
SET-UP PROCEDURE
(See Figures 1 - 3.)
1. Open lower front door. All hoses should be connected to their respective ports. (See Figure 1.)
2. Open the top door. The Pulmonex can use either of two different style Drierite cartridges. A
refillable plastic cartridge is shipped with your unit, it is located on the right side of the com-
partment. You can fill this cartridge to one-third full with loose Drierite (catalog #139-101).
Alternatively, you can remove this cartridge and replace it with the sample pre-filled single use
cartridge included in the sample Xenon Disposable Convenience Kit supplied. In either case,
the disposable blue Drierite cartridge prolongs charcoal trap life span. Failure to change the
moisture absorber regularly will significantly shorten the life span of the charcoal trap, there-
fore, it is recommended that you change it after every study. (See Figure 2.)
3. Located on the left side of the top compartment is the Soda-Lime cartridge. This cartridge
serves as a carbon dioxide trap. You can fill this cartridge to one-third full with loose Soda-
Lime (catalog #130-019). Alternatively, you can remove this cartridge and replace it with the
sample white pre-filled single use cartridge included in the sample Xenon Disposable
Convenience Kit supplied. In either case, the Soda-Lime cartridge helps stop the breakdown
of Soda-Lime granules that may possibly clog the 10-way valve and hinder patient air flow.
Failure to change the Soda-Lime regularly could cause the patient to hyperventilate, there-
fore, it is recommended that you change it after every study.
NOTE: The flow arrow on the pre-filled Soda-Lime cartridge should match the flow label in
the Pulmonex II cabinet.
4. Bring the unit to the area of operation. Ensure the Power Switch is OFF and the timer is on
"0". Plug into a near by electrical outlet.
5. At the rear of the unit, there are two black hose connections positioned side by side. (See
Figure 3.) The blue hose adapters in the open ends of hoses should be removed to connect
to these fittings. The blue adapters are required to connect to the Hans Rudolph Valve and to
old Pulmonex only.Attach the 1.125" bore hose and Hans Rudolph Valve assembly to the
hose connectors so that air flows into the Hans Rudolph Valve on the left and out from the
Hans Rudolph Valve on the right. Connect the bacteria filter directly to the Hans Rudolph
Valve. Attach either a face mask or mouthpiece directly to the bacteria filter.
NOTE: Keep the breathing tubes as short as possible. (See "Trouble-shooting" Strong Breathing
Resistance.) For supine patients, bring the system as close as possible to the bedside.
As a safety precaution you can connect a hose from your room vent to the exhaust port locat-
ed just below the overhang on the patient side of the Pulmonex II.
CAUTION: Some patients are sensitive to Oxygen. Consult a physician before using
Oxygen. If the physician prefers, substitute room air for Oxygen.
ATTENTION: Certains patients réagissent à l’oxygène. Se référer à un médecin avant
d’utiliser de l’oxygène. Utiliser l’air ambiant quand indiqué.
6. To add Oxygen, connect and clamp a .25" Oxygen hose from your supply to the Oxygen inlet
port on the Pulmonex II front panel.
CAUTION: Do NOT use humidified Oxygen.
ATTENTION: Ne pas utiliser d’oxygène contenant de la vapeur d’eau.
7. Ensure that the shipping cap is removed from the exhaust port located on the lower cabinet
on the rear of the unit (facing the patient).
-5- SET-UPPROCEDURE
Figure 1.
Ensure all hoses in lower cabinet
are properly connected.
Figure 2.
The Carbon Dioxide absorber
and moisture absorber cartridges
are located in the upper cabinet
Figure 3.
Connectors for the 1.125" bore
hose and Hans Rudolph Valve
assembly are located on the back
panel.