manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Co2meter
  6. •
  7. Measuring Instrument
  8. •
  9. Co2meter TIM-12 User manual

Co2meter TIM-12 User manual

1

OPERATIONMANUAL
DesktopCO2,Temperature&
RelativeHumidityMonitor 
TIM‐12



INTRODUCTION
ThankyouforpurchasingtheTIM12desktop
CO2monitor.Itmeasurescarbondioxidelevels,
airtemperatureandhumidity.Thismakesit
idealforindoorairquality(IAQ)diagnosisand
HVACsystemperformanceverification.

Poorindoorairqualityisconsideredunhealthy.
Itcanleadtodrowsiness,inabilityto
concentrate,andevenillness.

IAQmonitoringofCO2levelsiswidelyusedin
publicareassuchasoffices,classrooms,
factories,hospitalsandhotels.Itisalso
recommendedforindustrialhygieneinsome
countries.(Seeappendix).
ProductFeatures:
●LargedisplayshowsCO2level,temperature,
  humidity,dateandtime
●15degreetiltanglemakesthedisplayeasy
toseeandread
●Indoorairqualitylevelindication:
  Good,Normal,andPoor
●StableNDIRsensorforCO2detection
●VisibleandaudibleCO2warningalarm
●ABC(AutomaticBaselineCalibration)
  ormanualCO2calibration
●MaximumorminimumCO2levelrecall
●Samplingratefrom1sec.to4:59:59.

MATERIALSSUPPLIED
Thispackagecontains:
●Meter
●PowerSupply(5V+/‐10%,>=0.5A)
●USBcableandsoftwareCO
●Manual
POWERSUPPLY
ThemeterispoweredbyanACadaptor
(5V/0.5Aoutput).

LCDDISPLAY







Symbols
ppmCO2levelinparts‐per‐million
GOODCO2isingoodlevel
NORMALCO2isinnormallevel
POORCO2isinpoorlevel
AirTemp.Airtemperature
Humidity%Unitofairrelativehumidity
oE(C/F)Celsius/Fahrenheitoftemp.
MAX/MINMaximum/Minimumreadings
CO2alarmicon

NOTE: 
TWA/STEL/VENT.R/arevainiconsonthis
model.


KEYPAD
SET‐Entersetupmode
‐Saveandfinishsettings
ESC‐Exitsetuppage/mode
‐Terminatecalibration/datalogging
RESET‐ClearMAX/MINvalues
LOG▲‐Selectmode,increasevalueatsetup
‐Startdatalogging
MIN/MAX▼ ‐ActivateMAX,MINfunction
‐Selectmodeordecreasevalue
duringsetup
SET+▲+MIN/MAX▼ ‐CO2calibrationmode

OPERATION
POWERON/OFF
Plugthewalladaptorintoanappropriate
receptacle.Themeterwillturnonautomatically
afterashortbeep.ThelitLEDindicatesfull
power.Ifthevoltageistoohighorlow,“bAt”
willdisplayontheLCDandtheLEDwillflash.
Seepage14fortroubleshooting.

Duringnormaloperation,theLCDwilldisplay
thecurrentCO2levelinppm,airqualitylevel,
airtemperature,humidity,andthecurrentdate
andtime(Fig.1).






TAKINGMEASUREMENTS
Themeterstartstakingmeasurements
immediatelyafteritisturnedonandupdates
thereadingseverysecond.Intheeventthe
operatingenvironmentisquicklychanged(for
example,fromhightolowtemperature),itwill
takeapproximately2minutesfortheCO2and
temperaturetoadjustand10minutesfor
humiditytostabilize.
NOTE:Donotholdthemeterclosetoyourface
asyourbreathwillaffecttheCO2level.

MIN/MAX
Whilethemeterispoweredon,pressthe
“MIN/MAX▼“buttontoseetheminimumand
maximumCO2level,temperature,and
humidityinturn.Eachpressof“MIN/MAX“will
displaytheMINandMAXvaluesinorder,then
returnthemetertonormalmode(Fig.2).

Duringnormaloperationorwhileviewing
MIN/MAXlevels,pressandholdthe
“RESET“keyformorethan1secondtoclearthe
storedminimumandmaximumvalues.









