Dewetron DAQP-LV Product manual

TECHNICAL REFERENCE MANUAL
DAQP-LV Module
WELCOME TO
THE WORLD OF
DEWETRON!
Congratulations on your new device! It
will supply you with accurate, complete
and reproducible measurement results
for your decision making.
Look forward to the easy handling and
the flexible and modular use of your
DEWETRON product and draw upon
more than 30 years of DEWETRON
expertise in measurement engineering.
ISO9001
EN ISO 14001
THE MEASURABLE DIFFERENCE.

Copyright © DEWETRON GmbH
This document contains informaon which is protected by copyright. All rights are reserved. Reproducon,
adaptaon, or translaon without prior wrien permission is prohibited, except as allowed under the copyright
laws.
All trademarks and registered trademarks are acknowledged to be the property of their owners.

3
Notice
General Information, Safety Instructions
The information contained in this document is subject to change without notice.
DEWETRON GmbH (DEWETRON) shall not be liable for any errors contained in this document.
DEWETRON MAKES NO WARRANTIES OF ANY KIND WITH REGARD TO THIS DOCUMENT, WHETHER
EXPRESS OR IMPLIED. DEWETRON SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. DEWETRON shall not be liable for
any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any
other legal theory, in connection with the furnishing of this document or the use of the information in this
document.
Warranty Information:
A copy of the specic warranty terms applicable to your DEWETRON product and replacement parts can be
obtained from your local sales and service oce.
Restricted Rights Legend:
Use austrian law for duplication or disclosure.
DEWETRON GmbH
Parkring 4
8074 Grambach
Austria
Printing History:
Please refer to the page bottom for printing version.
Copyright © DEWETRON GmbH.
This document contains information which is protected by copyright. All rights are reserved. Reproduction,
adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright
laws.
All trademarks and registered trademarks are acknowledged to be the property of their owners.
450134 • DAQP/HSI/PAD Modules • Technical Reference Manual • Printing version 2.3.4 • November 18, 2020

4
Notice
Safety symbols in the manual
Indicates hazardous voltages.
WARNING Calls attention to a procedure, practice, or condition that could cause bodily
injury or death.
CAUTION Calls attention to a procedure, practice, or condition that could possibly cause
damage to equipment or permanent loss of data.
WARNINGS
The following general safety precautions must be observed during all phases of operation, service, and
repair of this product. Failure to comply with these precautions or with specic warnings elsewhere in this
manual violates safety standards of design, manufacture, and intended use of the product. DEWETRON
Elektronische Messgeraete Ges.m.b.H. assumes no liability for the customer’s failure to comply with these
requirements.
All accessories shown in this document are available as option
and will not be shipped as standard parts.

5
Safety Instructions
Safety instructions for DEWETRON amplifiers
The DEWETRON data acquisition systems and ampliers may only be installed by experts.
Read your manual carefully before operating.
Observe local laws when using the ampliers.
Ground the equipment: For Safety Class 1 equipment (equipment having a protective earth terminal), a
non interruptible safety earth ground must be provided from the mains power source to the product input
wiring terminals or supplied power cable.
DO NOT operate the product in an explosive atmosphere or in the presence of ammable gases
or fumes.
DO NOT operate damaged equipment: Whenever it is possible that the safety protection features built
into this product have been impaired, either through physical damage, excessive moisture, or any other
reason, REMOVE POWER and do not use the product until safe operation can be veried by service-
trained personnel. If necessary, return the product to a DEWETRON sales and service oce for service
and repair to ensure that safety features are maintained.
Keep away from live circuits: Operating personnel must not remove equipment covers or shields.
Procedures involving the removal of covers or shields are for use by service-trained personnel only.
Under certain conditions, dangerous voltages may exist even with the equipment switched o. To avoid
dangerous electrical shock, DO NOT perform procedures involving cover or shield removal unless you
are qualied to do so.
No modications are allowed at the ampliers.
DO NOT service or adjust alone. Do not attempt internal service or adjustment unless another person,
capable of rendering rst aid and resuscitation, is present.
DO NOT substitute parts or modify equipment: Because of the danger of introducing additional hazards,
do not install substitute parts or perform any unauthorized modication to the product. Return the product
to a DEWETRON sales and service oce for service and repair to ensure that safety features are
maintained.
DO NOT touch internal wiring!
DO NOT use higher supply voltage than specied!
Use only original plugs and cables for harnessing.
Safety of the operator and the unit depend on following these rules.
450134 • DAQP/HSI/PAD Modules • Technical Reference Manual • Printing version 2.3.4 • November 18, 2020

