ASG DTT-5 User manual

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898
Phone: (888) 486-6163 | Fax: (216) 481-4519
Email: asginfo@asg-jergens.com | Web: www.asg-jergens.com Page 1 of 23
© 2017 Jergens, Inc. All Rights Reserved | Revision Date: 04/10/17
DTT Series Digital Torque Testers
User Manual
for DTT Models above Serial Number: 2000

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 2 of 23
Table of Contents
Service and Warranty ............................................................................................... 4
Overload Capacity Caution ...................................................................................... 4
System Overview...................................................................................................... 4
Battery Charger............................................................................................... 5
Standard Equipment ................................................................................................ 6
Instructions for Using the Rundown Adapter ........................................................... 7
Usage....................................................................................................................... 8
Procedure for Manual Torque Screwdrivers..................................................... 8
Procedure for Power Screwdrivers................................................................... 8
Battery Indicator ...................................................................................................... 9
Basic Function.......................................................................................................... 9
Display of Clockwise/Counter-Clockwise......................................................... 9
Zeroing the Tester.......................................................................................... 10
Changing the Unit of Measure...................................................................... 10
Changing the Mode of Measure ................................................................... 10
Backlit Display............................................................................................... 11
Saved Reading to Memory............................................................................ 11
Output Signal................................................................................................ 11
Main Menu ............................................................................................................ 11
Setting........................................................................................................... 12
Auto Reset .................................................................................................... 12
Pass-Fail ........................................................................................................ 13
Auto Memory................................................................................................ 14
Auto Print...................................................................................................... 14
Factory Reset................................................................................................. 15
Memory......................................................................................................... 15
View.............................................................................................................. 15
Delete Last .................................................................................................... 16
Delete All ...................................................................................................... 16
Upload All ..................................................................................................... 16
DTT Series Digital Torque Testers
User Manual

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 3 of 23
Table of Contents (Continued)
DTT Series Digital Torque Testers
User Manual
Statistics........................................................................................................ 17
Calibration .................................................................................................... 17
Transducer State............................................................................................ 17
Date and Time............................................................................................... 18
About............................................................................................................ 18
Software ................................................................................................................ 19
TorqueLink 2.4 .............................................................................................. 19
Menu Bar ...................................................................................................... 19
Table ............................................................................................................. 20
Real Time ...................................................................................................... 21
Result............................................................................................................ 21
Pass/Fail ........................................................................................................ 21
Retrieving Information from the DTT............................................................. 22
Capturing Data in Real Time ......................................................................... 23
Error.............................................................................................................. 23

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 4 of 23
Service and Warranty
Overload Capacity Caution
ASG, Division of Jergens, Inc., warrants to the original purchaser buying the ASG DTT meter with the intention of use rather than
resale, for a period of 1 year. ASG will replace those items found to be defective or otherwise fail to conform, or at ASG’s option
repay the price paid for the item. The buyer’s remedies with respect to any item found to be defective or otherwise not conforming
shall be limited exclusively to the right of replacement or repayment. In no event shall ASG be liable for any incidental, special, or
consequential damages or for damages in the nature of penalties.
DISCLAIMER: Seller makes no other warranty what-so-ever, expressed or implied, and all implied warranties of merchantability and
tness for a particular purpose are disclaimed and excluded from this transaction and shall not apply to the goods sold hereunder.
The ASG DTT meter is an electronic instrument and should be treated with the same care given any sensitive electronic instrument.
Avoid dropping the unit or dropping items on the unit. Avoid high shock loads to the transducer. Use the proper driver adapter when
using clutch type power tools.
Do not deliberately exceed the capacity of the unit. The maximum capacity of the DTT meter can be found on the spacer ring
surrounding the transducer. The overload capacity of a DTT meter is 200% of its maximum load capacity. If the meter is taken over
capacity, a warning tone will sound, the word “OVERLOAD” will ash on the LCD, and the unit will LOCK until the RESET button is
pressed. Immediately stop applying torque when the overload warnings are observed. A unit subjected to overload or subjected to
harsh conditions, such as use with an impact tool, will not be covered under warranty.
System Overview
RESET ZERO UNIT MEM
MENU MODE PRINT
0000.00
Track lbf.in
M013
09:16:00
LCD
Display
USB Port
RS-232 Port
Charger
Connector
On/Off Switch
DTT Series Digital Torque Testers
User Manual

