Daavlin UV Series User manual

MNL-00007 [3] Proprietary and Confidential Page 1of 63
Flex™Controlled Clinical Operation Manual

MNL-00007 [3] Proprietary and Confidential Page 2of 63
Table of Contents
1.0 Introduction ....................................................................................5
2.0 Indications for Use .........................................................................6
3.0 Classifications .................................................................................6
4.0 Operating Specifications................................................................6
5.0 Transport and Storage Specifications..........................................7
6.0 EMC Precautions ...........................................................................7
7.0 Accuracy..........................................................................................7
8.0 Labels and Symbols........................................................................8
8.1 UV Risk Group 2 Labels (CE Devices Only)............................................................. 10
9.0 Delivery and Inspection...............................................................11
10.0 Lamp Inspection...........................................................................13
11.0 Site Selection .................................................................................13
12.0 Training Requirements................................................................13
13.0 Important Safeguards and Warnings ........................................14
14.0 Unlocking the Device....................................................................16
15.0 Changing or Disabling the Key Code.........................................16
16.0 Operating the Control Systems...................................................16
17.0 Flex Dosimeter Instructions: UV Series, 7 Series, M Series, and
4 Series....................................................................................................16
17.1 The Backup Time ....................................................................................................... 17
17.2 Measuring the Power Output of Your Device............................................................ 17
17.3 Setting up a Treatment................................................................................................ 19
17.4 Special Functions of the Integrated Dosimeter........................................................... 20
17.5 Special Notes.............................................................................................................. 20
17.6 Flex Dosimeter Range and Accuracy......................................................................... 20
18.0 Flex Dosimeter Instructions: 1 Series.........................................20
18.1 The Backup Time ....................................................................................................... 21
18.2 Measuring the Power Output of Your 1 Series........................................................... 21
18.3 Setting Up a Treatment............................................................................................... 23
18.4 Repeat Treatments...................................................................................................... 23
18.5 Special Functions of the Integrated Dosimeter........................................................... 24
18.6 Special Notes.............................................................................................................. 24
18.7 Flex Dosimeter Range and Accuracy......................................................................... 24
19.0 Flex Timer Instructions: UV Series, 7 Series, M Series, and 4
Series.......................................................................................................25
19.1 Determining a Treatment Time .................................................................................. 25
19.2 Setting Up a Treatment Time ..................................................................................... 26
19.3 Special Functions of the Digital Timer....................................................................... 26

MNL-00007 [3] Proprietary and Confidential Page 3of 63
19.4 Special Notes.............................................................................................................. 26
19.5 Flex Timer Range and Accuracy................................................................................ 27
20.0 Flex Timer Instructions: 1 Series ...............................................27
20.1 Determining a Treatment Time .................................................................................. 27
20.2 Setting Up a Treatment Time ..................................................................................... 28
20.3 Special Functions of the Digital Timer....................................................................... 28
20.4 Special Notes.............................................................................................................. 28
20.5 Flex Timer Range and Accuracy................................................................................ 29
21.0 UV Series X and UV Series T Information................................29
21.1 Domestic and Non-CE Electrical Requirements........................................................ 29
21.2 European CE Electrical Requirements....................................................................... 30
21.3 Circuit Breaker ........................................................................................................... 30
21.4 Unpacking and Assembly........................................................................................... 30
21.5 How to Position the Patient........................................................................................ 31
21.6 UV Series Lamp Removal and Replacement............................................................. 31
22.0 7 Series X and 7 Series T Information .......................................32
22.1 Electrical Requirements.............................................................................................. 32
22.2 Unpacking and Assembly........................................................................................... 33
22.3 How to Position the Patient........................................................................................ 34
22.4 7 Series Lamp Removal and Replacement................................................................. 35
23.0 M Series X and M Series T Information....................................35
23.1 Electrical Requirements.............................................................................................. 35
23.2 Unpacking and Assembly........................................................................................... 36
23.3 Special Notes.............................................................................................................. 36
23.4 How to Position the Patient........................................................................................ 37
23.5 M Series Lamp Removal and Replacement ............................................................... 37
24.0 4 Series Information.....................................................................38
24.1 Electrical Requirements.............................................................................................. 38
24.2 Unpacking and Assembly........................................................................................... 39
24.3 How to Position the Patient........................................................................................ 39
24.4 4 Series Lamp Removal and Replacement................................................................. 39
25.0 1 Series X and 1 Series T Information .......................................40
25.1 Electrical Requirements.............................................................................................. 40
25.2 Unpacking and Assembly........................................................................................... 41
25.3 How to Position the Patient........................................................................................ 41
25.3.1 Hand and Foot Treatments:..................................................................................... 41
25.3.2 Larger Area Treatment:........................................................................................... 41
25.4 1 Series Lamp Removal and Replacement................................................................. 42
25.5 Special Notes for 1 Series Units Equipped with Integrated Dosimetry ..................... 42
25.5.1 Treatment Distance................................................................................................. 42
25.5.2 Calibration Output Certificate................................................................................. 43
25.6 Special Notes for 1 Series Units Equipped with the Digital Timer............................ 43
25.6.1 Treatment Distances and Your Calibration Output Certificate............................... 43
26.0 Treatment Protocols.....................................................................45

