EnglishDeutsch
 logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Cole Parmer
  6. •
  7. Scale
  8. •
  9. Cole Parmer B2 Series User guide

Cole Parmer B2 Series User guide

This manual suits for next models

8

Other Cole Parmer Scale manuals

Cole Parmer IPSW Series User manual

Cole Parmer

Cole Parmer IPSW Series User manual

Cole Parmer IBS-200 User manual

Cole Parmer

Cole Parmer IBS-200 User manual

Cole Parmer MB-800 Series User manual

Cole Parmer

Cole Parmer MB-800 Series User manual

Cole Parmer LB-200 Series User manual

Cole Parmer

Cole Parmer LB-200 Series User manual

Cole Parmer Balance LA Series User guide

Cole Parmer

Cole Parmer Balance LA Series User guide

Cole Parmer TA-225.C User guide

Cole Parmer

Cole Parmer TA-225.C User guide

Cole Parmer PBL-602 User manual

Cole Parmer

Cole Parmer PBL-602 User manual

Cole Parmer TA-225.C User manual

Cole Parmer

Cole Parmer TA-225.C User manual

Cole Parmer CS-200 User manual

Cole Parmer

Cole Parmer CS-200 User manual

Cole Parmer IPSW-400 Series User manual

Cole Parmer

Cole Parmer IPSW-400 Series User manual

Popular Scale manuals by other brands

Omron HN-286 null

Omron

Omron HN-286 null

AWS ZEO-50 user manual

AWS

AWS ZEO-50 user manual

Gima Glass Body Fat Analyzer Use and maintenance book

Gima

Gima Glass Body Fat Analyzer Use and maintenance book

Seca 374 user manual

Seca

Seca 374 user manual

TEC TEC SL59 SERIES owner's manual

TEC

TEC TEC SL59 SERIES owner's manual

Axis B/E Series user manual

Axis

Axis B/E Series user manual

Alecto ACW-40 user manual

Alecto

Alecto ACW-40 user manual

Rice Lake MSI-4260 IS Technical manual

Rice Lake

Rice Lake MSI-4260 IS Technical manual

IKA RCT basic safety control operating instructions

IKA

IKA RCT basic safety control operating instructions

ADE M20011 operating manual

ADE

ADE M20011 operating manual

Duronic KS3000 manual

Duronic

Duronic KS3000 manual

Digi SM-720 Service manual

Digi

Digi SM-720 Service manual

Cardinal CS Series Operation manual

Cardinal

Cardinal CS Series Operation manual

Korona Kastella Instructions for use

Korona

Korona Kastella Instructions for use

A&D FC-50Ki instruction manual

A&D

A&D FC-50Ki instruction manual

BDA CAMARGUE installation guide

BDA

BDA CAMARGUE installation guide

Tanita HD-357 Instructions for use

Tanita

Tanita HD-357 Instructions for use

Leifheit 03173 quick start guide

Leifheit

Leifheit 03173 quick start guide

https://manuals.online logo
https://manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2026 Manuals.Online. All Rights Reserved.

B2-Series
Start up Guide
Balances
1. CONTENT
Weighing pan
x 1
Balance
x 1
Power adapter
x 1
Draft shield
x 1
Bottom insert
x 1
Models: VWR-64B2 | VWR-124B2 | VWR-214B2
Weighing pan
x 1
Power adapter
x 1
Grounding foot
x 1
Foot
x 3
Models: VWR-203B2 | VWR-403B2 | VWR-503B2
Models: VWR-4002B2 | VWR-5002B2
Balance
×1Weighing pan ×1Draft shield
1 ×1
Power adapter
x 1
Balance
x 1
Bottom insert
x 1
2. WORKROOM AND BASIC OPERATION GUIDELINES
3. UNPACKING
Place the balance away from heat
sources. Avoid exposing the balance
to the sunlight.
Avoid air drafts and air movements
at the workstation.
Avoid exposing the balance to a
magnetic field. Do not weigh
magnetic substances.
Make sure that the balance is placed
on an even surface.
Do not place the balance in a hazardous
area. Do not weigh explosive materials.
Do not place the balance on an unstable
ground exposed to shocks and vibrations.
+104oF
+40 oC
+50oF
+10 oC
Operate the device in a room where the temperature ranges between 10–40 °C
(50–104 °F) and where the relative humidity is below 80%.
Place the balance on a solid surface to ensure stability. To obtain stable and
repeatable weighing results, an anti-vibration table is recommended.
Keep the packaging should a warranty
claim or service be required.
TOP
1 2 3
Models: VWR-203B2 | VWR-403B2 | VWR-503B2
4. ACTIVITIES TO BE DONE PRIOR TO OPERATION
4.1. Remove transport lock
Models: VWR-64B2 | VWR-124B2 | VWR-214B2
Grounding foot
Keep the transport
lock should a
warranty claim or
service be required.
Check the grounding spring.
Make sure that the grounding
spring juts slightly out of the
hole.
Check the grounding spring.
Make sure that the grounding
spring juts slightly out of the
hole.
4.2. Grounding spring check
Models: VWR-64B2 | VWR-124B2 | VWR-214B2
12
5. COMPONENTS ASSEMBLY
1 2 3
Install the
grounding foot
in a freely
selected
opening.
Models: VWR-64B2 | VWR-124B2 | VWR-214B2
Models: VWR-203B2 | VWR-403B2 | VWR-503B2
1
Models: VWR-4002B2 | VWR-5002B2
2
1 23
6. PREPARING FOR WORK
6.1. Balance temperature stabilization time
Prior to switching the device on it is necessary
to ensure that it has reached a temperature
equal to room temperature.
Connect the power adapter to DC connector. Connect the power adapter to the mains.
To level the balance, turn its feet and keep
turning the feet until the air bubble takes
the central position.
Balance temperature stabilization time ranges
between 1 - 8 hours.
BUS 1 USB 2
Ethern
7. PANEL AND SCREEN
7.1. Panel keys
7.2. Home screen
8.1. Adjustment
8. BASIC OPERATIONS
Mass
standard [g]
5 000
VWR-4002B2
VWR-5002B2
Balance
model
Mass
standard [g]
50
100
200
200
200
500
VWR-64B2
VWR-124B2
VWR-214B2
VWR-203B2
VWR-403B2
VWR-503B2
Balance
model
Prior adjustment prepare the right
mass standard. You are recommended
to use mass standard of class F1 or F2.
View the table and select the mass
standard you need.
External adjustment is carried out using an external
mass standard of the right accuracy and weight
value, which value depends on balance type and
capacity.
2
1
4
3
Load the weighing pan with the
required mass standard (refer to
table above). Press key
2 000
Press key to start the adjustment
Remove all items from weighing
pan. Press key
Balance prepares for adjustment.
8.3. Taring
2
1
Unload the weighing pan and press
"Zeroing" button.
The balance has been zeroed.
The balance has been tared.
2
1
When the weighing result has been
displayed press "Tare" button.
8.2. Zeroing
Using keys select "Working mode"
and press key
2
8.4. Working modes accessibility
1
Press to enter "Setup" menu.
4
3
Using keys to select working modes for
which accessibility is to be set and press
key.
In the "Accessibility" position, press key
5
Using keys to set accessibility parameter
to "Yes" value for a selected working mode
and press key to confirm.