INTERNALDATALOGGING
Press“LOG▲”for2secondsundernormalmode
tostartlogging.ThegreenLEDblinkstoindicate
theloggingstatusandtheLCDshowsthereal
timeCO2valueand“rEC”.Lowerdisplaysare
therealtimetemperature,humidityandclock.

Toterminatedatalogging,press“ESC”for2
seconds,TheLEDlightwillstopblinkingandthe
displaywillshow“End”andCO2reading
interchangeably.Holddown“ESC”for2
secondsagain,anditwillreturntonormal
measurementmode. 

ThemetercansaveCO2,temperatureand
humiditylevelsovertimeininternalmemory.
Thememorycapacityis15,999points.The
samplingratecanbefrom1secondto4hours
59minutesand59seconds(SeeSETUP).
MinimumandMaximumrecallisstillavailable
duringlogging.





HIGHCO2ALARM
2

ALARM
Themeterfeaturesanaudiblealarmthatwarns
theuserwhentheCO2concentrationexceeds
theprogrammedlimit.Userscanset2limits:a
CO2levelthatrequiresventilation(SeeP1.3in
setupforsettingalarmlimits)andalowerCO2
levelwhenventilationisnolongerrequired
(SeeP1.2insetupCO2normallimit).

Duringhighalarm,themeteremitsan80dB
beepanddisplaysthefaniconontheLCD.The
alarmcanbetemporarilystoppedbypressing
anybutton,oritwillautomaticallystopwhen
theCO2levelfallsbelowthelowerlimit. 

Ifthealarmistemporarilysilenced,itwill
resumeiftheCO2levelfallbelowthelower
limitthenrisesabovetheupperlimitagain,orif
the“RESET“buttonispressedtore‐activatethe
alarm.

Thefaniconremainsflashingevenifthebeeps
aresilenced.Theicononlystopsflashingwhen
theCO2levelfallbelowthelowerlimit(Fig.3)








SETUP
Thereare5setupmodes:P1‐CO2alarmlevels,
P2‐temperature(CorF),P3‐date/time,P4‐
factoryresetandP5‐Logging.Whenyoupress
“SET”thesetupmodewillbeshowninthe
lowerrightcornerofthescreen.

Toentersetup,whileinnormalmode,hold
downthe“SET“buttonfor3seconds.Toexit
setup,press“ESC“whileP1.0,P2.0,P3.0,P4.0or
P5.0isdisplayed.

P1.1SETCO2GOODLEVEL
Press“SET“todisplayP1.0and”CO2”(Fig.4)on
theLCD.Press“SET“againtosettheCO2GOOD
level.Thecurrentsetvaluewillbeblinkingon
LCD(Fig.5).










Press“▲“toincreaseor“MIN/MAX▼“to
decreasethevalue.TheGOODalarmrangeis
from0to700ppmin100ppmincrements.
OncethepreferredCO2levelisselected,
press”SET“tocontinueontoP1.2tosetthe
NORMALlevel(Fig.6)orpress“ESC“toreturn
toP1.0withoutsavingchanges.

P1.2SETCO2NORMALLEVEL
P1.2istheupperlimitoftheCO2NORMAL
level.ThecurrentvaluewillflashontheLCD
(Fig.6).Press“▲“toincreaseor“MIN/MAX▼”
todecreasethevalue.TheNORMALlevelcan
besetbetween700and1,000ppm.Oncethe
preferredvalueisselected,press”SET“tosave
P1.2andcontinueontosettheP1.3POOR+
ALARMthresholdsetting(Fig.7)orpress
“ESC“toreturntoP1.0withoutsavingchanges.

P1.3SETCO2POORLEVEL+ALARM
P1.3isusedtosettheCO2alarmthreshold.The
currentsetvaluewillblinkontheLCD(Fig.7).
Press“▲“toincreaseor“MIN/MAX▼“to
decreasethevalue.ThePOORlevelcanbeset
between1,000to5,000ppm.Press“SET“to
confirmP1.3andcontinuetoP1.4(Fig.8)or
press“ESC“withoutsavingtoreturntoP1.0








P1.4SETALARMON/OFF
P1.4isusedtosetalarmonandoff.When
enteringP1.4,thedefault“On“willbeblinking
ontheLCD(Fig.8). 

