5400578 Rev. A
Message Reason Fix
ERROR 1
THERMISTOR OPEN THERMISTOR OPEN
CALL BIOLASE SERVICE
ERROR 2
THERMISTOR SHORTED THERMISTOR SHORTED
ERROR 3 SHUTDOWN TEMPERATURE SYSTEM TOO HOT ALLOW 5-10 MINUTES FOR LASER TO
COOL DOWN
ERROR 4
LASER CURRENT HIGH/LOW OUTPUT IS OUT OF SPECSCALL BIOLASE SERVICE
ERROR 5
FOOTSWITCH SHORTED IN STANDBY FOOTSWITCH IS PARTIALLY
PRESS
RELEASE FOOTSWITCH OR CALL
ERROR 6
ON/OFF BUTTON STUCK KEY STUCK PRESS FRONT KEY
ERROR 7
FLASH CORRUPTED MEMORY CORRUPTED CALL BIOLASE SERVICE
ERROR 8
NO FIBER FIBER NOT INSERTED PLUG IN TRUNK FIBER
ERROR 9
LOST FOOTSWITCH COMMUNICATION WIRELESS INTERFERENCE REPOSITION CONSOLE OR FOOTSWITCH
ERROR 10
EMERGENCY SWITCH E-SWITCH PRESSED PRESS E-SWITCH AGAIN
ERROR 11
REMOTE INTERLOCK REMOTE INTERLOCK OPEN CHECK REMOTE INTERLOCK CLOSED
ERROR 12
BATTERY CRITICALLY LOW BATTERY CRITICALLY LOW PLUG IN DC SUPPLY
ERROR 13
INTERNAL ERROR INTERNAL ERROR OCCURRED RESTART UNIT
ERROR 14
FOOTSWITCH BATTERY FOOTSWITCH BATTERY
REPLACE FOOTSWITCH BATTERY
WARNING 1
TEMP HIGH SYSTEM IS HOT ALLOW
MINUTES FOR LASER TO
WARNING 2
BATTERY LOW BATTERY IS LOW PLUG IN DC SUPPLY
WARNING 3
BATTERY NOT CONNECTED BATTERY NOT CONNECTED PLUG IN BATTERY
WARNING 4
FOOTSWITCH BATTERY LOW BATTERY ON THE FOOTSWITCH
REPLACE FOOTSWITCH BATTERY
WARNING 5
FOOTSWITCH FOOTSWITCH HELD RELEASE FOOTSWITCH
ALERT 1
WIRELESS NOT PAIRED NO WIRELESS CONNECT RE-ESTABLISH PAIRING (SEE SEC 4)
ALERT 2
SYSTEM MUST BE IN READY MODE TO
SYSTEM IS NOT IN READY
PRESS THE CONTROL BUTTON IN ANY