MNL-00007 [3] Proprietary and Confidential Page 4of 63
26.1 UVA Clearing Phase Protocols (Skin Types I - III)................................................... 45
26.2 UVA Clearing Phase Protocols (Skin Types IV - VI)................................................ 47
26.3 UVA Maintenance Phase Protocols: .......................................................................... 49
26.4 UVA Treatment Limits for Maintenance & As Required Protocols.......................... 50
26.5 LO Dose Broad Band UVB Clearing Phase Protocols:.............................................. 50
26.6 LO Dose Broad Band UVB Maintenance Phase Protocols:....................................... 50
26.7 LO Dose Broad Band UVB Treatment Limits for Maintenance & As Required
Protocols 51
26.8 HI Dose Broad Band UVB Clearing Phase Protocols: 11011, 3/week, 2/week....... 51
26.9 HI Dose Broad Band UVB Maintenance Phase Protocols: 1/WEEK ..................... 51
26.10 LO Dose Narrow Band UVB Clearing Phase Protocols: 11011, 3/week, 2/week.... 52
26.11 LO Dose Narrow Band UVB Maintenance Phase Protocols: 1/WEEK.................... 52
26.12 LO Dose Narrow Band UVB Treatment Limits for Maintenance & As Required
Protocols 52
26.13 HI Dose Narrow Band UVB Clearing Phase Protocols: ............................................ 53
26.14 HI Dose Narrow Band UVB Maintenance Phase Protocols: ..................................... 54
26.15 HI Dose Narrow Band UVB Treatment Limits for Maintenance & As Required
Protocols 54
27.0 Care of Your Phototherapy Unit................................................55
27.1 Recommended Maintenance Schedule....................................................................... 55
27.2 Cleaning/Disinfection................................................................................................. 55
27.2.1 General Cleaning .................................................................................................... 55
27.2.2 Low-Level Disinfection.......................................................................................... 55
27.2.3 High-Level Disinfection ......................................................................................... 56
27.3 Resetting Your Lamp Hours....................................................................................... 56
28.0 EMC Guidance .............................................................................56
29.0 Flex Controller Trouble Shooting Guide...................................58
29.1 Power-On-Self-Test.................................................................................................... 58
29.2 Error Code E8yz: yz = address of first location with range error.............................. 59
29.3 Sensing Errors............................................................................................................. 59
30.0 Warranty and Contact Information...........................................61
30.1 Limited Warranty Policy............................................................................................ 61
30.2 Warranty Coverage..................................................................................................... 61
30.3 Customer Responsibility............................................................................................. 62
30.4 Warranty Service........................................................................................................ 62
30.5 Disposal...................................................................................................................... 62
30.6 Other Services............................................................................................................. 62
30.7 Contact Information.................................................................................................... 63