Press“▲“or“MIN/MAX▼“toswitchthealarm
to“Off“.Thenpress“SET“toconfirmP1.4and
returntoP1.0(Fig.4)orpress“ESC“toreturnto
P1.0withoutsaving.

CAUTION:
AlthoughtheTIM12alarmcanbesetupto
5,000ppm,theaccuracyoftheCO2readingis
onlyspecifiedupto2,000ppm.Alarmvalues
higherthan2,000ppmareforreferenceonly
andshouldnotbeusedforpersonalsafety.

P2.0SETTEMPERATUREUNITS
Holddownthe“SET”buttonforatleast3
secondstoentersetupmode,thenpress“▲“in
P1.0modetoaccessP2.0tosetthetemperature
scale(Fig.9).Press“SET“againtoshow
temperatureinoCoroFonthedisplay(Fig.10).
Press“▲“or“MIN/MAX▼“tochangeunits,
thenpress“SET“tosaveor“ESC”toexit
withoutsavingandreturntoP2.0







P3.0SETREALTIMECLOCK
TheTIM10candisplaytimeineither24houror
12hour(am/pm)format.Holddownthe“SET”
buttonforatleast3secondstoentersetup
mode,thenpress“▲“twicetoaccesstheP3.0
real‐timeclock(rtc)setup(Fig.11).

Tostart,press“SET“todisplayP3.1with“rtc”
andtheblinkingyeardigits(Fig.12)inthelower
leftcornerofthedisplay.Tochangetheyear,
press“▲“or“MIN/MAX▼“tochangethevalue,
thenpress“SET“tosavetheyearandcontinue,
orpress“ESC“toexitwithoutsaving. 







Continuetopress“SET“tochangetothe
month,andthentheday.Afteryoupress“SET”,
press“▲“or“MIN/MAX▼”whiletheyear,
monthordayvaluesareflashingtosetthe
correctdate.Press“SET“tosaveandmove
forwardtosetthetime,orpress“ESC“to
returntoP3.0modewithoutsaving.

Tosettheclock,inP3.4modewiththehours
flashing,press“▲“or“MIN/MAX▼“tochange
from12to24hourtimeformat(Fig11).Press
“SET“tosaveandmoveforwardtoP3.5mode
tosetthetime,orpress“ESC“toreturntoP3.0
modewithoutsaving.







WhileinP3.5mode,thecurrenthoursetting
willblink(Fig.14).Tochangethesetting,
press“▲“or“MIN/MAX▼“toadjustthetime.
Press“SET“tosavethesettingorpress“ESC
“toreturntoP3.0.modewithoutsaving.

P4.0RESET
Holddownthe“SET”buttonforatleast3
secondstoentersetupmode,thenpress“▲“3
timestoaccessP4.0modetorevertthemeter
tothefactorydefaultstate(Fig.15).
Press“SET“togotoP4.1mode.“No”willbe
3

blinkingonthedisplay(Fig.16).Press“▲”
or”MIN/MAX▼”tochangeto“Yes”,thenpress
“SET“tosavethesetting.Press“ESC”toreturn
toP4.0Cwithoutsaving.







Ifyouchoose“Yes”,themeterisresettothe
followingfactorydefaults:






P5.0SETSAMPLINGRATE
GointoP5.0tosetthedataloggingsampling
rate(Fig.17).Therangeisfrom1secondto4
hours59minutesand59seconds. 

Press“SET”toseeP5.1blinkinghourdigitson
thelowerleftdisplay.Tochangethedigit,press
“▲“or“MIN/MAX▼“toadjustthehour
samplingrate.Press“SET“tosavethesetting. 

Afterthehoursareset,youwillseeP5.2and
theminutesdigitswillflash.Tochangethe
digits,press“▲“or“MIN/MAX▼“Press“SET“to
savethesettingandgotoP5.3.

Aftertheminutesareset,youwillseeP5.3and
thesecondsdigitswillflash(Fig.18).Tochange
thedigits,press“▲“or“MIN/MAX▼“Press
“SET“tosavethesettingandreturntoP5.0.

Press“ESC“atanytimetoexitwithoutsaving
yourlastchangesandreturntoP5.0.