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 5 of 23
System Overview (Continued)
Model DTT-5, DTT-10, DTT-30, DTT-50, DTT-100, DTT-200, DTT-500
Size 7.12” x 4.4” x 2.0” (180mm x 111mm x 50mm); Size with Mounting Plate*: 8.625” x 4.4” x 2.19” (219mm x 111mm x 56mm)
Weight 5 lb (2.3 kg) Weight with Mounting Plate*: 6.8 lb (3.1 kg)
Display Graphics LCD Display Size 128x64 dots
Battery 17 hour battery life (NiMH 7.2/v Rechargeable)
Battery Charger 110-120 VAC 60 Hz 12W input, 12 VDC 500ma output (Optional International Charger Available: 220-240 VAC input, ASG # 915966)
ADC 16 Bit
CPU 8 Bit
Communication Mini USB to USB, RS-232
Key Button 4 Buttons
Unit of Measure 6-7 User Selectable Torque Units (Depending on Model)
Mode of Measure 3 Modes: Track, Peak, and First Peak
Display Digits 6 Digits Maximum
Special Function Auto Reset, Auto Memory, Auto Zero, Real Time Clock, Pass/Fail, Statistics, Torque Bar, Battery Indicator and Charging Status, Transducer State
Memory 500 Readings with Alarm when Full
Accuracy ± 0.2% of Full Scale
DTT Series Digital Torque Testers
User Manual
Model Number ASG Number Torque Range
lbf.in N.m
DTT-5 66700 0.5 - 5.0 0.06 - 0.56
DTT-10 66701 1.0 - 10.0 0.11 - 1.13
DTT-30 66702 3.0 - 30.0 0.33 - 3.4
DTT-50 66703 5.0 - 50.0 0.56 - 5.6
DTT-100 66704 10.0 - 100.0 1.13 - 11.3
DTT-200 66705 20.0 - 200.0 2.3 - 22.6
DTT-500 66706 50.0 - 500.0 5.6 - 56.5
* DTT-200 and DTT-500 come with mounting plate
Battery Charger
110-120V US Power Adapter
(Supplied) ASG #66608
220-240V International Power Adapter
(Optional, Must Order Separately)

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 6 of 23
Standard Equipment
All DTT meters are supplied with the following standard items:
• DTT torque meter
• Rundown adapter (see chart below)
• Carrying case with insert
• Battery charger
• Mini USB to USB cable (ASG #66629)
Additional optional
rundown adapters can
be found on page 7
Included with DTT-5 (66700)
ASG Number Type Drive Size Torque Range Spring Color
lbf.in N.m
66620 Adapter w/ Spring 4mm HIOS 0.25 - 2.0 0.03 - 0.23 Black
CP66612 Adapter 1/4” Hex N/A N/A N/A
CP66614 Spring N/A 1.3 - 5.0 0.15 - 0.56 Yellow
Included with DTT-10 (66701) and DTT-30 (66702)
ASG Number Type Drive Size Torque Range Spring Color
lbf.in N.m
66620 Adapter w/ Spring 4mm HIOS 0.25 - 2.0 0.03 - 0.23 Black
CP66612 Adapter 1/4” Hex N/A N/A N/A
CP66614 Spring N/A 1.3 - 5.0 0.15 - 0.56 Yellow
CP66616 Spring N/A 4.5 - 26.0 0.51 - 2.9 Black
Included with DTT-50 (66703) and DTT-100 (66704)
ASG Number Type Drive Size Torque Range Spring Color
lbf.in N.m
66633 Adapter w/ Spring 4mm HIOS 4.5 - 26.0 0.51 - 2.9 Black
66617 Adapter w/ Spring 1/4” Hex 4.5 - 26.0 0.51 - 2.9 Black
66618 Adapter w/ Spring 1/4” Hex 25.0 - 80.0 2.8 - 9.03 Black
Included with DTT-200 (66705)
ASG Number Type Drive Size Torque Range Spring Color
lbf.in N.m
66618 Adapter w/ Spring 1/4” Hex 25.0 - 80.0 2.8 - 9.03 Black
66634 Adapter w/ Spring 3/8” Hex 50.0 - 200.0 5.6 - 22.6 Black
Included with DTT-500 (66706)
ASG Number Type Drive Size Torque Range Spring Color
lbf.in N.m
66618 Adapter w/ Spring 1/4” Hex 25.0 - 80.0 2.8 - 9.03 Black
66642 Adapter w/ Spring 3/8” Hex 50.0 - 500.0 5.6 - 56.5 Black
ASG #CP66614 and
#CP66616 are springs
to be used with
#CP66612 Adapter

