99-MAN 100119.v6 07/13
OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model MD304
Connect the black and red terminals to the Ampere AC range of your DMM or current measuring
instrument. Select the appropriate current range (10A AC range). Clamp the probe around the conductor
to be tested with the arrow pointed toward the load. Read the value display on the DMM and multiply it by
the probe ratio (100/1). (If reading = 4.59A, the current flowing through the probe is 4.59A x 100 = 459A
AC)
For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
For maintenance use only original replacement parts.
To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into
the probe.
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft, oil cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container.
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive Dover, NH 03820 USA
(800) 945-2362 (Ext. 360) • (603) 749-6434 (Ext. 360) • [email protected]
(Or contact your authorized distributor)
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or
application of this instrument, please contact our technical hotline:
(800) 343-1391 • (508) 698-2115 • [email protected]
LIMITED WARRANTY
The current probe is warranted to the owner for a period of one year from the date of original purchase
against defects in manufacture. This limited warranty is given by AEMC®Instruments, not by the
distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused
or if the defect is related to service not performed by AEMC®Instruments.
Full warranty coverage and product registration is available on our website at:
www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for your records.
AC Current Probe
Model MD304
User Manual
DESCRIPTION
The MD304 (Catalog #1201.37) is designed for use in industrial environments. The hook-shaped jaws
enable the user to “pry” into or “hook” onto cables (will accept 2 x 500 MCM) or even smaller bus bars.
The current output up to 500Arms continuous makes it the perfect tool for measurement with DMMs,
recorders or instruments with current ranges. The Model MD304 offers a 5 ft lead with safety 4mm
banana plugs.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument.
• Read the instruction manual completely and follow all the safety information before attempting
to use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may
pose a shock hazard.
• Read the Safety Specifications section prior to using the current probe. Never exceed the
maximum voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use.
Replace any defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 600V in overvoltage
category III (CAT III). Use extreme caution when clamping around bare conductors or bus bars.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type A current sensor. This symbol signifies that application around and removal
from HAZARDOUS LIVE conductors is permitted.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.
PACKAGING
The AC Current Probe MD304 is shipped with this instruction manual and a product warranty and
registration card. You can register online at www.aemc.com.
INSTRUMENT COMPATIBILITY
The Model MD304 is compatible with any AC ammeter, multimeter, or other current measurement
instruments with an input impedance lower than 0.1. To achieve the stated accuracy, use the MD304
with an Ammeter having an accuracy of 0.75% or better.
ELECTRICAL SPECIFICATIONS
Current Range:
1 to 600A AC
Transformation Ratio:
100:1
Output Signal:
10mA AC/A AC
Overload:
700A for 10 mn
Accuracy*:
Primary
current 25A 100A 250A;
500A 600A
Accuracy % 5% 2.5% 2% 3%
Phase shift 6° 5° 3° 3°
600A for 20 mn max.
(*Reference conditions: 23°C ± 5°K, 20 to 75%
RH, 45 to 65Hz, external magnetic field < 40 A/m, no
DC component, no external current carrying
conductor, test sample centered.) 0.1load.
Frequency Range:
48 to 1000Hz
(error: add 1% to ref.)
Load Impedance:
0.1max. non-inductive
Working Voltage: 600V rms
Common Mode Voltage: 30V rms
Influence of Adjacent Parallel Conductor:
< 30mA/A at 50Hz
Influence of Conductor in Jaw Opening:
< 1%
MECHANICAL SPECIFICATIONS
Operating Temperature:
-5° to 122°F (-15° to 50°C)
Storage Temperature:
Influence of Temperature:
< 0.1% per 10°K
Altitude:
Operating: 0 to 2000 m
Non-operating: 0 to 12,000 m
Jaw Opening:
1.3" (33mm)
Maximum Conductor Size:
1.18” (30mm)
Maximum Bus Bar Size:
2.48 x 0.20” (63 x 5mm)
Envelope Protection:
IP 20 (IEC 529)
Drop Test:
1.5 m (IEC 68-2-32)
Mechanical Shock:
100 g (IEC 68-2-27)
Vibration:
10/55/10Hz, 0.15mm (IEC 68-2-6)
Polycarbonate Material:
Handles: 10% fiberglass charged
polycarbonate UL 94 V0
Dimensions:
2.6 x 7.68 x 1.34" (66 x 195 x 34mm)
Weight:
14.82 oz. (420 g)
Colors:
Dark gray handles
Output:
Insulated 5 ft (1.5 m) lead with safety 4mm
banana plugs
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation
between the primary or secondary and the outer
case of the handle per IEC 1010-2-032.
- 600 V CAT III, Pollution: 2
- 300 V CAT IV, Pollution: 2
Electromagnetic Compatibility:
EN 50081-1 Class B
EN 50082-2 Electrostatic discharge
IEC 1000-4-2
Radiated field IEC 1000-4-3
Fast transients IEC 1000-4-4
Magnetic field at 50/60 Hz IEC 1000-4-8
ORDERING INFORMATION
Current Probe MD304..................Cat.# 1201.37
Accessories:
Banana plug adapter
(to nonrecessed plug) ...................Cat.# 1017.45
Black
S2
2.6"
(66 mm)
7.68"
(195 mm)
Ø =1.18" (30 mm)
Cable max.
2.5 x 0.19" (63 x 5 mm)
Bus Bar max.
Ø =0.98" (25 mm)
Dual Cable max.
I
1.34"
(34 mm)
Red
S1
ELECTRICAL SPECIFICATIONS