MNL-00007 [3] Proprietary and Confidential Page 5of 63
1.0 Introduction
Thank you for purchasing a Daavlin Phototherapy Unit. The use of light for the treatment of
photoresponsive skin disorders has been our passion since 1981. From the beginning we have
been devoted to providing our customers with the highest quality products coupled with industry
leading customer service.
At Daavlin, we are always keeping track of new developments and are doing our best to
implement the latest findings in our products. We appreciate receiving feedback from the
medical community and patients, so we can further improve our products. If you have any
comments or suggestions, please contact our Customer Service department and your input will
be channeled to the appropriate person.
The purpose of this manual is to instruct users on the proper methods of operation and general
maintenance. In addition to this, the manual also addresses important information regarding
device specifications, warnings, treatment protocols and warranty information. Please take a
moment to read the entire operation manual before operating your Daavlin phototherapy unit.
Here at Daavlin we are proud of our tradition of development and innovation in the field of
phototherapy and we are honored that you have chosen us for your phototherapy needs. Let’s be
clear. Our commitment to you starts…Now!
Sincerely,
Daavlin Distributing Company
205 West Bement Street
P.O. Box 626
Bryan, OH 43506-0626
Phone: (800) 322-8546, (419) 636-6304
Fax: (419) 636-1739
Email: [email protected]
Website: www.Daavlin.com

MNL-00007 [3] Proprietary and Confidential Page 6of 63
2.0 Indications for Use
The UV Series, 7 Series, and M Series, are medical ultraviolet devices, which are intended for
the treatment of psoriasis, vitiligo, and atopic dermatitis (eczema) on all skin types (I –VI).
The 1 Series Unit is a therapeutic product designed for individuals who require ultraviolet
radiation for diagnosed skin disorders. The 1 Series, equipped with PL-L36w/03 blue lamps, is
indicated for the treatment of mild to moderate acne vulgaris.
The 4 Series Unit is a therapeutic product designed for individuals who require ultraviolet
radiation for diagnosed skin disorders.
WARNING: Do not use these devices for anything other than their intended purposes.
CAUTION: Federal law restricts this device to be sold by or on the order of a physician or any
other practitioner licensed by the law of the state in which he/she practices.
3.0 Classifications
FDA:
Class II Device
93/42/EEC:
Class IIa Device
IEC 60601-1:
Class I Device
Pollution Degree:
Class II
Mode of Operation:
Non-continuous
IEC/EN
62471:2006 UV
Risk Group:
Risk Group 2 (Moderate-Risk)
WARNING: This device is designed for intermittent operation only and not for continuous use.
The device should not be cycled continuously for more than 10 minutes. After such time, the unit
should either be turned off or left idling for at least 10 minutes.
4.0 Operating Specifications
Ambient
Temperature:
15°C to 30°C (59°F to 86°F)
Relative Humidity:
10% to 95%, Non-condensing
Liquid Ingress
Rating:
IPX0 (This device does not have protection
against ingress of water.)
Ocular Hazard
Distance:
3 Meters (9.84 Feet)
Ambient
Luminance:
250 –500 lux

MNL-00007 [3] Proprietary and Confidential Page 7of 63
WARNING: Equipment not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide.
5.0 Transport and Storage Specifications
Ambient
Temperature:
-40°C to 70°C (-40°F to 158°F)
Relative Humidity:
10% to 95%, Non-condensing
Atmospheric
Pressure:
50 kPa to 106 kPa
6.0 EMC Precautions
The devices contained in this manual have been tested and found to comply with the EMC limits
of the international standard IEC 60601-1-2. These limits are designed to provide reasonable
protection against interference in a typical medical installation. The system can radiate radio
frequency energy if not installed in accordance with the instructions and may cause harmful
interference to other devices in the vicinity. There is no guarantee that interference will not
occur in a particular installation. If the system does cause interference with other devices, which
can be determined by turning the system off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
•Re-orient or relocate the receiving device
•Increase the separation between the equipment
•Connect the System into an outlet on a circuit different from that to which the other
device(s) is connected
•Consult the manufacturer or field service technician for help
7.0 Accuracy
The measuring device will maintain a 5% level of accuracy if the device is calibrated every one
hundred (100) hours of use as recommended. Please contact Daavlin for calibration.

