- -
13
4.5 AUTOMATIC OR MANUAL CONDUCTIVITY COMPENSATION
The conductivity measurement refers to a standard temperature, called reference temperature
T_REF: that is, the instrument proposes the conductivity you would get if the temperature were
T_REF. This temperature could be 20°C or 25°C according to the setting in the T_REF menu item
(see chapter 3).
The conductivity increase per each grade of temperature variation is a characteristic of the solution
and is indicated by the term "temperature coefficient αT": most solutions are within these values,
from 0.00 to 4.00%/°C. The default value is 2.00%/°C.
When a combined probe with temperature sensor is present, the instrument automatically applies
the temperature compensation function, and proposes the measurement using the reference
temperature T_REF according to the coefficient αTon the display.
In absence of the temperature probe, the lower display shows the manually set compensation
temperature (default=25°C).
To point this condition out, the °C or °F symbol blinks intermittently near the temperature value.
On the display an "m" (manual) is turned on near the battery symbol (if on).
How to manually change the temperature
To manually change the compensation temperature, proceed as follows:
•press the °C/°F key once: the indicated temperature value starts blinking;
•select the solution temperature value by using the arrows;
confirm by pressing DATA/ENTER: the display stops blinking, and the temperature displayed is
used for compensation.
•To change the unit of measurement between °C and °F, it is necessary to press twice the °C/°F
key.
4.6 CALIBRATION OF THE PROBE
The probe calibration can be carried out on one, two or three points using the standard solutions
automatically detected by the instrument (automatic calibration) or other solutions with known
value (manual calibration).
The CAL symbol will blink when the cell constant is modified using the menu (please see the
description of the K_CELL item, chapter 3).
4.6.1 Automatic calibration of conductivity using memorized buffer solutions
The instrument can recognize four standard calibration solutions:
•0.001 Molar KCl solution (147μS/cm @25°C),
•0.01 Molar KCl solution (1413μS/cm @25°C),
•0.1 Molar KCl solution (12880μS/cm @25°C),
•1 Molar KCl solution (111800μS/cm @25°C),
Using one of these solutions, the calibration is automatic; the procedure can be repeated with one of
the remaining standard solutions, up to a maximum of four different points.
The manual calibration is possible with a different conductivity solution from that used in the
automatic calibration.
The solution temperature for the automatic calibration must be between 15°C and 35°C: if
the solution is under 15°C, the solution is compensated at 15°C. When over 35°C, the solution
is compensated at 35°C.