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 7 of 23
Instructions for Using the Rundown Adapters
DTT Series Digital Torque Testers
User Manual
ASG Adapter Number Drive Size Torque Range
lbf.in N.m
66620 4 mm HIOS 0.25 - 2.0 0.03 - 0.23
66631 4 mm HIOS 1.3 - 5.0 0.15 - 0.56
66612 1/4" Hex 1.3 - 5.0, 4.5 - 26.0 0.15 - 0.56, 0.51 - 2.94
66617 1/4" Hex 4.5 - 26.0 0.51 - 2.94
66633 4 mm HIOS 4.5 - 26.0 0.51 - 2.94
66635 5 mm HIOS 4.5 - 26.0 0.51 - 2.94
66618 1/4" Hex 25.0 - 80.0 2.8 - 9.03
66639 5 mm Hex 25.0 - 80.0 2.8 - 9.03
66634 3/8" Hex 50.0 - 200.0 5.65 - 22.6
66642 3/8" Hex 50.0 - 500.0 5.65 - 56.5
Below is a list of all rundown adapters ASG offers for the DTT and DTT-L Series
• Rundown adapters are included with the DTT torque testers. Please see the charts on page 6 for the included adapters.
• Rundown adapters are not included with the DTT-L torque tester, choose the needed rundown adapters from the chart below
when ordering the tester.
* ASG Adapter #66612 comes with two springs covering torque range from 1.3-5.0 lbf.in (0.15-0.56) and 4.5-26.0 lbf.in (0.51-2.94 N.m)
* * *
ASG #66620 ASG #66631 ASG #66612 ASG #66633
ASG #66635 ASG #66618 ASG #66639 ASG #66634
ASG #66617
ASG #66642

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 8 of 23
Usage
Procedure for Manual Torque Drivers
• Attach a proper adapter to connect the driver to the 3/8” female square.
• NOTE: It is NOT recommended to use the power tool adapters to calibrate hand torque tools.
• Select FIRST PEAK mode using the MODE button. Press the RESET/ESC button if the display is not at 0.
• Turn the driver or wrench clockwise or counterclockwise until the “break” point is reached.
• In FIRST PEAK no other reading can then be displayed until the reset button is pressed.
• The LCD display will show the torque that the driver or wrench is set at. Note the reading or write it down. To enter it into the
DTT memory, press the MEM/ENTER button. If you are testing several tools note the memory numbers of the readings for the
various tools. This number is in the upper right hand corner of the LCD. The unit can save up to 500 readings. After that no
further readings can be entered until the memory is cleared. See the main menu section starting on page 11 for instructions on
downloading and clearing the readings in the memory.
• If the torque is not correct, adjust the tool. Repeat the test. When the reading is proper, perform the test several times to be sure
the readings are consistent. The proper reading may be an average of several readings.
Procedure for Power Screwdrivers
• Put the appropriate rundown adapter into the unit’s female 3/8” square socket. This is also called a joint simulator. It allows the
power tool to reach its working speed before tightening up and causing the clutch to operate.
• Run the tool in reverse to make sure the adapter is unwound. Do not disassemble the driver adapter.
• Make sure the unit is in PEAK mode. If it is not, use the MODE button to select PEAK mode.
• Press the RESET/ESC button to be sure the display is at 0.
• Run the tool forward until it shuts off. The reading on the display will be the torque setting of the tool’s clutch. Note the reading or
write it down.
• To enter it into the units memory, press the MEM/ENTER button. If you are testing several tools, note the memory numbers of the
readings for the various tools. This number is in the upper right hand corner of the LCD. The unit can save up to 500 readings.
After that no further readings can be entered until the memory is cleared. See the main menu section starting on page 11 for
instructions on downloading and clearing the readings in the memory.
• If the torque is not correct, adjust the clutch. Repeat the test. When the reading is proper, perform the test several times to be sure
the readings are consistent. The proper reading may be an average of several readings.
• Always unwind the rundown adapter after each test. Do NOT leave the spring under tension while not in use.
DTT Series Digital Torque Testers
User Manual