CO2CALIBRATION
Themeteriscalibratedat400ppm(parts‐per‐
million)CO2concentrationatthefactory.400
ppmiscomparabletofreshoutdoorair. 
Thereare2typesofcalibrationavailable:ABC
(AutomaticBaselineCalibration)ormanual
calibration.Oneortheothershouldbeused
regularlytomaintaingoodaccuracy.


Note:
IfTIM10accuracybecomesaconcernafterlong
useorotherspecialconditions,returnthe
metertothedealerforcalibration.

CAUTION:
Donotcalibratethemeterwithairorbottled
gasthatcontainsanunknownCO2level.The
meterwillassumeitis400ppmCO2andwill
giveinaccuratemeasurements.

ABCCalibration
ABC(AutomaticBaselineCalibration)is
designedtoeliminatethezerodriftovertimeof
theinfraredCO2sensor.TheABCfunctionis
enabledbydefaultfromthefactory.

ForABCtoworkproperly,themetermust
continuouslymonitorCO2levelsforatleast7.5
days(poweron).Duringthattime,itwillrecord
thelowestCO2levelandsetthatasthe400
ppmbaselineforfuturereadings.Therefore,it
iscriticalwhenusingABCcalibrationthatthe
meterbeexposedtofreshairatleastonce
every7.5days.Ifthisisnotpossible,then
manualcalibrationisrecommended.

ManualCalibration
Manualcalibrationrequiresfresh,dryoutdoor
airwhenCO2levelsareapproximately400
ppm.Donotcalibrateduringrainydays.High
humiditywillaffecttheCO2levelsinair. 

Duringmanualcalibration,avoidplacingthe
meterinareascrowdedwithpeopleorcloseto
structuresthatmightgiveoffCO2likeair
ventilationpipes,chimneys,orautomobiles.

Tomanuallycalibratethemeter,placeit
outdoors,andturnitonforatleast2minutes.
Thenholddown“SET“,"▲"and“MIN/MAX▼”
simultaneouslymorethan1secondtoenter
CO2calibrationmode.400ppmand“CO2”will
blinkduringcalibration.






Waitabout30minutesuntiltheblinkingstops.
Calibrationiscompletedautomatically.Tostop
manualcalibration,press“RESET“foratleast1
second.

Note:
Keepthemeterawayfromanyanimals,
humansorplantswhichmightaffecttheCO2
concentrationnearbyduringthecalibration.

RHCALIBRATION
ThishumidityaccuracyofthisdesktopCO2
meterissetatthefactory,andcannotbere‐
calibrated.


TROUBLESHOOTING
Cannotpoweron
Checkwhethertheadaptorhaspowerand
thatthemeterisfirmlypluggedin.
Slowresponse
Checkthattheairflowchannelson
therearofthemeterarenotblocked.
Nochangeinreadings
Checkwhetherthemeterisinmaximum
orminimummode.Resetfactorydefaults.
"BAt”andgreenLEDkeepflashing
Theadaptoroutputvoltageistoohighor
toolow.Usethesuppliedadaptorwith 
correct5V(+10%)>=0.5Aratings.

ErrorCodes






















SPECIFICATIONS












SUPPORT&WARRANTY


4

Support
Thequickestwaytoobtaintechnicalsupportis
viaemail.Pleasesendallsupportenquiresto
support@co2meter.com.Inyouremail,please
includeaclear,concisedefinitionofthe
problemandanyrelevanttroubleshooting
informationorstepstakensofar,sowecan
duplicatetheproblemandquicklyrespondto
yourinquiry.

Warranty
Thisunitcomeswitha1YEAR(warrantyperiod)
limitedmanufacturer’swarranty,startingfrom
thedatetheunitwasshippedtothebuyer.
Duringthisperiodoftime,CO2Meter.com
warrantsourproductstobefreefromdefects
inmaterialsandworkmanshipwhenusedfor
theirintendedpurposeandagreestofixor
replace(atourdiscretion)anypartorproduct
thatfailsundernormaluse. 
Totakeadvantageofthiswarranty,theproduct
mustbereturnedtoCO2Meter.comatyour
expense.If,afterexamination,wedetermine
theproductisdefective,wewillrepairor
replaceitatnoadditionalcosttoyou.

