AOIP CALYS 50 Use and care manual

CALIBRATION
CALYS 50
On-site multifunction calibrator
Simultaneous Measurement and Generation
Protected for on-site use
Easy connection system
Designed with a close collaboration with industrials, Calys 50 integrates all the necessary
functions for adjustment and maintenance of process.
Its ergonomic design and its embedded software allow Calys 50 to be a high performance
calibrator and very easy to use.
IP 54, full protected by an anti-shock sheath, with "easy-connect®" terminals integrated, a
very contrasted backlight display, it will be comfortable to use in all conditions
Its elastomer keypad protects it from dirty and grease marks, the keys allow to use Calys 50
with gloves. It has 10 working configurations recorded by user or user group, to always find
the good configuration for designed and repetitive jobs.
Ed 02/10/04 Document susceptible d'être modifié

CALIBRATION
Calys 50: presentation
Calys 50 is a portable calibrator able to measure and to generate simultaneously on 2 isolated channels.
It has a wide backlighted display with high contrast to be used for application in dark room.
Full protected by the sheath, a keypad in lexan protects it from dirties and numerical keypad knocked up is
usable even using protective gloves.
It is able to measure and generate voltage, current, frequency, resistance signals and also resistive probes and
thermocouples.
A
nti-shock sheath "easy-connec®t" terminals
Power supply
USB connector
Wide backlighted screen
Protected by a polycarbonat
A
lphanumerical keyboard
Navigator
Calys 50 is delivered in standard with a strap and a stand for desktop use, a quick battery charger, and a set of
6 measuring cables with crocodile clips
Ed 02/10/04 Document susceptible d'être modifié

CALIBRATION
"easy connect" system
This unique system is used by pushing on the
terminal’s top, by inserting:
- Wires with a diameter up to 3 mm,
- Compensated thermocouple connectors,
- Pin terminal on front panel,
and next by releasing.
Wires are tighten between 2 brass plates which
provide an great thermal gradient, so that allows a
very good cold junction compensation for
thermocouples.
Calys 50 allows 4mm connectors and also security
connectors to be connected on the front panel.
Display resolution:
Calys 50 allows the digit number after the dot to be selected: This function is justified by the needs of users
who want or not to display the best resolution for calibration or on the contrary limit it for simple verifications.
Functions:
Calys 50 allows the following physical values to be measured and simulated:
•
•
•
•
•
Voltage
Current
Resistance
Temperature by resistive probes and thermocouples
Frequency for frequency signals and for dry contacts
It is able to perform scaling of process signals and so to correct temperature probes.
It is compatible with HART transmitters by inserting a 250ohms resistance which doesn’t disturb digital data
transfer.
Display
Calys 50 dual display indicates permanently the
measurement value, and also the emitted value, the
gauge and the used functions.
On the top date, time and also external temperature
are also indicated.
During measuring average, maximum, minimum
and the number of measurements are displayed on
the left. While for emission this part of screen
displays all details of ramps, steps and constant
value emission functions.
Drop-down menus are used with the navigator, and
an on-line help is available to make easier
connections of probes and wires.
Ed 02/10/04 Document susceptible d'être modifié

CALIBRATION
Functions and performances:@23°C ±5°C
Accuracies are given in % of reading(a CALYS 50 display) + fixed value
Direct current: Measurement
Calys 50 is able to measure up to 50 mA, with or
without power supply of loop (24V power supply).
Range Resolution Accuracy / 1yr Remarks
±50mA 1 µA 0,018%R + 2 µA Rin < 25 Ω
For measurements of transmitters outputs, special
ranges give a dual display using mA and % of full
scale.
Calys 50 also allows to linearize following linear or
quadratic signal.
