manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. ChoiceMMed
  6. •
  7. Medical Equipment
  8. •
  9. ChoiceMMed MD300CF3 User manual

ChoiceMMed MD300CF3 User manual

Fingertip
Pulse Oximeter
USER MANUAL
Ver2.0CF3
General Description
Oxygen Saturation is a percentage of Oxyhemoglobin (HbO2) capacity, compounded with oxygen, by all combinative hemoglobin
(Hb) capacity in blood. In other words, it is consistency of Oxyhemoglobin in blood. It is a very important parameter for the
Respiratory Circulation System. Many respiratory diseases can result in oxygen saturation being lowered in human blood.
Additionally, the following factors can reduce oxygen saturation: Automatic regulation of organ dysfunction caused by Anesthesia,
Intensive Postoperative Trauma, injuries caused by some medical examinations. That situation might result in light-headedness,
asthenia, and vomiting. Therefore, it is very important to know the oxygen saturation of a patient so that doctors can find problems
in a timely manner.
The fingertip pulse oximeter features low power consumption, convenient operation and portability. Place one fingertip into the
photoelectric sensor for diagnosis and the pulse rate and oxygen saturation will appear on the display. It has been proven in clinical
experiments that it also features high precisionand repeatability.
Measurement Principle
Principle of the oximeter is as follows: A mathematical formula is established making use of Lambert Beer Law according to
Spectrum Absorption Characteristics of Reductive hemoglobin(RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared zones.
Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity
Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm glow and 940nm near
infrared light) can be focused onto a human nail tip through a clamping finger-type sensor. A measured signal obtained by a
photosensitive element, will be shown on the oximeter’s display through process in electronic circuits and microprocessor.
Diagram of Operation Principle
1. Red and Infrared-ray Emission Tube
2. Red and Infrared-ray Receipt Tube
Precautions For Use
1 Before use, carefully read the manual.
2 Operationof the fingertip pulse oximeter may be affected by the use of an electrosurgical unit (ESU).
3 The fingertip pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify
that nothing is hindering the pulse measurement before relying on the SpO2 measurement.
4 Do not use the fingertip pulse oximeter in an MRI or CT environment.
5 Do not use the fingertip pulse oximeter in situations where alarms are required. The device has no alarms. It is not for
continuous monitoring.
6 Do not use the fingertip pulse oximeter in an explosive atmosphere.
7 The fingertip pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunctionwith other
methods of assessing clinical signs and symptoms.
8 Check the pulse oximeter sensor application site every 4 hours to determine the positioning of the sensor and circulation
and skin sensitivity of the patient.
9 Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device in liquid. The device is not
intended for sterilization.
10 Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components,
including batteries.
11 This equipment complies with IEC 60601-1-2:2007 for electromagnetic compatibility for medical electrical equipment and/or
systems. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical
noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or
strength of a source might disrupt the performance of this device.
12 Portable and mobile RF communicationsequipment can affect medical electrical equipment.
Rx only: “Caution: Federal law restricts this device to sale by or on the order of a physician.”
Inaccurate measurements may be caused by
1 Significant levels of dysfunctional hemoglobin(such as carbonyl - hemoglobinor methemoglobin);
2 Intravascular dyes such as indocyanine green or methylene blue;
3 High ambient light. Shield the sensor area if necessary;
4 Excessive patient movement;
5 High-frequency electrosurgical interference and defibrillators;
6 Venous pulsations;
7 Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line;
8 The patient has hypotension, severe vasoconstriction,severe anemia, or hypothermia;
9 The patient is in cardiac arrest or is in shock;
10 Fingernail polishor false fingernails;
11 Weak pulse quality (low perfusion);
12 Low hemoglobin;
Product Properties
1 Operationof the product is simple and convenient.
2 The product is small in volume, light in weight and convenient to carry.
