4 Medical device description
4.1 First use
The Mini LED uses a Lithium-Ion battery. To guarantee optimum use, this battery must be fully charged before
use but must never be flat. To prepare your Mini LED, follow the steps below:
lclean the rigid protection shield with an alcohol wipe;
lclean the handpiece with an alcohol wipe;
lplug the charging base into the mains;
lcheck that the green indicator light on the base lights up twice and listen out for a beep sound;
linstall the battery on the handpiece;
lplace the Mini LED on its charging base;
lthe Mini LED is correctly installed when two beeps are heard and the green indicator light on the base
starts flashing;
lleave the Mini LED to fully charge;
lthe Mini LED is charged when the green indicator light on the base stops flashing and remains on
permanently;
lremove the protection plug from the handpiece nozzle;
linsert the sterilised optical guide into the handpiece nozzle;
lyou will hear a click when the optical guide is correctly inserted;
linstall the rigid protection shield;
lwear safety goggles and protective gloves;
lprovide your patient with safety goggles.
4.2 Using the Mini LED
The Mini LED is normally placed on its base. To use it, remove it from its base.
Prior to the day of use, check that you have enough sterilised optical guides and check the power output as
indicated in the chapter Checking the lamp power level page 19.
The patient and the practitioner must wear class II safety goggles when the Mini LED is in operation.
lRemove the Mini LED from the base;
lPress the on/off button.
lInstall the rigid protection shield;
lProgramme the Mini LED to select the required mode;
lPosition the end of the optical guide as close as possible to the surface of the material to be cured;
Do not allow the optical guide to touch the material to be cured.
lPress the on/off button to trigger the selected mode. A beep sound confirms cycle initiation;
lA second beep sound confirms cycle completion.
You can stop the cure cycle at any time by pressing the on/off button.
After three minutes of inactivity, the Mini LED switches to standby.
Depending on the cure material used, repeat the cure cycle as required.
4.3 Mini LED™
The Mini LED can only be used with the following accessories:
l7.5 mm opalescent optical guide [F02648]
l7.5 mm amber-coloured optical guide [F02620]
lMini LED titanium battery [F02520]
lMini LED handpiece [F02530]
lProtection plug [F57347]
lRigid protection shield [F02555]
lCharging base and mains adapter [F02544]
User Manual • Mini LED™ • J02541 • V5 • 07 • 01/2015 • NG13EN010E - Page 11/30