AEMC SR701 User manual

AC Current Probe
Models SR701 and SR704
User Manual
DESCRIPTION:
The SR701/SR704 (Catalog #2116.29 and #2116.30) are designed for use in industrial environments. The
ergonomic design allows them to easily attach to cables or small bus bars. The “circular” jaws guarantee a
very good accuracy and low phase shift. The probes have a measurement range up to 1000Arms and are
compatible with any AC ammeter, multimeter, or other current measurement instrument with an input
impedance lower than 5Ω. To achieve the stated accuracy, use the SR701/SR704 with an ammeter having
an accuracy of 0.75% or better.
WARNING
The safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument. Read the instruction completely.
• Use caution on any circuit: potentially high voltages and currents may be present and may pose a
shock hazard.
• Do not use the probe if damaged. Always connect the current probe to the measuring device before it
is connected around the conductor
• Do not use on non-insulated conductor with a potential to ground greater than 600V CAT III pollution
2. Use extreme caution when clamping around bare conductors or bus bars.
• Before each use, inspect the probe; look for cracks in housing or output cable insulation.
• Do not use clamp in wet environment or in locations that hazardous gases exist.
• Do not use the probe anywhere beyond the tactile barrier.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use only
factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before using
the instrument.
This is a type A current sensor. This symbol signifies that application around and removal from
HAZARDOUS LIVE conductors is permitted.
DEFINITION OF MEASUREMENT CATEGORIES
CAT IV: For measurements performed at the primary electrical supply (<1000V) such as on primary
overcurrent protection devices, ripple control units, or meters.
CAT III: For measurements performed in the building installation at the distribution level such as on
hardwired equipment in fixed installation and circuit breakers.
CAT II: For measurements performed on circuits directly connected to the electrical distribution system.
Examples are measurements on household appliances or portable tools.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your
distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the
carrier and notify your distributor at once, giving a detailed description of any damage.
ELECTRICAL SPECIFICATIONS
Current Range:
1mA to 1200A AC
Transformation Ratio: 1000:1
Output Signal: 1mA AC/A AC (1A AC at
1000A)
Accuracy and Phase Shift*:
Accuracy:
1mA to 100mA: ± 3% Reading ± 5mA
0.1 to 1A: ± 2% Reading ± 3mA
1 to 10A: ± 1% Reading
10 to 100A: ± 0.5% Reading
100 to 1200A: ± 0.3% Reading
Phase Shift:
1mA to 100mA: Not Specified
0.1 to 1A: Not Specified
1 to 10A: ≤ 2°
10 to 100A: ≤ 1°
100 to 1200A: ≤ 0.7°
*Reference conditions: 23°C±3°K, 20 to 75% RH,
48 to 65Hz, external magnetic field <40A/m, no
DC component, no external current carrying
conductor, test sample centered.
Load impedance ≤1Ω.
Overload: 1200A for 40mn on, 20mn off
Accuracy: Per IEC 185
SR701 SR704
Frequency Range: 30Hz to 5kHz; current
derating above 1kHz using the formula:
×

()
Load Impedance:
5Ω max
Working Voltage:
600V CAT III
Common Mode Voltage:
600V CAT III
Open Secondary Voltage:
< 25V by limiting circuit
Influence of Adjacent Conductor:
< 1mA/A AC
Influence of Conductor in Jaw Opening:
0.1% of reading
Influence of Frequency:
From 30 to 48Hz: < 0.5% of R
From 65 to 1000Hz: < 1% of R
From 1kHz to 5kHz: < 2% of R
MECHANICAL
SPECIFICATIONS
Operating Temperature:
14° to 122°F (-10° to 50°C)
Storage Temperature:
-4° to 158°F (-20° to 70°C)
Influence of Temperature:
< 0.15% per 10°K
Influence of Humidity:
From 10 to 90%: 0.1%
Jaw Opening:
2.25" (57mm) max
Maximum Conductor Size:
2.05” (52mm)
Envelope Protection: IP 40 (IEC 529)
Drop Test:
1m (IEC 68-2-32)
Mechanical Shock:
100g (IEC 68-2-27)
Vibration:
5 to 15Hz, 0.15mm (IEC 68-2-6)
15 to 25Hz, 1mm
25 to 55Hz, 0.25mm
Material:
Handles and jaws: Polycarbonate UL94V0
Dimensions:
4.37 x 8.50 x 1.77" (111 x 216 x 45mm)
Weight:
1.21 lbs (550g)
Output:
SR701: Two standard safety banana
jacks (4mm)
SR704: 5 ft (1.5m) lead with safety
4 mm banana plug
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation
between the primary or secondary and the
outer case of the handle conforms to IEC
1010-2-032.
Common Mode Voltage:
600V Category III, Pollution Degree 2
Dielectric Strength:
5550V, 50/60Hz between primary, secondary
and the outer case of the handle
Electromagnetic Compatibility:
Emission and immunity in an industrial
environment according to EN61326-1
ORDERING INFORMATION
AC Current Probe SR701 ..............Cat #2116.29
AC Current Probe SR704 ..............Cat #2116.30
Accessories:
Lead – Set of 2, 5 ft. PVC color-coded
(red/black) with 4mm straight/straight
banana plugs
(1000V CAT IV, 15A, UL V2)............ Cat #5100.07
Adapter BNC (Male) –Banana (Female)
(XM-BB) (600V CAT III) .................... Cat #2118.46
Accessories: Adapter- 4mm
non-insulated for safety leads ......... Cat #1017.45
99-MAN 100166.v1806/21
OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model SR701/SR704
•Connect the black lead of the current probe to “common” and the red lead to the AC current input on
your DMM or other current measuring instrument. Select the appropriate current range (2A AC range).
Clamp the probe around the conductor to be tested with the arrow pointed toward the load. If the
reading is less than 200mA, select the lower range until you obtain the best resolution. Read the value
display on the DMM and multiply it by the probe ratio (1000/1). (If reading = 0.659A, the current flowing
through the probe is 0.659A x 1000 = 659A AC).
•For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
•When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
•Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE
Warning:
•For maintenance use only original replacement parts.
•To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
•To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into
the probe
Cleaning:
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine 600)
to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will ensure
that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the
outside of the shipping container.
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive •Dover, NH 03820 USA
(800) 945-2362 (Ext. 360) • (603) 749-6434 (Ext. 360) • repair@aemc.com
(Or contact your authorized distributor)
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or application
of this instrument, please contact our technical hotline:
(800) 343-1391 • (508) 698-2115 • techsupport@aemc.com
LIMITED WARRANTY
The current probe is warranted to the owner for a period of two years from the date of original purchase
against defects in manufacture. This limited warranty is given by AEMC®Instruments, not by the distributor
from whom it was purchased. This warranty is void if the unit has been tampered with, abused or if the
defect is related to service not performed by AEMC®Instruments.
Full warranty coverage and product registration is available on our website at: www.aemc.com/warranty.html
Please print the online Warranty Coverage Information for your records.