3 Power consumptionof the product is low and the two AAA batteries can be operated continuously for 30 hours.
4 A low voltage warning will be indicated when battery voltage is low and normal operation of the oximeter might be influenced.
5 The product will automatically power off when there is no signal for longer than 8 seconds.
Intended Use
The fingertip pulse oximeter is a portable non-invasive device intended for spot checking of arterial hemoglobin oxygen saturation
(SpO2) and pulse rate of adult and pediatric patient at home and hospital (including clinical use in internist/surgery, anesthesia and
intensive care units). It is not for continuous monitoring.
Battery Installation
1. Lift the battery door cover.
2. Install two AAA batteries into the battery compartment.
Match the plus (+) and minus (-) signs in the compartment.
If the polarities are not matched, damage may be caused
to the oximeter.
3. Close the battery door cover.
Notes:
Install the batteries with the correct polarity. Incorrect placement may cause damage to the bracket.
Please remove the batteries if the pulse oximeter will not be used for long periods of time.
Using the Lanyard
1. Thread thinner end of the lanyard through the hanging hole.
2. Thread thicker end of the lanyard through the threaded end before pulling it tightly.
Warnings!
1. Keep the oximeter away from young children. Small items suchas the battery door, battery, and
lanyard are choking hazards.
2. Do not hang the lanyard from the device’s electrical wire.
Product Accessories
1. One lanyard
2. Two AAA batteries
3. One user manual
Operation Instructions
1 Install two AAAbatteries according to the Battery Installationinstructions.)
2 Open the clamp as illustrated in the picture below.
3 Fullyinsert one fingertip into the silicone hole of the oximeter before releasing the clamp.
4 Press the switch button once on front panel.
5 Keep your finger still during measurement.
6 Read corresponding data from display screen.
7 Press the button again to toggle between six display modes.
After turning on the Oximeter, each time you press the power switch, the Oximeter will switch to another display mode.
There are 6 display modes shown as follows:
1.
2.
3.
4.
5.
6.
Holding the power switch for longer than one second, will adjust the brightness of the oximeter. There are 10 levels of brightness.
The default level is level four.
8 PR tone modulation function
9 Warning indication:
Indicationrange: SpO2: <90%PR Rate: <60 bpm or >100bpm
Indicationmode:Audible and Visible
Indicationsilence:Whenthe audio indication occurs, the indication sound can pause for 30 seconds by pressing Power button
and there is only the visual indication. By pressing the power button again, the audio indication will resume.
NOTE: This function is only valuable for audio indication, but not for pulse rate sound.
Front Panel
Patient pulse quality signals areindicated by bar graph. The bar is graded as 10 levels, if the strengthis level 2 to 3, the pulse signal
is inadequate.
Maintenance and Storage
1. Replace the batteries in a timely manner when low voltage lamp is lighted.
2. Clean surface of the fingertip oximeter before it is used in diagnosis for patients.
3. Remove the batteries if the oximeter is not operated for a long time.
4. It is best to store the product in -20℃~+55℃and ≤93%
humidity
.
5. Keep in a dry place. Extreme moisture may affect oximeter lifetime and may cause damage.
6. Dispose of battery properly; follow any applicablelocal battery disposal laws.
Cleaning the fingertip pulse oximeter
Please use medical alcohol to clean the silicone touching the finger inside of oximeter with a soft cloth dampened with 70%
isopropyl alcohol.Also clean the being tested finger using alcohol before and after each test.
Do not pour or spray liquids ontothe oximeter, and do not allow any liquid to enter any openings in the device. Allow the oximeter to
dry thoroughly before reuse.
A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor. Clinical testing is used to establish
the SpO2 accuracy. The measured arterial hemoglobin saturation value (SpO2) of the sensors is compared to arterial hemoglobin
oxygen(SaO2) value, determinedfrom blood samples with a laboratory CO-oximeter. The accuracy of the sensors in comparisonto
the CO-oximeter samples measured over the SpO2 range of 70 –100%. Accuracy data is calculated using the
root-mean-squared(Arms value)for all subjects, per ISO 9919:2005,Medical Electrical
Equipment
–
Particular
requirements for the
basic safety and essential performance of pulse oximeter equipment for medical use.
