker Medical

Stryker Medical M Series Operation manual

Invacare ScanBed 750 user manual

Invacare

Invacare ScanBed 750 user manual

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Getinge

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics Sonicator 730 Maintenance manual

Mettler Electronics

Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care R1100 Mounting instruction

Pressalit Care

Pressalit Care R1100 Mounting instruction

Denas MS DENAS-T operating manual

Denas MS

Denas MS DENAS-T operating manual

bort medical ActiveColor quick guide

bort medical

bort medical ActiveColor quick guide

AccuVein AV400 user manual

AccuVein

AccuVein AV400 user manual

agiliti NP Trio user manual

agiliti

agiliti NP Trio user manual

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Rocket Medical

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Monteris Medical Neuroblate Instructions for use

Monteris Medical

Monteris Medical Neuroblate Instructions for use

Otto Block 8E33 7 manual

Otto Block

Otto Block 8E33 7 manual

ADVANCED HOME CARE Helios Guide

ADVANCED HOME CARE

ADVANCED HOME CARE Helios Guide

NeuLog NUL-208 Guide

NeuLog

NeuLog NUL-208 Guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Review Prior Night Summary
Reviewing Sleep Data
The RESmart®allows review of the prior night’s sleep data. To access
your data, make sure that the blower is not running by pressing the
stop/start button.
Press the Ramp Start button , and scroll through the display
using the +/- buttons. The data displayed consists of completion
time (Compl), P95 (95th percentile of the airway pressure over
time), AHI (apnea-hypopnea index), and SNI (snore index).
For a discussion on interpreting this data, contact your physician or
home care provider.
3B Medical, Inc.
21301 US Highway 27 | Lake Wales, FL 33859 | Phone: (863) 226-6285
Fax: (863) 226-6284 | [email protected] | www.3BProducts.com
Register Your New Account at www.iCodeConnect.com
With iCode Connect you can:
• Easily maintain patient records.
• Generate and archive compliance reports.
• Store patient information, including prescription information for device, mask,
and accessories; model numbers and size information; patient, physician, and
provider contact information; insurance information; reminders; and notes.
• Communicate and share information in real time with other members of your
patient’s sleep therapy team.
iCode and iCodeConnect are free of charge with all 3B Medical products.
Quick Reference Sheet
RESmart®BPAP & Auto-BPAP
1484_08.2014
Clinician Set-Up Menu
Display
Screen
User
Button (+)
Ramp Button
Pressure Start/Stop
Button
User
Button (–)
Heated
Humidifier
Button
To enter the Clinician’s Set-Up Menu:
Remove the power cord from the rear of the RESmart®unit. Press
and hold both the Start and Ramp buttons simultaneously, while
reconnecting the power cord.
To Select an Option:
Press the Ramp button once. The cursor will blink, and values may be
changed using the user buttons +/-.
Quickly press and release user buttons.
To Confirm a Setting:
Press the Ramp button again and the cursor will stop blinking
To Advance to the next Option:
Press the User + or - button to move through the Clinician Menu
To SAVE Menu Selections:
Press the Ramp button when the display screen prompts “SAVE”
NOTE: Changes to any options will be final only when you have
confirmed them by pressing the RAMP button when you are at the SAVE
selection on the display screen.
Treat P
Init P
Ramp
epap
ipap
Ramp
Init ipap
Rise Time
isense
esense
Ti Min
CLINIC MENU
CPAP S Mode
Auto
S Mode
Ti Max
Unit
Use Time
Date
Save
epap
ipap
Ramp
Max ipap
init epap
Rise Time
isense
esense
Ti Min
Ti Max
Treat P (CPAP Mode only): Indicates CPAP Treatment Pressure. Pressure Range
from 4 to 20 cm H2O in 0.5 cm H2O increment.
Init P (CPAP Mode only): The starting pressure if a Ramp time is set. Setting range
is 4 cm H2O to Treatment P (Prescription Pressure). Default setting is 4 cm H2O.
EPAP (S and Auto S Mode):
Expiratory Positive Airway Pressure.
Setting range is 4-25 cm H2O.
Cannot go above IPAP Pressure.
Default setting is 6 cm H2O.
IPAP (S and Auto S Mode):
Inspiratory Positive Airway Pressure.
Setting range is 4-25 cm H2O..
Default setting is 8 cm H2O.
Ramp P: the starting pressure
after blower is turned on when
RAMP Time is set at anything above
0 minutes. The setting range is
4cmH2O to Treat P or Init EPAP
Init EPAP (S and Auto S Mode):
When a Ramp time is set, this will
be the initial pressure to begin ramp
pressure rise.
Max IPAP (Auto S Mode only):
This setting is the maximum IPAP
Pressure. Default setting is 25 cm
H2O.
Ti Max (S and Auto S Mode):
Maximum Inspiration Time. It refers
to the maximum duration of IPAP in
each respiratory cycle. Setting range
is 0.3-3.0 Seconds. Adjustment step is
0.1 S. Default setting is 2.0 S.
Ti Min (S and Auto S Mode): Minimum Inspiration Time. It refers to the
minimum duration of IPAP in each respiratory cycle. Setting range is 0.1-Ti Max.
Adjustment step is 0.1 S. Default setting is 0.5 S.
Rise Time (S and Auto S mode): is the time it takes for the device to transition
from EPAP to IPAP once inspiration has triggered. It is adjustable by the provider
from 0 to 3, default setting is 2.
I Sense (S and Auto S mode): This Setting can adjust the trigger sensitivity when
patient’s breath switches into inspiratory phase. It can be adjusted from 1 (less
sensitive) to 8 (most sensitive).Default setting is 6
E Sense (S and Auto S mode): This Setting adjusts the trigger sensitivity when
a patient’s breath switches into expiration phase. It can be adjusted from 1 (less
sensitive) to 8 (most sensitive). Default setting is 7.