6
Support
For any support please contact your local distributor rst or DEWETRON directly.
For Asia and Europe, please contact:
DEWETRON GmbH
Parkring 4
8074 Grambach
AUSTRIA
Tel.: +43 316 3070
Fax: +43 316 307090
Web: http://www.dewetron.com
The telephone hotline is available Monday to Friday between 08:00 and 12:00 CET (GMT -1:00) and Monday
to Thursday between 13:00 and 17:00 CET.
For the Americas, please contact:
DEWETRON, Inc.
2850 South County Trail, Unit 1
East Greenwich, RI 02818
U.S.A.
Tel.: +1 401 284 3750
Toll-free: +1 866 598 3393
Fax: +1 401 284 3755
Web: http://www.dewetron.us
The telephone hotline is available Monday to Friday between 08:00 and 17:00 GST (GMT +5:00)

12345
6789
7
General Module Information
9-pin male SUB-D connector
Interface pin assignment:
1 Module output (±5 V)
2 RS-485 (A)
3 RS-485 (B)
4 GND
5 +9 V power supply
6 +12 V power / sensor supply
7 Module input (from D/A
converter of the A/D board)1)
8 reserved
9 -9 V power supply
1) Triggerout at DAQP-FREQ-A
Calibration information
All DEWETRON modules are calibrated at 25 °C after a warmup time of 30 minutes and meet their
specications when leaving the factory.
The time interval for recalibration depends on environmental conditions. Typically, the calibration should be
checked once a year.
Calibration certicates are available from DEWETRON as an option. DEWETRON oers two types:
ISO traceable DEWETRON certicate
Calibration certicate according to ÖKD (equivalent to DKD)
This manual contains no calibration information. For self calibration, there is a separate calibration kit for the
DAQ series modules available. The CAL-KIT contains the required cables, software and instructions.
General module specifications
Module dimensions: 20 x 65 x 105 mm (0.79 x 2.56 x 4.13 in.)
(W x H x D without front cover and connectors)
Frontcover: 20 x 87 x 2 mm (0.79 x 3.43 x 0.08 in.)
(W x H x D without connector)
Environmental:
Temp. range storage: -30 °C to +85 °C (-22 °F to 185 °F)
Temp. range operating: -5 °C to +60 °C (23 °F to 140 °F)
Relative humidity
(MIL202): 0 to 95 % at 60 °C, non-condensing (unless otherwise noticed)
RFI susceptibility: ±0.5 % span error at 400 MHz, 5 W, 3 m
All specications within this manual are valid at 25 °C!
All modules are produced according ISO9001 and ISO14001.
Module connectors
Frontpanel connector: Accessable to the user. The connector type and pin assignment varies
from module to module. Detailed pin assignment of each module is
shown in the appropriate module description.
Rear connector: 9-pin male SUB-D, interface to the DEWE-System, not accessable to the user.
HSI/DAQx and PAD module
rear view
450134 • DAQP/HSI/PAD Modules • Technical Reference Manual • Printing version 2.3.4 • November 18, 2020