MNL-00007 [3] Proprietary and Confidential Page 8of 63
8.0 Labels and Symbols
Every UV series comes affixed with an orange “safety key code” label located near the
Integrated Dosimetry Control System. This label will provide you with important information
about the factory default key code needed to unlock the unit.
If your unit has been equipped with our FlexRx™exposure limiting software there will be a pink
label, in addition to the orange “safety key code” label, informing you that your unit has been
equipped with this feature.
Figure 1 Safety Key Code
A warning and danger labels are affixed to your Daavlin phototherapy device, see Figure 2
Warning and Danger Labels in a prominent and easily readable position. Please read the label
carefully as they contain important safety information for you. In addition to the warning and
danger labels, an identification label indicates the serial number and date of manufacture that is
specific to your device. A chart detailing all symbols located on your unit can be found on in
Table 1 Symbols on page 8.

MNL-00007 [3] Proprietary and Confidential Page 9of 63
Figure 2 Warning and Danger Labels
The following is a chart detailing all symbols located on your device and their meanings:
Table 1 Symbols
Symbol
Description
DANGEROUS VOLTAGE
NON-IONIZING RADIATION
EARTH (ground)
PROTECTIVE EARTH (ground)
OPERATING INSTRUCTIONS
KEEP DRY
CAUTION, CONSULT ACCOMPANYING DOCUMENTS
NBUVB LAMP/TIMER LABEL
UVA LAMP/TIMER LABEL

MNL-00007 [3] Proprietary and Confidential Page 10 of 63
Symbol
Description
PACKAGING LABEL (CE ONLY)
MAIN CIRCUIT BREAKER SWITCH (UV SERIES)
EU REP LABEL (CE ONLY DEVICE)
CE ONLY LABEL
8.1 UV Risk Group 2 Labels (CE Devices Only)
According to IEC/EN 62471:2006, sources of optical radiation are classified into risk groups
subject to their potential photobiological hazard. This device falls under Risk Group 2
(Moderate-Risk) and does not pose a severe hazard due to the natural aversion response to bright
light or thermal discomfort. A Risk Group 2 warning label, which also includes the primary
emission range and potential output range, is affixed to your device in a prominent and easily
readable position. CE devices have this label in English and French. See Figure 3 English and
French UV Risk Group Labels
Figure 3 English and French UV Risk Group Labels

MNL-00007 [3] Proprietary and Confidential Page 11 of 63
Figure 4 Wiring Materials label (7 & UV Series, Japan & CE Devices Only)
Figure 5 7 Series Package Contents Label (7 Series ONLY)
Figure 6 Time-Lag 250V Label (Japan 7 Series Only)
Figure 7 On/Off Key Switch Label (Japan Only)
9.0 Delivery and Inspection
When you first receive the device, please inspect the shipping crate or box.
Any signs of shipping damage must be noted on the delivery receipt that you will be asked to
sign by the delivery driver. Be sure to open the crate and verify that there is no damage to your
unit before the driver leaves. If it is not possible for you to inspect your unit before the driver
leaves, we recommend you write “Concealed damage possible. Further inspection required” on

MNL-00007 [3] Proprietary and Confidential Page 12 of 63
the delivery receipt. If damage is discovered after unpacking the unit, save all packing materials
and call Daavlin at 1 800 322 8546 for inspection and repair.
Note: The delivering carrier must be notified of any shipping damage within twenty-four (24)
hours to protect your right to an insurance claim.
Note: As part of the claims process the delivering carrier may require that a damage inspection
be conducted. The delivering carrier may request to conduct the inspection at the delivery site,
provided that a mutually agreed upon date and time can be established, or they may elect to
collect the package for inspection at their facilities.