The use life of the device is five years when it is used for 15 measurements every day and 10 minutes per one
measurement. Stop using and contact local service center if one of the following cases occurs:
An error in the Possible Problems and solutions is displayed on screen.
The oximeter cannot be powered on in any case and not the reasons of battery.
There is a crack on the oximeter or damage on the display resulting readings cannot be identified; the spring is invalid; or the
key is unresponsive or unavailable.
Specifications
1. Display Type
OLED display
2. SpO2
Display range: 0-99%
Measurement range: 70-99%
Accuracy: 70%-99%:
±
3%;
0%~69% no definition
Resolution: 1%
3. Pulse Rate
Display range: 0~254BPM
Measure range: 30-235 BPM
Accuracy: 30~99bpm, ±2bpm; 100~235bpm,
±
2%
Resolution: 1BPM
4. Probe LED Specifications
Wavelength
Radiant Power
RED
660
±
2nm
1.8mW
IR
940
±
10nm
2.0mW
5. Power Requirements
TwoAAA alkaline Batteries
Power consumption: Less than 40mA
Low
power
indication:
Battery Life: TwoAAA1.5V, 600mAh alkaline batteries could be continuously operated as long as 30 hours.
6. Dimension
Length: 60mm
Width: 32mm
Guidance and Manufacturer’s declaration - electromagnetic emission
The MD300CF3 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or
the user of MD300CF3 Pulse Oximeter should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic Environment – guidance
RF emissions CISPR 11
Group 1
The MD300CF3 Pulse Oximeter uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Guidance and Manufacturer’s declaration - electromagnetic immunity
The MD300CF3 Pulse Oximeter is intended for use in the electromagnetic environment specified below.
The customer or the user of the MD300CF3 Pulse Oximeter should assure that it is used in such an environment.
Immunity test IEC 60601 test level
Electromagnetic Environment – guidance
Electrostatic Discharge
(ESD) IEC 61000-4-2 +/- 6kV
+/- 8kV +/- 6kV
+/- 8kV
Floors should be wood, concrete or ceramic tile. If
floor are covered with synthetic material, the relative
humidity should be at least 30%.
Electrostatic transient /
burst IEC 61000-4-4
±2kV for power supply
lines
±1kV for input/output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-4-5
1 kV differential mode
2 kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips,
interruptions
voltage variations
power supply
lines IEC
< 5 % UT (>95 % dip
UT) for 0.5
40 % UT (60 % dip
UT) for 5
70 % UT (30 % dip
UT) for 25
< 5 % UT (>95 % dip
UT) for 5
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the MD300CF3 Pulse Oximeter requires continued
operation during power mains interruptions, it is
recommended that the MD300CF3 Pulse Oximeter
be powered from an uninterruptible power supply or
a battery.
Power
(50/60 Hz)
field IEC
Power frequency magnetic fields should be at levels
characteristics of a typical location in a typical
commercial or hospital environment.
NOTE UTis the a. c. mains voltage prior to applicationof the test level.
Symbol Definition Symbol Definition Symbol Definition
Type BF applied part.
Low power indication Manufacturer’s
information
SN
Serial No.
Date of Manufacture
Protected against
dripping water.
SpO2﹪
Oxygen saturation
PR bpm
Pulse rate (BPM)
No SpO2 Alarm
Storage temperature
and relative humidity
European union
approval
Attention, consult
accompanying
documents.
Authorized
representative in the
European community
Guidance and Manufacturer’s declaration - electromagnetic immunity
The MD300CF3 Pulse Oximeter is intended for use in the electromagnetic environment specified below.
The customer or the user of the MD300CF3 Pulse Oximeter should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Electromagnetic Environment – guidance
Conducted
RF IEC
61000-4-6
Radiated
RF IEC
61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
N/A
3 V/m
Portable and mobile RF communications equipment should be used no closer to
any part of the MD300CF3 Pulse Oximeter, including cables, than the
recommended separation distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHzto 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
Problems
Possible
Solution
SpO2 or PR can not
be shown normally
1. Finger is not inserted correctly
2. Patient’s SpO2 value is too low to be measured
1. Retry by inserting the finger
2. There is excessive illumination
3. Try some more times. If you can make sure no
problem is existing in the product, please go
to a hospital timely for exact diagnosis.
SpO
2
or PR is shown
unstably
1. Finger might not be inserted deep enough.
2. Excessive patient movement
1. Retry by inserting the finger
2. Be calmness
The oximeter can not
