ChoiceMMed Fingertip Pulse Oximeter User manual

Fingertip
Pulse Oximeter
USER MANUAL
General Description
Oxygen binds to hemoglobin in red blood cells when moving through the lungs. It is transported throughout the body as arterial
blood. A pulse oximeter uses two frequencies of light (red and infrared) to determine the percentage (%) of hemoglobin in the blood
that is saturated with oxygen. The percentage is called blood oxygen saturation, or SpO2. A pulse oximeter also measures and
displays the pulse rate at the same time it measures the SpO2level.
Measurement Principle
Principle of the oximeter is as follows: The pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The
sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is
905nm, which is infrared-red light. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The
photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The
arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and
decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This
measurement is referred to as SpO2.
Diagram of Operation Principle
1. Red and Infrared-ray Emission Tube
2. Red and Infrared-ray Receipt Tube
Precautions For Use
1.Before use, carefully read the manual.
2.Operation of the fingertip pulse oximeter may be affected by the use of an electrosurgical unit (ESU).
3.The fingertip pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO2measurement. Verify that
nothing is hindering the pulse measurement before relying on the SpO2measurement.
4.Do not use the fingertip pulse oximeter in an MRI or CT environment.
5.Do not use the fingertip pulse oximeter in situations where alarms are required. The device has no alarms. It is not for
continuous monitoring.
6.Do not use the fingertip pulse oximeter in an explosive atmosphere.
7.The fingertip pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other
methods of assessing clinical signs and symptoms.
8.In order to ensure correct sensor alignment and skin integrity, the maximum application time at a single site for our device
should be less than half an hour.
9.Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device in liquid. The device is not
intended for sterilization.
10.Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components,
including batteries.
11.This equipment complies with IEC 60601-1-2:2014 for electromagnetic compatibility for medical electrical equipment and/or
systems. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in
healthcare and other environments, it is possible that high levels of such interference due to close proximity or strength of a source
might disrupt the performance of this device.
12.Portable and mobile RF communications equipment can affect medical electrical equipment
13.This equipment is not intended for use during patient transport outside the healthcare facility.
14.This equipment should not be used adjacent to or stacked with other equipment.
15.It may be unsafe to:
—use accessories、detachable parts and materials not described in the instructions for use
—interconnect this equipment with other equipment not described in the instructions for use
—disassemble, repair or modify the equipment
16.These materials that contact with the patient’s skin contain medical silicone and ABS plastic enclosure are all pass the
ISO10993-5 Tests for invitro cytotoxicity and ISO10993-10 Tests for irritation and delayed-type hypersensitivity.
17.When the signal is not stable, the reading may inaccurate. Please do not reference.
Rx only: “Caution: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.”
Contraindication
It is not for continuous monitoring.
Inaccurate measurements may be caused by
1. Significant levels of dysfunctional hemoglobin (such as carbonyl - hemoglobin or methemoglobin).
2. Intravascular dyes such as indocyanine green or methylene blue.
3. High ambient light. Shield the sensor area if necessary.
4. Excessive patient movement.
5. High-frequency electrosurgical interference and defibrillators.
6. Venous pulsations.
7. Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line.
8. The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia.
9. The patient is in cardiac arrest or is in shock.
10. Fingernail polish or false fingernails.
11. Weak pulse quality (low perfusion).
12. Low hemoglobin.
Product Features
1. Simple to operate and convenient to carry.
2. Small volume, light weight and low power consumption.
3. Dual color OLED displays SpO2, PR, Pulse bar, and waveform.
4. Level 1-10 adjustable brightness.
5. 6 display modes.
6. 2pcs AAA-size alkaline batteries; battery-low indicator.
7. When it shows "Finger out", the pulse oximeter will power off automatically in 8 seconds.
Intended Use
The Fingertip Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial
hemoglobin (SpO2) and Pulse Rate of adult, adolescent and child patients in hospitals, hospital-type facilities and homecare.
Operation Instructions
