manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. CPS
  6. •
  7. Measuring Instrument
  8. •
  9. CPS HVP-252 User manual

CPS HVP-252 User manual

Other CPS Measuring Instrument manuals

CPS VG200 User manual

CPS

CPS VG200 User manual

CPS ELIMINATOR LS3000 User manual

CPS

CPS ELIMINATOR LS3000 User manual

CPS IAQPRO SmartAir User manual

CPS

CPS IAQPRO SmartAir User manual

CPS BlackMax 2 Valve Series User manual

CPS

CPS BlackMax 2 Valve Series User manual

CPS IAQPRO SmartAir User manual

CPS

CPS IAQPRO SmartAir User manual

CPS AC750 User manual

CPS

CPS AC750 User manual

CPS TEMP-SEEKER TM360A User manual

CPS

CPS TEMP-SEEKER TM360A User manual

CPS Leak-Seeker LSCG User manual

CPS

CPS Leak-Seeker LSCG User manual

CPS UV User manual

CPS

CPS UV User manual

CPS VORTECH 2 User manual

CPS

CPS VORTECH 2 User manual

CPS PT200W User manual

CPS

CPS PT200W User manual

CPS AC750 User manual

CPS

CPS AC750 User manual

CPS TEMP-SEEKER TM360 User manual

CPS

CPS TEMP-SEEKER TM360 User manual

CPS AC750 User manual

CPS

CPS AC750 User manual

CPS Pro-Set LDA1000H User manual

CPS

CPS Pro-Set LDA1000H User manual

CPS IAQPRO SmartAir User manual

CPS

CPS IAQPRO SmartAir User manual

CPS FlexOM-PP User manual

CPS

CPS FlexOM-PP User manual

Popular Measuring Instrument manuals by other brands

Suaoki D40 user manual

Suaoki

Suaoki D40 user manual

Elster BK-G1.6 operating instructions

Elster

Elster BK-G1.6 operating instructions

Endress+Hauser Deltabar FMD71 technical information

Endress+Hauser

Endress+Hauser Deltabar FMD71 technical information

Duratool D03128 manual

Duratool

Duratool D03128 manual

Vega VEGAPULS 65 Quick setup guide

Vega

Vega VEGAPULS 65 Quick setup guide

Tonghui TH2618B Operation manual

Tonghui

Tonghui TH2618B Operation manual

Tenmars TM-801 user manual

Tenmars

Tenmars TM-801 user manual

MSR 145 user manual

MSR

MSR 145 user manual

TPS WP-82 manual

TPS

TPS WP-82 manual

PCB Piezotronics 003C30 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 003C30 Installation and operating manual

Tenma 72-10465 manual

Tenma

Tenma 72-10465 manual

Endress+Hauser Proline Promag 50 Brief operating instructions

Endress+Hauser

Endress+Hauser Proline Promag 50 Brief operating instructions

Siemens SITRANS F Coriolis FCT030 Function manual

Siemens

Siemens SITRANS F Coriolis FCT030 Function manual

KLINGER CMF V Series instruction manual

KLINGER

KLINGER CMF V Series instruction manual

Keysight M8290A Getting started guide

Keysight

Keysight M8290A Getting started guide

Acterna MS1400 Operation manual

Acterna

Acterna MS1400 Operation manual

R&S ZPH manual

R&S

R&S ZPH manual

Datacolor ColorReader DC10-2 user guide

Datacolor

Datacolor ColorReader DC10-2 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

INTRODUCTION & SPECIFICATIONS
1.1 INTRODUCTION
The CPS HVP- 252 is a compact, accurate tool for precise mea-
surements of high voltage. The suggested complement to the
divider probe is a differential voltmeter or high input impedance
DVM so that virtually no current is drawn from the measure-
ment circuit. The probe can also be calibrated (factory only) for
use with a digital voltmeter with 10 MΩinput impedance.
WARNING: many digital voltmeters change input impedance
when the range selection is changed.
1.2 ELECTRICAL SPECIFICATION
INPUT VOLTAGE: 50 kV maximum
VOLTAGE RATIO: 1000:1 / 10000:1
INPUT IMPEDANCE: 1000 MΩ (+/− 15%)
OUTPUT LOADING: >10 GΩ for high Z calibration
10 MΩ(W/10 MΩ option )
FREQ. RESPONSE: .25dB @ 60Hz (typical)
ACCURACY: < 0.05% 0 to 50kV
TIPS: CORONA BALL & POINTED
OUTPUT CONNECTOR: BNC
OPERATING INSTRUCTIONS
1. Before making voltage measurements, ensure that the probe’s
insulated ground wire is electrically connected to earth ground.
Periodically inspect the lead for damage due to misuse or other
damage. WARNING: The full measured voltage could be
present at the probe BNC output connector if the probe is not
solidly connected to earth ground.
2. With the probe properly grounded, connect the probe output
to a DVM with an input Z > 10 GΩ. If using a 10 MΩcalibra-
tion probe, be sure to use only a 10 MΩinput Z digital voltme-
ter. CAUTION: When using a digital voltmeter with the HVP-
250, be sure to avoid meter ranges that present loading other
than the specified meter matching calibration impedance.
Check the DVM manual to be sure. Most DVM’s present an
input Z of 10 MΩon input ranges over 10 volts.
3. Mount the selected probe tip to the probe. Turn the tip clock-
wise until it seats against the probe body.
4. Set the voltmeter to the proper range. The divider ratio is
1000:1. Ten volts on the DVM represents 10kV, the divider
ratio is 10000:1 One volt on the DVM represents 10kV at the
test point. Proceed with the measurement.
5. If desired, the probe may be semi-permanently mounted in a
measurement system by means of the two 8-32 inserts in the
handle guard. These inserts are electrically isolated. Do not use
these inserts as a ground connection.
6. Your divider probe is a precision instrument. Please store the
probe, accessories, and manual in the instrument case. If you do
not have an instrument case we suggest you order one through
customer service at (503) 598-9595.
WARRANTY
COMPUTER POWER SUPPLY, Inc. warrants equip-
ment of its manufacture against defective materials or
workmanship for a period of two years from the date of
shipment.
CPS will repair or replace any defective product which
was not damaged by negligence, misuse, improper
installation, accident or unauthorized repair or alteration
by the buyer.
This warranty is applicable to the original Buyer only
and constitutes the sole and exclusive warranty of Seller.
No other warranty is made, expressed or implied. INSTRUCTION MANUAL
HVP-252
HIGH VOLTAGE PROBE
7313 SW Tech Center Drive
Portland, OR 97223
(503) 684-8026
Document No. 0252-90-0001 Rev. A Date: 1/15/2012
10 20 30 40 50 kV
0.05
0.00
% DIFF (TEST-REF)
0.04
0.03
0.02
0.01
-0.01
-0.02
-0.03
-0.04
-0.05
0
UPPERLIMIT
LOWERLIMIT
SPECIFICATION TEST LIMITS