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 9 of 23
Battery Indicator
Basic Function
Display of Clockwise/Counter-Clockwise
Battery Basic Function
TRACK
0.003 kgf.cm
CW
005
TRACK
0.003 kgf.cm
CCW
005
F-PEAK
0.003 kgf.cm
CCW
005
PEAK
0.003 kgf.cm
CCW
005
Battery Level > 8.00V
7.99V > Battery Level > 7.60V
7.59V > Battery Level > 7.30V
7.29V > Battery Level > 7.00V
Battery Level < 7.00V
If the battery level is less than 6.9 V, the “Battery Empty” message will be displayed and the tester
will power down automatically. When battery charger is connected and battery is charging, the
plug icon will blink. Charging time is about 2.3 hours for full charge. NOTE: Only use the 12V
power adapter/charger supplied.
Clockwise torque is displayed on the DTT and recognized by the symbol “CW.” Counter-clockwise
torque is displayed on the DTT and recognized by the symbol “CCW.”
Clockwise symbol
Charging symbol
Load indicator bar
for clockwise
Load indicator
bar for counter-
clockwise
Counter-clockwise symbol
A load indicator bar alerts the operator as to how much torque load is being applied to the
transducer as related to its full scale rating.
DTT Series Digital Torque Testers
User Manual

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 10 of 23
The display automatically zeros when the unit is turned on. The zero key can be pressed to set the meter to zero.
You can choose from the following modes of measure: Track, First Peak and Peak Torque. To change the display mode press the
MODE key. Each time the MODE key is pressed, it will select the next available mode until the tester returns to its original setting.
You can choose from the following units of measure depending on the capacity of your tester: N.mm (not available on the DTT-500),
N.cm, N.m, kgf.cm, kgf.m, lbf.in, and lbf.ft. To change the display units, press the UNIT key. Each time the UNIT key is pressed, it will
select the next available unit until the tester returns to its original setting. The DTT automatically converts readings as new units of
measure are selected. NOTE: All units may not be displayed depending on tester capacity.
Basic Function (Continued)
Zeroing the Tester
Changing the Mode of Measure
Changing the Unit of Measure
Press the MODE key until “Track” appears on
the display. The display will now indicate the
torque applied in either direction as it is applied
to the transducer.
Press the MODE key until “Peak” appears on
the display. The display will show the maximum
torque applied to the transducer during a cycle.
Press the MODE key until “F-Peak” appears on
the display. The display will show the torque
level applied to the transducer. Applying more
torque will not change the reading on the
display. Example: F-Peak records the rst click
and ignores additional clicks.
Track Mode
Peak Mode
First Peak Mode
FIRST PEAK
DTT Series Digital Torque Testers
User Manual

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 11 of 23
Any reading can be saved at anytime by pressing the MEM key. A total of 500 readings can be stored in the database including the
torque reading, unit, direction, date, and time. When reading 501 is attempted, an alarm will beep twice. The DTT will not accept
any new values. The records need to be deleted manually to create space in order to record any new values.
The DTT meter can output data using the supplied mini USB cable (or the optional RS-232 cable, not supplied) and the TorqueLink 2.4
software. TorqueLink 2.4 can be downloaded at www.asg-jergens.com on the Literature and Downloads page.
Basic Function (Continued)
Saved Reading to Memory
Output Signal
Command Action
Pressing
PRINT key
If current mode is track mode, send live reading value with unit
If current mode is rst peak torque mode, send rst peak torque value with unit
If current mode is peak torque mode, send peak torque value with unit
Upload All Send all memory
When you press any key or apply torque to the transducer greater than 0.5% of full scale, the backlight will go on for 60 seconds.
Backlit Display
DTT Series Digital Torque Testers
User Manual
Main Menu
• Press and hold the MENU key to access the main menu
• To move between the options listed on the main menu page, press the UP or
DOWN arrow keys to move the cursor
• Press ENTER to select the sub-menus, activate feature and enter values. Within
sub-menus, UP and DOWN arrow keys will also change numerical values
• Sub-menus include setting, memory, calibration, transducer state, date and
time, and about
• Press MENU/ESC to return to the main menu page