1. Install two AAA batteries according to the Battery Installation instructions.
2. Place one of your fingers into the rubber opening of the pulse oximeter.
3. Press the switch button one time on front panel to turn the pulse oximeter on.
4. Keep your hands still for the reading. Do not shake your finger during the test. It is recommended that you do not move your
body while taking a reading.
5. Read the data from the display screen
6. Press the power switch for longer than one second, will adjust the brightness of the oximeter. There are 10 levels of brightness.
The default is level four.
After turning on the Oximeter, each time you press the power switch, the Oximeter will switch to another display mode.
There are 6 display modes shown as follows:
1. 2. 3. 4. 5. 6.
Front Panel
Note:
1. If the screen display “?”, it means the signal is unstable, please keep your hands still and retry.
2. The pulse bar less than 30% indicates signal inadequacy and the displayed SpO2and pulse rate value is potentially incorrect.
Battery Installation
1. Install two AAA batteries into the battery compartment. Match the plus (+) and minus (-) signs in the compartment. If the
polarities are not matched, damage may be caused to the oximeter.
2. Slide the battery door cover horizontally along the arrow shown as the picture.
Notes:
Please remove the batteries if the pulse oximeter will not be used for long periods of time.
Please replace the battery when the power indicator starting flickering.
Using the Lanyard
1. Thread thinner end of the lanyard through the hanging hole.
2. Thread thicker end of the lanyard through the threaded end before pulling it tightly.
Warnings!
Keep the oximeter away from young children. Small items such as the battery door, battery, and lanyard are choking hazards.
Do not hang the lanyard from the device’s electrical wire.
Please notice that the lanyard which is tied to the oximeter may cause strangulation due to excessive length.
Maintenance and Storage
1. Replace the batteries in a timely manner when low voltage lamp is lighted.
2. Clean surface of the fingertip oximeter before it is used in diagnosis for patients.
3. Remove the batteries if the oximeter is not operated for a long time.
4. It is best to store the product in -25℃~+70℃and ≤93% humidity.
5. Keep in a dry place. Extreme moisture may affect oximeter lifetime and may cause damage.
6. Dispose of battery properly; follow any applicable local battery disposal laws.
Cleaning the fingertip pulse oximeter
Please use medical alcohol to clean the silicone touching the finger inside of oximeter with a soft cloth dampened with 70%
isopropyl alcohol. Also clean the being tested finger using alcohol before and after each test.
Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any openings in the device. Allow the oximeter to
dry thoroughly before reuse.
The fingertip pulse oximeter requires no routine calibration or maintenance other than replacement of batteries.
The use life of the device is five years when it is used for 15 measurements every day and 10 minutes per one
measurement. Stop using and contact local service center if one of the following cases occurs:
An error in the Possible Problems and solutions is displayed on screen.
The oximeter cannot be powered on in any case and not the reasons of battery.
There is a crack on the oximeter or damage on the display resulting readings cannot be identified; the spring is invalid; or the
key is unresponsive or unavailable.
Disinfecting
The applied parts touching the patients’ body are required to be disinfected once after each use. The recommended disinfectants
include: ethanol 70%, isopropanol 70%, glutaraldehyde-type 2% liquid disinfectants.
Disinfection may cause damage to the equipment and is therefore not recommended for this pulse oximeter unless otherwise
indicated in your hospital’s servicing schedule. Clean the pulse oximeter before disinfecting it.
CAUTION: Never use EtO or formaldehyde for disinfection.
Specifications
1. Display Type
OLED display
2. SpO2
Display range: 0%~100%
Measurement range: 70%~100%
Accuracy: 70%~100%±2%; 0%~69% no definition
Resolution: 1%
Note:A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor. Clinical testing is used to
establish the SpO2accuracy. The measured arterial hemoglobin saturation value (SpO2) of the sensors is compared to arterial
hemoglobin oxygen (SaO2) value, determined from blood samples with a laboratory CO-oximeter. The accuracy of the sensors in
comparison to the CO-oximeter samples measured over the SpO2range of 70%~100%. Accuracy data is calculated using the
root-mean-squared (Arms value) for all subjects, per ISO 9919:2005, Medical Electrical Equipment–Particular requirements for the
basic safety and essential performance of pulse oximeter equipment for medical use.
A functional tester is used to measure how accurately Fingertip Pulse Oximeter is reproducing the specified calibration curve and
the PR accuracy.