Current Range:
1 to 600A AC
Transformation Ratio:
100:1
Output Signal:
10mA AC/A AC
Overload:
700A for 10 mn
Accuracy*:
Primary
current 25A 100A 250A;
500A 600A
Accuracy % 5% 2.5% 2% 3%
Phase shift 6° 5° 3° 3°
600A for 20 mn max.
(*Reference conditions: 23°C ± 5°K, 20 to 75%
RH, 45 to 65Hz, external magnetic field < 40 A/m, no
DC component, no external current carrying
conductor, test sample centered.) 0.1load.
Frequency Range:
48 to 1000Hz
(error: add 1% to ref.)
Load Impedance:
0.1max. non-inductive
Working Voltage: 600V rms
Common Mode Voltage: 30V rms
Influence of Adjacent Parallel Conductor:
< 30mA/A at 50Hz
Influence of Conductor in Jaw Opening:
< 1%
MECHANICAL SPECIFICATIONS
Operating Temperature:
-5° to 122°F (-15° to 50°C)
Storage Temperature:
Influence of Temperature:
< 0.1% per 10°K
Altitude:
Operating: 0 to 2000 m
Non-operating: 0 to 12,000 m
Jaw Opening:
1.3" (33mm)
Maximum Conductor Size:
1.18” (30mm)
Maximum Bus Bar Size:
2.48 x 0.20” (63 x 5mm)
Envelope Protection:
IP 20 (IEC 529)
Drop Test:
1.5 m (IEC 68-2-32)
Mechanical Shock:
100 g (IEC 68-2-27)
Vibration:
10/55/10Hz, 0.15mm (IEC 68-2-6)
Polycarbonate Material:
Handles: 10% fiberglass charged
polycarbonate UL 94 V0
Dimensions:
2.6 x 7.68 x 1.34" (66 x 195 x 34mm)
Weight:
14.82 oz. (420 g)
Colors:
Dark gray handles
Output:
Insulated 5 ft (1.5 m) lead with safety 4mm
banana plugs
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation
between the primary or secondary and the outer
case of the handle per IEC 1010-2-032.
- 600 V CAT III, Pollution: 2
- 300 V CAT IV, Pollution: 2
Electromagnetic Compatibility:
EN 50081-1 Class B
EN 50082-2 Electrostatic discharge
IEC 1000-4-2
Radiated field IEC 1000-4-3
Fast transients IEC 1000-4-4
Magnetic field at 50/60 Hz IEC 1000-4-8
ORDERING INFORMATION
Current Probe MD304..................Cat.# 1201.37
Accessories:
Banana plug adapter
(to nonrecessed plug) ...................Cat.# 1017.45
Black
S2
2.6"
(66 mm)
7.68"
(195 mm)
Ø =1.18" (30 mm)
Cable max.
2.5 x 0.19" (63 x 5 mm)
Bus Bar max.
Ø =0.98" (25 mm)
Dual Cable max.
I
1.34"
(34 mm)
Red
S1
99-MAN 100119.v703/18
OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model MD304
Connect the black and red terminals to the Ampere AC range of your DMM or current measuring
instrument. Select the appropriate current range (10A AC range). Clamp the probe around the conductor
to be tested with the arrow pointed toward the load. Read the value display on the DMM and multiply it by
the probe ratio (100/1). (If reading = 4.59A, the current flowing through the probe is 4.59A x 100 = 459A
AC)
For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
For maintenance use only original replacement parts.
To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into
the probe.
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft, oil cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container.
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive Dover, NH 03820 USA
(800) 945-2362 (Ext. 360) • (603) 749-6434 (Ext. 360) • [email protected]
(Or contact your authorized distributor)
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or
application of this instrument, please contact our technical hotline:
(800) 945-2362 (Ext. 351) • (603) 749-6434 (Ext. 351) • [email protected]
LIMITED WARRANTY
The current probe is warranted to the owner for a period of two yearsfrom the date of original pur-
chase against defects in manufacture. This limited warranty is given by AEMC®Instruments, not
by the distributor from whom it was purchased. This warranty is void if the unit has been tampered
with, abused or if the defect is related to service not performed by AEMC®Instruments.
Full warranty coverage and product registration is available on our website at:
www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for your records.
AC Current Probe
Model MD304
User Manual
DESCRIPTION
The MD304 (Catalog #1201.37) is designed for use in industrial environments. The hook-shaped jaws
enable the user to “pry” into or “hook” onto cables (will accept 2 x 500 MCM) or even smaller bus bars.
The current output up to 500Arms continuous makes it the perfect tool for measurement with DMMs,
recorders or instruments with current ranges. The Model MD304 offers a 5 ft lead with safety 4mm
banana plugs.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument.
• Read the instruction manual completely and follow all the safety information before attempting
to use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may
pose a shock hazard.
• Read the Safety Specifications section prior to using the current probe. Never exceed the
maximum voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use.
Replace any defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 600V in overvoltage
category III (CAT III). Use extreme caution when clamping around bare conductors or bus bars.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type A current sensor. This symbol signifies that application around and removal
from HAZARDOUS LIVE conductors is permitted.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.
PACKAGING
The AC Current Probe MD304 is shipped with this instruction manual and a product warranty and
registration card. You can register online at www.aemc.com.
INSTRUMENT COMPATIBILITY
The Model MD304 is compatible with any AC ammeter, multimeter, or other current measurement
instruments with an input impedance lower than 0.1. To achieve the stated accuracy, use the MD304
with an Ammeter having an accuracy of 0.75% or better.