DEWETRON
data acquisition
system
RS-232/485
conversion on
DEWETRON-
motherboard
DAQx or PAD
module
RS-232
interface
RS-485
businterface
8
RS-232/485 interface
HSI/DAQP modules can be congured via
RS-485 interface, PAD modules require
this interface for all data transfers.
For all DEWETRON systems, an internal
RS-232/485 converter is available
(standard with DEWE-800, -2000, -2500, -3000, -4000, -5000 series systems). This converter allows
communication with HSI/DAQP and PAD modules.
To communicate with the modules, the RS-232 interface has to be set to the following parameters:
baud rate: 9600 bps
data bits: 8
parity: no parity
stop bits: 1
handshake: not required
HSI/DAQP module configuration
1. Push button selection
All ranges and lters can be selected directly by pressing the push buttons on the module. Approx. 15 sec.
after changing range and / or lter, the range and lter information is stored in an EEPROM. This procedure
increases the livetime of the EEPROM.
The current input range setting is shown all the time by LED. To change the range just press RANGE button
a few times until the required range is displayed.
To see the current lter setting just press the FILTER button once. The corresponding LED is ashing for
approx. 3 seconds. Within this time, the lter can be selected by pressing the FILTER button again. Approx.
3 seconds after the last key activity, the information will be stored, the LED stops ashing and shows the
input range again.
CAUTION: Power loss during this time leaves the module in the former settings.
2. RS-232/485 programming
All ranges and lters can also be selected via RS-232/485 interface. All new DEWE-800, -2000, -2500,
-3000, -4000, -5000 series systems are prepared as a standard to work with HSI/DAQP modules.
The easiest way to change the conguration is to use the DEWECong software, which comes as a standard
with the DEWETRON data acquisition system.
Detailed information about HSI/DAQP modules programming for customer applications is available in the
DEWE-Modules Programmers Reference Manual.
CAUTION: All range and lter changes which are done via RS-232/485 interface are not stored in the
EEPROM of the HSI/DAQP modules! You have to store this information in a separat initialisation le to
keep settings information for next system start!
PAD module communication
All PAD modules are only working through the RS-232/485 interface. All new DEWE-800, -2000, -2500,
-3000, -4000, -5000 series systems are prepared as a standard to work with PAD modules. The easiest
way to change the conguration is to use the DEWECong software, which comes as a standard with the
DEWETRON data acquisition system.
Detailed information about PAD modules programming for customer applications is available in the
DEWE-Modules Programmers Reference Manual.
General Module Information

9
nical Reference Manual • Prin
450134 • DAQP/HSI/PAD Modules • Tech ti ng version 2.3.4 • November 18, 2020
DAQP-LV
Isolated voltage amplifier
Filter bandwidth: 300 kHz, 10 selectable lowpass lters
Input ranges: 12 ranges (10 mV to 50 V)
Input type: AC and DC coupling software selectable
TEDS: Supports electronic data sheet sensors
Signal connection:
DAQP-LV-B: Banana plugs
DAQP-LV-BNC: BNC connector
DAQP-LV-D: 9-pin D-SUB connector
DAQP-LV-LEMO: 7-pin LEMO connector (non-standard version)
DAQP-LV-SC: Screw terminals (non-standard version)
Module specifications
DAQP-LV
Input ranges unipolar and bipolar 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2.5 V, 5 V, 10 V, 25 V, 50 V
Push button selectable ranges 10 mV, 50 mV, 200 mV, 1 V, 5 V, 10 V, 50 V
Rated input voltage 33 VRMS, 46.7 VPEAK, 70 VDC according to EN-61010-1 and EN-61010-2-30
Accuracy(1)
Unipolar
10 mV to 50 mV DC ±0.04 % of reading ±40 µV
100 mV to 50 V DC ±0.04 % of reading ±0.05 % of range
Bipolar
10 mV to 50 mV DC ±0.02 % of reading ±40 µV
1 Hz to 1 kHz ±0.12 % of reading ±0.1 % of range ±40 µV
1 kHz to 10 kHz ±0.32 % of reading ±0.1 % of range ±40 µV
100 mV to 50 V DC ±0.02 % of reading ±0.05 of range
1 Hz to 1 kHz ±0.12 % of reading ±0.15 of range
1 kHz to 10 kHz ±0.32 % of reading ±0.15 of range
Input coupling DC or AC software selectable (1.5 Hz standard, custom on request down to 0.01 Hz)
Gain linearity 0.01 % of full scale
Gain drift range Typically 10 ppm/°K (max. 20 ppm/°K)
O set drift
10 mV to 200 mV
500 mV to 50 V
Uni- and bipolar
3 µV/°K
10 ppm of Range/°K
Long term stability 100 ppm/sqrt (1000 hrs)
Input resistance 1 MOhm
Bandwidth (-3 dB) 300 kHz
Filter selection Push button or software
Filters (low pass) 10 Hz, 30 Hz, 100 Hz, 300 Hz, 1 kHz, 3 kHz, 10 kHz, 30 kHz, 100 kHz
Filter characteristics 10 Hz to 100 kHz: Butterworth or Bessel 40 dB/dec (2nd order; ±1.5 dB @ f0)
300 kHz: Bessel 60 dB/dec (3rd order; 0 to -3 dB @ 300kHz)
Typical SFDR and SNR:
20 mV
1 V
50 V
300 kHz bandwidth 100 kHz bandwidth 10 kHz bandwidth
SFDR SNR SFDR SNR SFDR SNR
100 dB 72 dB 98 dB 76 dB 97 dB 84 dB
102 dB 82 dB 99 dB 93 dB 97 dB 96 dB
102 dB 82 dB 99 dB 93 dB 97 dB 96 dB
Typical CMRR 10 mV to 1 V range: 2.5 V to 50 V range:
>100 dB @ 50 Hz 90 dB @ 50 Hz
>100 dB @ 1 kHz 65 dB @ 1 kHz
83 dB @ 10 kHz 55 dB @ 10 kHz
Input overvoltage protection 350 VDC
Isolation voltage 1 kVRMS
2)
Sensor supply ±9 V (±1 %), 12 V (±5 %), 200 mA resettable fuse protected2)
Output voltage ±5 V
Output resistance <10 Ohm
Maximum output current 5 mA
Output protection Short to ground for 10 sec.
Power On default settings Software programable
Power supply ±9 VDC ±1 %
Power consumption 1.05 W without sensor supply
RS-485 interface Yes
TEDS Hardware support for TEDS (Transducer Electronic Data Sheet)
Supported TEDS chips DS2406, DS2430A, DS2432, DS2433, DS2431
Supported MSI MSI-V-ACC; MSI-V-RTD, MSI-V-CH-50
1) Conditions for accuracy: Module temperature is calibration temperature ±5 °C; humidity is 30 to 90 RH;
AC accuracy: the highest lter (300 kHz) must be activated.
For the 2 year accuracy multiply all % of range and % of reading values by 1.5
2) Although the rated input voltage is 33 VRMS, 46.7 VPEAK or 70 VDC according to EN-61010-1 and EN-61010-2-30, the galvanic isolation has been tested with
1 kVRMS for 1 min.
3) Overall current should not exceed DEWE-30-xx maximum power.
continued on next page ...