The model of functional tester is Index2 FLUKE simulator and the version is 2.1.3.
3. Pulse Rate
Display range: 0bpm~250bpm
Measure range: 30bpm~250bpm
Accuracy: 30bpm~99bpm, ±2bpm; 100bpm~250bpm, ±2%
Resolution: 1bpm
4. Probe LED Specifications
Wavelength Radiant Power
RED 660±3nm 3.2mw
IR 905±10nm 2.4mw
NOTE: The information about wavelength range can be especially useful to clinicians.
5. Power Requirements
Two AAA alkaline Batteries
Power consumption: Less than 40mA
6. Environment Requirements
Operation Temperature: 5 ~40℃℃
Storage Temperature: -25 ~+70℃℃
Ambient Humidity: 15%~93% no condensation in operation; ≤93% no condensation in storage/transport
Atmosphere pressure: 70kPa~106kPa
7. Equipment data update period
As shown in the following figure. Data update period of slower average is 8s.
8. Classification
According to the type of protection against electric shock: INTERNALLY POWERED EQUIPMENT;
According to the degree of protection against electric shock: TYPE BF APPLIED PART, (applied part: the rubber hole of the device);
According to the degree of protection against ingress of water: IP22
According to the mode of operation: CONTINUOUS OPERATION
Clinical Study Summary
The following details are provided to disclose actual performance observed in the clinical validation study of healthy adult
volunteers. The ARMS value analysis statement and Bland-Altman plot of data is shown as following:
ARMS Value Analysis Statement
Item 90--100 80--<90 70--<80
#pts 78 66 63
Bias 1.02 0.40 -0.48
ARMS 1.66 1.46 1.93
Bland-Altman Plot Graphic
Declaration
Requirement-Test Result/Comments Verdict
Clause 7-Emissions
Classification -- —
Class A or B………………………………………………….....: Class B —
Group 1 or 2………………………………………………........: Group 1 —
CISPR 11, 14-1, 32 or ISO 7137…………………………......: CISPR 11 —
Conducted RF Emissions………………………………..........: N/A N/A
Radiated RF Emissions………………………………………..: -- P
Disturbance Power (if applicable)………………...................: N/A N/A
Harmonic Distortion per IEC61000-3-2 (Class A, B, C, D)…: N/A N/A
Voltage Fluctuations and Flicker per IEC61000-3-3……......: N/A N/A
Clause 8 - Immunity
Electrostatic Discharges………………………………………..: IEC 61000-4-2 P
Radiated RF EM Fields and Proximity Wireless fields……...: IEC 61000-4-3 P
Electrical Fast Transients and bursts………………...............: IEC 61000-4-4 N/A
Surges………………………………………….........................: IEC 61000-4-5 N/A
Conducted Disturbances, Induced by RF fields…….............: IEC 61000-4-6 N/A
Voltage Dips and Interruptions………………………..............: IEC 61000-4-11 N/A
Rated Power-frequency Magnetic Field………......................: IEC 61000-4-8 P
Possible Problems and Solutions
Problems Possible reason Solution
SpO2or PR can not
be shown normally
1. Finger is not inserted correctly
2. Patient’s SpO2value is too low to be measured
1. Retry by inserting the finge
r
2. There is excessive illumination
3. Try some more times. If you can make sure no
problem exist in the product, please go to a
hospital timely for exact diagnosis.
SpO2or PR is shown
unstably
1. Finger might not be inserted deep enough.
2. Excessive patient movement
1. Retry by inserting the finge
r
2. Be calmness
The oximeter cannot
be powered on
1. No battery or low power of battery
2. Batteries might be installed incorrectly
3. The oximeter might be damaged
1. Please replace batteries
2. Please reinstall the batteries
3. Please contact with local customer service
centre
Indication lamps are
suddenly off
1. The product is automatically powered off when no
signal is detected longer than 8 seconds
2. The battery power is too low to work
1. Normal
2. Replace the batteries
“Err7” is displayed on
screen
Err 7 means all the emission LED or reception diode is
damaged.
Please contact with local customer service centre
Symbol Definitions
Symbol Definition Symbol Definition
Type BF applied part. Attention
IP22 Protected against dripping
water. ﹪SpO2Oxygen saturation
PR bpm Pulse rate (BPM) Low power indication
No SpO2Alarm Serial No.
Storage temperature and
relative humidity Follow instruction for use
Date of Manufacture Authorized representative in the
European community
European union approval Manufacturer’s information
Conformity to WEEE Directive ?Indicate the signal is not stable
Box Contents
1. Fingertip pulse oximeter
2. One lanyard
3. Two AAA batteries
4. One instruction manual
Applicable Models
MD300C2 MD300C21 MD300C21C MD300C22 MD300C23 MD300C25 MD300C26 MD300C29 MD300C2A
MD300C2B MD300C2D MD300C2E MD300C2F MD300C2G MD300C2H MD300C2I MD300C203 MD300C21-P
MD300C20
Pay attention:
Only the device of MD300C203 is single-color OLED screen.
Only the device of MD300C21-P has the feature of automatically power on.
Notes:
1. The illustrations used in this manual may differ slightly from the appearance of the actual product.
2. The specifications are subject to change without prior notice.
ALL RIGHTS RESERVED
Revised Date: April 01, 2020
Version: Ver3.0

