CAPETTI ELETTRONICA s.r.l. - Strada Stratta, 57
10090 CASTIGLIONE TORINESE - TORINO - ITALY
VAT 04837130014 - Phone 011.981.98.11 - Fax 011.981.98.210
WA0241E_WSD12T-CO_User Manual_R02
V4.8
The following COMMAND table describes the available commands:
Shows the device STATUS. As answer the led perform a flash sequence as reported
in the “STATUS” table. If the device is performing the TEST (refer to TEST command)
this command stops it.
Enter in TEST mode and transmits status and measurements every 5 seconds. If
the device is in STANDBY mode or it is out of radio range, this command forces the
connection procedure to the WSN and the return to the operative mode. The TEST
stops after 120 seconds. During TEST, the led continuously shows the STATUS to
monitor the received radio signal quality.
CAUTION: Measures acquired during TEST phase are NOT saved.
Association to the network: must be used when the device has not yet been
included in a network, starts the entry and association procedure to the
basestation (refer to “WineCap System - User Manual R30”).
Temporary device deactivation: the device is stopped. The sampling process and
the radio are/is turned off losing the connection to the network. To reactivate, a
TEST command is necessary. The STANDBY command must be given twice to
confirm it: at the first sequence the led flashes alternating RED and GREEN lights,
waiting for the second confirm sequence within 15 seconds. At the command
execution the led flashes as the STANDBY status (refer to “WIRELESS MODE STATUS
Table”).
The device performs the memory deleting procedure and goes in STOP status. All
samples, configuration and wireless network data associated are LOST. To
reactivate the device a new association and configuration procedure is necessary
(ENROLL command). Also in this case, the FACTORY RESET command must be given
twice to confirm it. At the command execution the led flashes as the
“PROBE/DATALOGGER NOT ASSOCIATED” status refer to “WIRELESS MODE STATUS
Table”).
As the previous command but performs only the WSN deleting procedure and
disassociate from the basestation. The device enters in LOGGER STAND ALONE
mode: data are kept, and the sampling activity CONTINUES with previous setup.
Command must be given with 2 sequences: 5 flashes and then 3 flashes. At the
command execution wait for the device reboot. At the STATUS command,
“LOGGER” will be the answer (refer to “STAND-ALONE MODE STATUS Table”). A new
association (ENROLL command) is possible to a new basestation.
4. Enrolling the device.
Not necessary if performed in factory before delivery.
Enroll the device to the network referring to the “WineCap System - User Manual R30”. In case the device is already enrolled but in
STANDBY status, a TEST command must be issued (refer to Picture 4 - Commands table).
5. Installation procedure.
After installing the basestation in appropriate place in charge, (refer to “WineCap System - User Manual R30”), be sure that the device is
enrolled to the basestation and activated.
Head for the environment to be monitored. On the way, to check the quality of the radio coverage, use the "Field Measurer" function.
This function is activated issuing the TEST (refer to Picture 4 - Commands table)command: position the WineCapKey in the spot indicated
in Picture 3 - WineCapKey positioning and wait for two AMBER flashes, then remove the WineCapKey from device. The "Field
Measurer" function lasts enabled for two minutes.
To issue commands to the device, place the WineCapKey where indicated.
Once the WineCapKey is detected, the led periodically emits AMBER flashes with a 2 second cadence.
For each flash, a different command is associated; to confirm the command the WineCapKey must be removed from the sensible
area immediately after the number of flashes corresponding at the desired command. The TEST corresponds to the second pulse and
activate the "Field Measurer" function.
Picture 4 - Commands table