Beka Hospitec AVERO Premium Plus User manual

BEKA Hospitec Ltd.
Edition V2.4 E
Subject to technical modifications
2018.02.14
User Manual
for
AVERO Premium Plus Lifting Tub

BEKA Hospitec Ltd.
AVERO Premium Plus User Manual V2.4 E - Page Page 2of 58

BEKA Hospitec Ltd.
AVERO Premium Plus User Manual V2.4 E - Page Page 3of 58
Dear Customer,
We thank you for the confidence you have placed in us by purchasing this BEKA Hospitec
Ltd. product. Our products are manufactured and tested under strict quality controls.
Proper use of the unit is essential for safe operation. Therefore, thoroughly familiarize
yourself with the contents of this user manual before operating the unit.
Please note the safety instructions in particular!
Slight deviations of the actual delivered units from the pictures shown here are possible. If
you have questions or want more information about your device, please call us or fax us.
Our staff will be happy to help.
Beka Hospitec Ltd. has been certified by TÜV SÜD Product Service Ltd. in
accordance with DIN EN ISO 9001 and DIN EN ISO 13485.
With this, we set a high standard of quality in development, manufacturing, quality
assurance and service for our entire product range.

Contents
AVERO Premium Plus User Manual V2.4 E - Page Page 4of 58
1. Foreword ........................................................................................................................6
2. Signs and symbols used...............................................................................................7
Type label...............................................................................................................................7
3. Unpacking and assembly..............................................................................................8
Tools Required.......................................................................................................................8
Remove the cardboard...........................................................................................................8
Loosen from the pallet............................................................................................................9
Installation on site...................................................................................................................9
4. Installation....................................................................................................................10
4.1 Note before initial operation........................................................................................11
4.2 Electrical system ........................................................................................................11
4.3 AVERO Premium Plus 175, Floor installation.............................................................12
4.4 AVERO Premium Plus 175, Wall installation ..............................................................13
4.5 Adjusting the thermostat mixer (scald protection) .......................................................15
4.6 Thermal rinse and subsequent adjustment of the outlet temperature..........................17
5. Operating the AVERO Premium Plus Lifting Tub ......................................................20
Application areas..................................................................................................................20
Duration of use.....................................................................................................................20
Typical indications................................................................................................................20
Contraindications..................................................................................................................20
6. Safety Information .......................................................................................................21
7. AVERO Premium Plus Lifting Tub Operating Elements............................................24
7.1 Operating elements....................................................................................................24
7.2 Tub interior.................................................................................................................24
7.3 10 button control panel...............................................................................................26
8. Operating the AVERO Premium Plus Lifting Tub ......................................................27
9. Using the AVERO Premium Plus Lifting Tub.............................................................27
10. Advanced Lifting Tub Functions ................................................................................28
10.1 Button disinfection (1) .............................................................................................28
10.2 Display (2) / Display (3) / Display (4).......................................................................29
10.3 Handheld showerhead button (5)............................................................................29
10.4 Fill level 1 button (6) + fill level 2 button (7) (option)................................................29
10.5 Air spa system button (8)........................................................................................30
10.6 Color light therapy button........................................................................................31
10.7 Power / Function LED Display (14) .........................................................................31
10.8 Music - Sound system.............................................................................................32
10.9 Emergency drain function during power failure (option) ..........................................33

AVERO Premium Plus User Manual V2.4 E - Page Page 5of 58
10.10 Aroma-therapy-system (option)...............................................................................35
11. Cleaning / Disinfection ................................................................................................37
Cleaning the AVERO Premium Plus Lifting Tub ...................................................................37
Disinfecting the AVERO Premium Plus Lifting Tub...............................................................37
Sterilization...........................................................................................................................39
Contact for technical information ..........................................................................................39
12. Waste disposal.............................................................................................................39
Disposing of packaging waste..............................................................................................39
Disposal of product...............................................................................................................39
13. Applicable test procedures & controls.......................................................................40
14. Troubleshooting and customer service .....................................................................41
Help Troubleshooting ...........................................................................................................41
Customer service .................................................................................................................42
15. Appendix ......................................................................................................................43
15.1 Technical data ........................................................................................................43
15.2 Dimensions AVERO Premium Plus 190..................................................................45
15.3 Dimensions AVERO Premium Plus 170..................................................................46
15.4 Installation plan AVERO Premium ..........................................................................47
15.5 Declaration of Conformity........................................................................................48
15.6 AVERO Premium Plus Optional extras / Accessories .............................................49
15.7 Spare Parts / Consumables ....................................................................................49
15.8 Electromagnetic compatibility..................................................................................50
15.9 Inventory.................................................................................................................54

