Anritsu MS2830A User manual

Document No.: M-W3448AE-2.0
ANRITSU CORPORATION
MX283087A
TRX Sweep Calibration
Operation Manual
Operation
Second Edition
•For safety and warning information, please read this
manual before attempting to use the equipment.
•
Additional safety and warning information is provided
within the MS2830A
Signal Analyzer Operation Manual
(Mainframe: Operation). Please also refer to this
document before using the equipment.
•
Keep this manual with the equipment.

ii
Safety Symbols
To prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Corporation uses the
following safety symbols to indicate safety-related information. Ensure that you clearly understand the meanings of
the symbols BEFORE using the equipment. Some or all of the following symbols may be used on all Anritsu
equipment. In addition, there may be other labels attached to products that are not shown in the diagrams in this
manual.
Symbols used in manual
This indicates a very dangerous procedure that could result in serious injury or
death if not performed properly.
This indicates a hazardous procedure that could result in serious injury or death if
not performed properly.
This indicates a hazardous procedure or danger that could result in light-to-severe
injury, or loss related to equipment malfunction, if proper precautions are not taken.
Safety Symbols Used on Equipment and in Manual
The following safety symbols are used inside or on the equipment near operation locations to provide information
about safety items and operation precautions. Ensure that you clearly understand the meanings of the symbols and
take the necessary precautions BEFORE using the equipment.
This indicates a prohibited operation. The prohibited operation is indicated
symbolically in or near the barred circle.
This indicates an obligatory safety precaution. The obligatory operation is
indicated symbolically in or near the circle.
This indicates a warning or caution. The contents are indicated symbolically in or
near the triangle.
This indicates a note. The contents are described in the box.
These indicate that the marked part should be recycled.
MX283087A
TRX Sweep Calibration
Operation Manual Operation
15 October 2010 (First Edition)
29 May 2013 (Second Edition)
Copyright © 2010-2013, ANRITSU CORPORATION.
All rights reserved. No part of this manual may be reproduced without the prior written permission of the
publisher.
The contents of this manual may be changed without prior notice.
Printed in Japan
DANGER
CAUTION
WARNING

iii
Equipment Certificate
Anritsu Corporation guarantees that this equipment was inspected at
shipment and meets the published specifications.
Anritsu Warranty
•During the warranty period, Anritsu Corporation will repair or exchange this
software free-of-charge if it proves defective when used as described in the
operation manual.
•The warranty period is 6 months from the purchase date.
•The warranty period after repair or exchange will remain 6 months from
the original purchase date, or 30 days from the date of repair or
exchange, depending on whichever is longer.
•This warranty does not cover damage to this software caused by Acts of
God, natural disasters, and misuse or mishandling by the customer.
In addition, this warranty is valid only for the original equipment purchaser. It
is not transferable if the equipment is resold.
Anritsu Corporation shall assume no liability for injury or financial loss of the
customer due to the use of or a failure to be able to use this equipment.
Anritsu Corporation Contact
In the event that this equipment malfunctions, contact an Anritsu Service and
Sales office. Contact information can be found on the last page of the printed
version of this manual, and is available in a separate file on the CD version.

iv
Notes On Export Management
This product and its manuals may require an Export License/Approval by
the Government of the product's country of origin for re
-
export from your
country.
Before re
-exporting th
e product or manuals, please contact us to confirm
whet
her they are export-controlled items or not.
When you dispose of export
-
controlled items, the products/manuals need
to be broken/shredded so as not to be unlawfully used for military purpose.

