Apeks WTX INFLATOR User manual

TECHNICAL MAINTENANCE MANUAL
WTX INFLATOR

PAGE # REV. DATE TITLE OR DESCRIPTION CHANGE MADE BY
All March 20, 2010 Remove section on replacing
brass button Aqua Lung America
All Nov 1, 2014 Add WTX-D & PSD Inator Aqua Lung America
8 Feb 1, 2017 Installation change to PSD inator Spectra cord Aqua Lung America
6 Sept 1, 2017 Change to Service Procedures (QD Fitting) Aqua Lung America
CHANGE RECORD
2

COPYRIGHT NOTICE............................................................................................. 4
GENERAL GUIDELINES.........................................................................................4
GENERAL CONVENTIONS.................................................................................... 4
DISASSEMBLY PROCEDURE............................................................................... 5
REASSEMBLY PROCEDURE................................................................................ 7
FINAL TESTING......................................................................................................10
TABLE 1: TROUBLESHOOTING GUIDE............................................................... 12
TABLE 2: LIST OF TOOLS .....................................................................................13
TABLE 3: CHECKING SPECIFICATIONS.............................................................. 13
TABLE 4: RECOMMENDED CLEANERS AND LUBRICANTS............................. 14
PROCEDURE A: CLEANING AND LUBRICATING................................................15
MAINTENANCE NOTES..........................................................................................16
WTX INFLATOR, EXPLODED VIEW...................................................................... 17
CONTENTS
3
WTX Inator Technical Maintenance Manual

WARNINGS, CAUTIONS, & NOTES
Pay special attention to information provided in warnings, cautions and
notes that are accompanied by one of these symbols:
SCHEDULED SERVICE
It is recommended that the Inator and Over-Pressure valves should be
rinsed in fresh water after each use, and should be disassembled and
serviced annually.
However, if at all unsure about the correct functioning of the
Inator, then it must be inspected immediately.
WARNINGS indicate a procedure or situation that may result in
serious injury or death if instructions are not followed correctly.
CAUTIONS indicate any situation or technique that will result in
potential damage to the product, or render the product unsafe
if instructions are not followed correctly.
NOTES are used to emphasize important points, tips and reminders.
An Ofcial Inspection consists of:
1. Check that all parts are assembled correctly and that no parts are loose.
2. Inspect for signs of corrosion, cracks, damage to sealing surfaces and
check the general condition of the ination assemblies.
3. Inspect ribbed hose for holes or tears and conrm that it is securely
clamped on both ends.
4. Follow FINAL TESTING instructions located in this manual.
If a inator fails any of the 4 steps it should be fully serviced.
Pinch Method
Press upwards on sides of o-ring to
create a protrusion. Grab o-ring or
insert o-ring tool at protrusion.
COPYRIGHT NOTICE
This manual is copyrighted, all rights reserved. It may not, in whole or in
part, be copied, photocopied, reproduced, translated or reduced to any
electronic medium or machine-readable form without prior consent in
writing from Aqua Lung International. It may not be distributed through the
internet or computer bulletin board systems without prior consent in writing
from Aqua Lung International.
©2017 Aqua Lung America, Inc.
WTX InatorTechnical Maintenance Manual
INTRODUCTION
This manual provides factory prescribed procedures for the correct service
and repair of the Aqua Lung or Apeks products described in this manual. It is
not intended to be used as an instructional manual for untrained personnel.
The procedures outlined within this manual are to be performed only by
personnel who have received Factory Authorized training through an
Aqua Lung or Apeks Service & Repair Seminar. If you do not completely
understand all of the procedures outlined in this manual, contact Aqua Lung
to speak directly with a Technical Advisor before proceeding any further.
GENERAL GUIDELINES
1. In order to correctly perform the procedures outlined in this manual, it
is important to follow each step exactly in the order given. Read over the
entire manual to become familiar with all procedures before attempting to
disassemble the product in this manual, and to learn which specialty tools
and replacement parts will be required. Keep the manual open beside you
for reference while performing each procedure. Do not rely on memory.
2. All service and repair should be carried out in a work area
specically set up and equipped for the task. Adequate lighting, cleanliness,
and easy access to all required tools are essential for an efcient repair facilit y.
3. As the product is disassembled, reusable components should be seg-
regated and not allowed to intermix with nonreusable parts or parts from
other units. Delicate parts, which contain critical sealing surfaces, must
be protected and isolated from other parts to prevent damage during the
cleaning procedure.
4. Use only genuine Aqua Lung or Apeks parts for the service of this product.
DO NOT attempt to substitute an original part with another manufacturer’s,
regardless of any similarity in shape or size.
5. Do not attempt to reuse mandatory replacement parts under any circum-
stances, regardless of the amount of use the product has received since it
was manufactured or last serviced.
6. When reassembling, it is important to follow every torque
specication prescribed in this manual, using a calibrated torque wrench.
Most parts are made of either marine brass or plastic, and can be perma-
nently damaged by undue stress.
7. In order to make the product compatible with nitrox up to 40%
O2 (EAN40), the product must be properly cleaned, lubricated
and assembled using genuine Aqua Lung or Apeks replace-
ment parts. In addition, assembly must be carried out in a clean
environment using powderless, latex gloves or equivalent. For more detailed
information, be sure to read Procedure A: Cleaning and Lubricating.
GENERAL CONVENTIONS
Unless otherwise instructed, the following terminology and techniques
are assumed:
1. When instructed to remove, unscrew, or loosen a threaded part, turn
the part counter-clockwise.
2. When instructed to install, screw in, or tighten a threaded part, turn
the part clockwise.
3. When instructed to remove an o-ring, use the pinch method
(see illustration below) if possible, or use a brass or plastic o-ring removal
tool. Avoid using hardened steel picks, as they may damage the o-ring
sealing surface. All o-rings that are removed are discarded and replaced
with brand new o-rings.
4. The following acronyms are used throughout the manual: MP is Medium
Pressure; HP is High Pressure; LP is Low Pressure.
5. Numbers in parentheses reference the key numbers on the
exploded parts schematics. For example, in the statement,
“...remove the o-ring (7) from the crown (8)...”, the number 7 is
the key number to the crown o-ring.
4