Foreword
AVERO Premium Plus User Manual V2.4 E - Page Page 6of 58
1. Foreword
Dear Customer,
Please note the following points:
1. Please read the instruction manual carefully.
2. Use the device only for its intended use.
3. Only trained personnel are permitted to operate the device.
4. Electrical installations may only be carried out by specialist companies, according to the
currently valid VDE regulations.
5. Maintenance and safety checks must be carried out by BEKA Hospitec or by specialist
companies authorized by BEKA Hospitec.
6. Assembly, extensions, modifications or repairs must be performed by BEKA Hospitec or by
specialist companies authorized by BEKA Hospitec.
In case of technical interventions, such as additions or modifications to our equipment, which
are not carried out by the BEKA Hospitec or by BEKA Hospitec authorized specialist
companies, the warranty is voided for the changes as well as for the device or for device
functions which are related to the change.
Liability and warranty
The manufacturer of the unit is only responsible for the safety and reliability of the device, if:
a. the installation is carried out by authorized persons
b. the space in question complies with the current VDE regulations
and it was observed that:
c. it will be used only for the purposes intended
d. regular function tests are performed
Manufacturer Address:
BEKA Hospitec Ltd.
Am Rübenmorgen 3 ● 35582 Wetzlar
Phone: +49(0)641-9 22 20 - 0
Fax: +49(0)641-9 22 20 - 20
info@beka-hospitec.de
www.beka-hospitec.de

Signs and symbols used
AVERO Premium Plus User Manual V2.4 E - Page Page 7of 58
2. Signs and symbols used
Consult accompanying documents / user manual
Warning of potential danger
Upgrades prohibited,
Feet should not be placed on the base frame.
Application Part "Type B" according to DIN EN60601-1
Not household waste
Do not put the treatment system and the packaging materials
in the household trash.
CE marking according to the EC Directive concerning
medical devices
can only be used indoors
Protection class I
IPX4
Protection against penetration of liquids: Protection against
splashing water on all sides
Type label
Picture 4
This figure shows the type label.
The type label is located on the side panel.
The serial number shown here (SN) 0001.05.15 is only
an example. For further inquiries, please always
indicate the serial number located on your unit’s type
label.
Note: For legal reasons, the item number and serial
number may be required to be computer-readable and
are therefore additionally applied as a bar-code on the
type label.

Unpacking and assembly
AVERO Premium Plus User Manual V2.4 E - Page Page 8of 58
3. Unpacking and assembly
Transport the lifting tub into the vicinity of the intended installation site.
Use a pallet or similar for transporting.
Arrange the tub, so that it is secure and cannot tip.
Tools Required
To remove the packaging you need:
Sharp-
bladed cutter
Remove the cardboard
Cut the straps with the cutter and remove
them.
Note: Do not cut into the box, as the
surface of the tub may be damaged!
Lift the box up and off and set the box aside.

Unpacking and assembly
AVERO Premium Plus User Manual V2.4 E - Page Page 9of 58
Loosen from the pallet
AVERO Premium Plus
The tub is mounted on the pallet through
the foot by means of a screw.
Unscrew this screw using a cordless
screwdriver and a matching bit, and put
them aside with the washer. Repeat this
on all 4 feet.
After all fasteners are released, you can remove the lifting tub from the pallet and position it in
the intended place of use.
Remove the bubble wrap and stretch film. Make sure that you do not damage the surface of the
tub through the use of tools.
Installation on site
Tools Required:
Spirit level
Adjustable wrench SW 10
With the spirit level and the wrench (SW 10) align the tub so that it is perpendicular. The bath
feet can be adjusted in height by screwing or unscrewing. Make sure that all feet touch the
ground and the tub has a firm footing.

Auspacken und Montieren
4. Installation
Dear Customer,
Please check your electrical installation according to the applicable VDE
regulations 0100 and 0750, or check with the applicable regulations in your
country before operating our equipment.
As the operator of this system, you need to have the functionality of the fault
current circuit breaker checked regularly. Please ask your electrician.
These checks are necessary for the safety of your patients and for your own
safety.
This device is not approved for operation in potentially explosive atmospheres.
Maintenance of the unit must be carried out only by us or by a party expressly
authorized by us, otherwise the warranty shall lapse.
The connection of the AVERO Premium Plus must be done in accordance with the
technical drawings of BEKA Hospitec and the guidelines for the establishment of
wet areas and the current standards in force in your country.
All inlet and outlet hoses and electrical cables must be arranged so that there is no
risk of tripping.
Before installing the tub, the power cables must be adequately rinsed.
Before commissioning the thermostatic mixer must be set by a suitable
person.

Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E Seite 11 von 58
4.1 Note before initial operation
WARNING:
The device may only be used in accordance with the
accompanying documents.
Only under these conditions is the manufacturer considered
responsible for the safety effects, reliability and performance of
the device.
If the AVERO Premium Plus lifting tub is newly connected, the
technical specifications must be observed.
WARNING:
To prevent scalding, the hot / cold water mixer tap must be set to
the on-site water supply parameters prior to first use.
Note:
In order to ensure safe functioning and safe operation, we
recommend regular maintenance and adjustment of the
thermostat mixer. Setting up the AVERO Premium Plus tub in
another space or changes to the local water supply may require
readjustment of the thermostatic mixer.
4.2 Electrical system
Connection to the power grid:
When connecting the device to an electric power grid / AC Power supply / AC outlet
the particular national specifications must be observed (e.g. in Germany DIN VDE
0100-710).
The connection of the AVERO Premium Plus via a junction box with legally required
splash protection (IP class), 230V AC / 50Hz, 3x1.5 mm², 1.5 KW, 16 amps, earth
leakage circuit breaker 30 mA, potential equalization (in Germany) according to DIN
VDE 0107.
Please note that the connection must be performed by a qualified electrician.
Protective conductor:
When using protection class I devices, the quality of the protective conductor of the
installation is important. The specified stipulations that have been specified in many
countries by national authorities are to be observed.
WARNING:
To avoid the risk of electric shock, this unit must be connected to
a supply network with a ground wire.

Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E Seite 12 von 58
4.3 AVERO Premium Plus 175, Floor installation
In order to avoid tilting of the tub, the tub has to be additionally secured.
Note they refer to the following pictures:
Illustration 1:
Illustration 2:
Illustration 3:
Illustration 4:
Illustration 5:
Illustration 6:
Illustration 7:
Illustration 8:
Illustration 9:
Illustration 10:
Illustration 11:
Illustration 12:
Illustration 13:
Illustration 14:
Components 15:
Tools 16:

Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E Seite 13 von 58
4.4 AVERO Premium Plus 175, Wall installation
Illustration 1:
Illustration 2:
Illustration 3:
Illustration 4:
Illustration 5:
Illustration 6:
Illustration 7:
Illustration 8:
Illustration 9:
Illustration 10:
Illustration 11:
Illustration 12:
Illustration 13:
Illustration 14:
Illustration 15:
Illustration 16:

Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E Seite 14 von 58
Illustration 17:
Components
18:
Tools18:

Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E Seite 15 von 58
4.5 Adjusting the thermostat mixer (scald protection)
Figure 1:
Tools you need:
1 x screwdriver flat,
0,6 x 3,5 mm,
1 Screwdriver Cross-
hatch,
1 x thermometer
Figure 2:
Turn the knob to the
left until it stops (Snap
button not pressed)
Figure 3:
Use the screwdriver to
remove the cap
Figure 4:
Unscrew the cross
screw from the rotary
actuator and pull it
forwards.
Figure 5:
Open a solenoid valve
to check the water
temperature
(In the pictures shown
handbrake).
Figure 6:
Check the outlet
temperature with the
external thermometer!
Figure 7:
Adjust the water to the
desired maximum
discharge
temperature.
Figure 8:
The picture shows the
desired maximum
temperature, which
must agree with the
manually measured
temperature.
Figure 9:
Slide the knob to the
maximum position.
Figure 10:
Switch off the solenoid
valve.
Figure 11:
Install the rotary knob
with the cross bolt
Figure 12:
Reinsert the cover.
!!!!
!

Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E Seite 16 von 58
Figure 13:
Switch on a solenoid
valve (hand shower or
water spout) to check
the setting again.
Figure 14:
Check the outlet
temperature with the
external thermometer!
The process is now
complete.
If the temperature is set at 42 ° C (anti-scalding), the corresponding display
"HOT" alternates with the temperature and the water flow is switched off.
The handshower can then be switched on again with the following key
combination to continue with the setting:
Hold down key 5 and then press 8 and 10 together.
Should the setting go over the tank filling, please press key 6 or 7 (depending
on the selected filling height), then additionally press 8 and 10 together.
41,5

Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E Seite 17 von 58
4.6 Thermal rinse and subsequent adjustment of the outlet temperature
Note: Follow the steps carefully. If not, the disinfecting device may be blocked.
You can only remove this blocking by disconnecting from the mains voltage!
Figure 1:
Tools you need:
1 x screwdriver flat,
0,6 x 3,5 mm,
1 Screwdriver Cross-
hatch,
1 x thermometer
Figure 2:
Turn the knob to the
left until it stops (Snap
button not pressed)
Figure 3:
Use the screwdriver to
remove the cap
Figure 4:
Unscrew the cross
screw from the rotary
actuator and pull it
forwards.
Figure 5:
Open the solenoid
valve of the hand
shower
Figure 6:
Turn the knurled knob
to the left to increase
the temperature
Figure 7:
If the temperature
exceeds 42 ° C, the
control switches off for
safety reasons. The
display shows "hot"
alternating with the
temperature.
Figure 8:
Press and hold
button 5, than press
the buttons 8 and 10
together
This bridges the
protection circuit and
allows you to set a
higher temperature.

Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E Seite 18 von 58
Figure 9:
Adjust the required
temperature for
thermal rinsing by
turning the knurled
knob. Check the
temperature with the
thermometer
knurled knob
Perform the thermal rinse of the hand shower according to the valid laws of your country.
Observe the temperature setting and the time.
Do not turn off the water flow after completion of the thermal rinse.
Continue with figure 10 of this manual!
Figure 10:
Turn off the solenoid
valve of the hand
shower.
Press button 5,
1 time!
Figure 11:
Activate the thermal
flushing for the water
inlet.
Press and hold
button 6, than press
the buttons 8 and 10
together
Perform the thermal rinse of the water spout according to the valid laws of your country.
Observe the temperature setting and the time.
Do not turn off the water flow after completion of the thermal rinse.
Continue with figure 12 of this manual!
Figure 12:
Adjust the water
temperature below 42
° C. Observe the
display on the control
panel
Figure 13:
The picture shows the
temperature below 42
° C
Figure 14:
Switch off the
magnetic valve for the
water inlet.
Press button 6
2 times!
Press the button for the first time to switch off
the "overtemperature" mode.
With the second actuation switch off the
solenoid valve for the water inlet.

Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E Seite 19 von 58
Switch on the water supply via hand shower or via the water inlet
Figure 15:
Adjust the water to the
desired maximum
discharge
temperature. Measure
the temperature with
the thermometer
Figure 16:
The picture shows the
desired maximum
temperature, which
must agree with the
manually measured
temperature.
Figure 17:
Slide the knob to the
maximum position.
Figure 18:
Switch off the solenoid
valve.
Figure 19:
Install the rotary knob
with the cross bolt
Figure 20:
Reinsert the cover.
Figure 21:
Switch on a solenoid
valve (hand shower or
water spout) to check
the setting again.
Figure 22:
Check the outlet
temperature with the
external thermometer!
The process is now
complete.
41,5

Operating
User Manual AVERO Premium Plus V2.4 E Seite 20 von 58
5. Operating the AVERO Premium Plus Lifting Tub
Application areas
The AVERO Premium Plus lifting tub is used for bathing and care of dependent persons.
Duration of use
The duration of the AVERO Premium Plus lifting tub application unit is set at a maximum of 30
minutes.
Note
Side effects from usages are not known.
Note the contraindications in Section 5.4
Typical indications
For bathing and care of dependent patients in hospitals, nursing homes, facilities for the
disabled and private households.
Contraindications
WARNING:
Patients with:
Heart disease / arrhythmias
Tumors in the application area
Leukemia
Bacterial infection
Rheumatism in the advanced stages
may only be treated with the lifting tub after consultation
with their attending physician.
WARNING:
Patients with active implants such as infusion or insulin pumps
as well as patients with febrile conditions or inflammatory skin
diseases may not be treated with the AVERO Premium Plus
lifting tub. There should be a warning that is highly visible in the
entrance area of the treatment room.
Please also observe the safety instructions in the following section.
Other manuals for AVERO Premium Plus
2
Table of contents
Other Beka Hospitec Medical Equipment manuals

Beka Hospitec
Beka Hospitec NORA Pro Operation instructions

Beka Hospitec
Beka Hospitec AVERO Phoenix User manual

Beka Hospitec
Beka Hospitec NORA Alu User manual

Beka Hospitec
Beka Hospitec EVE! User manual

Beka Hospitec
Beka Hospitec NORA Alu User manual

Beka Hospitec
Beka Hospitec SENTA PUR L Installation instructions

Beka Hospitec
Beka Hospitec CARLO Alu Classic 230 User manual

Beka Hospitec
Beka Hospitec AVERO Premium tub 190 User manual