MNL-00007 [3] Proprietary and Confidential Page 13 of 63
10.0 Lamp Inspection
Your device can be equipped with a variety of different lamps, each having a different effect on
the skin. While you are unpacking and setting up your device it is important to check that the
proper lamps are installed in the device. Consult the Lamp Specification Guide below and if
there is any question as to whether or not your device it is equipped with the lamps that you have
ordered, contact Daavlin immediately. The code numbers shown are generally located at the base
of the lamp.
Applicable
Units
UVA
UVA-1
Narrowband UVB
Broadband
UVB
Blue
UV Series:
7 Series:
-F72T12-BL-
HO-RDC
-
TL100W/1
0R
-TL-01/100W-FS72 –
FS72T12/NBUVB/HO
/RDC/100W
-FS72T12-
UVB-HO-
TL100W/12
N/A
M Series:
-PLL-36W/P-
F24T12/BL/
HO/RDC
-PLL-
36W/10/4P
-PLL-36W/01/4P
-TL20W/01-RS
-060PLL-
36W-06-
FS24T12-
UVB-HO
N/A
4 Series:
-PLL-36W/P
-PLL-
36W/10/4P
-PLL-36W/01/4P
-060PLL-
36W-06
PLL-
36W/03
1 Series:
-PLL-36W/P
-PLL-
36W/10/4P
-PLL-36W/01/4P
-060PLL-
36W-06
PLL-
36W/03
11.0 Site Selection
A site should be chosen within reach of the specified electrical connection (refer to the Electrical
Requirements section of your specific device) and where the unit can be left in place without
obstructing traffic flow. The device should be positioned in such a way the power inlet or circuit
breakers on the device are easily accessible. It is important that the unit be properly grounded.
Extension cords are not recommended. The site should not be in any area where water or
moisture might collect and should be protected from access by children and other unintended
users.
WARNING: To avoid the risk of electric shock this equipment must only be connected to a
supply main with a protective earth.
12.0 Training Requirements
Phototherapy services require staff that have appropriate training, knowledge, and clinical skills
in order to ensure effective outcomes and quality care for patients. Staff must be assessed as
being competent and safe in order to provide phototherapy treatments that maximize benefit and
minimize adverse effects. Recommended training requirements include:

MNL-00007 [3] Proprietary and Confidential Page 14 of 63
•Staff should be either a physician, nurse, or any other practitioner licensed by the law of
the state or country in which he/she practices
•Training and experience in dermatology is important to provide holistic patient care. This
knowledge includes:
oAnatomy and the Physiology of the skin
oRecognition and understanding of skin diseases
oSkin assessment
oUnderstanding of photoresponsive diseases
•Theoretical knowledge of phototherapy and its use
•A period of supervised practice for approximately 3 months with a competent practitioner
•Managers should ensure that staff members have, or are provided with, the appropriate
level of training and education needed prior to administering phototherapy services.
13.0 Important Safeguards and Warnings
•To protect the eyes during operation, the operator and anyone in view of the device must
wear the provided UV blocking glasses or goggles designed to block 100% of all UVA and
UVB light from the eye area when worn. Always use Daavlin approved eyewear.
•Do not remove protective eyewear, or any other protective equipment, during treatment.
•If psoralens (photosensitizing drugs) are being used as part of treatment (“PUVA”), the eyes
should be protected from exposure to ultraviolet (sunlight) for 24 hours after taking the drug.
Ultraviolet blocking glasses are provided with all Daavlin devices.
•Do not use over skin eruptions without express consent from the attending physician.
•If a patient experiences burning, never treat the patient until the noticeable effects of burning
subside, and always reduce the subsequent treatment time.
•To protect unaffected skin during operation, the operator, patient, and anyone in view of the
device must generously apply UV blocking skin protection to all exposed skin that the
physician does not intend to treat.
•During extremely long treatments, some surfaces may become hot. Remove any body parts
from those surfaces if it becomes uncomfortable. The unit should have a 50% duty cycle of
10 minutes on, 10 minutes off.
•Erythema can result in as little as 15 seconds of exposure, or approximately 200 millijoules
of dose to UVB light. If using the unit at home, contact your prescribing physician for
specific treatment instructions and dosing information.
•Center patients between the lamps during treatment to avoid over exposure to isolated areas
of the body.
•All people and pets should leave the area to avoid exposure to ultraviolet light.
•Ask your doctor about protecting areas of the body that have not been exposed to sunlight.
•Do not use this device for anything other than its intended purpose.
•This device is only to be used by authorized users.