DISASSEMBLY PROCEDURE
1Loosen the collar (1) and remove the inator assembly from
the air cell. Remove the gasket (2) from the air cell manifold.
2Using side cutters, snip the two lower clamps (11) at the buckle.
Carefully remove the ribbed hose (12) off the inator body.
4Using side cutters, snip the upper clamp (11) at the buckle.
Carefully remove the ribbed hose (12) from the elbow (4).
Remove the collar (1) from the elbow (4). Using a small punch or
screwdriver, carefully push the retaining pin (5) out of the elbow
and remove the cable (13).
NOTE: Before performing any disassembly, refer to the exploded
parts drawing, which references all mandatory replacement parts.
These parts should be replaced with new, and must not be reused
under any circumstances - regardless of the age of the product or
how much use it has received since it was last serviced.
CAUTION: Use only a plastic or brass o-ring removal tool
when removing o-rings to prevent damage to the sealing sur-
face. Even a small scratch across an o-ring sealing surface
could result in leakage. Once an o-ring sealing surface has
been damaged, the part must be replaced with new. DO NOT
use a dental pick or any other steel instrument.
NOTE: Inspect and clean the inside clamping areas of the ribbed hose and
the connection point on the inator body and elbow.
Using a small punch or screwdriver, carefully push the retaining pin (5) out
of the inator body. The Spectra cord (13) will slide off the retaining pin.
Using a small punch or screwdriver, carefully push the retaining pin (5)
out of the inator body. The cable (13) will slide off the retaining pin.
3 WTX INFLATOR
Separate the ribbed hose (12) from the elbow (4). Slide the collar
(1) off the elbow.
NOTE: Inspect and clean the inside clamping areas of the ribbed hose and
the connection point on the inator body and elbow.
NOTE: Move forward to step 11
PSD/BC-1 INFLATOR
3
5 WTX INFLATOR
PSD/BC-1 INFLATOR
6
5
WTX Inator Technical Maintenance Manual