be powered on
1. No battery or low power of battery
2. Batteries might be installedincorrectly
3. The oximeter might be damaged
1. Please replace batteries
2. Please reinstall the batteries
3. Please contact with local customer service
centre
Indicationlamps are
suddenly off
1. The product is automatically powered off when no
signal is detected longer than 8 seconds
2. The battery power is too low to work
1. Normal
2. Replace the batteries
“Error3” or
“Error4” is displayed
on screen
1. Low power
2. Receiving tube being shielded or damaged together
with broken connector.
3. Mechanical Misplace for receive-emissiontube
4. Amp circuit malfunctions.
1. Change batteries
2. Please contact local customer service center
3. Please contact local customer service center
4. Please contact local customer service center
Error 6
Err 6 means the screen is failure
Change the screen
“Error7” is displayed
on screen
Err 7 means all the emissionLED or receptiondioxide is
damaged.
Check the emissionLED and reception dioxide.
Height: 34mm
Weight: 60g (including two AAA batteries)
Note: The dimensions of each device may differ slightly from the size that mentioned above.
7. Environment Requirements
OperationTemperature: 5~40℃
Storage Temperature: -20~+55℃
Ambient Humidity: ≤80% in operation
≤93%in storage
8. Equipment Response Time
As shown in the following figure.
Response time of slower average is 12.4s.
9. Interference Resistance Capacity against Ambient Light:
Device works normally when mixed noise produced by BIO-TEK INDEX Pulse Oximeter tester.
10. Classification
According to the type of protectionagainst electric shock: INTERNALLYPOWEREDEQUIPMENT;
According to the degree of protection against electric shock: TYPE BF APPLIED PART;
According to the method(s) of sterilization or disinfection recommended by the manufacturer: Equipment with method(s) of
sterilizationor disinfection recommended by the manufacturer;
According to the degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR or WITH
OXYGEN OR NITROUS OXIDE: EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANAESTHETIC MIXTURE.
According to the degree of protection against ingress of water: IPX1
According to the mode of operation: CONTINUOUSOPERATION
Declaration
Guidance and Manufacturer’s declaration – electromagnetic emissions-For all EQUIPMENT and SYSTEMS
site surveya, should be less than the compliance level ineach frequency range. b
Interference may occur in the vicinity of equipment marked with following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations, Electromagneticpropagationis affected by absorption and reflection
structures, objects and people.
a Field strengths from fixed transmitters, such as base station for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Pulse Oximeter (MD300CF3) should be observed to verify normal operation.
If abnormal performance is observed, additional measurements may be necessary, such as reorienting of the relocating the
Pulse Oximeter (MD300CF3).
b Over the frequency range 150 kHzto 80 MHz, fields strengths should be less than 3 V/m
Recommended separation distances between portable and mobile RF communications equipment and
the EQUIPMENT or SYSTEMS - For all EQUIPMENTand SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances
between
portable and mobile RF communications equipment and the MD300CF3 Pulse
Oximeter
The MD300CF3 Pulse Oximeter is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the MD300CF3 Pulse Oximeter can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
MD300CF3 Pulse Oximeter as recommended below, according to the maximum output power of the communications
equipment
Rated maximum output of
transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHzto 2.5 GHz
0.01
/
0.12
0.23
0.1
/
0.38
0.73
1
/
1.2
2.3
10
/
3.8
7.3
100
/
12
23
For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separationdistance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Possible Problems and Solutions
Guidance and Manufacturer’s declaration – electromagnetic immunity-For all EQUIPMENTand SYSTEMS
Symbol Definitions
Guidance and Manufacturer’s declaration – electromagnetic immunity
–
For EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING
Applicable Models
MD300CF3 MD300CF31 MD300CF32 MD300CF33 MD300CF34 MD300CF35 MD300CF36 MD300CF37
MD300CF38 MD300CF39 MD300CF30 MD300CF3A MD300CF3B MD300CF3C MD300CF3D MD300CF3E
MD300CF3F MD300CF3G MD300CF3H
Note: The illustrations used in this manual may differ slightly from the appearance of the actual product.
ALL RIGHTS
RESERVED
Issue Date: 28/June
2011