MNL-00007 [3] Proprietary and Confidential Page 15 of 63
•Do not operate this device with a damaged cord or plug.
•If appliance coupler or mains plug is used as the mains disconnect, do not position the ME
equipment so that it is difficult to operate the disconnection device.
•To avoid the risk of electric shock, this equipment must only be connected to a supply main
with a protective earth.
•To prevent electric shock, remove power to the device prior to cleaning and servicing.
•To eliminate the risk of fire when replacing the fuse, replace ONLY with a fuse of the same
type and rating.
•NO MODIFICATION OF THIS EQUIPMENT IS PERMITTED. Unauthorized
modification will void the warranty and may result in hazardous or improper device
operation.
•DANGER - ULTRAVIOLET RADIATION. As with natural sunlight, overexposure can
cause eye and skin injury, and allergic reactions. Repeated exposure may cause premature
aging of the skin and/or skin cancer. ALWAYS WEAR PROTECTIVE EYEWEAR:
FAILURE TO DO SO MAY RESULT IN SEVERE ERYTHEMA OR LONG-TERM
INJURY TO THE EYES. Medications or cosmetics may increase skin sensitivity to
ultraviolet radiation. Inform physician before using this device if you are using medications
or have a history of skin problems or sensitivity to light.
•If the device malfunctions, cease operation immediately. If the device is placed close to other
equipment, it is possible that the cause is interference by external noise sources and fields, in
which case you should follow the remedies found under EMC Precautions. If the device
continues to malfunction cease operation and contact the Daavlin Service Department.
•If necessary, operation of equipment may be terminated by disconnecting unit from AC
mains. This can be accomplished by unplugging device or, if hard wired, turning off quick
disconnect.
•Prior to each use, always verify that the device is in correct working order and operating
condition and plugs, sockets, lamps, and electrical cables and connections are not worn or
damaged.
•Only original components and accessories should be used with the device to avoid damage.
•Before opening the device casing to perform maintenance or service, disconnect the device
from the power source.
•The device must never be directly exposed to flowing or splashing liquids of any kind. If the
device is inadvertently exposed to liquid, it must be tested for safe function before being
placed in operation again.
•The device contains glass lamps. Avoid excessive force to the device to prevent lamp
damage.
•The device control system display is susceptible to damage from excessive force. Avoid
excessive force to the control system to prevent damage.
•All treatments must be administered under the direction of a licensed physician only.

MNL-00007 [3] Proprietary and Confidential Page 16 of 63
•The room that the device is placed in should be vented to allow free air displacement to
ensure the device and surroundings remain cool.
•The device should be placed a minimum of 1 inch (2.5 cm) away from surrounding walls,
devices, and furniture to ensure proper cooling airflow.
•A qualified, licensed electrician must wire the service for all devices in accordance with all
national and local codes and the electrical instructions provided in the accompanying Service
& Installation Instructions manual. Unauthorized personnel should not open the panels. The
Daavlin Service Department should be consulted before any service is performed.
14.0 Unlocking the Device
To prevent unauthorized use, the device will “self-lock” when left idle for twenty (20) minutes.
To unlock your unit, follow the instructions below.
1. Press any key to awaken your unit.
2. Your device will display the word “CODE”.
3. The factory preset key code is “0007”. Press the ▲ key seven times. Your unit will now
display “0007”.
4. Press the “ENTER” key to unlock your unit.
15.0 Changing or Disabling the Key Code
If you would like to disable or personalize your factory default key code, please contact the
Daavlin service department at 1-800-322-8546 for step by step instructions.
The Daavlin service department is available to assist you Monday through Thursday 7:30am-
8:00pm EST and Friday 8:30am-5:00pm EST.
16.0 Operating the Control Systems
Daavlin Flex controlled phototherapy equipment is available with two distinct control systems, a
Flex Timer and a Flex Dosimeter. Please verify the type of control system incorporated in your
unit and proceed to the applicable operating instructions.
17.0 Flex Dosimeter Instructions: UV Series, 7 Series, M Series, and
4 Series
Your Daavlin phototherapy unit features a built-in power meter that measures energy output.
Once a dose and backup time have been selected the unit will use the information gathered by
the sensor to determine the exact amount of time that each treatment will last. The dose is
entered in milliJoules for UVB and Joules for UVA.