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 12 of 23
• Press and hold the MENU key to access the main menu. Use the UP or DOWN
arrow keys to move the cursor to SETTING and press ENTER.
• To move between the options listed on the setting page, use the UP or DOWN
arrow keys to move the cursor. The options on the setting page are auto reset,
pass-fail, auto memory, auto print and factory rest.
• Press ENTER to select the sub-menus, activate feature and enter values. Within
sub-menus the UP and DOWN arrow keys will also change numerical values.
• Press MENU/ESC to return to the setting menu page
The Auto Reset feature is used to automatically reset the reading value in peak
mode. This mode works when the reading value is higher than the setting value,
causing the peak value to change to a new held value. The user does not need
to press the RESET key, since the peak value will automatically reset. This feature
works in peak mode only. If this feature is activated, the icon “AR” will display on
main display.
• To access the Auto Reset menu, press and hold the MENU key to access the main
menu. Use the UP or DOWN arrow keys to move the cursor to SETTING and press
ENTER. Use the UP or DOWN arrow keys to move the cursor to AUTO RESET and
press ENTER.
• The display will show the set auto-reset menu page. Press the MESU/ESC key
to return the settings page.
• Use UP and DOWN keys to change the value. Press and hold UP key to
change units. Press SAVE to set new value and return to setting menu. Press
and hold SAVE key to reset value to zero to set a new value. Press ESC to
return to setting menu.
• Select a reset value 10% of the expected torque reading on the tool being
tested to avoid erroneous readings
• Auto Reset feature will automatically be disabled if you set AUTO RESET = 0
or Auto Memory feature is on
• Press and hold the RESET/SAVE key to reset to zero
• Function of button will change depending on the menu
Setting
Auto Reset
Main Menu (Continued)
DTT Series Digital Torque Testers
User Manual

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 13 of 23
Main Menu (Continued)
The Pass-Fail feature is used to set a dened acceptable maximum and minimum
torque range for measuring. It is activated by setting the lower level and upper
level torque limit. As long as the torque value is within this range, the display will
show a PASS message. Any reading values outside this range (higher or lower), the
display will show a FAIL message. If this feature is activated, the PF symbol will
display on the main display.
• To access the Pass-Fail menu, press and hold the MENU key to access the main
menu. Use the UP or DOWN arrow keys to move the cursor to SETTING and
press ENTER. Use the UP or DOWN arrow keys to move the cursor to PASS-FAIL
and press ENTER.
• Press and hold the DOWN arrow key to toggle between the upper and lower
limits. Press and hold the UP arrow key to change units. Press the UP and
DOWN keys to change the numerics. Press and hold the SAVE key to reset
value to 0 to select a new value. Press ESC key to return to the setting menu.
• To reset, press and hold the RESET button
NOTE: Pass-Fail feature will be automatically disabled if you set LOWER LEVEL and
UPPER LEVEL = 0.000
NOTE: Lower level must be less than the upper level
Pass-Fail
DTT Series Digital Torque Testers
User Manual
Example:
Lower Level
Upper Level
= 0 N.m
= 20 N.m
Upper Torque
Level
Lower Torque
Level
Upper Torque
Level
Lower Torque
Level
LOAD LOAD LOAD
TIME TIME TIME
Example:
Lower Level
Upper Level
= 20 N.m
= 0 N.m
Example:
Lower Level
Upper Level
= 10 N.m
= 20 N.m

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 14 of 23
Main Menu (Continued)
Auto Memory
The Auto Memory feature is used to automatically store the reading in the memory
and reset the display to 0. The user does not need to press RESET key or MEM
key. When you get the peak value, the reading will automatically be stored in
the memory included the torque reading, unit, date, and time. If you activate this
feature, the AM symbol will be displayed on the main display.
• To access the Auto Memory menu, press and hold the MENU key to access the
main menu. Use the UP or DOWN arrow keys to move the cursor to SETTING
and press ENTER. Use the UP or DOWN arrow keys to move the cursor to AUTO
MEMORY and press ENTER.
• If OFF is selected in this mode, the auto memory feature will be turned off
• If CLOCKWISE is selected in this mode, tester will only record the clockwise
torque reading. If counter-clockwise torque is applied, the tester will reset the
reading and the tester will be ready to take new reading.
• If COUNTER-CLOCKWISE is selected in this mode, the tester will only record
counter-clockwise torque reading. If clockwise torque is applied, the tester
will reset the reading and the tester will be ready to take new reading.
DTT Series Digital Torque Testers
User Manual
Auto Print
The Auto Print feature is used to automatically send measured value to ports
connection. The user does not need to press the PRINT key. When you receive the
measured value will automatically send to ports connection. If you activate this
feature, the AP symbol will display on the main display.
• To access the Auto Print menu, press and hold the MENU key to access the main
menu. Use the UP or DOWN arrow keys to move the cursor to SETTING and
press ENTER. Use the UP or DOWN arrow keys to move the cursor to AUTO
PRINT and press ENTER.
• If OFF is selected in this mode, turns auto print feature off
• If CLOCKWISE is selected in this mode, the auto print feature will work in
specic clockwise direction only
• If COUNTER-CLOCKWISE is selected in this mode, the auto print feature will
work in specic counter-clockwise direction only

