SPECIFICATIONS
Capacity 20 kg, 44 lb:oz
Readability 0.005 kg 1 oz.
Weighing units Kg, and Lb:Oz
Power 6V, 500ma adaptor
Batteries 4 x AA size
Pan Size 560 x 285 mm
(22” x 11.2”)
Overall Size (w x h x d) 560 x 350 x 140 mm
(22" x 13.75" x 5.5")
Net weight 3.2 kg (7 lb.)
OPERATION
Install batteries, ensuring the polarity is as shown in the battery holder, or insert the power supply into
the connector on the rear of the scale.
Place the scale on a flat, very stable surface, and clear the scale platform.
Press the [ON/ZERO] key, turning on the scale.
Display will show 0.000 kg or 0 0.00lb:oz. depending on the last weighing unit selected.
It is now ready to weigh.
Put the subject on the scale platform, the display shows the weight.
Press the [ON/ZERO] key briefly to tare the scale, setting the display to 0.00.
Any additional weight added will displayed.
SELECT WEIGHING UNIT
Press the [MODE] key to change from kilograms to pounds:ounces and back again.
The display will show the weighing unit selected.
STORE THE WEIGHT
To store the displayed weight Press the [MEM] key when the LCD display is stable.
RECALL THE STORED THE WEIGHT
To recall the stored weight Press the [MEM] key when the LCD display is at zero
Press it a second time to return to weighing.
To turn off the scale.
Press and hold the [ON/ZERO] key
The scale will turn off automatically after 5 minutes if it is not being used.
CALIBRATION
Place the scale on a flat, very stable surface, and clear the scale platform.
When the scale is off press the [MODE] key, then press the [ON/ZERO] key. Hold both keys down
until the display shows “CAL” then release both keys.
The display will show a number representing the internal count from the sensor
Press the [ON/ZERO] key briefly.
The display will flash “C10 “ and then show “ 0”.
Place a 10kg mass on the platform.
The display will show a number from the sensor.
Press the [ON/ZERO] key.
The display will flash “C20 “ and then show “ 0000”.
Press the [ON/ZERO] key again, the display will show “End” then turn off.
The calibration is complete.
Remove the 10kg mass from the platform and turn on the scale.
Place the 10kg on the scale, it should display 10.000kg ±0.010kg. If it is not within acceptable
values repeat the calibration
.
WARNING MESSAGES
“Loo”: Low power, please change the batteries.
“o_Ld”: Over loading, please remove the item immediately to protect the scale.