MNL-00007 [3] Proprietary and Confidential Page 17 of 63
The output of fluorescent lamps varies with age, use and other factors such as room temperature.
Integrated Dosimetry measures for such variations and then integrates them into the dose
treatment time resulting in a very effective way to ensure the precision and consistency of each
treatment.
Figure 8 UVA J and UVB mJ
17.1 The Backup Time
An important safety feature of Daavlin’s Integrated Dosimetry System is the use of a safety
backup time. While the dose is entered in Joules or milliJoules, a backup time must also be
entered as a safety check. The treatment will stop when the dose has been delivered, OR, when
the backup time has elapsed, whichever comes first. At the end of the treatment, the dose
delivered and the time elapsed are alternately displayed so that you will know which limit ended
the treatment.
When preparing for a treatment, the back-up time is entered first. To determine an appropriate
back-up time, it is necessary to know the power output of your device.
17.2 Measuring the Power Output of Your Device
The power output value of your device, as measured at the factory, is shown on the Power
Output Certificate included with your device literature.
Note: It is not necessary to take a power output reading before each treatment. We suggest once
every couple of weeks or earlier if the backup time ends your treatment before the full dose has
been delivered.
1. When just a decimal point is displayed, the device is “asleep“(energy saving feature).
Press MODE or any key to “wake up” the device.
2. The word “CODE” will appear. Enter in your key code. Unless you have changed it, the
factory preset key code will be 0007.
3. Once the safety key code, if activated, has been entered, press the ▲ key five times. The
screen will read “00:05”.

MNL-00007 [3] Proprietary and Confidential Page 18 of 63
4. Press the ◄ key two times. The screen will now read “05:00”.
5. Press the ENTER key one time to lock in the backup time.
6. Press the MODE key one time to switch to dose entry mode.
7. Press the ▲ key five times. The screen will now read “0005”.
8. Press the ◄ key until 5000 is displayed if checking a UVB device or 50.0 is displayed if
checking a UVA device.
9. Press the ENTER key one time to lock in the dose.
10. Press the ENTER key a second time to begin the treatment. The lamps will now come
on. When measuring ultraviolet be sure to apply SPF 30 sunscreen to exposed areas of
the skin and wear goggles while monitoring the device.
11. Once the treatment begins press the MODE key one time to view the time counting
down. Allow the device to run until 30 seconds remains. While waiting step away from
the unit to limit unnecessary exposure to UV light.
12. With 30 seconds remaining press the POWER key one time. The system will display the
power output of the lamps. A value such as P9.0 (example) will display. Write down the
value.
13. When the machine shuts off, press the END key to clear the treatment and return the unit
to its energy saving mode.
Once you know your power output (9.0 mW/cm2 in the example above) consult the
laminated Backup Time Chart, a partial example of which is included Table 2 Partial
UVB Backup Time Chart below. The numbers in the top row of the chart (horizontal
axis) represent a range of power output values. Find the value that is closest to the power
output of your device.
The numbers in the left most column (vertical axis) represent a range of treatment doses.
Find your desired dose and trace that row to the right until it intersects with the column
corresponding to the current power output of your device. The backup time is shown
where the two lines intersect. The times are rounded up to the nearest 10-second
increment. On the partial time chart thirty (30) seconds is the backup time for a power
output of 9.0 mW/cm2 and a dose of 280 mJ.
Table 2 Partial UVB Backup Time Chart
The UVA Backup Time Chart (not shown here) is configured the same as the UVB Backup Time
Chart however the doses are listed in Joules and the energy output levels and backup times are
different than the UVB output levels. Each chart is clearly marked.
mW/cm2
mJoules
2.8
3.0
3.6
4.0
4.5
5.0
6.0
7.0
9.0
240
01:50
01:20
01:30
01:20
01:20
01:10
01:00
01:00
00:30
280
02:00
01:30
01:40
01:30
01:30
01:20
01:10
01:00
00:30*
320
02:20
01:50
01:50
01:40
01:30
01:30
01:20
01:10
00:40
→