12Using a brass or plastic o-ring removal tool, remove the o-
ring (15) and the o-ring (16) from the valve module (14).
11 Using the Apeks Inator
Tool (pn 15762), remove
the ination valve module (14)
from the inator body.
CAUTION: DO NOT attempt to disassemble the valve module.
The valve module is assembled with permanent LocTite and
can not be disassembled. No internal service parts are available
for the module. Attempting to disassemble the valve module
will cause permanent damage to the module requiring replace-
ment of the module assembly.
8
Pull down on the Spectra cord (13) to remove the poppet guide
assembly (6-10) from the elbow (4). Once the poppet guide as-
sembly is out, remove the poppet dump valve (6).
9
Remove the Spectra cord (13) from the inator cable hook (10).
10
Use needle nose pliers to gently grip the end of the poppet
stem (7). Separate the inator cable hook (10), poppet stem,
pull dump spring (8) and poppet guide (9).
7
Pull and keep slight tension on the Spectra cord (13). Use a
3mm hex key to push in the legs of the poppet guide (9) one at
a time to release it.
13Remove the mouthpiece
(17) from the body.
THIS CONCLUDES DISASSEMBLY
CAUTION: DO NOT attempt to remove the QD tting. The QD
tting is assembled with permanent LocTite and is not remove-
able. Attempting to remove the QD tting can cause permanent
damage to the inator requiring replacement of the complete
inator assembly.
CAUTION: DO NOT attempt to remove the oral button. The oral
button is assembled with permanent LocTite and is not remove-
able. No service parts are available for the oral button assem-
bly. Attempting to remove the oral button can cause permanent
damage to the inator requiring replacement of the complete
inator assembly.
6

NOTE: Before beginning reassembly, perform
parts cleaning and lubrication in accordance
with Procedure A: Cleaning and Lubricating.
NOTE: Before performing any reassembly, it is important to inspect
all parts, both new and those that are being reused, to ensure that
every part and component is perfectly clean and free of any dust,
corrosion, or blemishes. Before dressing each o-ring with Christo-
Lube®, check to ensure it is clean, supple, and free of any blemish.
WARNING: Use only genuine Aqua Lung/Apeks parts, subas-
semblies, and components whenever assembling any Aqua
Lung/Apeks product. DO NOT attempt to substitute an Aqua
Lung/Apeks part with another manufacturer’s, regardless of any
similarity in shape, size or appearance. Doing so may render
the product unsafe, and could result in serious injury or death.
REASSEMBLY PROCEDURE
Push the retaining pin (5) halfway into the elbow (4). Slide the cable
(13) onto the pin. Push the pin the rest of the way through the el-
bow, capturing the cable. Make sure the pin is ush on both sides
of the elbow.
NOTE: If servicing the PSD/BC-1 model inator,
Move forward to step 6
2Slide the collar (1) onto the
elbow (4).
3Slide the ribbed hose (12)
over the cable (13).
4Brush a thin layer of Gasgacinch onto the elbow (4). Slide the
ribbed hose (12) onto the elbow up to the line. Align the ribbed
hose so the at side is against your body.
5Lightly fasten a clamp (11) over the ribbed hose (12). Position
the buckle of the clamp to the side. Pull tight by hand. Using
side cutters, trim the excess strap.
NOTE: Move forward to step 16
1 WTX INFLATOR
Slide the spring (8) onto the
poppet stem (7). Insert the
narrow end of poppet stem into
the poppet guide (9). Install the
dump valve poppet (6) onto the
poppet stem.
PSD/BC-1 INFLATOR
6
7
WTX Inator Technical Maintenance Manual

8
With the Spectra cord (13) doubled up, make sure the knot is
in the center of the cord. Take one end of the cord and thread it
through the double-looped end of the hook (10). Take the other end
of the cord and bring it through the loop in the cord creating a Larks
Head knot. Pull the knot snug.
9
Install poppet assembly into elbow (4). The cut-out on the poppet
guide (9) faces the bottom of the elbow.
12
Pull on Spectra cord (13) to conrm actuation of pull dump.
Install collar (1) onto elbow (4).
13
Insert the Spectra cord (13) through the ribbed hose (12). If
needed, use a large paperclip as a weight to get the cord to
slide through the ribbed hose.
15Lightly fasten a clamp (11) over the ribbed hose (12). Posi-
tion the buckle of the clamp to the side. Pull tight by hand.
Using side cutters, trim the excess strap.
14Brush a thin layer of Gasgacinch onto the elbow (4). Slide
the ribbed hose (12) onto the elbow up to the line.
7
Compress the poppet spring
(8) and install the inflator
cable hook (10) onto the poppet
stem (7). The hook should be held
at the single-wire end.
10
Squeeze poppet guide (9) feet together and push into elbow
(4).
11
Conrm poppet guide (9)
is locked into place and
poppet guide feet are seated ush
with the elbow (4) on both sides.
8

