BALANCE SD PLATFORM ACTUATOR CALIBRATION
1. Tap the touch screen to turn on the display.
2. Touch <UTILITIES> on the screen.
Note: The ext step requires pressi g the hidde keypads i the upper right a d left cor ers of the
display's touch scree .
3. Touch Top-Right Corner, Top-Left Corner, Top-Right Corner successively, to access the
<SYSTEM MAINTENANCE> menu on the display.
4. Touch <SYSTEM MAINTENANCE>.
5. Touch <DIAGNOSTICS>
6. Touch <PLATFORM ACTUATOR> on the screen.
7. ress the <DOWN ARROW> on the screen until the motor stalls (bottoms out). Then look at
the READING in the bottom right of screen and bring motor up 3 digits.
8. Touch <SET NEW MIN> on the screen to calibrate level 1.
9. Touch and hold the <UP> arrow on the screen until the motor stalls.
10. Touch <SET NEW LOCKED> on the screen to calibrate the locked position.
11. Touch <COMPENSATION> at the bottom of the screen and listen for the actuator. The
“arrow” indicator will move also.
12. Touch <BACK> on the screen four times to exit and return to the main screen.
TILT SENSOR CALIBRATION PROCEDURE
1. Install the platform if it’s not already installed.
2. Tap the touch screen to turn on the display.
3. Touch UTILITIES” on the screen.
Note: The ext step requires pressi g the hidde keypads i the upper right a d left cor ers of the
display's touch scree .
4. Touch Top-Right Corner, Top-Left Corner, Top-Right Corner successively, to access the
SYSTEM MAINTENANCE” menu on the display.
5. Touch SYSTEM MAINTENANCE”.
6. Touch DIAGNOSTICS”
CautioN!: The platform will release whe you perform the ext step
atteNtioN!: La plate-forme libérera qua d vous exécutez la prochai e étape.
7. Touch TILT SENSOR” to release the platform.
8. Depress the platform to the maximum 9 o’clock position. The LEFT (min)”
window will turn green signifying proper calibration.
9. Depress the platform to the maximum 3 o’clock position. The RIGHT (max)”
window will turn green signifying proper calibration.
10. Depress the latform to the maximum 6 o’clock position. The BACKWARD (min)”
window will turn green signifying proper calibration.
11. Depress the platform to the Maximum 12 o’clock position. The FORWARD (max)”
window will turn green signifying proper calibration.
12.YOU MUST press the BACK key 4 times to save and exit the calibration screen.