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 15 of 23
Main Menu (Continued)
Memory
Factory Reset
This is used to view the saved record, delete the last record, delete all records and
calculate the statistics value of all saved memory records.
• To access the Memory menu, press and hold the MENU/ESC key to access the
main menu. Use the UP or DOWN arrow keys to move the cursor to MEMORY
menu and press ENTER.
• The display will show the MEMORY menu page. The options on the memory
page are view, delete last, delete all, and statistics.
• Press the MENU/ESC key to return to the main menu page
Factory Reset will turn off AUTO RESET, PASS-FAIL, and AUTO MEMORY at the same
time without having to go into each option individually.
• To access the Factory Reset menu, press and hold the MENU key to access the
main menu. Use the UP or DOWN arrow keys to move the cursor to SETTING
and press ENTER. Use the UP or DOWN arrow keys to move the cursor to
FACTORY RESET and press ENTER.
• Press ENTER to activate factory reset or press MENU/ESC to return to the
SETTING menu
View
This is used to view all saved records in memory. The detail of each saved record
consists of mode, reading value with unit, direction, date, and time.
• To access the View menu, press and hold the MENU/ESC key to access the
main menu. Use the UP or DOWN arrow keys to move the cursor to MEMORY
and press ENTER. Use the UP or DOWN arrow keys to move the cursor to
VIEW and press ENTER. The display will show the view menu page.
• Press the MENU/ESC key to return to the memory menu page
DTT Series Digital Torque Testers
User Manual
ESC
0.0000
0.0000
0.0000
0.0000
000.00
DATE AND TIME
ESC ENTERDEC- INC+
PREV NEXT
mm/dd/yy: 00 00 00
hh:mm:ss: 00 00 00
FACTORY RESET
ESC ENTER
AUTO RESET OFF:
:
:
:
OFF
OFF
OFF
PASS-FAIL
AUTO MEMORY
AUTO PRINT

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 16 of 23
Main Menu (Continued)
Delete All
Upload All
This is used to delete all saved records.
• To access the Delete All menu, press and hold the MENU/ESC key to access the
main menu. Use the UP or DOWN arrow keys to move the cursor to MEMORY
and press ENTER. Use the UP or DOWN arrow keys to move the cursor to
DELETE ALL and press ENTER.
• Press ESC key to return to the memory menu page
• Press the UP or DOWN arrow keys to select NO or YES. If you select NO and
press the ENTER key, the meter will return to the memory menu page. If you
select YES and press the ENTER key, the meter will delete all saved records and
return to the memory menu page.
This is used to transfer memory to a computer.
• To transfer memory to a computer, press and hold the MENU/ESC key to
access the main menu. Use the UP or DOWN arrow keys to move the cursor
to MEMORY and press ENTER. Use the UP or DOWN arrow keys to move the
cursor to UPLOAD ALL and press ENTER.
• The display will show the transfer memory menu page
• Press the ESC key to return to the memory menu page
DTT Series Digital Torque Testers
User Manual
Delete Last
This is used to delete the last saved record.
• To access the Delete Last menu, press and hold the MENU key to access the
main menu. Use the UP or DOWN arrow keys to move the cursor to MEMORY
and press ENTER. Use the UP or DOWN arrow keys to move the cursor to
DELETE LAST and press ENTER.
• The display will show DELETE LAST menu page
• Press the ESC key to return to the memory menu page
• Press the UP and DOWN arrow keys to select NO or YES. If you select NO and
press the ENTER key, the monitor will return to the memory menu page. If you
select YES and press the ENTER key, the meter will delete last saved record and
return to the memory menu page.