16Install o-rings (15 & 16)
to the valve module (14).
17Insert the valve module (14) into the body. Tighten down
using the Apeks Inator Tool (pn 15762) until snug.
18Replace the retainer loop (3) if removed during disassem-
bly. Install the retaining pin (5) halfway into the body.
Slide the cable (13) over the pin (5). Slide the pin completely into
the body to capture the cable. Make sure the pin is ush on both
sides of the body.
WTX INFLATOR
19
PSD/BC-1 INFLATOR
Create a Larks Head knot in the
Spectra cord (13) and slide the
loop over the pin (5). Slide the pin
completely into the body to cap-
ture the cord. Make sure the pin
is ush on both sides of the body.
20Brush a thin layer of Gasgacinch onto the body. Slide the
ribbed hose (12) onto the body up to the edge. Align the
ribbed hose and body so there is no twist.
21Take note of raised edge
on body. Lightly fasten two
clamps (11) over the ribbed hose
(12), one below and one above
the raised edge of the body. Posi-
tion both buckles to the side. Pull
tight by hand. Using side cutters,
trim the excess strap.
22Slide the mouthpiece (17)
onto the body.
9
WTX Inator Technical Maintenance Manual

THIS CONCLUDES REASSEMBLY
23Install a new gasket (2)
into the air cell manifold.
2Turn the collar (1) over and
hold next to the manfold
of the air cell. Take note of the
12-sided placement notches on
the manifold and inside the col-
lar. These must align together
while tightening down the collar.
3Place the collar (1) over
the manifold of the air cell.
Gently turn the collar to engage
the threads, taking care not to
cross-thread. Line up the ina-
tor assembly in the desired posi-
tion while tightening the collar by
hand until snug.
4Install a MP inator hose to a properly adjusted rst stage. At-
tach the rst stage to a cylinder lled to 3000 psi (206 bar). At-
tach the hose to the quick disconnect tting on the inator.
5Slowly open the cylinder
valve and pressurize the
regulator, listen for any leaks.
There should be no leaking of
gas. If leakage occurs, deter-
mine the source and de-pres-
surize. Refer to Table 1: Trou-
bleshooting Guide.
CAUTION:Before pressurizing the rst stage, it is important
to have a properly adjusted second stage attached to the rst
stage. This will provide a safety relief valve if the MP exceeds
145 psi (10 bar). Failure to relieve increasing MP may result in
damage to the MP hose.
FINAL TESTING
1While holding the upper elbow (4) secure, rmly grasp the in-
ator and pull it in a straight line directly away from the elbow.
Check both of the attachment points of the ribbed hose (12). Check
the ribbed hose for signs of damage or decay, if found you will need
to replace the ribbed hose before proceeding.
10

6Depress the inator but-
ton several times to ensure
that airow is unobstructed, DO
NOT completely ll the air cell.
After releasing the button, listen
carefully to ensure that the air-
ow has completely stopped. If
airow continues, de-pressurize
and refer to Table 1: Trouble-
shooting Guide.
7Orally inate the air cell with
several deep exhalations
to conrm air is entering the air
cell. If air is not owing into air
cell refer to Table 1: Trouble-
shooting Guide.
8Hold the inator button de-
pressed to fully inate the
air cell until an over-pressure
relief valve opens to release the
excess pressure.
CAUTION:Before performing this test, conrm the aircell has
functioning over-pressure relief valves. Failure to do this could
cause permanent damage to the air cell that will require replace-
ment of the aircell.
NOTE: The WTX inator does not have a built in over-pressure
relief valve.
PSD/BC-1 Inator Only: Pull straight down on the inator to vent
air from the inator exhaust valve.
Conrm both methods will rapidly deate the air cell.
10With the air cell fully
inated, listen for any
leaks. Submerge the entire in-
ator assembly in fresh water
and check for any bubbles. If a
leak is found, depressurize and
refer to Table 1: Troubleshoot-
ing Guide.
11 If leaks were not detect-
ed, conrm the air cell
is fully inated. Disconnect the
MP inator hose from the quick
disconnect tting. Set a timer
for one hour to ensure there are
no leaks. If air loss occurs, de-
pressurize and refer to Table 1:
Troubleshooting Guide.
THIS CONCLUDES THE SERVICE
OF THE WTX INFLATOR
12If the air cell has begun to deate within one hour, fully in-
ate the air cell once again and hold the entire unit com-
pletely submerged in fresh water for at least one minute to deter-
mine the source of the air loss. Carefully observe the inator for
any signs of bubble formation indicating a leak, especially around
the inator button, ribbed hose & manifold connection. If a leak is
found, refer to Table 1: Troubleshooting Guide.
9With the air cell fully inated, Press the oral button briey to
vent air.
11
WTX Inator Technical Maintenance Manual