Other ChoiceMMed Medical Equipment manuals

ChoiceMMed MD100C Pro User manual

ChoiceMMed

ChoiceMMed MD100C Pro User manual

ChoiceMMed OxyWatch C20 User manual

ChoiceMMed

ChoiceMMed OxyWatch C20 User manual

ChoiceMMed LungBoost MD8000 User manual

ChoiceMMed

ChoiceMMed LungBoost MD8000 User manual

ChoiceMMed MD300CF3 User manual

ChoiceMMed

ChoiceMMed MD300CF3 User manual

ChoiceMMed MD300C318 User manual

ChoiceMMed

ChoiceMMed MD300C318 User manual

ChoiceMMed MD300 User manual

ChoiceMMed

ChoiceMMed MD300 User manual

ChoiceMMed MD300C312 User manual

ChoiceMMed

ChoiceMMed MD300C312 User manual

ChoiceMMed MD300W User manual

ChoiceMMed

ChoiceMMed MD300W User manual

ChoiceMMed MDTS100 User manual

ChoiceMMed

ChoiceMMed MDTS100 User manual

ChoiceMMed MD300C2 User manual

ChoiceMMed

ChoiceMMed MD300C2 User manual

ChoiceMMed MD300K User manual

ChoiceMMed

ChoiceMMed MD300K User manual

ChoiceMMed MD300C5 User manual

ChoiceMMed

ChoiceMMed MD300C5 User manual

ChoiceMMed MD300I User manual

ChoiceMMed

ChoiceMMed MD300I User manual

ChoiceMMed MD300C228 User manual

ChoiceMMed

ChoiceMMed MD300C228 User manual

ChoiceMMed Fingertip Pulse Oximeter User manual

ChoiceMMed

ChoiceMMed Fingertip Pulse Oximeter User manual

ChoiceMMed MD300C2ES User manual

ChoiceMMed

ChoiceMMed MD300C2ES User manual

ChoiceMMed MD300-D User manual

ChoiceMMed

ChoiceMMed MD300-D User manual

Popular Medical Equipment manuals by other brands

Portascope PVSM6150 instruction manual

Portascope

Portascope PVSM6150 instruction manual

Labomed eVA 500 user manual

Labomed

Labomed eVA 500 user manual

EchoNous KOSMOS quick guide

EchoNous

EchoNous KOSMOS quick guide

Masimo Radius PPG manual

Masimo

Masimo Radius PPG manual

Roche CoaguChek XS Pro Operator's manual

Roche

Roche CoaguChek XS Pro Operator's manual

OCENCO M-40 SCSR Instruction booklet

OCENCO

OCENCO M-40 SCSR Instruction booklet

Gima OXY-3 user manual

Gima

Gima OXY-3 user manual

TECHNO-GAZ Master Flux Plus Instructions for use

TECHNO-GAZ

TECHNO-GAZ Master Flux Plus Instructions for use

natus Camino ICP user manual

natus

natus Camino ICP user manual

EKO Device CORE user guide

EKO Device

EKO Device CORE user guide

Dräger Apollo Quick reference guide

Dräger

Dräger Apollo Quick reference guide

Kid-Man HL-7100 user manual

Kid-Man

Kid-Man HL-7100 user manual

GE LOGIQ P3 PRO Service manual

GE

GE LOGIQ P3 PRO Service manual

Cogvis cogvisAI installation guide

Cogvis

Cogvis cogvisAI installation guide

Skytron F-PTB2SKY Instructions for use

Skytron

Skytron F-PTB2SKY Instructions for use

College Park venture Technical instructions

College Park

College Park venture Technical instructions

CORETECH ROM ELBOW BRACE quick start guide

CORETECH

CORETECH ROM ELBOW BRACE quick start guide

Ofa Bamberg Dynamics Knee immobilisationsplint Important notes

Ofa Bamberg

Ofa Bamberg Dynamics Knee immobilisationsplint Important notes

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.