
PMC PROCEDURE (continued)
NM MRI
PM Certification Procedure for Narkomed MRI Anesthesia System
1. Use the PM Certification form for Narkomed MRI Anesthesia Systems (P/N 4114551).
2. Completely fill in the header information.
3. Replace the VENTILATOR RELIEF VALVE DIAPHRAGM in accordance with SP
00075. Place a check mark and the replacement date at “VENT VALVE REPLACEMENT”
line on the Periodic Manufacturer’s Certification form.
4. Check all vapor 19 and 19.1 vaporizers for correct labeling. All vaporizers must have a label
stating "THE CONCENTRATION OUTPUT OF THIS VAPORIZER SHALL BE VERIFIED
AFTER IT HAS BEEN ATTACHED TO THE ANESTHESIA MACHINE" (part # S010015).
This label shall be attached to the rear of the vaporizer directly below the mount.
5. All Key Index Safety Systems vaporizers, (K.I.S.S.) must have a label stating "CAUTION:
AFTER FILLING HAS BEEN COMPLETED, REINSERT PLUG INTO UPPER FILLER
PORT AND TIGHTEN LOCKING SCREW" (part # 4112520-001). This label shall be attached
to the vaporizer directly above the keyed filler. Place a check mark at "K.I.S.S. LABEL" on the
PM Certification form.
6. If machine is equipped with a HALOTHANE Dräger Vapor 19 or 19.1 vaporizer, determine if
vaporizer must be inspected for soil condition one. Check the serial number plate located on
the rear of the vaporizer for a plus (+) preceding the serial number. A HALOTHANE vaporizer
serial number not preceded with a (+) must be tested for soil in accordance with SP00073. If
vaporizer does not need to be inspected, indicate so with a plus (+) next to the “Vapor
Inspection (H)” line on the Vigilance Audit form. If vaporizer is soil condition 0, indicate so
with “SOIL 0” written next to the “Vapor Inspection (H)” line on the Vigilance Audit form. If
vaporizer is soil condition one, indicate so with “SOIL 1” writen next to the “Vapor Inspection
(H)” line on the Vigilance Audit form. Place a “CAUTION DO NOT USE” label (part #
4114327) on the vaporizer, and issue a departmental alert. The TSR shall also seek
permission from the equipment operator to remove the failed vaporizer from the machine and
apply a replacement vaporizer or an adapter block onto the mount. All “SOIL 1” vaporizers
must be removed from service for machine to receive certification.
7. Perform the vapor concentration test on all Dräger vapor vaporizers in accordance with
SP00073. For every vaporizer tested, fill out a "VAPOR VAPORIZER CALIBRATION
CHECK" label (part # S010016). Information on this label shall include your signature, type
of agent, date tested, test results @ 1%, 2.5%, 4%, and a PASS or FAIL indication. This label
shall be attached to the upper right side of the vaporizer. If vaporizer fails the concentration
test, check "NO" in the "RECOMMENDED FOR USE" section on the PM Certification form.
Place a "CAUTION DO NOT USE" label (part # 4114327) on the vaporizer, and issue a
departmental alert. The TSR shall also seek permission from the equipment operator to
remove the failed vaporizer from the machine and install a replacement vaporizer or an
adapter block onto the mount. All nonfunctional Dräger vapor vaporizers must be removed
from service for machine to receive certification.
6-2J
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS