AEMC SR661 User manual

AC CURRENT
OSCILLOSCOPE PROBE
SR661
ENGLISH User Manual

Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
certifies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be
requested at the time of purchase, or obtained
by returning the instrument to our repair and
calibration facility, for a nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer. For recalibration, please
use our calibration services. Refer to our repair
and calibration section at www.aemc.com.
Serial #: ________________________________
Catalog #: 2113.49
Model #: SR661
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.
d.b.a AEMC®Instruments
www.aemc.com

Table of Contents
1. INTRODUCTION............................................................................... 2
1.1 International Electrical Symbols................................................3
1.2 DenitionofMeasurementCategories .....................................3
1.3 ReceivingYourShipment..........................................................3
1.4 OrderingInformation.................................................................4
1.4.1 AccessoriesandReplacementParts............................4
2. PRODUCT FEATURES ...................................................................... 5
2.1 Description................................................................................5
2.2 Compatibility .............................................................................5
2.3 Features....................................................................................6
3. SPECIFICATIONS............................................................................. 7
3.1 ElectricalSpecications ............................................................7
3.2 MechanicalSpecications ........................................................8
3.3 EnvironmentalSpecications....................................................9
3.4 SafetySpecications ................................................................9
3.5 TypicalResponseCurves .......................................................10
4. OPERATION .................................................................................. 13
4.1 CurrentMeasurement.............................................................13
5. MAINTENANCE ............................................................................. 14
5.1 Maintenance ...........................................................................14
5.2 Cleaning..................................................................................14
RepairandCalibration...........................................................................15
TechnicalandSalesAssistance ............................................................15
Limited Warranty ...................................................................................16
WarrantyRepairs...................................................................................16

2
AC Current Oscilloscope Probe Model SR661
CHAPTER 1
INTRODUCTION
WARNING
The safety warnings are provided to ensure the safety of
personnel and proper operation of the instrument. Read the
instruction completely.
• Usecautiononanycircuit:potentiallyhighvoltagesandcurrents
maybepresentandmayposeashockhazard.
• Donot usethe probeif damaged.Alwaysconnectthecurrent
probetothemeasuringdevicebeforeitisconnectedaroundthe
conductor
• Donotuseonnoninsulatedconductorwithapotentialtoground
greaterthan600VCATIIIpollution2.Useextremecautionwhen
clampingaroundbareconductorsorbusbars.
• Beforeeachuse,inspecttheprobe;lookforcracksinhousingor
outputcableinsulation.
• Donotuseclampinwetenvironmentorinlocationsthathazard-
ousgasesexist.
• Donotusetheprobeanywherebeyondthetactilebarrier.

AC Current Oscilloscope Probe Model SR661
3
1.1 International Electrical Symbols
This symbol signifies that the instrument is protected by double or rein-
forced insulation.
This symbol on the instrument indicates a WARNING and that the operator
must refer to the user manual for instructions before operating the
instrument. In this manual, the symbol preceding instructions indicates that
if the instructions are not followed, bodily injury, installation/sample and
product damage may result.
Risk of electric shock. The voltage at the parts marked with this symbol may
be dangerous.
This symbol refers to a type A current sensor. This symbol signifies that
application around and removal from HAZARDOUS LIVE conductors is per-
mitted.
Probe fitted with an electronic output limiter providing protection against
voltage surges caused by the accidental opening of the probe secondary
circuit to 30V max. peak.
In conformity with WEEE 2002/96/EC.
1.2 Definition of Measurement Categories
CAT II: For measurements performed on circuits directly connected to
the electrical distribution system. Examples are measurements
onhouseholdappliancesorportabletools.
CAT III: For measurements performed in the building installation at
the distribution level such as on hardwired equipment in xed
installationandcircuitbreakers.
CAT IV: For measurements performed at the primary electrical supply
(<1000V) such as on primary overcurrent protection devices,
ripplecontrolunits,ormeters.
1.3 Receiving Your Shipment
Uponreceivingyourshipment,makesurethatthecontentsareconsistent
with the packing list. Notify your distributor of any missing items. If the
equipmentappearstobedamaged,leaclaimimmediatelywiththecar-
rierandnotifyyourdistributoratonce,givingadetaileddescriptionofany
damage.Savethedamagedpackingcontainertosubstantiateyourclaim.