v
Software End-User License Agreement (EULA)
Please read this Software End-User License Agreement (hereafter this EULA) carefully before using
(includes executing, copying, registering, etc.) this software (includes programs, databases, scenarios, etc.,
used to operate, set, etc., Anritsu electronic equipment). By reading this EULA and using this software, you
are agreeing to be bound by the terms of its contents and Anritsu Corporation (hereafter Anritsu) hereby
grants you the right to use this Software with the Anritsu-specified equipment (hereafter Equipment) for the
purposes set out in this EULA.
1. Grant of License and Limitations
1. Regardless of whether this Software was
purchased from or provided free-of-charge by
Anritsu, you agree not to rent, lease, lend, or
otherwise distribute this Software to third
parties and further agree not to disassemble,
recompile, reverse engineer, modify, or create
derivative works of this Software.
2. You may make one copy of this Software for
backup purposes only.
3. You are not permitted to reverse engineer
this software.
4. This EULA allows you to install one copy of
this Software on one piece of Equipment.
2. Disclaimers
To the extent not prohibited by law, in no
event shall Anritsu be liable for personal
injury, or any incidental, special, indirect or
consequential damages whatsoever,
including, without limitation, damages for
loss of profits, loss of data, business
interruption or any other commercial
damages or losses, arising out of or related
to your use or inability to use this Software.
3. Limitation of Liability
a. If a fault (bug) is discovered in this Software,
preventing operation as described in the
operation manual or specifications whether
or not the customer uses this software as
described in the manual, Anritsu shall at its
own discretion, fix the bug, or exchange the
software, or suggest a workaround,
free-of-charge. However, notwithstanding
the above, the following items shall be
excluded from repair and warranty.
i) If this Software is deemed to be used for
purposes not described in the operation
manual or specifications.
ii) If this Software is used in conjunction with
other non-Anritsu-approved software.
iii) Recovery of lost or damaged data.
iv) If this Software or the Equipment has been
modified, repaired, or otherwise altered
without Anritsu's prior approval.
v) For any other reasons out of Anritsu's direct
control and responsibility, such as but not
limited to, natural disasters, software virus
infections, etc.
b. Expenses incurred for transport, hotel, daily
allowance, etc., for on-site repairs by Anritsu
engineers necessitated by the above faults
shall be borne by you.
c. The warranty period for faults listed in
article 3a above covered by this EULA shall
be either 6 months from the date of purchase
of this Software or 30 days after the date of
repair, whichever is longer.

vi
4. Export Restrictions
You may not use or otherwise export or
re-export directly or indirectly this Software
except as authorized by Japanese and
United States law. In particular, this
software may not be exported or re-exported
(a) into any Japanese or US embargoed
countries or (b) to anyone on the Japanese or
US Treasury Department's list of Specially
Designated Nationals or the US Department
of Commerce Denied Persons List or Entity
List. By using this Software, you warrant
that you are not located in any such country
or on any such list. You also agree that you
will not use this Software for any purposes
prohibited by Japanese and US law,
including, without limitation, the
development, design and manufacture or
production of missiles or nuclear, chemical or
biological weapons of mass destruction.
5. Termination
Anritsu shall deem this EULA terminated if
you violate any conditions described herein.
This EULA shall also be terminated if the
conditions herein cannot be continued for
any good reason, such as violation of
copyrights, patents, or other laws and
ordinances.
6. Reparations
If Anritsu suffers any loss, financial or
otherwise, due to your violation of the terms
of this EULA, Anritsu shall have the right to
seek proportional damages from you.
7. Responsibility after Termination
Upon termination of this EULA in
accordance with item 5, you shall cease all
use of this Software immediately and shall
as directed by Anritsu either destroy or
return this Software and any backup copies,
full or partial, to Anritsu.
8. Dispute Resolution
If matters of dispute or items not covered by
this EULA arise, they shall be resolved by
negotiations in good faith between you and
Anritsu.
9. Court of Jurisdiction
This EULA shall be interpreted in
accordance with Japanese law and any
disputes that cannot be resolved by
negotiation described in Article 8 shall be
settled by the Japanese courts.

vii
Cautions against computer virus infection
•Copying files and data
Only files that have been provided directly from Anritsu or generated
using Anritsu equipment should be copied to the instrument.
All other required files should be transferred by means of USB or
CompactFlash media after undergoing a thorough virus check.
•Adding software
Do not download or install software that has not been specifically
recommended or licensed by Anritsu.
•Network connections
Ensure that the network has sufficient anti-
virus security protection in
place.

viii
CE Conformity Marking
Anritsu affixes the CE conformity marking on the following product(s) in
accordance with the Council Directive 93/68/EEC to indicate that they
conform to the EMC and LVD directive of the European Union (EU).
CE marking
1. Product Model
Software: MX283087A TRX Sweep Calibration
2. Applied Directive and Standards
When the MX283087A TRX Calibration Software is installed in the
MS2830A, the applied directive and standards of this unit conform to
those of the MS2830A main frame.
PS: About main frame
Please contact Anritsu for the latest information on the main frame
types that MX283087A can be used with.

