WRISTPULSEOXIMETER V2.0W1
4
zUse the Oximeter only within the specified temperature ranges: 5℃~40℃
humidity≤80% for operating, and -20℃~50℃humidity≤93% for storage and
transportation.
zWhen you simultaneously measure blood pressure, attach the sensor to a different
arm’s finger.
zDo not stretch the adhesive tale while applying the pulse oximeter sensor.
zEnsure that the wrist band fits comfortably on the patient’s arm. Do not over-tighten the
wrist band.
zPulse oximeter readings might be affected while patients are being defibrillated.
zDo not use the unit under conditions of shocks and vibrations. Do not use it, either, with
the patient connected to such medical electrical equipment as a cardiac pacemaker and
other electrical stimulators. And keep away from MRI, CT etc. It may cause burning and
adversely affect MRI measurement.
zSpO2measurements may be adversely affected in the presence of high ambient light.
Shield the sensor area (with a surgical towel, or direct sunlight, for example) if
necessary.
zThe device should not be used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the device should be observed carefully to verify normal
operation.
zThe use of accessories, sensors, and cables other than those listed in this manual may
result in increased emission and /or decreased immunity of this device.
zDO NOT USE THE OXIMETER FOR TREATMENT, we are not responsible for
contretemps happened during measuring.