4
AC Current Oscilloscope Probe Model SR661
1.4 Ordering Information
ACCurrentOscilloscopeProbeModelSR661.................... Cat. #2113.49
Includes a user manual.
1.4.1 Accessories and Replacement Parts
MaleBanana-FemaleBNCAdaptor(XF-SS) .................... Cat. #2111.32

AC Current Oscilloscope Probe Model SR661
5
CHAPTER 2
PRODUCT FEATURES
2.1 Description
TheACCurrentOscilloscopeProbeModelSR661expandsoscilloscope
applicationsinindustrialorpowerenvironments,andisidealforanalysis
andmeasurementofdistortedcurrentwaveformsandharmonics.
TheModelSR661permitsaccuratedisplayandmeasurementofcurrents
from100mAto1000Arms,1Hzto100kHz(withcurrentderating)without
breaking into the circuit.A passive lter eliminates noise, ring on rapid
rising(di/dt)waveforms,andensuresaccuratescreendisplays.
Theprobeconnectsdirectlytoanoscilloscopethroughaninsulatedcoax-
ialcablewithaninsulatedBNC.
2.2 Compatibility
TheModelSR661iscompatiblewithanyanalogordigitaloscilloscopeor
othervoltage-measuringinstrumentwhichhasthefollowingfeatures:
• BNCinputconnector
• Rangecapableofdisplaying0.2to0.5Vperdivision,or2Vrange
• Minimuminputimpedanceof1MΩ

6
AC Current Oscilloscope Probe Model SR661
2.3 Features
4.37" (111mm)
3.98" (101mm)
2.25" (57mm)
1.22" (31mm)
8.50" (216mm)
1.71"
(43.5 mm)
1.38"
(35 mm)
1.77"
(45 mm)
3.90" (99 mm)
2.13"
(54.1mm)
1
2
3
5
4
6
1. Jawopeningof2.25"
2. Maximumdimensionofconductis2.13"(54mm)
3. TacticalBarrier.Alwayskeephandsbelowthisbarrieratalltimes.
4. Arrow indicates direction of current ow. Current ows in a positive
directionwhenittravelsfromsupplytoload.
5. Threepositionswitch
6. 6.5ft(2m)outputcable

AC Current Oscilloscope Probe Model SR661
7
CHAPTER 3
SPECIFICATIONS
3.1 Electrical Specifications
*Referenceconditions:23°C±3°K,20to75%RH,48to65Hz,externalmagneticeld
<40A/m,noDCcomponent,noexternalcurrentcarryingconductor,testsamplecentered.
Operating Range: 0.1to2000Apeak
Measurement Range: 100mAto1200Arms(2000Apeak)
Output Signal:mVoutputsignal(2Vpeakmax)
100mV/A: 10mAto20Apeak
10mV/A: 0.1Ato200Apeak
1mV/A: 1Ato2000Apeak
1mV/A, (1V at 1000A)
Range Accuracy Phase Shift
1 to 50A
≤1% ± 1mV
N/A
50 to 200A ≤3°
200 to 1000A ≤2°
1000 to 1200A ≤1°
Overload: 1200Afor40mnON,20mnOFF
10 mV/A, (1V at 100A)
Range Accuracy Phase Shift
0.1 to 5A
≤2% ± 5mV
N/A
5 to 20A ≤15°
20 to 100A ≤10°
100 to 120A ≤5°
Overload:120Acontinuous
100mV/A, (1V at 10 Arms)
Range Accuracy Phase Shift
0.1 to 0.5A
≤3% ± 10mV
N/A
0.5 to 2A N/A
2 to 10A ≤15°
10 to 12A
Overload: 12Acontinuous

8
AC Current Oscilloscope Probe Model SR661
Frequency Range (with current derating):
1Hzto100kHz(@-3dB)
Seetypicalresponsecurves(pages10-11)
Crest Factor:≤6
Load Impedance: ≥1MΩ@≤47pF
Rise/Fall Time: <40µS
Working Voltage:600VCATIII
Common Mode Voltage:600VCATIII
Inuence of Adjacent Conductor: <0.2mA/AAC
Inuence of Conductor Position in Jaw Opening:
0.02%ofReadingunder400Hz
Inuence of Frequency:
Range1mV/A: From10to1000Hz:<1%ofReading
From1to10kHz:<2%ofReading
From10to50kHz:<10%ofReading
From50to100kHz:-3dB
Range10mV/A: From10to1000Hz:<5%ofReading
From1to10kHz:<3%ofReading
From10to50kHz:<20%ofReading
From50to100kHz:-3dB
Range100mV/A: From10to1000Hz:<10%ofReading
From1to10kHz:<5%ofReading
From10to50kHz:<20%ofReading
From50to100kHz:-3dB
3.2 Mechanical Specifications
Dimensions: 4.4x8.5x1.8"(111x216x45mm)
Weight:1.21lbs(550g)
Jaw Opening:2.25"(57mm)max
Maximum Cable Diameter:2.13"(54mm)
Maximum Conductor Size:
Cable:2.05"(52mm)
BusBar:1.95x.19"(50x5mm)
Case Protection:IP40(IEC529)

