7.0 OPERATING THE INSTRUMENT
Switching on: Press for 3 seconds.
Changing the calibration curve: or .
Setting the time: 2 times - Options – Date / Time
Set date and time using the button , according to the format indicated
(JJ.MM.TT). After entering the year, press the button for entering the month and
again for entering the day. For changing from date to time also press the button
. After finishing, press OK for saving the entered data.
Datalog: Select your desired interval in the menu Options – Log Time using the
arrow keys, and confirm by pressing OK. Now in the store menu appears the
symbol . By pressing this
symbol you can activate the AutoLog.
Info: In order to save battery power the device switches off automatically at a
log interval of 1 minute or longer, and activates again for saving the logs!
For completing the AutoLog, switch on the device (if necessary) and press the
button. If you want to add supplier´s data please press the button. Supplier´s data
can also be entered on the PC subsequently.
Switching on the display lighting: Press the key briefly; the display lighting
switches off automatically after approx. 20 seconds. Pressing any key activates the
display lighting again, and the period for switching off again is prolonged to four
minutes (The display lighting time can be modified in menu level Options – BL On
Time).
Switching off: Press the key for 5 seconds. The instrument switches off after
releasing the key. The instrument switches off automatically after approx. four
minutes. (The turn-off time can be modified in menu level Options – Auto Off Time).
Other instrument functions – overview
•Manual saving of single measuring values in a measurement series
•Display of measuring series and measuring values directly on the instrument
•Printing the saved measuring series (only with PC interface and printer)
•Transfer and saving of measuring series on a PC (only with PC interface)
•Automatic single-point adjustment at 50% humidity standard
•Selection of menu language (DE, EN, FR, IT, ES, RU)
•Display of temperature in Celsius or Fahrenheit
Electromatic Equipment Co., Inc.
-15-
18.0 TECHNICAL DATA
Measurement Meas. Range / Resolution / Accuracy
rel. humidity 0 to 100% rh / 0,1%
calibration 10 to 90% / ±1,5% rh (at 25°C)
temperature °C -10 to +60°C / 0,1°C / ±0,3°C (at 25°C)
temperature °F 14 to 140°F / 0,3°F / ±0,5°F (at 77°F)
dew point °C -55 to +60°C / 0,1°C
dew point °F -67 to 140°F / 0,3°F
Operating temp. range -10°C to 60°C / 14 to 140°F
Storage temp -20°C to 60°C / -4 to 140°F
Temperature compensation Automatic
Data storage approx. 10.000 measuring values
Menu languages English, German, French, Italian, Spanish,
Russian
Power supply 4 pcs. 1.5Volt AA Alkaline batteries
(for approx. 1800 measurements)
Auto Off time after approx. 4 minutes
Power consumption 30 mA (with display lighting)
Display 128 x 64 matrix display with LED backlighting
Dimensions housing 145 x 63 x 29mm
Dimensions sword sensor 295 x 20 x 4mm
Weight approx. 260g (incl. batteries)
Protection class IP 40
Scope of supply wooden case, rubber protection cover, 4 pcs.
1.5 Volt AA Alkaline batteries, user manual