Other ChoiceMMed Medical Equipment manuals

ChoiceMMed MD300I User manual

ChoiceMMed

ChoiceMMed MD300I User manual

ChoiceMMed MD300C228 User manual

ChoiceMMed

ChoiceMMed MD300C228 User manual

ChoiceMMed MD300C312 User manual

ChoiceMMed

ChoiceMMed MD300C312 User manual

ChoiceMMed MD100C Pro User manual

ChoiceMMed

ChoiceMMed MD100C Pro User manual

ChoiceMMed MD300C5 User manual

ChoiceMMed

ChoiceMMed MD300C5 User manual

ChoiceMMed MD300K User manual

ChoiceMMed

ChoiceMMed MD300K User manual

ChoiceMMed LungBoost MD8000 User manual

ChoiceMMed

ChoiceMMed LungBoost MD8000 User manual

ChoiceMMed Fingertip Pulse Oximeter User manual

ChoiceMMed

ChoiceMMed Fingertip Pulse Oximeter User manual

ChoiceMMed OxyWatch C20 User manual

ChoiceMMed

ChoiceMMed OxyWatch C20 User manual

ChoiceMMed MD300W User manual

ChoiceMMed

ChoiceMMed MD300W User manual

ChoiceMMed MDTS100 User manual

ChoiceMMed

ChoiceMMed MDTS100 User manual

ChoiceMMed MD300-D User manual

ChoiceMMed

ChoiceMMed MD300-D User manual

ChoiceMMed Fingertip Pulse Oximeter User manual

ChoiceMMed

ChoiceMMed Fingertip Pulse Oximeter User manual

ChoiceMMed MD300C2ES User manual

ChoiceMMed

ChoiceMMed MD300C2ES User manual

ChoiceMMed MD300C318 User manual

ChoiceMMed

ChoiceMMed MD300C318 User manual

ChoiceMMed MD300 User manual

ChoiceMMed

ChoiceMMed MD300 User manual

ChoiceMMed MD300C2 User manual

ChoiceMMed

ChoiceMMed MD300C2 User manual

Popular Medical Equipment manuals by other brands

Orthofix Oscar Instruction leaflet

Orthofix

Orthofix Oscar Instruction leaflet

Alltion AC-2000 Series user manual

Alltion

Alltion AC-2000 Series user manual

SENZIME TetraGraph Philips Interface operating instructions

SENZIME

SENZIME TetraGraph Philips Interface operating instructions

natus Photic Stimulator User & service manual

natus

natus Photic Stimulator User & service manual

Respirex MFC WR0219 product manual

Respirex

Respirex MFC WR0219 product manual

THERMODx Warm Blocks 360 user manual

THERMODx

THERMODx Warm Blocks 360 user manual

3D Systems Touch user guide

3D Systems

3D Systems Touch user guide

Johnson & Johnson ETHICON Megadyne quick start guide

Johnson & Johnson

Johnson & Johnson ETHICON Megadyne quick start guide

Optimum Medical Ugo Fix 3006 user guide

Optimum Medical

Optimum Medical Ugo Fix 3006 user guide

Portascope PVSM6150 instruction manual

Portascope

Portascope PVSM6150 instruction manual

FujiFilm CR-IR341 Service manual

FujiFilm

FujiFilm CR-IR341 Service manual

Mizuho SMART FIX quick start guide

Mizuho

Mizuho SMART FIX quick start guide

Roche CoaguChek XS Pro Operator's manual

Roche

Roche CoaguChek XS Pro Operator's manual

OAKWORKS Clodagh Gemini user manual

OAKWORKS

OAKWORKS Clodagh Gemini user manual

Orliman MANUTEC MF-50 Instructions for use

Orliman

Orliman MANUTEC MF-50 Instructions for use

PHILIPS Respironics Wisp SE manual

PHILIPS Respironics

PHILIPS Respironics Wisp SE manual

Thomashilfen Thevo-Adapt user manual

Thomashilfen

Thomashilfen Thevo-Adapt user manual

natus Camino ICP user manual

natus

natus Camino ICP user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.