ix
C-Tick Conformity Marking
Anritsu affixes the C-Tick mark on the following product(s) in accordance
with the regulation to indicate that they conform to the EMC framework of
Australia/New Zealand.
C-Tick mark
1. Product Model
Software: MX283087A TRX Sweep Calibration
2. Applied Directive and Standards
When the MX283087A TRX Sweep Calibration is installed in the
MS2830A, the applied directive and standards of this unit conform to
those of the MS2830A main frame.
PS: About main frame
Please contact Anritsu for the latest information on the main frame
types that MX283087A can be used with.

x

I
About This Manual
About this document
This operation manual is for MS283087A TRX Sweep Calibration
(Operation).
MX283087A
TRX Sweep Calibration
Operation Manual (Operation)
MS2830A
Signal Analyzer Operation Manual (Mainframe Operation)
MS2690A/MS2691A/MS2692A and MS2830A
Signal Analyzer Operation Manual (Mainframe Remote Control)
MX283087A
TRX Sweep Calibration
Operation Manual (Remote Control)
Related manuals
Each related manual describes the following operations.
Refer to each operation manual for details.
Signal Analyzer Operation Manual (Mainframe Operation)
Signal Analyzer Operation Manual (Mainframe Remote Control)
These describe basic operations, maintenance procedure, common
functions and common remote functions of the signal analyzer.
MS283087A TRX Sweep Calibration Operation Manual (Operation) [This
document]
This manual describes how to operate the MS283087A TRX Sweep
Calibration . As for MS2830A signal analyzer hardware and its basic
functions and operation outline, refer to “MS2830A Signal Analyzer
Operation Manual (Mainframe Operation)”.
TRX Sweep Calibration Operation Manual (Remote Control)
This manual describes how to use the MS283087A TRX Sweep
Calibration via remote command. As for MS2830A signal analyzer
application's basic remote control functions and its definitions of
common commands, refer to “MS2690A/MS2691A/MS2692A and
MS2830A Signal Analyzer Operation Manual (Mainframe Remote
Control)”.

II
Document Conventions
This indicates the Signal Analyzer panel keys.
This indicates pages and sections to be referred to.
Boldface This indicates message that appears on the screen.
‘ ’This indicates reference that does not pertain to screen
messages.
“ ” Thisindicatesreference,or jump to other section of the
manual.

1
III
2
3
Appendix Index
Table of Contents
About This Manual........................................ I
Chapter 1 Outline ....................................... 1-1
1.1 Product Outline .......................................................... 1-2
1.2 Product Composition ................................................. 1-3
1.3 Product Specifications ............................................... 1-4
Chapter 2 Preparation................................ 2-1
2.1 Part Names ................................................................ 2-2
2.2 Signal Path Setup ...................................................... 2-10
2.3 Loading and Switching Applications .......................... 2-11
2.4 Initialization and Calibration....................................... 2-12
Chapter 3 Measurement ............................ 3-1
3.1 Usage Precautions .................................................... .. 3-2
3.2 Overview of Measurement......................................... .. 3-4
3.3 Measurement Flow .................................................... .. 3-8
3.4 Measurement Screen ................................................ 3-13

IV.
Appendix A Default Value List ................... A-1
Index .......................................................... Index-1

Chapter 1 Outline
1-1
1
Outline
This chapter provides an overview and describes the product
configuration of the MX283087A TRX Sweep Calibration Software.
1.1 Product Outline ............................................................. 1-2
1.2 Product Composition .................................................... 1-3
1.2.1 Standard composition....................................... 1-3
1.2.2 Applicable parts ................................................ 1-3
1.3 Product Specifications .................................................. 1-4

Chapter 1 Outline
1-2
1.1 Product Outline
The MS2830A Signal Analyzer enables high-speed, high-accuracy, and
simple measurements of transmission characteristics of base stations
and mobile stations for various types of mobile communications. The
MS2830A has high-performance signal analyzer and spectrum analyzer
functions as standard, with optional measurement software allowing
modulation analysis functionality supporting various digital modulation
modes.
The MX283087A TRX Sweep Calibration (hereinafter, “MX283087A”) is a
software option used to measure the transmission and reception power
characteristics of wireless transceivers at high speed.
Chapter 3 describes how to perform major measurements with the
MX283087A by using remote commands.
The MX283087A supports the following measurements.
•TRX Sweep Measurement
•RX Sweep Measurement