AC Current Oscilloscope Probe Model SR661
9
Drop Test:1m(IEC68-2-32)
Mechanical Shock:100g(IEC68-2-27)
Vibration: 5to15Hz,0.15mm(IEC68-2-6)
15to25Hz,1mm
25to55Hz,0.25mm
Output: InsulatedleadwithinsulatedBNCConnector,6ft(2m)
3.3 Environmental Specifications
Operating Temperature/RH:14°to122°F(-10°to50°C)
Storage Temperature/RH:-4°to158°F(-20to70°C)
Altitude:Non-operating:0to12,000m
Operating:0to2000m
Operating Relative Humidity:0to85%@35°C
Inuence of Temperature: ≤0.15%/10K
Inuence of Humidity:10to90%:0.1%
3.4 Safety Specifications
E192383
Electrical:Doubleinsulationorreinforcedinsulationbetweentheprimary
orsecondaryandtheoutercaseofthehandleperEN61010-2-032.
600V,CATIII,PollutionDegree:2
Dielectric Strength: 5550V, 50/60Hz between primary, secondary, and
theoutercaseofthehandle
Electromagnetic Compatibility: EN 50081-1 Class B
EN50082-2Electrostaticdischarge
IEC 1000-4-2
RadiatedeldIEC1000-4-3
Fast transients IEC 1000-4-4
Magneticeldat50/60HzIEC1000-4-8
*Specications are subject to change without notice

10
AC Current Oscilloscope Probe Model SR661
3.5 Typical Response Curves
1mV/A Range:
10 100 1000 10,000
100,000
-4.00
-5.00
1.00
0.00
-1.00
-2.00
-3.00
10 100 1000 10,000
100,000
-35.00
-40.00
5.00
-15.00
-20.00
-25.00
-30.00
-10.00
-5.00
0.00
Accuracy in %
Phase in °
Frequency (Hz)
Frequency (Hz)
Accuracy vs. Frequency
Phase Shift vs. Frequency
@ 10 A
@ 50 A
@ 10 A
@ 50 A

AC Current Oscilloscope Probe Model SR661
11
10mV/A Range:
10 100 1000 10,000
100,000
-25.00
-30.00
0.00
-5.00
-10.00
-15.00
-20.00
10 100 1000 10,000 100,000
-60.00
-70.00
10.00
-20.00
-30.00
-40.00
-50.00
-10.00
0.00
5.00
Accuracy in %Phase in °
@ 1 A
@ 5 A
@ 1 A
@ 5 A
Frequency (Hz)
Frequency (Hz)
Accuracy vs. Frequency
Phase Shift vs. Frequency

12
AC Current Oscilloscope Probe Model SR661
100mV/A Range:
10 100 1000 10,000 100,000
10 100 1000 10,000 100,000
-40.00
-45.00
-15.00
-20.00
-25.00
-30.00
-35.00
5.00
-10.00
-5.00
0.00
-40.00
-50.00
10.00
0.00
-10.00
-20.00
-30.00
20.00
30.00
40.00
@ 0.1 A
@ 0.5 A
@ 0.1 A
@ 0.5 A
Accuracy in %
Phase in °
Frequency (Hz)
Frequency (Hz)
Accuracy vs. Frequency
Phase Shift vs. Frequency

AC Current Oscilloscope Probe Model SR661
13
CHAPTER 4
OPERATION
4.1 Current Measurement
WARNING: Always connect the probe to the instrument before clamping onto
the circuit under test.
• Connect the Model SR661 to the proper input channel on the
oscilloscope.
•
Beginwiththeleastsensitiverangeonthecurrentprobe(1mV/A)
• Selectthe0.5V/Divisionrangeontheoscilloscope.
• Clamptheprobeontheconductortobemeasuredandreadthe
currentowingdirectlyonyouroscilloscope.
(Remember to unclamp the probe from the conductor before discon-
necting it from your meter or instrument.)
You may also use your oscilloscope to amplify the signal while using
the1mV/Aproberange(whichoersthebestaccuracyandleastphase
shift).
NOTE: It is possible to change the range on the current probe without remov-
ing the probe from the current carrying conductor, but it is important to
remember not to exceed the permissible peak ratings of 2000mV peak or
4000 mV peak to peak maximum. The peak ratings by range are: 20A peak
@ 100mV/A, 200A peak @ 10mV/A and 2000A peak @ 1mV/A.

