Treat P (CPAP Mode only): Indicates CPAP Treatment Pressure. Pressure Range
from 4 to 20 cm H2O in 0.5 cm H2O increment.
Init P (CPAP Mode only): The starting pressure if a Ramp time is set. Setting range
is 4 cm H2O to Treatment P (Prescription Pressure). Default setting is 4 cm H2O.
EPAP (S and Auto S Mode):
Expiratory Positive Airway Pressure.
Setting range is 4-25 cm H2O.
Cannot go above IPAP Pressure.
Default setting is 6 cm H2O.
IPAP (S and Auto S Mode):
Inspiratory Positive Airway Pressure.
Setting range is 4-25 cm H2O..
Default setting is 8 cm H2O.
Ramp P: the starting pressure
after blower is turned on when
RAMP Time is set at anything above
0 minutes. The setting range is
4cmH2O to Treat P or Init EPAP
Init EPAP (S and Auto S Mode):
When a Ramp time is set, this will
be the initial pressure to begin ramp
pressure rise.
Max IPAP (Auto S Mode only):
This setting is the maximum IPAP
Pressure. Default setting is 25 cm
H2O.
Ti Max (S and Auto S Mode):
Maximum Inspiration Time. It refers
to the maximum duration of IPAP in
each respiratory cycle. Setting range
is 0.3-3.0 Seconds. Adjustment step is
0.1 S. Default setting is 2.0 S.
Ti Min (S and Auto S Mode): Minimum Inspiration Time. It refers to the
minimum duration of IPAP in each respiratory cycle. Setting range is 0.1-Ti Max.
Adjustment step is 0.1 S. Default setting is 0.5 S.
Rise Time (S and Auto S mode): is the time it takes for the device to transition
from EPAP to IPAP once inspiration has triggered. It is adjustable by the provider
from 0 to 3, default setting is 2.
I Sense (S and Auto S mode): This Setting can adjust the trigger sensitivity when
patient’s breath switches into inspiratory phase. It can be adjusted from 1 (less
sensitive) to 8 (most sensitive).Default setting is 6
E Sense (S and Auto S mode): This Setting adjusts the trigger sensitivity when
a patient’s breath switches into expiration phase. It can be adjusted from 1 (less
sensitive) to 8 (most sensitive). Default setting is 7.