TABLE 1: TROUBLESHOOTING GUIDE
CAUTION: Recommended treatments which require disassem-
bly of the regulator must be performed during a complete over-
haul, according to the prescribed procedures for scheduled,
annual service. Do not attempt to perform partial service.
NOTE: This is a partial list of possible problems and recommended
treatments. For more information, contact Aqua Lung’s Technical Ser-
vices Department for assistance with problems not described here.
SYMPTOM POSSIBLE CAUSE TREATMENT
Restricted airow or
Aircell inates slowly
(with full tank, stable MP)
1. MP hose is obstructed 1. Clean or replace hose
2. Dirt/salt deposits are present within the
inator assembly 2. Flush with warm fresh water
3. MP of supply gas too low 3. Ensure MP is set properly
External air leak from
inator button
1. Dirt /salt deposits are present on internal
sealing surfaces of valve module
1. Remove valve module and thoroughly
clean all parts before reassembly
2. The o-ring (16) is damaged and worn 2. Replace o-ring
3. Damage to o-ring grooves of valve module (14) 3. Replace valve module
4. The inator body internal sealing surfaces
for valve module are damaged 4. Replace lower inator assembly
Air leak from mouthpiece
1. Dirt/salt deposits present on internal sealing
surfaces of oral button 1. Flush with warm fresh water
2. Damage to oral button assembly 2. Replace lower inator assembly
Air leak from ribbed hose
or connection points
1. Cuts, tears or pinholes in hose 1. Replace ribbed hose
2. Hose connection points not clean or
damage to sealing surfaces
2. Inspect all hose connection points &
clean or replace damaged parts
BC auto-inating
1. Dirt/salt deposits are present on internal sealing
surfaces of valve module (14) 1. Flush with warm fresh water
2. Damage to internal components of valve
module (14) 2. Replace valve module
Cannot orally inate the
buoyancy cell
1. Dirt/salt deposits are present on internal
sealing surfaces 1. Flush with warm fresh water
2. Internal parts of oral inflator worn or
damaged 2. Replace lower inator assembly
3. Tear/hole in ribbed hose (12) 3. Replace ribbed hose
4. Collar (1) not tightened properly 4. Remove collar from aircell & reinstall
5. Gasket (2) missing or damaged 5. Replace gasket
Air does not vent when
rapid exhaust valve cable
is pulled (PSD/BC-1)
1. Rapid exhaust cable (13) is not properly
connected to the inator or elbow, or is damaged
1. Check condition and connections of
cable, and correct as needed
2. Incorrect rapid exhaust valve cable (13)
installed (to long) 2. Replace cable
Air leaks continuously
from Upper-Valve when
BC is inated (PSD/BC-1)
1. Dump plug (6) is worn or damaged 1. Replace dump plug
2. Dump plug spring (8) is damaged 2. Replace spring
3. Incorrect rapid exhaust valve cord (13)
installed (to short) 3. Replace cord
12

TABLE 2: LIST OF TOOLS
PART NO. DESCRIPTION APPLICATION
15762 Apeks Inator Tool Removing and installing valve module (16)
944022
103102
Brass O-ring
Tool Set
O-ring Tool
(Plastic)
Removing and installing o-rings and gasket
N/A Side Cutters Removing/Trimming panduit clamps (11)
N/A 1/8” Punch Removing retaining pins (5)
N/A 3mm Hex Key Release poppet guide legs (9)
N/A Christo-Lube Lubrication of o-rings
N/A Gasgacinch Sealing of ribbed hose (12)
N/A
Magnier with illumination
Sealing surface inspection
N/A Ultrasonic cleaner Brass and stainless steel parts cleaning
TABLE 3: CHECKING SPECIFICATIONS
TEST SPECIFICATION
Leak Test No Leaks Permitted
13
WTX Inator Technical Maintenance Manual