14
AC Current Oscilloscope Probe Model SR661
CHAPTER 5
MAINTENANCE
5.1 Maintenance
WARNING
• Toensureoptimumperformance,itisimportanttokeeptheprobe
jawmatingsurfacescleanatalltimes.Failuretodosomayresult
inerrorinreadings.
• Formaintenanceuseonlyspeciedfactoryreplacementparts.
• Toavoidelectricalshock,donotattempttoperformanyservicing
unlessyouarequaliedtodoso.
• Toavoidelectricalshockand/ordamagetotheinstrument,donot
allowwaterorotherforeignsubstancesintothecase.
• Disconnecttheunitfromallcircuitsandtestcablesbeforeopen-
ingthecase.
• Usecautionwithmetallictoolsthatmayshortbatterypacks,power
supplies,etc.
5.2 Cleaning
• Tocleantheprobebody,useasoftclothdampenedinasolution
ofmilddetergentandwater.Tocleanthecore,openthejawand
cleanthe exposedcoresurfaceswithacottonswabdampened
with isopropyl alcohol (isopropanol) or ethyl alcohol (fotocol or
ethanol).Lubricatethejawsmatingsurfaceswithalightoil.
• Do not use chemicals containing benzine, benzene, toluene,
xylene,acetone,orsimilarsolvents.
• Donotimmersetheprobeinliquidsoruseabrasivecleaners.

AC Current Oscilloscope Probe Model SR661
15
Repair and Calibration
Toensurethatyourinstrumentmeetsfactoryspecications,werecommend
thatitbescheduledbacktoourfactoryServiceCenteratone-yearintervals
forrecalibration,orasrequiredbyotherstandardsorinternalprocedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer ServiceAuthorization
Number(CSA#).Thiswillensurethatwhenyourinstrumentarrives,itwillbe
trackedandprocessedpromptly.PleasewritetheCSA#ontheoutsideofthe
shipping container. If the instrument is returned for calibration, we need to
knowifyouwantastandardcalibration,oracalibrationtraceabletoN.I.S.T.
(Includescalibrationcerticateplusrecordedcalibrationdata).
Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments
15 Faraday Drive
Dover,NH03820USA
Phone:(800)945-2362(Ext.360)
(603)749-6434(Ext.360)
Fax: (603)742-2346or(603)749-6309
E-mail:[email protected]
(Orcontactyourauthorizeddistributor)
Costsforrepair,standardcalibration,andcalibrationtraceabletoN.I.S.T.are
available.
NOTE: You must obtain a CSA# before returning any instrument.
Technical and Sales Assistance
Ifyouareexperiencinganytechnicalproblems,orrequireanyassistancewith
theproperoperationorapplicationofyourinstrument,pleasecall,mail,faxor
e-mailourtechnicalsupportteam:
ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments
200FoxboroughBoulevard
Foxborough,MA02035USA
Phone:(800)343-1391
(508)698-2115
Fax: (508)698-2118
E-mail:[email protected]
www.aemc.com
NOTE: Do not ship Instruments to our Foxborough, MA address.

16
AC Current Oscilloscope Probe Model SR661
Limited Warranty
TheACCurrentOscilloscopeProbeModelSR661iswarrantedtotheowner
foraperiodoftwoyearsfromthedateoforiginalpurchaseagainstdefectsin
manufacture.ThislimitedwarrantyisgivenbyAEMC®Instruments,notbythe
distributorfromwhomitwaspurchased.Thiswarrantyisvoidiftheunithas
beentamperedwith,abusedorifthedefectisrelatedtoservicenotperformed
byAEMC®Instruments.
Full warranty coverage and product registration is available on our
website at www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for your records.
What AEMC®Instruments will do:
Ifamalfunctionoccurswithinthewarranty period,youmayreturntheinstrument
tousforrepair,providedwehaveyourwarrantyregistrationinformation on
file or a proof of purchase.AEMC®Instruments will, at its option, repair or
replacethefaultymaterial.
REGISTER ONLINE AT: www.aemc.com
Warranty Repairs
What you must do to return an Instrument for Warranty Repair:
First, request a Customer ServiceAuthorization Number (CSA#) by phone
orbyfaxfromourServiceDepartment(seeaddressbelow),thenreturnthe
instrumentalong withthe signedCSAForm. Pleasewrite theCSA# onthe
outsideoftheshippingcontainer.Returntheinstrument,postageorshipment
pre-paidto:
Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments
15FaradayDrive•Dover,NH03820USA
Phone:(800)945-2362(Ext.360)
(603)749-6434(Ext.360)
Fax: (603)742-2346or(603)749-6309
E-mail:[email protected]
Caution:Toprotectyourselfagainstin-transitloss,werecommendyouinsure
yourreturnedmaterial.
NOTE: You must obtain a CSA# before returning any instrument.

AC Current Oscilloscope Probe Model SR661
17
Notes:

10/17
99-MAN100130v13
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15FaradayDrive•Dover,NH03820USA•Phone:(603)749-6434•Fax:(603)742-2346
www.aemc.com
This manual suits for next models