
B270 User Guide
19
amscope.com
2.4: Locang the Specimen, and Focusing
With the specimen mounted on the stage, and the illuminator turned on, use the stage’s X-Y control knobs under the
right side of the stage to adjust the posion of the specimen. The larger knob will move the specimen forward and
backward, while the smaller knob will move the specimen le and right. Compound microscopes produce inverted
images due to the nature of the opcal system. Sliding the specimen towards the front of the microscope will result in
the viewed image moving in the opposite direcon. Use the knobs unl the specimen appears to be centered under the
objecve lens. This will be the starng point.
Observaon should typically begin at the lowest magnicaon. This provides the broadest view of the specimen. Rotate
the objecve turret to set the lowest magnicaon lens in place. The condenser is typically designed with a range of
aperture values ranging from approximately 0.10 to 1.25. If using a 4X objecve lens with a 0.10 aperture, adjust the con-
denser’s aperture using the lever on the outside of the condenser to reduce the size of the aperture. This will maximize
the focus depth. When higher magnicaons are used, the aperture should be adjusted more open to beer match the
objecve lens’ aperture.
Look through the eyepieces, adjusng the interpupillary adjusters as needed to achieve proper alignment. Both eyes
should see the same image with a bright, round background. If the image appears oset or obscured, you should conn-
ue to adjust the interpupillary distance unl the image appears correct. The le ocular tube also has a dioptric adjuster
which can be used if the eyes do not focus the same. If the eyepieces are properly aligned, but the light-source appears
oset, use the two centering screws aached to the condenser holder to adjust the condenser’s posion unl it is cen-
tered. For trinocular models, the photo tube can be adjusted to parfocal by rst obtaining focus with the right eye, then
adjusng the length of photo tube by unscrewing the top poron of the tube from the boom poron unl the camera
obtains focus, then ghtening the locking ring in the middle unl it engages against the boom poron of the photo
tube.
[image: adjusng the coarse/ne focus]
The stage should be set to a low posion when mounng the specimen. Gradually raise the stage using the coarse-focus
knob. If you have already performed the height-limit adjustment, the specimen should come into focus near the maxi-
mum height. Once you begin to see shapes through the eyepieces, use the ne-focus knobs to adjust the focus unl you
can see details with clarity. Use the X-Y control knobs to reposion the specimen unl the area of interest is in view. You
can increase magnicaon as needed by rotang the objecve turret. When doing so, it is advisable to watch the lens as
it is set in place to ensure it will not collide with the specimen. Slight adjustments to focus should be made at each mag-
nicaon, as well as to the condenser’s aperture and the illuminator’s brightness.