In current measurement Hart compatibility can
selected to measure currents coming from Hart
protocol transmitters
DC current: Emission
Range Resolution Accuracy / 1yr
24mA 1 µA 0,018%R + 2 µA
Emission with or without loop supply (24V)
Preprogrammed steps
0% 25% 50
%
75% 100%
4-20mA linear 4 8 12 16 20
0-20mA quad 0 5 10 15 20
4-20mA linear 4 5 8 13 20
0-20Ma quad 0 1,25 5 11,25 20
4-20mA valves 3,8-4 -4,2 12 19,20,21
Direct voltage: Measurement
Range Resolution Accuracy / 1yr Remarks
±100mV 1 µV 0,013%R + 3 µV Rin > 10 MΩ
±1V 10 µV 0,013%R + 20 µV Rin > 10 MΩ
±10V 100 µV 0,015%R + 200 µV Rin = 1MΩ
±50V 1 mV 0,015%R + 2 mV Rin = 1MΩ
Rin: input resistance
Direct voltage: Emission
Range Resolution Accuracy / 1yr Remarks
100mV 1 µV 0,013% R+ 3 µV Load 1KΩ
2V 10 µV 0,013% R+ 30 µV Load 2KΩ
20V 100 µV 0,015% R+ 300 µV Load 4KΩ
Frequency and counting: Measurement
Range Resolution Accuracy / 1yr
20 kHz < 0.01 Hz 0.005%R
Threshold triggering: 1V
Unit scale: pulse/min or Hz
Measurement on frequency signal and on dry
contacts
Measurement for counting will be done on defined
time or on infinite time.
Frequency and pulses: Emission
Range Resolution Accuracy / 1yr
1000 Hz < 0.01 Hz 0.005% R
10 kHz 10 Hz 0.005% R
Unit scale: pulse/min or Hz
Pulse emissions
Dry contact simulation
Max amplitude: 20V selectable by user
Resistance: Measurement
Range Resolution Accuracy / 1yr Remarks
400 Ω1 mΩ0.012% R+ 10 mΩMeasureme
nt current =
0.25 mA
4000 Ω10 mΩ0.012% R+ 100 mΩMeasureme
nt current =
0.25 mA
2, 3 or 4 wires resistance measurement :automatic
recognition of number of connected wires, with
indication on screen
Resistance: Emission
Range Resolution Accuracy / 1yr Remarks
400 Ω10 mΩ0.014% + 30 mΩIext from
0.1 to 10
mA
4000 Ω100 mΩ0.014%R + 300 mΩIext from
0.01 to 0.1
mA
Resistance emission: establishing time <1ms for
compatibility with smart transmitters type
Ed 02/10/04 Document susceptible d'être modifié

CALIBRATION
Temperature
Resistive probes: Measurement and emission
Probe type Range Resolution
Measurement
Accuracy / 1yr
Measurement
Resolution
Emission
Accuracy / 1yr
Emission
Pt 50(α= 3850) - 220°C + 1 200°C 0,01°C 0,012 % L+ 0,06°C 0,03°C 0,014 % L+ 0,18°C
Pt 100 (α= 3850) - 220°C + 1 200°C 0,01°C 0,012 % L+ 0,05°C 0,02°C 0,014 % L+ 0,12°C
JPt 100 (α= 3916) - 200°C + 510°C 0,01°C 0,012 % L+ 0,05°C 0,02°C 0,014 % L+ 0,12°C
Pt 100 (α= 3926) - 210°C + 850°C 0,01°C 0,012 % L+ 0,05°C 0,02°C 0,014 % L+ 0,12°C
Pt 200 (α= 3851) - 220°C + 600°C 0,01°C 0,012 % L+ 0,12°C 0,10°C 0,014 % L+ 0,33°C
Pt 500 (α= 3850) - 220°C + 1 200°C 0,01°C 0,012 % L+ 0,07°C 0,03°C 0,014 % L+ 0,18°C
Pt 1 000 (α= 3851) - 220°C + 1 200°C 0,01°C 0,012 % L+ 0,05°C 0,02°C 0,014 % L+ 0,08°C
Ni 100 (α= 618) - 60°C + 180°C 0,01°C 0,012 % L+ 0,03°C 0,01°C 0,014 % L+ 0,08°C
Ni 120 (α= 672) - 40°C + 205°C 0,01°C 0,012 % L+ 0,03°C 0,01°C 0,014 % L+ 0,08°C
Ni 1 000 (α= 618) - 60°C + 180°C 0,01°C 0,012 % L+ 0,03°C 0,01°C 0,014 % L+ 0,08°C
Cu 10 (α= 427) - 70°C + 150°C 0,1°C 0,012 % L+ 0,18°C 0,01°C 0,014 % L+ 0,10°C
Cu 50 (α= 428) - 50°C + 150°C 0,01°C 0,012 % L+ 0,06°C 0,03°C 0,014 % L+ 0,15°C
Resistive probes measurements in 2,3 or 4 wires: automatic recognition of number of connected wires, with indication on
screen
• Temperature coefficient: < 10 % of accuracy /°C.