This manual suits for next models

3

Other AEMC Measuring Instrument manuals

AEMC PowerPad 8436 User manual

AEMC

AEMC PowerPad 8436 User manual

AEMC MN352 User manual

AEMC

AEMC MN352 User manual

AEMC MN375 User manual

AEMC

AEMC MN375 User manual

AEMC 3710 User manual

AEMC

AEMC 3710 User manual

AEMC MN312 User manual

AEMC

AEMC MN312 User manual

AEMC 1227 User manual

AEMC

AEMC 1227 User manual

AEMC MN379 User manual

AEMC

AEMC MN379 User manual

AEMC 6527 User manual

AEMC

AEMC 6527 User manual

AEMC 603 User manual

AEMC

AEMC 603 User manual

AEMC 565 User manual

AEMC

AEMC 565 User manual

AEMC 502 User manual

AEMC

AEMC 502 User manual

AEMC 300-24-2-10 User manual

AEMC

AEMC 300-24-2-10 User manual

AEMC 8230 User manual

AEMC

AEMC 8230 User manual

AEMC SR701 User manual

AEMC

AEMC SR701 User manual

AEMC PowerPad 8335 Setup guide

AEMC

AEMC PowerPad 8335 Setup guide

AEMC 6526 User manual

AEMC

AEMC 6526 User manual

AEMC 6550 Setup guide

AEMC

AEMC 6550 Setup guide

AEMC PowerPad IV User manual

AEMC

AEMC PowerPad IV User manual

AEMC 6611 User manual

AEMC

AEMC 6611 User manual

AEMC 193-24-BK User manual

AEMC

AEMC 193-24-BK User manual

AEMC 511 User manual

AEMC

AEMC 511 User manual

AEMC MN125 User manual

AEMC

AEMC MN125 User manual

AEMC 1510 User manual

AEMC

AEMC 1510 User manual

AEMC PowerPad 3945-B User manual

AEMC

AEMC PowerPad 3945-B User manual

Popular Measuring Instrument manuals by other brands

GHM Delta OHM LPPYRA12 operating manual

GHM

GHM Delta OHM LPPYRA12 operating manual

Martindale Electric VT12 MK2 instruction manual

Martindale Electric

Martindale Electric VT12 MK2 instruction manual

DNV VacuDAP 2002 Operation manual

DNV

DNV VacuDAP 2002 Operation manual

Endress+Hauser LT5 Operating	 instruction

Endress+Hauser

Endress+Hauser LT5 Operating instruction

Vetus LNVOLT12 manual

Vetus

Vetus LNVOLT12 manual

Meggitt VM600 manual

Meggitt

Meggitt VM600 manual

JUMO AQUIS touch P Interface description

JUMO

JUMO AQUIS touch P Interface description

Cantium Scientific MicroBio MB2-RSH Installation and Operation Supplement

Cantium Scientific

Cantium Scientific MicroBio MB2-RSH Installation and Operation Supplement

Hanna Instruments HI 96702 instruction manual

Hanna Instruments

Hanna Instruments HI 96702 instruction manual

Keithley 2461 Reference manual

Keithley

Keithley 2461 Reference manual

Chauvin Arnoux PEL 106 user manual

Chauvin Arnoux

Chauvin Arnoux PEL 106 user manual

Bender ISOMETER isoHV425 manual

Bender

Bender ISOMETER isoHV425 manual

Red Ion APLI quick start guide

Red Ion

Red Ion APLI quick start guide

Skye APOLLO SKA 400 manual

Skye

Skye APOLLO SKA 400 manual

horiba D-72 instruction manual

horiba

horiba D-72 instruction manual

Simpson Electric 14510-2 Operator's manual

Simpson Electric

Simpson Electric 14510-2 Operator's manual

Schonstedt Instrument GA-52Cx instruction manual

Schonstedt Instrument

Schonstedt Instrument GA-52Cx instruction manual

sauermann CO2ST-S quick start guide

sauermann

sauermann CO2ST-S quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.