Thiswarrantydoesnotcoveranyproductsthat
havebeensubjectedtomisuse,neglect,
accident,modificationsorrepairsbyyouorbya
thirdparty.Noemployeeorresellerof
CO2Meter.com’sproductsmayalterthis
warrantyverballyorinwriting.

Returns
Iftheproductfailsundernormaluseduringthe
warrantyperiod,anRMA(ReturnMaterial
Authorization)numbermustbeobtainedfrom
CO2Meter.com.Aftertheitemisreceived,
CO2Meter.comwillrepairorreplacetheitemat
ourdiscretion.

ToobtainanRMAnumber,pleasecall
CO2Meter.comat(877)678‐4259TollFreeor
(386)310‐4933.WhenrequestinganRMA
number,pleaseprovidethereasonforreturn
andoriginalordernumber.

Ifwedeterminethattheproductfaileddueto
improperuse(waterdamage,dropping,
tampering,electricaldamageetc.)orabuse,or
ifitisbeyondthewarrantyperiod,wewill
informyouofthecosttofixorreplaceyour
device.

Ifyouarereturningyourdeviceduetoa
warrantyclaim(withanRMAnumber)andyou
stillhavetheunitoriginalpackage,pleaseuseit
toshipyourunittous.Pleasemakesureto
includetheprovidedRMAnumberonthe
outsideofthebox,preferablyontheshipping
label.Makesureyousecuretheunitinsidethe
packageproperlytopreventanydamageduring
transitthatcouldvoidyourdevice’swarranty.
Finally,pleaseshipyourdevicetotheaddress
shownunderthe“ContactUs”sectionbelow.
CO2Meterwillnot,underanycircumstances,be
responsibleforyourshipmentexpensesandno
refundwillbeissuedforshippingcharges
necessaryforyoutoshiptheunittous.




Liability
Allliabilitiesunderthisagreementshallbe
limitedtotheactualcostoftheproductpaidto
CO2Meter.com.Innoeventshallwebeliable
foranyincidentalorconsequentialdamages,
lostprofits,lossoftime,lostsalesorlossor
damagetodata,injurytopersonorpersonal
propertyoranyotherindirectdamagesasthe
resultofuseofourproducts.

Contactus:We’reheretohelp!
Ifthetroubleshootingguideabovedoesn’thelp
yousolvingyourproblemorformore
information,pleasecontactususingthe
informationbelow.

Support@CO2Meter.com
(386)256‐4910TechnicalSupport
(386)872‐7665Sales 
Monday‐Friday9:00am‐6:00pmEST

CO2Meter,Inc. 
131BusinessCenterDrive
OrmondBeach,FL32174USA
www.CO2Meter.com







































Appendix
CO2LEVELSANDGUIDELINES
Non‐EnforcedReferencelevels
NIOSHrecommendations
250‐350ppm:normaloutdoorambient
concentrations
600ppm:minimalairqualitycomplaints
600‐1000ppm:lessclearlyinterpreted
1000ppm:indicatesinadequateventilation;
complaintssuchasheadaches,
fatigue,andeye/throatirritation
willbemorewidespread.1,000ppm
shouldbeusedasanupperlimit
forindoorlevels.
EPATaiwan:600ppmand1,000ppm
Type1indoorareassuchasdepartment
stores,theaters,restaurants,libraries,
theacceptableCO2concentrationforan
8hoursaverage(TWA)is1,000ppm.
Type2indoorareaswithspecialrequirements
forgoodairqualitysuchasschools,
hospitals,anddaycarecenters,the
suggestedCO2levelis600ppm.

RegulatoryExposurelimit
ASHRAEStandard62‐1989:1,000ppm
CO2concentrationinoccupiedbuildingshould 
notexceed1,000ppm.
Buildingbulletin101(Bb101):1,500ppm
UKstandardsforschoolsstatethatCO2levels
overaday(i.e.9amto3.30pm)shouldnot
exceed1,500ppm.
OSHA:5000ppm
Timeweightedaverageoveran8‐hourwork
dayshouldnotexceed5,000ppm.
Germany,Japan,Australia,UK:5,000ppm
Occupationalexposurelimitis5,000ppmover
an8hourworkday.