TABLE 4: RECOMMENDED CLEANERS & LUBRICANTS
LUBRICANT/CLEANER APPLICATION SOURCE
Christo-Lube®MCG 111
PerouroLube 20/1
Lubricant for all o-rings
Aqua Lung, PN 820466,
or
Lubrication Technologies
310 Morton Street
Jackson, OH 45640
(800) 477-8704
Performance Fluids
Ste 101 Lomeshaye Business Park
Turner Road
Nelson Lancashire BBP 7DR
White distilled vinegar
(diluted with water)
Bath for reusable stainless
steel and brass parts. “Household” grade
Oxygen Compatible Solution
Promoclean TP108
Janitol Plus
Nitrox/O2 Cleaning
As Per Training
INVENTEC PERFORMANCE
CHEMICALS SA.
20, Rue de bourgogne
BP 211
69802 SAINT-PRIEST cedex
JOHN LAWSON DIST.
SCOTSHAW BROOK HOUSE
BRANCH ROAD
LOWER DARWEN
LANCASHIRE BB3 0PR
Liquid Dishwashing Detergent
(diluted with warm water)
Degreaser for brass and stain-
less steel parts; general clean-
ing solution for plastic and
rubber
“Household” grade
CAUTION: Silicone rubber requires no lubrication or preservative treatment. DO NOT apply grease or spray to silicone rub-
ber parts. Doing so may cause a chemical breakdown and premature deterioration of the material.
CAUTION: Do not use muriatic acid for the cleaning of any parts. Even if strongly diluted, muriatic acid can harm chrome
plating and may leave a residue that is harmful to o-ring seals and other parts.
14

PROCEDURE A: CLEANING & LUBRICATING
Cleaning Brass and Stainless Steel Parts
1. Preclean in warm, soapy water* using a soft nylon bristle brush.
2. Thoroughly clean parts in an ultrasonic cleaner lled with soapy water. If there are stubborn deposits, household white distilled
vinegar (acetic acid) in an ultrasonic cleaner will work well. DO NOT place plastic, rubber, silicone or anodized aluminum parts
in vinegar.
3. Remove parts from the ultrasonic cleaner and rinse with fresh water. If tap water is extremely “hard,” place the parts in a bath
of distilled water to prevent any mineral residue. Agitate lightly, and allow to soak for 5-10 minutes. Remove and blow dry with
low pressure (25 psi) ltered air, and inspect closely to ensure proper cleaning and like-new condition.
Cleaning Anodized Aluminum, Plastic & Rubber Parts
Anodized aluminum parts and parts made of plastic or rubber, such as box bottoms, box tops, dust caps, etc., may be soaked and
cleaned in a solution of warm water mixed with mild dish soap. Use only a soft nylon toothbrush to scrub away any deposits. Rinse
in fresh water and thoroughly blow dry, using low pressure ltered air.
Cleaning MP Hoses (Air use Only)
Follow Hose Inspection & Cleaning Guidelines for more detailed information
1. Hose ttings: Ultrasonically clean with soapy water; Use soft nylon bristle brush. If corrosion is evident, use a brass bristle brush.
2. Run water through hose if needed
3. Thoroughly rinse with fresh water
4. Blow out hose before installing
Lubrication and Dressing
Wear powderless, latex gloves when handling and lubricating o-rings. Keeping internal parts free from skin oils and other contaminates
is important when running enriched air nitrox through a rst stage. All o-rings should be lubricated with Christo-Lube®M CG -111. D re ss
the o-rings with a very light lm of grease, and remove any visible excess by running the o-ring between thumb and forenger. Avoid
applying excessive amounts of Christo-Lube grease, as this will attract particulate matter that may cause damage to the o-ring.
*Soapy water is dened as “household” grade liquid dishwashing detergent diluted in warm water.
When it comes to issues of nitrox safety and compatibility, the concerns lie primarily with the regulator’s rst stage as it is subjected to
high inlet pressures. High inlet pressures lead to adiabatic compression or heating of the gas. The Aqua Lung or Apeks regulator product
described in this manual, when properly cleaned and assembled, is authorized for use with enriched air nitrox (EAN) that does not ex-
ceed 40% (EAN 40). It is authorized because it has undergone adiabatic compression testing and the authorized service kit components
and lubricants are compatible in elevated oxygen environments. During cleaning, a mild detergent must be used to remove condensed
hydrocarbons (compressor oils) from the inside passageways of the rst stage. For the rst stage to remain EAN40 compatible, only
use hyperltered compressed gas (hydrocarbons < 0.1 mg/m3). Ordinary compressed breathing air (Grade E) usually does not meet
this criterion. Once ordinary breathing air is used, the rst stage is no longer EAN40 compatible until it is cleaned and serviced again.
Although regulator second stage components are not exposed to high pressure EAN, Aqua Lung and Apeks recommend that the same
cleaning procedures be followed for the complete regulator. This prevents the possibility of cross contamination and guarantees the
cleanliness of the entire regulator.
AQUA LUNG AND APEKS REGULATORS AND NITROX
CAUTION: Do not place complete hose length in contact with acid solu-
tions. This could alter their physical properties and cause degradation
and premature breakdown.
CAUTION: Do not place plastic and rubber parts in contact with acid solu-
tions. This could alter their physical properties and cause degradation and
premature breakdown.
15
WTX Inator Technical Maintenance Manual