1.2 Product Composition
1-3
1
Outline
1.2 Product Composition
1.2.1 Standard composition
Table 1.2.2-1 lists the application parts for the MX283087A.
Table 1.2.1-1 Standard composition
Item Model/
Symbol Name Quantity Remarks
Application MX283087A TRX Sweep Calibration 1
Accessories ―Installation CD-ROM 1
Application software,
operation manual
CD-ROM
1.2.2 Applicable parts
Table 1.2.2-1 lists the application parts for the MX283087A.
Table 1.2.2-1 Applicable parts
Model/Symbol Name Remarks
W3448AE MX283087A TRX Sweep Calibration
Operation Manual Operation
English, Printed
Version
W3449AE MX283087A TRX Sweep Calibration
Operation Manual Remote Control
English, Printed
Version

Chapter 1 Outline
1-4
1.3 Product Specifications
Table 1.3-1 lists the specifications for the MX283087A.
Table 1.3-1 Product specifications
Item Specification
Function Performs measurement while switching both the level for each
measurement unit (segment) according to the level list and the
frequency for each measurement unit group (sequence) according
to the frequency list.
Measurement mode TRX mode:
Performs transmission measurement and reception measurement
at the same time.
RX mode:
Performs reception measurement only. In the RX mode, a trigger
signal, which consists of output On (active) and Off (inactive)
intervals, can be output before a measurement signal.
Items common to transmission and reception
Frequency Range
Setting range of segment
length
Setting range of segment
Setting range of sequence
400 to 3500 MHz
10 ms, 20 ms
1 to 80
1 to 20
Transmitter Power Measurement
Analysis bandwidth
Measurement time range
Trigger
Trigger mode
Trigger setting range
Measured level range
Transmitter power accuracy
2.5 MHz, 5 MHz, 10 MHz, 25 MHz
Symmetric about the center of the segment and 20 to 90% of the
specified segment length
Free Run (Trig Off), Video (Trig On)
–30 to –10 dB (compared to the measurement level specified for
the first segment)
–30 dBm to +30 dBm
After CAL execution at 18 to 28°C, the input signal level is within
the measurement level range, and
the input level is as follows:
±0.7 dB
The transmitter power accuracy is calculated from an input
attenuator switching error, a measured linearity error, and a root
sum square (RSS) error of the absolute amplitude accuracy and
in-band frequency characteristics.

1.3 Product Specifications
1-5
1
Outline
Table 1.3-1 Product specifications (Continued)
Item Specification
Reception Power measurement
Output level range
Output level accuracy
Level error from CW during
vector modulation
Trigger signal
Short burst interval
setting range
Time offset interval
setting range
–120 dBm to –5 dBm
CW, at 18 to 28°C
±0.5 dB (Output level ≥–110 dBm)
±1 dB (Output level < 110 dBm)
The output level accuracy is based on that of the
MS2830A-020/021 Vector Signal Generator Option.
AWGN signal whose bandwidth is 5 MHz, at 18 to 28°C, with an
output frequency of 100 MHz or higher
±0.2 dB
Based on the level error from CW during vector modulation with
MS2830A-020/021 Vector Signal Generator Option.
Output On interval (short burst) and Off interval (time offset)
1 to 100 Segment
1 to 100 Segment

Chapter 1 Outline
1-6.
Other manuals for MS2830A
8
This manual suits for next models
1
Table of contents
Other Anritsu Analytical Instrument manuals
Popular Analytical Instrument manuals by other brands

XTline
XTline XT130512 Original instruction manual

VOLTCRAFT
VOLTCRAFT 10 27 31 operating instructions

Powerfix Profi
Powerfix Profi PEK 2.4 B1 User manual and service information

Silverline
Silverline 725352 manual

National Instruments
National Instruments AT-MIO E Series Programmer's manual

Leica Geosystems
Leica Geosystems DISTO D110 manual