2.4 X-MIND TRIUM CONFIGURATIONS
X-MIND trium can be sold in these congurations:
PAN only
• The equipment can carry out uniquely exams of the PAN group (PAN, TMJ, Sinus);
• The image detector is xed to the PAN bay;
• All the exams are carried out with the same extension in height (146 mm height on image receptor), primary collimator
being of xed dimension;
• The equipment can be upgraded in a later stage by replacing and adding several parts.
PAN and CBCT
• The equipment can carry out both exams of the PAN group (PAN, TMJ, Sinus) and of the CBCT group;
• There are two image detectors, one for PAN and one for CBCT exam; the equipment automatically exposes in front of X-Ray
source the right detector, depending on the selected exam;
• PAN exams can be carried out with different extension in height (up to 146 mm height on image receptor), since the
collimator height can be adjusted depending on the exam selected by the user;
• CBCT exams can be carried out at different FOV dimensions as specied in the technical data appendix;
• The equipment can be upgraded in the eld to CEPH conguration by simple procedure, adding CEPH arm with CEPH image
detector replacing the existing PAN image detector.
PAN / CEPH
• The equipment can carry out both exams of the PAN group (PAN, TMJ, Sinus) and of the CEPH group (AP/PA, LL, carpus);
• The image detector can optionally be unique, both for PAN and CEPH exams, and in this case it must be manually moved by
the operator from PAN bay to CEPH bay and viceversa, depending on the selected exam; otherwise two different detectors
can be dedicated: one for CEPH exams and the other one for PAN exams;
• The exams can be carried out with different extension in height (up to 146 mm height on image receptor for PAN, up to 220
mm height on image receptor for CEPH), since the primary collimator can be automatically adjusted in height depending on
exam selected by the user;
• The equipment can be upgraded in the eld to CBCT conguration by a simple procedure, adding CBCT image detector and
bite block holder with vertical movement.
PAN / CBCT / CEPH
• The equipment can carry out all feasible exams of the PAN group (PAN, TMJ, Sinus), of the CEPH group (AP/PA, LL, carpus)
and of the CBCT group;
• There are optionally two image detectors, one for PAN/CEPH and one for CBCT exam; in this case the PAN/CEPH image
detector must be manually moved by the operator from PAN bay to CEPH bay and viceversa; otherwise three image detectors,
one for PAN, one for CEPH and one for CBCT exams;
• To carry out PAN or CEPH or CBCT exam the equipment automatically exposes in front of X-Ray source the right detector,
depending on the selected exam;
• PAN and CEPH exams can be carried out with different extension in height (up to 146 mm height on image receptor for
PAN, up to 220 mm height on image receptor for CEPH), since the primary collimator can be automatically adjusted in height
depending on exam selected by the user;
• CBCT exams can be carried out at different FOV dimensions as specied in the technical data appendix.
2.5 EXAMS
In its different congurations X-MIND trium equipment carries out the following exams:
Panoramic group
The panoramic group includes all the 2D radiographs of the jaws, dentition, temporomandibular joints and maxillary sinuses.
1) Standard panoramic
· Full scan
· Right side only
· Left side only