Other AEMC Measuring Instrument manuals

AEMC PowerPad 8436 User manual

AEMC

AEMC PowerPad 8436 User manual

AEMC SL261 User manual

AEMC

AEMC SL261 User manual

AEMC 1227 User manual

AEMC

AEMC 1227 User manual

AEMC 6527 User manual

AEMC

AEMC 6527 User manual

AEMC MN186 User manual

AEMC

AEMC MN186 User manual

AEMC 1050 User manual

AEMC

AEMC 1050 User manual

AEMC 8230 User manual

AEMC

AEMC 8230 User manual

AEMC 6526 User manual

AEMC

AEMC 6526 User manual

AEMC A196-24-BK User manual

AEMC

AEMC A196-24-BK User manual

AEMC 500 User manual

AEMC

AEMC 500 User manual

AEMC 501N User manual

AEMC

AEMC 501N User manual

AEMC MR417 User manual

AEMC

AEMC MR417 User manual

AEMC JM875 User manual

AEMC

AEMC JM875 User manual

AEMC SR701 User manual

AEMC

AEMC SR701 User manual

AEMC F09 User manual

AEMC

AEMC F09 User manual

AEMC 6505 User manual

AEMC

AEMC 6505 User manual

AEMC MR521 User manual

AEMC

AEMC MR521 User manual

AEMC PowerPad 8335 Setup guide

AEMC

AEMC PowerPad 8335 Setup guide

AEMC MN373 User manual

AEMC

AEMC MN373 User manual

AEMC MN375 User manual

AEMC

AEMC MN375 User manual

AEMC CA7027 User manual

AEMC

AEMC CA7027 User manual

AEMC 6416 User manual

AEMC

AEMC 6416 User manual

AEMC PowerPad 8436 User manual

AEMC

AEMC PowerPad 8436 User manual

AEMC K100 User manual

AEMC

AEMC K100 User manual

Popular Measuring Instrument manuals by other brands

Sanwa DCM2000DR instruction manual

Sanwa

Sanwa DCM2000DR instruction manual

GROCO SSA-3 Series Installation and operating instructions

GROCO

GROCO SSA-3 Series Installation and operating instructions

Hanna Instruments HI96738 user manual

Hanna Instruments

Hanna Instruments HI96738 user manual

Bosch PLR 50 Original instructions

Bosch

Bosch PLR 50 Original instructions

METREL MI 3205 instruction manual

METREL

METREL MI 3205 instruction manual

BEA LZR-MICROSCAN U920 user guide

BEA

BEA LZR-MICROSCAN U920 user guide

BizLine BIZ 700 058 instruction manual

BizLine

BizLine BIZ 700 058 instruction manual

HT HT701 user manual

HT

HT HT701 user manual

MGL Avionics Blaze ASI-5 operating manual

MGL Avionics

MGL Avionics Blaze ASI-5 operating manual

Chauvin Arnoux C.A 1725 user manual

Chauvin Arnoux

Chauvin Arnoux C.A 1725 user manual

Olympus 45MG Getting started guide

Olympus

Olympus 45MG Getting started guide

Ametek Reichert 13104190 instruction manual

Ametek

Ametek Reichert 13104190 instruction manual

PANCONTROL PAN KLM-2000 manual

PANCONTROL

PANCONTROL PAN KLM-2000 manual

PP Systems TARGAS-1 Operation manuals

PP Systems

PP Systems TARGAS-1 Operation manuals

ThermoPro TP359 quick start guide

ThermoPro

ThermoPro TP359 quick start guide

Water I.D. PoolLab 2.0 user manual

Water I.D.

Water I.D. PoolLab 2.0 user manual

LumaSense IMPAC IN 500 Series Operation manual

LumaSense

LumaSense IMPAC IN 500 Series Operation manual

Ludlum Measurements 3-97 manual

Ludlum Measurements

Ludlum Measurements 3-97 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.