10
DAQP-LV
LED state
The DAQP-LV modules have a set of 8 LEDs showing the current input range (constant active), the lter
range (ashing) and the input mode setting. Seven of the dierent input ranges are indicated directly by
the LEDs. The additional four input ranges (only software selectable) are indicated by lightening the two
“neighbour” ranges. For example the 25 V input range is shown by the LEDs 50 V and 10 V.
Due to the large number of low pass lters, two LEDs are used to display the current frequency. The left LED
indicates the multiplier, the right one shows the exponent with the base of 10. Example: for the 10 kHz
frequency range, the lower left and the upper right LED are ashing (1x 104Hz = 10 000 Hz).
The U/B LED shows the input mode: If this LED is o, the bipolar input range is selected, otherwise the
unipolar mode is selected.
Input range and filter selection
The DAQP-LV module has two push buttons with multiple functions.
Range button: Push the RANGE button several times until the LED displays the desired input
range.
Pressing the RANGE button for more than three seconds changes the input
mode from bipolar to unipolar.
Filter button: Push the FILTER button once - the LEDs will ash for approx. 3 seconds and
display the current lter setting. Push the FILTER button within the three
seconds several times until the ashing LED displays the desired lter range.
DAQP-LV
Input ranges unipolar and bipolar 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2.5 V, 5 V, 10 V, 25 V, 50 V
Push button selectable ranges 10 mV, 50 mV, 200 mV, 1 V, 5 V, 10 V, 50 V
Rated input voltage 33 VRMS, 46.7 VPEAK, 70 VDC according to EN-61010-1 and EN-61010-2-30
Accuracy(1)
Unipolar
10 mV to 50 mV DC ±0.04 % of reading ±40 µV
100 mV to 50 V DC ±0.04 % of reading ±0.05 % of range
Bipolar
10 mV to 50 mV DC ±0.02 % of reading ±40 µV
1 Hz to 1 kHz ±0.12 % of reading ±0.1 % of range ±40 µV
1 kHz to 10 kHz ±0.32 % of reading ±0.1 % of range ±40 µV
100 mV to 50 V DC ±0.02 % of reading ±0.05 of range
1 Hz to 1 kHz ±0.12 % of reading ±0.15 of range
1 kHz to 10 kHz ±0.32 % of reading ±0.15 of range
Input coupling DC or AC software selectable (1.5 Hz standard, custom on request down to 0.01 Hz)
Gain linearity 0.01 % of full scale
Gain drift range Typically 10 ppm/°K (max. 20 ppm/°K)
Oset drift
10 mV to 200 mV
500 mV to 50 V
Uni- and bipolar
3 µV/°K
10 ppm of Range/°K
Long term stability 100 ppm/sqrt (1000 hrs)
Input resistance 1 MOhm
Bandwidth (-3 dB) 300 kHz
Filter selection Push button or software
Filters (low pass) 10 Hz, 30 Hz, 100 Hz, 300 Hz, 1 kHz, 3 kHz, 10 kHz, 30 kHz, 100 kHz
Filter characteristics 10 Hz to 100 kHz: Butterworth or Bessel 40 dB/dec (2nd order; ±1.5 dB @ f0)
300 kHz: Bessel 60 dB/dec (3rd order; 0 to -3 dB @ 300kHz)
Typical SFDR and SNR:
20 mV
1 V
50 V
300 kHz bandwidth 100 kHz bandwidth 10 kHz bandwidth
SFDR SNR SFDR SNR SFDR SNR
100 dB 72 dB 98 dB 76 dB 97 dB 84 dB
102 dB 82 dB 99 dB 93 dB 97 dB 96 dB
102 dB 82 dB 99 dB 93 dB 97 dB 96 dB
Typical CMRR 10 mV to 1 V range: 2.5 V to 50 V range:
>100 dB @ 50 Hz 90 dB @ 50 Hz
>100 dB @ 1 kHz 65 dB @ 1 kHz
83 dB @ 10 kHz 55 dB @ 10 kHz
Input overvoltage protection 350 VDC
Isolation voltage 1 kVRMS
2)
Sensor supply ±9 V (±1 %), 12 V (±5 %), 200 mA resettable fuse protected2)
Output voltage ±5 V
Output resistance <10 Ohm
Maximum output current 5 mA
Output protection Short to ground for 10 sec.
Power On default settings Software programable
Power supply ±9 VDC ±1 %
Power consumption 1.05 W without sensor supply
RS-485 interface Yes
TEDS Hardware support for TEDS (Transducer Electronic Data Sheet)
Supported TEDS chips DS2406, DS2430A, DS2432, DS2433, DS2431
Supported MSI MSI-V-ACC; MSI-V-RTD, MSI-V-CH-50
1) Conditions for accuracy: Module temperature is calibration temperature ±5 °C; humidity is 30 to 90 RH;
AC accuracy: the highest lter (300 kHz) must be activated.
For the 2 year accuracy multiply all % of range and % of reading values by 1.5
2) Although the rated input voltage is 33 VRMS, 46.7 VPEAK or 70 VDC according to EN-61010-1 and EN-61010-2-30, the galvanic isolation has been tested with
1 kVRMS for 1 min.
3) Overall current should not exceed DEWE-30-xx maximum power.
continued from previous page ...

