HSI-HV
Isolated high voltage amplifier
Input ranges: 1400 V to 20 V
Bandwidth: 2 MHz
Isolation: 1.8 kVRMS line to line
1.4 kVRMS line to ground
Input resistance: 10 MΩ
Protection: ±4 kV surge / burst
Module specifications
Input ranges ±20 V1), ±50 V1), ±100 V, ±200 V, ±400 V, ±800 V, ±1400 V
1 year accuracy2) Range Signal frequency Accuracy
20 V; 50 V DC ±0.05 % of reading ±60 mV
100 V to 1400 V DC ±0.05 % of reading ±0.05 % of range
0.1Hz to 500 Hz ±0.05 % of reading ±0.01 % of range
>500 Hz to 5 kHz ±0.1 % of reading ±0.05 % of range
>5 kHz to 50 kHz ±0.4 % of reading ±0.05 % of range
>50 kHz to 100 kHz ±(0.016*f) % of reading ±0.1 % of range
>100 kHz to 1 MHz ±(0.010*f) % of reading ±1 % of range
>1 MHz to 2 MHz ±(0.014*f) % of reading ±3 % of range
f = signal frequency in kHz
Gain linearity 0.05 %
Gain drift range Typically 20 ppm/°C (max. 50 ppm/°C)
O set drift
20 V to 100 V
200 V to 1400 V
typical 1.5 mV/°C of range max. 4 mV/°C
typical 5 ppm/°C max. 20 ppm of range/°C
Long term stability 100 ppm/sqrt (1000 hrs)
Input resistance 10 MΩ || 2.2 pF
-3 dB Bandwidth 2 MHz
Signal delay @ full bandwidth approx. 390 ns
Filter selection Push button or software
Filter (lowpass) 100, 300, 1k, 3k, 10k, 30k, 100k, 300 kHz, 1 MHz, 2 MHz3)
Filter type Bessel or Butterworth 40 dB/decade
Filter characteristics
100 Hz to 1 MHz
2 MHz
Butterworth or Bessel 40 dB/dec (2nd order; ±1.5 dB @ f0)
Butterworth 60 dB/dec (3rd order; 0 to -3 dB @ 2 MHz)
Typical SFDR and SNR
50 V
400 V
1400 V
10kHz bandwidth 100kHz bandwidth 1MHz bandwidth 2 MHz bandwidth
SFDR SNR SFDR SNR SFDR SNR SFDR SNR
110 dB 91 dB 110 dB 82 dB 94 dB 76 dB 84 dB 73 dB
110 dB 95 dB 110 dB 92 dB 94 dB 82 dB 84 dB 77 dB
110 dB 95 dB 110 dB 95 dB 94 dB 82 dB 84 dB 77 dB
Typical CMRR >80 dB @ 50 Hz 60 dB @ 1 kHz
70 dB @ 400 Hz 48 dB @ 10 kHz
Isolation voltage Line to Ground 1.4 kVrms
Line to Line 1.8 kVrms
Protection
Surge (1.2/50)
Burst (5 kHz)
CAT III 600
CAT IV 300
±4000 V
±4000 V
Output voltage ±5 V
Output resistance 10 Ohm
Output current maximum 35 mA CAUTION: do not exceed maximum output current!
Power supply ±9 VDC ± 1%
Power consumption 1.2 W
Power On default settings Software programable
Special functions Integrated temperature sensor
Programming interface RS-485
1) 20 V and 50 V are auxiliary ranges and have a limited bandwidth.
20 V range typically 0.9 Mhz
50 V range typically 1.9 Mhz
2) Conditions for accuracy: Module temperature is calibration temperature ±5 °C; humidity is 30 to 90 RH;
AC accuracy: the highest lter (2 Mhz) has to be activated. f = signal frequency in kHz.
for the 2 year accuarcy multiply all % of range and % of reading values by 1.5.
3) 2 MHz lter: exclusively Butterworth 60 dB/decade. Please consider possible bandwidth limitation of further components in the measuring chain, e.g. A/D cards or
signal conditioning mainframe
450134 • DAQP/HSI/PAD Modules • Technical Reference Manual • Printing version 2.3.4 • November 18, 2020