7
IR425-D4_D00039_06_M_DEEN/07.2017
ISOMETER® IR425-D4
Supply voltage
IR425-D4-1, IR425-D4W-1:
Supply voltage Us........................................................................... AC 16…72 V / DC 9.6…94 V
Frequency range Us............................................................................................. 15…460 Hz / DC
IR425-D4-2, IR425-D4W-2:
Supply voltage Us.............................................................................................. AC/DC 70…300 V
Frequency range Us.............................................................................................. 15…460 Hz, DC
Power consumption...............................................................................................................≤4VA
IT System being monitored
Nominal system voltage Un.............................................................................. AC / DC 0… 300 V
Nominal frequency fn.................................................................................................. 15…460 Hz
Response values
Response value Ran1 (ALARM 1)................................................................................... 1…200 kΩ
Response value Ran1 (ALARM 2)................................................................................... 1…200 kΩ
Preset function:
Un≤72 V: Ran1 (ALARM 1)/ Ran2 (ALARM 2) ........................................................... 20 kΩ / 10 kΩ
Un>72 V: Ran1 (ALARM 1)/ Ran2 (ALARM 2) ........................................................... 46 kΩ / 23 kΩ
Operating error (1…5 kΩ) / (5…200 kΩ) ..........................................±0.5 kΩ
/
±15 %
Hysteresis (1…5 kΩ) / (5…200 kΩ) .................................................................. +1 kΩ
/
+25 %
Time response
Response time tan at RF= 0,5 x Ran and Ce= 1 μF ................................................................ ≤ 2 s
Starting delay t ................................................................................................................... 0…10 s
Response delay ton ..............................................................................................................0…99 s
Measuring circuit
Measuring voltage Um............................................................................................................ ±12 V
Measuring current Im(RF= 0Ω)......................................................................................≤ 200 μA
Internal d.c. resistance Ri...................................................................................................... ≥62 kΩ
Internal impedance Zi(50 Hz) ............................................................................................. ≥60 kΩ
Admissible extraneous d.c. voltage Ufg ......................................................................... ≤DC 300 V
System leakage capacitance Ce.............................................................................................≤ 20 μF
Displays, memory
Display ..................................................................... LC display, multi-functional, non-illuminated
Display range, measuring value ................................................................................ 1 kΩ…1 MΩ
Operating error (1…5 kΩ) ................................................................................................ ±0.5 kΩ
Percentage operating error (5 kΩ…1 MΩ) ........................................................................ ±15 %
Password ..................................................................................................................... off / 0…999
Fault memory (alarm relay) ................................................................................................. on / off
Inputs
Cable length external test / reset button.............................................................................. ≤ 10 m
Switching elements
Number of....................................................................................... 2 (changeover contacts K1, K2)
Operating principle........................................................................ (N/O operation)(N/C operation)
Electrical endurance ............................................................................ 10000 switching operations
Contact data according IEC 60947-5-1
Rated operational voltage AC ............................................................................... 230 V ....... 230 V
Utilization category AC ........................................................................................ AC 13 ....... AC 14
Rated operational current AC .................................................................................. 5 A ........... 3 A
Rated operational voltage DC ............................................................ 220 V....... 110 V ......... 24 V
Utilization category DC ...................................................................... DC 12........DC 12 ....... DC 12
Rated operational current DC ............................................................ 0.1 A........ 0.2 A ........... 1 A
Minimum current ..........................................................................................1 mA at AC/DC ≥ 10 V
Environment/EMC
EMC ....................................................................................................................... acc. to IEC 61326
Operating temperature ........................................................................................ -25 °C…+55 °C
Climatic categories acc. to IEC 60721:
Stationary use (IEC 60721-3-3) (except condensation and formation of ice) ......................... 3K5
Transport (IEC 60721-3-2) (except condensation and formation of ice) ................................. 2K3
Storage (IEC 60721-3-1) (except condensation and formation of ice) .................................... 1K4
Classification of mechanical conditions acc. to IEC 60721:
Stationary use (IEC 60721-3-3) .................................................................................................3M4
Transport (IEC 60721-3-2) .........................................................................................................2M2