MNL-00007 [3] Proprietary and Confidential Page 19 of 63
Note: For your convenience, and as an alternative to the UVB backup time chart, the latest
version of our free iPhone and Android application “Phototherapy Math” is available for
download at the iTunes app store and Android market. It quickly calculates treatment times,
backup safety times, and makes the calculation of dose and time increases simple. If using a PC,
visit www.phototherapymath.com for a web-based calculator and additional information.
17.3 Setting up a Treatment
You will first enter a backup time (steps 1 –6). Based on the example above the back-up time
will be 00:30.
1. Press any key to awaken the unit. The word “CODE” will appear.
2. Enter in your key code. Unless you have changed it, the factory preset key code will be
0007.
3. Once the key code is entered, the display should read 00:00.
4. Press the ▲key until the first digit of the backup time is displayed (“3” in this case).
5. Press the ◄ key to move the “3” one space to the left, display should now read 00:30.
6. Press ENTER to lock in your back up time.
Now you will enter the actual dose. Continuing our example from above the dose will be 280
milliJoules. Therefore, the first digit will be the number 2.
1. Press the MODE key to switch to dose mode. The display should now read 0000.
2. Press the ▲key until the number “2” is displayed.
3. Press the ◄key to move the “2” one space to the left.
4. Press the ▲key until the number “8” is displayed.
5. Press the ◄ key to move the “8” one space to the left and repeat until all desired digits
are shown on the display. In this example the last digit is a zero, so once you have moved
the “8” into position, the “0” already shows and you can move on to step 6.
6. When the desired dose is displayed, press ENTER to lock in your dose.
7. Put on your goggles and position yourself per the instructions.
8. Begin the treatment by again pressing ENTER
To repeat the same dose for another exposure, press the END key to clear the display. Then press
the MODE and POWER keys at the same time and release. This will automatically bring up the
last treatment administered. Press ENTER to start the treatment.
Note: The system beeps three times to signal the end of the treatment and then flashes the
delivered dose on the display alternating with the elapsed time. If the complete dose is not
delivered, meaning that the backup time was reached before the dose was fully delivered, this is
an indication that the backup times being entered are not long enough. Verify that the sensor is
clean, unobstructed and centered in the hole of the reflector. Then, re-take your power output
reading and adjust the back-up time. It is normal to have to increase back-up times due to aging
lamps and/or because of slow starting lamps, low room temperatures or low line voltage.

MNL-00007 [3] Proprietary and Confidential Page 20 of 63
17.4 Special Functions of the Integrated Dosimeter
To pause an active treatment press either the END or ENTER key.
•When a treatment is paused the screen will read “HOLD”. To resume a paused treatment,
press the ENTER key.
•When a treatment is in progress press the MODE key to toggle between watching the
dose administered count up or the treatment time count down.
•To check the power output while a treatment is in progress press the POWER key.
•To end or cancel an active treatment press the END key two times.
•If you make a mistake while entering a dose or backup time, repeatedly press the ◄ key
until all zeros are displayed.
•To check the age (in hours) of your lamps press and release the MODE and ◄ keys
together at the same time. The lamps must be off to do this.
17.5 Special Notes
It is not possible to enter a backup time of 60 seconds. You must enter 1 minute instead.
•A short beep is sounded for every valid key entered. A long beep is sounded when an
invalid key is entered.
•If power goes off during a treatment, the system will remember how much of the
treatment has elapsed. When power is restored, your device will be on “HOLD”. To
continue the treatment, press START. To cancel the treatment press END.
•The small orange light above the display will be lit when a treatment is in process. It will
blink when the treatment is paused.
17.6 Flex Dosimeter Range and Accuracy
Backup Time Range:
1 Second - 59 Minutes 59 Seconds (00:01 - 59:59)
Dose Range:
UVA: 0.1 Joules - 999.8 Joules UVB: 1 Millijoule - 9998 Millijoules
Accuracy:
+ / - 5%
Calibration:
1 year or every 100 hours of use
18.0 Flex Dosimeter Instructions: 1 Series
Your Daavlin 1 Series features a built-in power meter that measures energy output. Once a dose
and backup time have been selected the unit will use the information gathered by the sensor to
determine the exact amount of time that each treatment will last. The dose is entered in
milliJoules for UVB and Joules for UVA.
The output of fluorescent lamps varies with age, use and other factors such as room temperature.
Integrated Dosimetry measures for such variations and then integrates them into the dose
treatment time. This is a very effective way to ensure the precision of each treatment.
Other manuals for UV Series
1
This manual suits for next models
4
Table of contents
Other Daavlin Medical Equipment manuals