Important Safety Information
13
Operating Instructions Savina SW 2.1n 1. US ed.
S mmary of WARNINGS and CAUTIONS
WARNING!
Draeger cannot warrant or endorse the safe performance
of third party humidifiers for use with the Savina ventila
tor. Specifically, the user has to assess the risks of deliv
ery of breathing gas not maintained at a proper
temperature associated with different humidifier designs.
It is strongly recommended to use the electronic temper
ature monitoring feature of the ventilator if no proximal
airway temperature monitoring is performed by the
humidifier used.
Increased pneumatic resistance in the inspiratory line
caused by a humidifier may result in less accurate airway
pressure readings.
We recommend contacting the manufacturers/ distribu
tors of third party humidifier devices about compliance of
their products with the requested performance character
istics.
WARNING!
Do not use a heat/moisture exchanger (HME) simulta
neously with a heated humidifier!
Risk of increased breathing resistance due to conden
sation.
WARNING!
In order to avoid any risk of electric shock in the event of
faulty grounding of patient monitoring equipment, do not
use antistatic or electrically conductive patient circuits
.
WARNING!
Do not use inspiratory circuits with a total length of less
than 1.2 m (4 feet) between humidifier and wye.
Otherwise risk of excessive breathing gas temperatures!
WARNING !
Do not use humidifiers without temperature regulation.
Otherwise risk of excessive breathing gas temperature!
WARNING !
This device is to be used only in rooms with line power
installations complying with national safety standards
for hospital patient rooms. (e.g., IEC 601.1, "Safety of
Medical Equipment).
To maintain grounding integrity, connect only to a
"hospital grade" receptacle. Always disconnect supply
before servicing.
WARNING !
Connecting other devices to the same extension power
strip may, in the event of grounding failure, cause the
leakage current to the patient to increase beyond the
permissible values.
In this case, the risk of electric shock cannot be safely
excluded.
WARNING !
Use rechargeable battteries ONLY.
Because of the charging feature of the Savina DC power
supply, connecting non rechargeable batteries might lead
to a risk of explosion when operating the ventilator from
line power.
WARNING !
Always use medical grade oxygen that is dry and free
from contaminations. Contaminated gas may cause
ventilator malfunction.
Ensure a reliable source of 100% oxygen. Brief fluctua
tions in the concentration may lead to an incorrect cali
bration of the O2 sensor.
WARNING !
The ventilator is ready for operation only when:
●
it is completely assembled with all required auxiliary
equipment in place,
●
all sensors are calibrated (O
2
, Flow),
●
all tests of readiness for operation have been
completed successfully.
WARNING !
●Installation of the Savina nurse call kit should only be
performed by DraegerService or factory trained and
authorized service personnel.
WARNING !
The operator of the ventilator must still assume full
responsibility for ventilation monitoring via the Savina
screen when the nurse call is connected.
Only warning messages (!!!), i. e. alarm messages with
top priority, will activate the nurse call.
●Regularly check on screen messages.