General warnings.
•The following information must be read and understood before proceeding with the installation, commissioning and maintenance of
the devices described in this document.
•ATTENTION! Any omission or failure to follow these instructions scrupulously can cause danger.
•ATTENTION! Explosion hazard. In case of batteries substitution, make sure that the type is compatible and complies with the
specifications indicated by the manufacturer.
•In case of batteries substitution, DO NOT disconnect the flat cable which connect the electronic boards without removing batteries
before
•ALWAYS substitute all the batteries, also in case of one single battery exhausted.
•If the device is powered by a fixed network, make sure you have disconnected the power supply before carrying out any type of
intervention. Failure to comply with this indication can cause damage to people and/or property.
•Follow the manufacturer's suggested warm-up time (
time required to obtain a reliable measurement
) of the transducer.
•Follow the manufacturer's suggested electric wiring of the transducer to measure (
ground shields at a single point, cable length and
section
); voltage measures on distances exceed 15/20 meters are subjected to electromagnetic disturbances. 0÷25mA inputs have a
superior electromagnetic compatibility (
EMC
).
•Avoid passage in cavities with power or high voltage cables.
•The protection and safety measures and the warranty provided by the Manufacturer with the equipment may be compromised if it’s
used in a manner that does not comply with this user manual.
•This equipment complies with CE regulations.
•Modifications or tampering not expressly approved by the Manufacturer could void the user's authorization to operate the equipment.
•This equipment must be installed by qualified personnel and in accordance with national regulations and/or related local requirements.
•Make sure that the object is properly fixed to supports/infrastructures capable of withstanding this load. Make sure proper methods
and materials are used when fixing the equipment to a wall.
•Only personnel expressly authorized by the manufacturer can open the container. There are no user serviceable parts inside.
1. Description.
The WSD12-VW is a datalogger designed to measure 4 input channels to acquire frequency and temperature on vibrating wires, with storage
functionality of samples acquired.
2. Device pre-set and use mode.
a. Wireless Mode:
In this use mode, the datalogger records and transmits each measure sampled to the linked
gateway. Typically, the monitoring system is configured in factory, so, if the device is already
associated to the system gateway is in STANDBY mode (
refer to Picture 4 - Status table –
Wireless mode
). In this case, it is necessary to start it with the TEST command (
refer to Picture
3 –Wireless devices user interface
). After this operation, the datalogger reactivates, resumes
measure activity, and performs the connection to the gateway or to a
repeater WR12.
Connection The onboard LED shows the radio signal quality for 2 minutes (
refer to Picture
4 - Status table –Wireless mode
).
Otherwise, in case the device is in FACTORY RESET mode (
refer to refer to Picture 4 - Status
table –Wireless mode
), that means it's ready for connecting to an existing system, in order
to associate it using the ENROL command, Refer to the
“WineCap System - User Manual
R30”
software manual. It is necessary to use the
WineCapManager
software on the PC
connected to the gateway that will be linked with the device.
b. USB Logger Mode:
For this operation mode, the dataloggers works in stand-alone, recording data in its internal
memory and without any wireless connection. This is not the factory setup so, to select it, the connection with the PC and the
WineCapManager
running on it is necessary to modify the device 's operation mode. The device must be set in STAND-ALONE
(refer to 6
- Stand-alone USB datalogger installation.)
mode and the sample time must be configured; automatically, the device 's clock is aligned with
the PC's clock, in order to assure the temporal reference of the sample.