• The accuracy in table above is given for a sensor connection in 4 wires
• Take into account peculiar error of temperature sensor used and implementation conditions
• Measurement current: 0.01mA to 1mA
• Establishing time: <1ms for simulation (simulation on quick transmitters)
Thermocouples: Measurement and Emission
Measurement Emission
Type Range Resolution Accuracy/1 year OUT range Resolution Accuracy / 1 yr
K
- 250 to - 200°C
- 200 to - 120°C
- 120 to - 0°C
+ 0 to + 1 372°C
0,2°C
0,1°C
0,05°C
0,05°C
0,80°C
0,25°C
0.1°C
0,013 % R +0,08°C
- 240 to - 50°C
- 50 to - 0°C
+ 0 to + 1 372°C
0,2°C
0,1°C
0,05°C
0,60°C
0,10°C
0,013 % R + 0,08°C
T
- 250 to - 200°C
- 200 to - 120°C
- 120 to - 50°C
- 50 to + 400°C
0,2°C
0,05°C
0,05°C
0,05°C
0,70°C
0,25°C
0.10°C
0,013 % R +0,08°C
- 240 to - 100°C
- 100 to - 0°C
+ 0 to + 400°C
0,2°C
0,05°C
0,05°C
0,40°C
0,10°C
0,013 % R + 0,08°C
J
- 210 to - 120°C
- 120 to - 0°C
+ 0 to + 1 200°C
0,05°C
0,05°C
0,05°C
0,25°C
0.09°C
0,013 % R +0,07°C
- 210 to - 0°C
+ 0 to + 1 200°C
0,05°C
0,05°C
0,20°C
0,013 % R + 0,07°C
E
- 250 to - 200°C
- 200 to - 100°C
- 100 to - 0°C
+ 0 to + 1 000°C
0,1°C
0,05°C
0,05°C
0,05°C
0,45°C
0,15°C
0.07°C
0,013 % R +0,05°C
- 240 to - 100°C
- 100 to + 40°C
+ 40 to + 1 000°C
0,10°C
0,10°C
0,05°C
0,25°C
0,10°C
0,013 % R + 0,05°C
R
- 50 to + 150°C
+ 150 to + 550°C
+ 550 to + 1 768°C
0,5°C
0,2°C
0,1°C
0,80°C
0,013 % R +
0,35°C
0,013 % R +0,2°C
- 50 to + 350°C
+ 350 to + 900°C
+ 900 to + 1 768°C
0,5°C
0,2°C
0,1°C
0,5°C
0,013 % R + 0,35°C
0,013 % R + 0,20°C
S
- 50 to + 150°C
+ 150 to + 550°C
+ 550 to + 1 768°C
0,5°C
0,2°C
0,1°C
0,80°C
0,013 % R +0,35°C
0,013 % R +0,25°C
- 50 to + 120°C
+ 120 to + 450°C
+ 450 to + 1 768°C
0,5°C
0,2°C
0,1°C
0,8°C
0,013 % R + 0,35°C
0,013 % R + 0,25°C
B + 400 to + 900°C
+ 900 to + 1 820°C
0,2°C
0,1°C
0,013 % R + 0,4°C
0,013 % R+ 0,2°C
+ 400 to + 850°C
+ 850 to + 1 820°C
0,2°C
0,1°C
0,013 % R + 0,4°C
0,013 % R+ 0,2°C
U - 200 to + 660°C 0,05°C 0,15°C - 200 to + 600°C 0,05°C 0,15°C
L - 200 to + 900°C 0,05°C 0,2°C - 200 to + 900°C 0,05°C 0,2°C
C - 20 to + 900°C
+ 900 to + 2 310°C
0,1°C
0,1°C
0,25°C
0,013 % R+ 0,15°C
- 20 to + 900°C
+ 900 to + 2 310°C
0,1°C
0,1°C
0,25°C
0,013 % R+ 0,15°C
N
- 240 to - 190°C
- 190 to - 110°C
- 110 to – 0°C
+ 0 to + 1 300°C
0,2°C
0,1°C
0,05°C
0,05°C
0,5°C
0,15°C
0,08°C
0,013 % R +0,06°C
- 240 to - 190°C
- 190 to - 110°C
- 110 to – 0°C
+ 0 to + 1 300°C
0,2°C
0,1°C
0,05°C
0,05°C
0,3°C
0,15°C
0,08°C
0,013 % R + 0,06°C
PR - 100 to + 1 400°C 0,05°C 0,3°C - 100 to + 1 400°C 0,05°C 0,3°C
Mo 0 to + 1 375°C 0,05°C 0,013 %R + 0,06°C 0 to + 1 375°C 0,05°C 0,013 %R + 0,06°C
NiMo
/NiCo
- 50 to + 1 410°C 0,05°C 0,013 %R + 0,30°C - 50 to + 1 410°C 0,05°C 0,013 %R + 0,30°C
Accuracy is warranted for reference junction (RJ) at 0°C