Storage (IEC 60721-3-1) ............................................................................................................1M3
Versorgungsspannung
IR425-D4-1, IR425-D4W-1:
Versorgungsspannung Us.............................................................. AC 16…72 V / DC 9,6…94 V
Frequenzbereich Us............................................................................................. 15…460 Hz / DC
IR425-D4-2, IR425-D4W-2:
Versorgungsspannung Us.................................................................................. AC/DC 70…300 V
Frequenzbereich Us............................................................................................... 15…460 Hz, DC
Eigenverbrauch ..................................................................................................................... ≤ 4 VA
Überwachtes IT-System
Netznennspannung Un..................................................................................... AC / DC 0… 300 V
Nennfrequenz fn.......................................................................................................... 15…460 Hz
Ansprechwerte
Ansprechwert Ran1 (ALARM 1) .................................................................................... 1…200 kΩ
Ansprechwert Ran2 (ALARM 2) .................................................................................... 1…200 kΩ
Preset-Funktion:
Un≤72 V: Ran1 (ALARM 1)/ Ran2 (ALARM 2) .......................................................... 20 kΩ / 10 kΩ
Un>72 V: Ran1 (ALARM 1)/ Ran2 (ALARM 2) .......................................................... 46 kΩ / 23 kΩ
Ansprechabweichung (1…5 kΩ) / (5…200 kΩ) ............................................ ±0,5 kΩ
/
±15 %
Hysterese (1…5 kΩ) / (5…200 kΩ) ................................................................. +1 kΩ
/
+25 %
Zeitverhalten
Ansprechzeit tan bei RF= 0,5 x Ran und Ce= 1 μF................................................................. ≤ 2 s
Anlaufverzögerung t .......................................................................................................... 0…10 s
Ansprechverzögerung ton .................................................................................................. 0…99 s
Messkreis
Messspannung Um.................................................................................................................. ±12 V
Messstrom Im(bei RF= 0Ω).............................................................................................≤ 200 μA
Innenwiderstand DC Ri........................................................................................................ ≥62 kΩ
Impedanz Zibei 50 Hz......................................................................................................... ≥60 kΩ
Zulässige Fremdgleichspannung Ufg ..............................................................................≤ DC 300 V
Zulässige Netzableitkapazität Ce.......................................................................................... ≤ 20 μF
Anzeigen, Speicher
Anzeige ......................................................................... LC-Display, multifunktional, unbeleuchtet
Anzeigebereich Messwert.......................................................................................... 1 kΩ…1 MΩ
Betriebsmessabweichung (1…5 kΩ) ............................................................................... ±0,5 kΩ
Betriebsmessabweichung (5 kΩ…1 MΩ) ......................................................................... ±15 %
Passwort ...................................................................................................................... off / 0…999
Fehlerspeicher (Alarmrelais) ................................................................................................ on / off
Eingänge
Leitungslänge externe Test- / Reset-Taste ..........................................................................≤ 10 m
Schaltglieder
Anzahl .............................................................................................................. 2 (Wechsler K1, K2)
Arbeitsweise ........................................................................................... Ruhestrom / Arbeitsstrom
Elektrische Lebensdauer bei Bemessungsbedingungen .................................. 10000 Schaltspiele
Kontaktdaten nach IEC 60947-5-1:
Bemessungsbetriebsspannung AC ....................................................................... 230 V....... 230 V
Gebrauchskategorie AC ........................................................................................ AC 13....... AC 14
Bemessungsbetriebsstrom AC ................................................................................. 5 A............ 3 A
Bemessungsbetriebsspannung DC ................................................... 220 V....... 110 V.......... 24 V
Gebrauchskategorie DC ..................................................................... DC 12........DC 12....... DC 12
Bemessungsbetriebsstrom DC .......................................................... 0.1 A........ 0.2 A............ 1 A
Mindeststrom................................................................................................ 1 mA at AC/DC ≥ 10 V
Umwelt/EMV
EMV........................................................................................................................... nach IEC 61326
Arbeitstemperatur ............................................................................................... -25 ºC…+55 ºC
Klimaklassen nach IEC 60721:
Ortsfester Einsatz (IEC 60721-3-3) (ohne Betauung und Eisbildung)....................................... 3K5
Transport (IEC 60721-3-2) (ohne Betauung und Eisbildung)................................................... 2K3
Langzeitlagerung (IEC 60721-3-1) (ohne Betauung und Eisbildung)...................................... 1K4
Mechanische Beanspruchung nach IEC 60721:
Ortsfester Einsatz (IEC 60721-3-3) .......................................................................................... 3M4
Transport (IEC 60721-3-2) ....................................................................................................... 2M2
Langzeitlagerung (IEC 60721-3-1) .......................................................................................... 1M3