11
DAQP-LV
Programmable
amplifier
2nd order
lowpass filter
Bessel/Butterworth
(10 Hz to 100 kHz)
3rd order
350 kHz filter
100nF 980k
20k
AC/DCAC/DC
Isolation
TEDS
controller
TEDS
IN+
IN-
+12V
GND
+9V
-9V
+Vs
-Vs
1
2
7
6
4
5
9
3
8
7
1
2
6
5
3
4
9-pin DSUB
7-pin LEMO
INPUT
CONNECTORS
RS-485(A)
RS-485(B)
reserved
reserved
±5 V out
+12V
GND
+9V
-9V
2
3
7
8
1
6
4
5
9
9-pin DSUB
OUTPUT
CONNECTOR
10 mV .. 1 V
2.5 V .. 50 V
Calibration
memory
optional 2W DC DC
±5 V, ±15 V
DAQP-LV-module
resettable
200 mA fuse
Frequency additional error with activated
Butterworth lter
additional error with activated
Bessel lter
f/f0% of reading % of reading
<0.1 0 0
0.01 0.00 0.00
0.02 0.00 0.02
0.03 0.00 0.04
0.05 0.00 0.11
0.1 0.01 0.47
0.2 0.14 1.9
0.3 0.73 4.3
0.5 5.24 12
0.75 20.34 25
1 40.45 40.45
CAUTION: TEDS terminal is not isolated!
Block diagram
The base block diagram of the DAQP-LV gives an idea of the internal structure.
AC accuracy with activated filter
With activated hardware lter an additional % of reading error has to be considered due to the damping of
the lter. This error depends on the signal frequency f and the selected ler frequency f0.
450134 • DAQP/HSI/PAD Modules • Technical Reference Manual • Printing version 2.3.4 • November 18, 2020