Other Co2meter Measuring Instrument manuals

Co2meter Aranet4 PRO User manual

Co2meter

Co2meter Aranet4 PRO User manual

Co2meter AZ-0002 User manual

Co2meter

Co2meter AZ-0002 User manual

Co2meter AZ 77535 User manual

Co2meter

Co2meter AZ 77535 User manual

Co2meter SAN-10 User manual

Co2meter

Co2meter SAN-10 User manual

Co2meter RAD-0002-ZR User manual

Co2meter

Co2meter RAD-0002-ZR User manual

Co2meter AZ-0004 User manual

Co2meter

Co2meter AZ-0004 User manual

Co2meter 7755 User manual

Co2meter

Co2meter 7755 User manual

Co2meter eSENSE Outdoor SE-0036 User manual

Co2meter

Co2meter eSENSE Outdoor SE-0036 User manual

Co2meter 77597 User manual

Co2meter

Co2meter 77597 User manual

Co2meter CM-1650 User manual

Co2meter

Co2meter CM-1650 User manual

Co2meter SAN-20 User manual

Co2meter

Co2meter SAN-20 User manual

Co2meter Gastrack GAP-100 User manual

Co2meter

Co2meter Gastrack GAP-100 User manual

Co2meter CM-1650 User manual

Co2meter

Co2meter CM-1650 User manual

Co2meter SAN-10 User manual

Co2meter

Co2meter SAN-10 User manual

Co2meter CM-650 User manual

Co2meter

Co2meter CM-650 User manual

Co2meter XT-10 User manual

Co2meter

Co2meter XT-10 User manual

Co2meter TON-0007 User manual

Co2meter

Co2meter TON-0007 User manual

Co2meter AQ500 User manual

Co2meter

Co2meter AQ500 User manual

Co2meter CM-1650 User manual

Co2meter

Co2meter CM-1650 User manual

Co2meter TecPen MAP TS-200 O2 User manual

Co2meter

Co2meter TecPen MAP TS-200 O2 User manual

Co2meter RAD-0002 User manual

Co2meter

Co2meter RAD-0002 User manual

Co2meter SAN-0001 User manual

Co2meter

Co2meter SAN-0001 User manual

Popular Measuring Instrument manuals by other brands

PCB Piezotronics U357B69 Installation and operating manual

PCB Piezotronics

PCB Piezotronics U357B69 Installation and operating manual

AccuEnergy AcuRev 1300 user manual

AccuEnergy

AccuEnergy AcuRev 1300 user manual

KVANT Spectrum Series Operation manual

KVANT

KVANT Spectrum Series Operation manual

Tenmars ST-109R user manual

Tenmars

Tenmars ST-109R user manual

Sandberg 133-74 user guide

Sandberg

Sandberg 133-74 user guide

Hach DR2400 manual

Hach

Hach DR2400 manual

Thies CLIMA 4.3223.50 Series Instructions for use

Thies CLIMA

Thies CLIMA 4.3223.50 Series Instructions for use

HoMedics Tri-Fitness SC-560 instruction manual

HoMedics

HoMedics Tri-Fitness SC-560 instruction manual

Hanna Instruments HI 713 instruction manual

Hanna Instruments

Hanna Instruments HI 713 instruction manual

Advanced Instruments OsmoTECH PRO Installation & Training Guide

Advanced Instruments

Advanced Instruments OsmoTECH PRO Installation & Training Guide

TESTO Saveris quick start guide

TESTO

TESTO Saveris quick start guide

Atdec ATD-5544 owner's manual

Atdec

Atdec ATD-5544 owner's manual

Steinberg SBS-CT-123 user manual

Steinberg

Steinberg SBS-CT-123 user manual

LOVATO ELECTRIC DME D305T2MID instruction manual

LOVATO ELECTRIC

LOVATO ELECTRIC DME D305T2MID instruction manual

Garnet 900D6 DUAL manual

Garnet

Garnet 900D6 DUAL manual

HEIDENHAIN QUADRA-CHEK 2000 operating instructions

HEIDENHAIN

HEIDENHAIN QUADRA-CHEK 2000 operating instructions

REED R8000 instruction manual

REED

REED R8000 instruction manual

Cirrus CK675 user manual

Cirrus

Cirrus CK675 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.