With use of internal RJ (except couple B) add a additional uncertainty of 0.3°C
CJC localisation can be selected by keypad programming, except for couple B:
• External at 0°C, internal (temperature compensation of instrument’s terminals) or by temperature programming
• Temperature coefficient: <10% of accuracy /°C. Display unit: °C and F.
Ed 02/10/04 Document susceptible d'être modifié

CALIBRATION
Additional functions
File Menu:
User can save up to 10 full configurations of the
instruments and recall them whenever.
Configurations can be saved and recalled in
function of user and of use. Configurations include
all programming done on instrument, as the range.
Contrast adjustment:
Screen’s contrast can be adjusted whenever to fit
with measurement environment
Screen Backlighting:
Time of backlighting can be programmed to save
battery
Autonomy:
Calys 50 autonomy is 8 hours in the worst condition
of use
Scaling:
In measurement and simulation, scaling allows
process signals to be displayed in % of FS or in all
other unit
This function also allows sensors to be corrected
after a calibration
Relative measurement:
AOIP
BP 182
91006 EVRY CEDEX
•
•
•
•
•
•
Programming of a reference value different
from the one of the instrument (NUL function)
Substracting of constant value by measuring or
programming it from a measured value(TARE
function)
Square root:
In current measurement and simulation, this
function allows to take into account a quadratic
signal coming from transmitter of type ∆P
Statistical functions:
Average, minimum, maximum and also number of
measurements done are always displayed.
Reset key allows values to be updated.
Simulation Menu:
Simulation value is set by entering value on keypad
or by changing the according digit with the cursor
Ramps generation:
Starting, ending and length time values of simple or
cyclic ramps can be set to do simulation.
Number of ramps can also be adjusted in case of
cyclic ramps for any signals.
Steps simulation:
2 modes are proposed.
Program mode: Starting value, number of
steps and the length time have to be set
Manual mode: User has about a hundred of
preset values
In current simulation, user will have some additional
preset values in function of range and according to
0%, 25%, 50%, 75% and 100% from selected
gauge. Choice is done between gauges:
0-20mA: linear or quadratic
4-20mA: linear or quadratic
Synthesizer:
With 100 values manually set, Calys 50 allows
curve generation to be remade.
Transmitter function:
CALYS 50 is able to be used as a transmitter.
Measurement input is copied on the output with
scaling.
Calys 50 is supplied in standard with 6 testing leads, a quick battery charging system, an instruction manual
Instructions to order:
On-site calibrator CALYS 50
Protective carrying Case: ACL6050
Ed 02/10/04 Document susceptible d'être modifié