MAINTENANCE NOTES
16

WTX INFLATOR
Key # Part # Description
------ 42662 Airway, Complete, K (WTX/WTX-D)
------ 15350 Airway, Complete, K (WTX-PSD)
------ 15778 Lower Inator Assembly, K (WTX/WTX-D/PSD)
1 ----- 15706 Collar (WTX & WTX-D)
------ 15763 Collar (WTX-PSD/BC-1)
2 ----- 15309 Gasket
3 ----- 42783 Retainer Loop
4----- 15304 Elbow (WTX & WTX-D)
------ 15161 Elbow (WTX-PSD/BC-1)
5 ----- 15610 Retaining Pin
6 ----- 778559 Dump Valve, Poppet (WTX-D PSD/BC-1)
Key # Part # Description
7----- 15164 Poppet Stem (WTX-PSD/BC-1)
8 ----- 15215 Spring (WTX-PSD/BC-1)
9 ----- 15162 Poppet Guide (WTX-PSD/BC-1)
10 ---- 15178 Cable Hook (WTX-PSD/BC-1)
11 ---- 15719 Clamp
12 ---- 15756 Ribbed Hose (Flat)
------ 15151 Ribbed Hose, 18” (WTX-PSD/BC-1)
13 ---- 15757 Cable Assembly
------ N/A Spectra Cord, (WTX-PSD/BC-1)
14 ---- 15717 Valve Module w/Brass Button
------ 15761 Valve Module w/Plastic Button
15 ---- 820312P O-ring (10 pk)
16 ---- 820120P O-ring (10 pk)
17 ---- 15759 Mouthpiece
2
43
5
13
1
12
14
1
4
WTX-PSD/BC-1 MODEL ONLY
6
7
8
9
10
11
17
5
11
17
14 15 16
15
16
13
L-STYLE BODY
NLA
K-STYLE BODY
17
WTX Inator Technical Maintenance Manual

Rev 09 -2017
Aqua Lung USA
2340 Cousteau Court, Vista CA 92081
760-597-5000
www.aqualung.com
AUTHORIZED TECHNICIAN
TECHNICAL MAINTENANCE MANUAL
WTX INFLATOR
WTX INFLATOR
Table of contents
Other Apeks Diving Instrument manuals

Apeks
Apeks 388300 User manual

Apeks
Apeks NS158000 User manual

Apeks
Apeks WTX User manual

Apeks
Apeks Black Ice Twin Cylinder Kit User manual

Apeks
Apeks TX Series User manual

Apeks
Apeks ATX 200 User manual

Apeks
Apeks ThermiQ Dry Advanced User manual

Apeks
Apeks Black Ice User manual

Apeks
Apeks ThermiQ Dry Advanced User manual

Apeks
Apeks wtx series User manual

Apeks
Apeks WTX-D User manual

Apeks
Apeks Quantum User manual

Apeks
Apeks wtx series User manual

Apeks
Apeks MTX-R User manual

Apeks
Apeks QuantumX User manual

Apeks
Apeks 427106 User manual

Apeks
Apeks BLACK SAPPHIRE User manual

Apeks
Apeks Regulator User manual

Apeks
Apeks XTX200 User manual

Apeks
Apeks wtx series User manual