12
DAQP-LV
1
2
3
6
5
4
7
IN +
IN -
Red
Black
DAQP-LV-LEMO module (EoL: 01/2017)
Voltage measurement via LEMO cord
1 IN +
2 IN -
3 +9 V (200 mA max.)
4 -9 V (200 mA max.)
5 GND
6 +12 V (200 mA max.; +15 V in conjunction with a DEWE-30-4)
7 TEDS
7-pin LEMO connector female
EGG.1B.307
1 TEDS
2 IN +
3 Reserved for custom sensor supplies
4 GND (not isolated)
5 +9 V (200 mA max.)
6 +12 V (200 mA max.; +15 V in conjunction with a DEWE-30-4)
7 IN -
8 Reserved for custom sensor supplies
9 -9 V (200 mA max.)
Voltage measurement via D-SUB cord
Hot: IN +
Shield: IN -
DAQP-LV-BNC module
Voltage measurement via BNC cord
DAQP-LV-D module
If signals above 60 V may appear, don’t use the metal housing of D-SUB connector!
Signal connection
DAQP-LV-B module
Voltage measurement via banana plug cords

13
DAQP-LV
TEDS
GND
IN+
IN-
+9V
-9V
TEDS
Output
+
-
Supply
+
-
1
4
2
7
5
9
7
5
1
2
3
4
9-pin DSUB
Connectors Sensor
7-pin LEMO
Sensor
1
4
2
7
5
9
7
5
1
2
3
4
9-pin DSUB
Connectors
7-pin LEMO
TEDS
GND
IN+
IN-
+9V
-9V
4..20 mA
4..20 mA
output
50 OHM
6
2
7
4
6
1
2
5
9-pin DSUB
Connectors Sensor
2-wire
4..20 mA
7-pin LEMO
GND
IN+
IN-
50 OHM
+12V
Output
+
-
6
2
7
4
6
1
2
5
9-pin DSUB
Connectors Sensor
7-pin LEMO
GND
IN+
IN-
+12V
Output
GND
VCC
Typical sensor connection
Sensor with dierential output
powered by the module
Loop powered sensor Sensor with common ground
Current measurement
450134 • DAQP/HSI/PAD Modules • Technical Reference Manual • Printing version 2.3.4 • November 18, 2020

14
DAQP-LV
Notes

C1
L
E
V
M
C
Dipl.-Ing. Roland Jeutter / Managing director
Manufacturer:
Address:
Name of product:
Kind of product:
CE-Certificate of conformity
Graz, April 28, 2010
Place / Date of the CE-marking
DEWETRON GmbH
Parkring 4
8074 Grambach, Austria
Tel.: +43 316 3070 0
Fax: +43 316 3070 90
e-mail: [email protected]
http://www.dewetron.com
DEWE-MODULES
Signal conditioning modules
The product meets the regulations of the following EC-directives:
73/23/EEC
"Directive on the approximation of the laws of the Member States relating to electrical
equipment designed for use within certain voltage limits amended by the directive 93/68/
EEC"
89/336/EEC
"Directive on the approximation of the laws of the Member States relating to
electromagnetic compatibility amended by the directives 91/263/EEC, 92/31/EEC, 93/68/
EEC and 93/97/EEC
The accordance is proved by the observance of the following standards:
Safety
Emissions
Immunity
IEC/EN 61010-1:1992/93 IEC 61010-1:1992/300 V CATIII Pol. D. 2
IEC/EN 61010-2-031 IEC 1010-2-031
EN 61000-6-4 EN 55011 Class B
EN 61000-6-2 Group standard
CE-Certificate of conformity
450134 • DAQP/HSI/PAD Modules • Technical Reference Manual • Printing version 2.3.4 • November 18, 2020

C2
Notes
Table of contents
Other Dewetron Control Unit manuals

Dewetron
Dewetron DAQP-LA Series Product manual

Dewetron
Dewetron HSI-LV Product manual

Dewetron
Dewetron EPAD2 Series Product manual

Dewetron
Dewetron DAQP-STG Product manual

Dewetron
Dewetron DAQP-HV Product manual

Dewetron
Dewetron XR-RTD8 Product manual

Dewetron
Dewetron DAQP-ACC-A Product manual

Dewetron
Dewetron DAQP-CFB2 Product manual

Dewetron
Dewetron DAQP-MULTI Product manual

Dewetron
Dewetron DAQP-CHARGE-B Series Product manual