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 17 of 23
Main Menu (Continued)
Transducer State
Statistics
This is used to check the status of the transducer. If you suspect that your transducer
has sustained an overload, check the status of the transducer immediately.
• To access the Transducer State menu, press and hold the MENU/ESC key to
access the main menu. Use the UP or DOWN arrow keys to move the cursor to
TRANSDUCER STATE and press ENTER.
• Place the tester horizontally on a at level surface and go to the main menu page
• Use the UP and DOWN arrow keys to move the cursor point to TRANSDUCER
STATE and press the ENTER key
• The display will show TRANSDUCER STATE menu page. Press the ESC key to
return to the main menu page.
• If the % offset is between 5%-10%, please contact ASG to arrange a
recalibration of your tester
• If the % offset is greater than 10%, please contact ASG to arrange for repair or
replacement
• These values are given only as an indicator – the need for repair may vary
according to the individual characteristics of the transducer
The statistics are calculated for all of the readings currently in the memory and
include: maximum, minimum, mean, std, cov., based on all units in memory.
• To access the Statistics menu, press and hold the MENU key to access the
main menu. Use the UP or DOWN arrow keys to move the cursor to MEMORY
and press ENTER. Use the UP or DOWN arrow keys to move the cursor to
STATISTICS and press ENTER.
• Press the ESC key to return to the memory menu page
DTT Series Digital Torque Testers
User Manual
Calibration
ASG service technicians to calibrate the tester; please contact ASG for additional
calibration information.
ESC
0.0000
0.0000
0.0000
0.0000
000.00
DATE AND TIME
ESC ENTERDEC- INC+
PREV NEXT
mm/dd/yy: 00 00 00
hh:mm:ss: 00 00 00
FACTORY RESET
ESC ENTER
AUTO RESET OFF:
:
:
:
OFF
OFF
OFF
PASS-FAIL
AUTO MEMORY
AUTO PRINT

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 18 of 23
Main Menu (Continued)
Date and Time
About
This allows you to set the current date and time.
• To access the Date and Time menu, press and hold the MENU key to access the
main menu. Use the UP or DOWN arrow keys to move the cursor to DATE AND
TIME and press ENTER.
• Use the DEC- and INC+ arrow keys to change the numerical values. Press and hold
these keys (PREV and NEXT) to toggle between the category.
• Press the ENTER key to save information
• Press the ESC key to return to the main menu page
This shows the information regarding your tester (rmware revision, model, capacity,
serial number).
• To access the About menu, press and hold the MENU key to access the main menu.
Use the UP or DOWN arrow keys to move the cursor to ABOUT and press ENTER.
• Press the ESC key to return to the main menu page
DTT Series Digital Torque Testers
User Manual
ESC
0.0000
0.0000
0.0000
0.0000
000.00
DATE AND TIME
ESC ENTERDEC- INC+
PREV NEXT
mm/dd/yy: 00 00 00
hh:mm:ss: 00 00 00
FACTORY RESET
ESC ENTER
AUTO RESET OFF:
:
:
:
OFF
OFF
OFF
PASS-FAIL
AUTO MEMORY
AUTO PRINT

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 19 of 23
Software
Torque Link 2.4
Menu Bar
DTT Series Digital Torque Testers
User Manual
This program is designed to enhance the functionality of DTT. The program will retrieve memory from the DTT meter and analyze it
as well as display and analyze readings in real time.
• New: Open a new data sheet
• Open: Open saved “*.tl” data les to the table
• Save: For data storage in the form of torque le “* .tl”
• Print: Command to print directly to the printer
• Export to Excel: Save and export the data to Excel spreadsheet
• Click To Connect: Connect to DTT meter
• Download: For downloading the data in the memory of DTT
• Delete Last: To delete data at the end of the table
Menu Bar:
Displays all the
program’s menu
Table: Shows
the information
retrieved from
DTT memory
and reading
taken in real
time
Real Time: The area
that shows the data
retrieval in real time
Result: Shows the
analyzed result
Pass/Fail: The area
that shows the result
of the set conditions
of the test tools where
the upper and lower
limits are entered and
enabled

ASG, Division of Jergens, Inc.
15700 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 20 of 23
Software (Continued)
Table
DTT Series Digital Torque Testers
User Manual
• No.: Shows the sequence of data
• Mode: Indicates the mode of operation when the reading was taken
• Direction: Indicates the direction of the force, CW represents the clockwise rotation, CCW represents the counter-clockwise rotation
• Torque: Indicates the torque value taken
• Units: Represents the unit of measurement
• Data & Time: The date and time the reading was taken
* Edit the table by double-clicking the cell
and then changing it
This manual suits for next models
6
Table of contents
Other ASG Test Equipment manuals